| Literature DB >> 17592454 |
Peter Dunscombe1, Harry Johnson, Clement Aresnault, George Mawko, Jean-Pierre Bissonnette, Jan Seuntjens.
Abstract
Among the essential components of a comprehensive quality assurance program in radiotherapy are the quality control protocols to be used on the equipment and, in particular, the performance objectives and criteria. In the present work, we describe the development of a suite of quality control documents for use across Canada. Following a generic format, we are generating concise, clear standards for the most commonly used equipment in radiotherapy, with the emphasis on performance measures. The final standards of performance are confirmed following cross-country consultation facilitated by the availability of draft documents on the Canadian Medical Physics web site.Entities:
Mesh:
Year: 2006 PMID: 17592454 PMCID: PMC5722407 DOI: 10.1120/jacmp.v8i1.2380
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Quality control tests for medical linear accelerators (tolerances and action levels are specified in millimeters unless otherwise stated)
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
|
| |||
| DL1 | Door interlock/last person out | Functional | |
| DL2 | Motion interlock | Functional | |
| DL3 | Couch brakes | Functional | |
| DL4 | Beam status indicators | Functional | |
| DL5 | Patient audiovisual monitors | Functional | |
| DL6 | Room radiation monitors | Functional | |
| DL7 | Beam interrupt/counters | Functional | |
| DL8 | Lasers/crosswires | 1 | 2 |
| DL9 | Optical distance indicator | 1 | 2 |
| DL10 | Optical back pointer | 2 | 3 |
| DL11 | Field size indicator | 1 | 2 |
| DL12 | Output constancy—photons | 2% | 3% |
| DL13 | Dynamic wedge factors | 1% | 2% |
| DL14 | Output constancy—electrons | 2% | 3% |
|
| |||
| ML1 | Emergency off | Functional | |
| ML2 | Wedge, tray cone interlocks | Functional | |
| ML3 | Accessories integrity and centering | Functional | |
| ML4 | Gantry angle readouts | 0.5° | 1° |
| ML5 | Collimator angle readouts | 0.5° | 1° |
| ML6 | Couch position readouts | 1 | 2 |
| ML7 | Couch isocenter | 1 | 2 |
| ML8 | Couch angle | 0.5° | 1° |
| ML9 | Optical distance indicator | 1 | 2 |
| ML10 | Crosswire centering | 1 | 2 |
| ML11 | Light/radiation coincidence | 1 | 2 |
| ML12 | Field size indicator | 1 | 2 |
| ML13 | Relative dosimetry | 1% | 2% |
| ML14 | Central axis depth dose reproducibility | 1 (2%) | 2 (3%) |
| ML15 | Beam flatness | 2% | 3% |
| ML16 | Beam symmetry | 2% | 3% |
| ML17 | Records | Complete | |
|
| |||
| AL1 | Reference dosimetry—TG51 | 1% | 2% |
| AL2 | Relative output factor reproducibility | 1% | 2% |
| AL3 | Wedge transmission factor reproducibility | 1% | 2% |
| AL4 | Accessory transmission factor reproducibility | 1% | 2% |
| AL5 | Output reproducibility vs. gantry angle | 1% | 2% |
| AL6 | Beam symmetry reproducibility vs. gantry angle | 2% | 3% |
| AL7 | Monitor chamber linearity | 1% | 2% |
| AL8 | End monitor effect | 0.1 MU | 0.2 MU |
| AL9 | Collimator rotation isocenter | 1 | 2 |
| AL10 | Gantry rotation isocenter | 1 | 2 |
| AL11 | Couch rotation isocenter | 1 | 2 |
| AL12 | Coincidence of collimator, gantry, couch axes | 1 | 2 |
| AL13 | Coincidence of isocenters | 1 | 2 |
| AL14 | Couch deflection | 3 | 5 |
| AL15 | Independent quality control review | Complete | |
Quality control tests for electronic portal imaging devices (tolerances and action levels are specified in millimeters unless otherwise stated)
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
|
| |||
| DE1 | Mechanical integrity | Functional | |
| DE2 | Electrical integrity | Functional | |
| DE3 | Collision interlocks | Functional | |
| DE4 | Image quality | Reproducibility | |
