| Literature DB >> 15770195 |
Mark J Rivard1, Dee-Ann Radford Evans, Ian Kay.
Abstract
The Fully Integrated Real-time Seed Treatment (FIRST) system by Nucletron has been available in Europe since November 2001 and is being used more and more in Canada and the United States. Like the conventional transrectal ultrasound implant procedure, the FIRST system utilizes an ultrasound probe, needles, and brachytherapy seeds. However, this system is unique in that it (1) utilizes a low-dose-rate brachytherapy seed remote afterloader (the seedSelectron), (2) utilizes 3D image reconstruction acquired from electromechanically controlled, nonstepping rotation of the ultrasound probe, (3) integrates the control of a remote afterloader with electromechanical control of the ultrasound probe for integrating the clinical procedure into a single system, and (4) automates the transfer of planning information and seed delivery to improve quality assurance and radiation safety. This automated delivery system is specifically intended to address reproducibility and accuracy of seed positioning during implantation. The FIRST computer system includes two software environments: SPOT PRO and seedSelectron; both are used to facilitate treatment planning and brachytherapy seed implantation from beginning to completion of the entire procedure. In addition to these features, the system is reported to meet certain product specifications for seed delivery positioning accuracy and reproducibility, seed calibration accuracy and reliability, and brachytherapy dosimetry calculations. Consequently, a technical evaluation of the FIRST system was performed to determine adherence to manufacturer specifications and to the American Association of Physicists in Medicine (AAPM) Task Group Reports 43, 53, 56, 59, and 64 and recommendations of the American Brachytherapy Society (ABS). The United States Nuclear Regulatory Commission (NRC) has recently added Licensing Guidance for the seedSelectron system under 10 CFR 35.1000. Adherence to licensing guidance is made by referencing applicable AAPM Task Group recommendations. In general, results of this evaluation indicated that the system met its claimed specifications as well as the applicable recommendations outlined in the AAPM and ABS reports.Entities:
Mesh:
Year: 2005 PMID: 15770195 PMCID: PMC5723507 DOI: 10.1120/jacmp.v6i1.1985
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1(a) The complete assembly of the image acquisition system and seedSelectron. Left to right: template, bimodal ultrasound probe, ultrasound probe stepper assembly, endocavity rotational mover (ECRM), and the seedSelectron. (b) closeup of the Nucletron seedSelectron.
Figure 2Nucletron selectSeed and selectSpacer cartridges oriented to mate with the delivery drivewire.
Seed X/Spacer O configurations tested to determine whether the seedSelectron prevents configurations with lengths greater than 80 mm. Unacceptable build indicates that the system prevented the configuration from being built because the nominal length exceeded 80 mm.
| Build tube position | ||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Needle configuration | 15 | 14 | 13 | 12 | 11 | 10 | 9 | 8 | 7 | 6 | 5 | 4 | 3 | 2 | 1 | 0 | Length (mm) | Acceptable build ? |
| 1 | X | O | O | O | O | O | O | O | O | O | O | O | O | O | X | 80.5 | yes | |
| 2 | X | X | O | O | O | O | O | O | O | O | O | O | O | O | X | 79.5 | yes | |
| 3 | X | X | X | O | O | O | O | O | O | O | O | O | O | O | X | 78.5 | yes | |
| 4 | X | O | O | O | O | O | O | O | O | O | O | O | O | O | X | 80.5 | yes | |
| 5 | X | X | O | O | O | O | O | O | O | O | O | O | O | O | X | 79.5 | yes | |
| 6 | X | X | X | O | O | O | O | O | O | O | O | O | O | O | X | 78.5 | yes | |
| 7 | X | O | O | O | O | O | O | O | O | O | O | O | O | O | O | X | 86.0 | no |
| 8 | X | X | O | O | O | O | O | O | O | O | O | O | O | O | O | X | 85.0 | no |
| 9 | X | X | X | O | O | O | O | O | O | O | O | O | O | O | O | X | 84.0 | no |
| 10 | X | X | X | X | O | O | O | O | O | O | O | O | O | O | O | X | 83.0 | no |
| 11 | X | X | X | X | X | O | O | O | O | O | O | O | O | O | O | X | 82.0 | no |
| 12 | X | X | X | X | X | X | O | O | O | O | O | O | O | O | O | X | 81.0 | no |
| 13 | X | X | X | X | X | X | X | O | O | O | O | O | O | O | O | X | 80.0 | yes |
| 14 | X | X | X | X | X | X | X | X | O | O | O | O | O | O | O | X | 79.0 | yes |
| 15 | X | X | X | X | X | X | X | X | X | O | O | O | O | O | O | X | 78.0 | yes |
| 16 | X | X | X | X | X | X | X | X | X | X | O | O | O | O | O | X | 77.0 | yes |
| 17 | X | X | X | X | X | X | X | X | X | X | X | O | O | O | O | X | 76.0 | yes |
| 18 | X | X | X | X | X | X | X | X | X | X | X | X | O | O | O | X | 75.0 | yes |
| 19 | X | X | X | X | X | X | X | X | X | X | X | X | X | O | O | X | 74.0 | yes |
| 20 | X | X | X | X | X | X | X | X | X | X | X | X | X | X | O | X | 73.0 | yes |
| 21 | X | X | X | X | X | X | X | X | X | X | X | X | X | X | X | X | 72.0 | yes |
Figure 3Nucletron QA tool regularly used for verifying positioning accuracy of the delivery drivewire and accuracy of seed and spacer delivery.
Measurements of seed positioning accuracy; results for all positions are in millimeters.
| Variable needle position | Fixed needle offset | Nucletron‐calculated offset | Pythagorean‐calculated offset | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Row | A | B | C | D | A | B | C | D | A | B | C | D | A | B | C | D |
| 4.0 | 1.5 | 1.0 | 0.5 | 0.0 | 1.7 | 1.3 | 0.7 | 0.0 | 2.0 | 1.2 | 0.5 | 0.0 | 1.5 | 0.7 | 0.2 | 0.0 |
| 3.5 | 1.8 | 1.0 | 0.5 | 0.0 | 1.8 | 1.3 | 1.0 | 0.0 | 2.0 | 1.2 | 0.5 | 0.2 | 1.5 | 0.7 | 0.2 | 0.0 |
| 3.0 | 2.0 | 1.0 | 0.8 | 0.3 | 2.0 | 1.5 | 1.2 | 0.5 | 2.0 | 1.2 | 0.9 | 0.5 | 1.7 | 0.8 | 0.3 | 0.2 |
| 2.5 | 2.5 | 1.5 | 1.0 | 0.5 | 2.3 | 1.8 | 1.5 | 1.0 | 2.5 | 1.5 | 1.2 | 0.9 | 1.9 | 1.0 | 0.5 | 0.4 |
| 2.0 | 2.7 | 1.5 | 1.0 | 0.8 | 2.5 | 2.0 | 1.7 | 1.5 | 2.5 | 2.0 | 1.2 | 1.2 | 2.2 | 1.3 | 0.8 | 0.7 |
| 1.5 | 2.8 | 1.7 | 1.3 | 1.0 | 3.0 | 2.3 | 1.8 | 1