|
| |||
| ME1 | Positioning in the imaging plane | 1 | 2 |
| ME2 | Positioning perpendicular to the imaging plane | 10 | 20 |
| ME3 | Image quality | Reproducibility | |
| ME4 | Artifacts | Reproducibility | |
| ME5 | Spatial distortion | 1 | 2 |
| ME6 | Monitor controls | Reproducibility | |
| ME7 | Records | Complete | |
|
| |||
| SE1 | Spatial resolution | Reproducibility | |
| SE2 | Noise | Reproducibility | |
| SE3 | On‐screen measurement tools | 0.5 | 1 |
| SE4 | Setup verification tools | 0.5 (0.5°) | 1 (1°) |
|
| |||
| AE1 | Independent quality control review | Complete | |
Quality control tests for conventional simulators (tolerances and action levels are specified in millimeters unless otherwise stated)
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
|
| |||
| DS1 | Door interlock | Functional | |
| DS2 | Motion interlock | Functional | |
| DS3 | Beam status indicators | Functional | |
| DS4 | Emergency off buttons | Functional | |
| DS5 | Collision avoidance | Functional | |
| DS6 | Lasers/crosswires | 1 | 2 |
| DS7 | Optical distance indicator | 1 | 2 |
| DS8 | Crosswires/reticle/block tray | 1 | 2 |
| DS9 | Light/radiation coincidence | 1 | 2 |
| DS10 | Field size indicators | 1 | 2 |
|
| |||
| MS1 | Gantry angle readouts | 0.5° | 1° |
| MS2 | Collimator angle readouts | 0.5° | 1° |
| MS3 | Couch position readouts | 1 | 2 |
| MS4 | Alignment of FAD movement | 1 | 2 |
| MS5 | Couch isocenter | 2 | 3 |
| MS6 | Couch parallelism | 1 | 2 |
| MS7 | Laser/crosswire isocentricity | 1 | 2 |
| MS8 | Optical distance indicator | 1 | 2 |
| MS9 | Crosswire centering | 1 | 2 |
| MS10 | Light/radiation coincidence | 1 | 2 |
| MS11 | Field size indicators | 1 | 2 |
| MS12 | Records | Complete | |
|
| |||
| SS1 | Lead apron | Functional | |
| SS2 |
| 5% | 10% |
| SS3 | Reference dosimetry | 5% | 10% |
| SS4 | Beam quality (HVL) | 5% | 10% |
| SS5 | Automatic exposure control | 5% | 10% |
| SS6 | Focal spot | Reproducible | |
| SS7 | Contrast | Reproducible | |
| SS8 | Resolution | Reproducible | |
| SS9 | Fluoroscopic timer | 5% | 10% |
|
| |||
| AS1 | Redefine isocenter | 1 | 2 |
| AS2 | Couch deflection | 3 | 5 |
| AS3 | Alignment of focal spots | 0.5 | 1 |
| AS4 | Independent quality control review | Complete | |
Quality control tests for kilovoltage radiotherapy units (tolerances and action levels are specified in millimeters unless otherwise stated)
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
|
| |||
| DK1 | Patient monitoring audiovisual devices | Functional | |
| DK2 | Door closing mechanism and interlock | Functional | |
| DK3 | Couch movement and brakes | Functional | |
| DK4 | Unit motions and motion stops | Functional | |
| DK5 | Interlocks for added filters/kV‐filter choice | Functional | |
| DK6 | Beam status indicators | Functional | |
| DK7 | Beam‐off at key‐off test | Functional | |
| DK8 | Emergency off test | Functional | |
| DK9 | kV and mA indicators | Functional | |
| DK10 | Backup timer/monitor unit channel check | 1% | 2% |
| DK11 | Dosimetric test: output check | 3% | 5% |
|
| |||
| MK1 | Mechanical stability and safety | Functional | |
| MK2 | Cone selection and competency | Functional | |
| MK3 | Physical distance indicators | 2 | 3 |
| MK4 | Accuracy of head tilt and rotation readouts | 1° | 1.5° |
| MK5 | Light/x‐ray field coincidence | 2 | 3 |
| MK6 | Light field size | 2 | 3 |
| MK7 | X‐ray field size indicator | 2 | 3 |
| MK8 | X‐ray field uniformity/filter integrity | 5% | 8% |
| MK9 | Timer and end effect error | Characterize |
|
| MK10 | Output linearity | 1% | |
| MK11 | Output reproducibility | Characterize | <.03 CoV |
| MK12 | Beam quality | 10% | 15% |
| MK13 | Output calibration verification | 2% | 3% |
| MK14 | Timer accuracy verification | 2% | 3% |
| MK15 | Records | Complete | |
|
| |||
| AK1 | Reference dosimetry | 1% | 2% |
| AK2 | Alignment of focal spots | 0.5 | 1 |
| AK3 | kVp measurement | 5% | 10% |
| AK4 | Focal spot size | Reproducible | |
| AK5 | Independent quality control review | Complete | |
Quality control tests for teletherapy units (tolerances and action levels are specified in millimeters unless otherwise stated)
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
|
| |||
| DCO1 | Door interlock/last person out | Functional | |
| DCO2 | Motion interlock | Functional | |
| DCO3 | Couch brakes | Functional | |
| DCO4 | Beam status indicators | Functional | |
| DCO5 | Patient audiovisual monitors | Functional | |
| DCO6 | Room radiation monitors | Functional | |
| DCO7 | Emergency off | Functional | |
| DCO8 | Beam interrupt/counters | Functional | |
| DCO9 | Head swivel lock | Functional | |
| DCO10 | Lasers/crosswires | 1 | 2 |
| DCO11 | Optical distance indicator | 1 | 2 |
| DCO12 | Optical back pointer | 2 | 3 |
| DCO13 | Field size indicator | 1 | 2 |
|
| |||
| MCO1 | Latching of wedges, trays | Functional | |
| MCO2 | Wedge interlocks | Functional | |
| MCO3 | Gantry angle readouts | 0.5° | 1° |
| MCO4 | Collimator angle readouts | 0.5° | 1° |
| MCO5 | Couch position readouts | 1 | 2 |
| MCO6 | Couch rotation isocenter | 2 | 3 |
| MCO7 | Optical distance indicator | 1 | 2 |
| MCO8 | Crosswire centering | 1 | 2 |
| MCO9 | Light/Radiation coincidence | 2 | 3 |
| MCO10 | Field size indicator | 1 | 2 |
| MCO11 | Relative dosimetry | 1% | 2% |
| MCO12 | Shutter error | Reproducible | |
| MCO13 | Beam symmetry (source position) | 2% | 3% |
| MCO14 | Records | Complete | |
|
| |||
| ACO1 | Reference dosimetry | 1% | 2% |
| ACO2 | Relative output factor reproducibility | 1% | 2% |
| ACO3 | Central axis depth dose reproducibility | 1% | 2% |
| ACO4 | Wedge transmission factor reproducibility | 1% | 2% |
| ACO5 | Accessory transmission factor reproducibility | 1% | 2% |
| ACO6 | Output reproducibility vs. gantry angle | 1% | 2% |
| ACO7 | Beam symmetry reproducibility vs. gantry angle | 2% | 3% |
| ACO8 | Timer linearity | 1% | 2% |
| ACO9 | Shutter error | 0.03 min. | 0.05 min. |
| ACO10 | Collimator rotation isocenter | 2 | 3 |
| ACO11 | Gantry rotation isocenter | 2 | 3 |
| ACO12 | Couch rotation isocenter | 2 | 3 |
| ACO13 | Coincidence of collimator, gantry, couch axes | 2 | 3 |
| ACO14 | Coincidence of isocenters | 2 | 3 |
| ACO15 | Couch deflection | 3 | 5 |
| ACO16 | Independent quality control review | Complete | |
Quality control tests for multileaf collimators (tolerances and action levels are specified in millimeters unless otherwise stated)
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
|
| |||
| PM1 | Verification of transferred data vs. printed template | 1 | 2 |
| PM2 | Daily verification of correct data | Reproducibility | |
| PM3 | Verification of record and verify programming | Reproducibility | |
|
| |||
| MM1 | Digitizer check (if used) | Functional | |
| MM2 | Light and radiation field coincidence | 1 | 2 |
| MM3 | Leaf positions for standard field template | 1 | 2 |
| MM4 | Electron field interlocks | Functional | |
| MM5 | Leaf alignment | 1 | |
| MM6 | Records | Complete | |
| Yearly | |||
| AM1 | Leaf transmission (all energies) | Reproducibility | |
| AM2 | Leakage between leaves (all energies) | Reproducibility | |
| AM3 | Transmission through abutting leaves | Reproducibility | |
| AM4 | Stability with gantry rotation | Reproducibility | |
| AM5 | Alignment with jaws | 1 | |
| AM6 | Independent quality control review | Complete | |
May not apply to all designs.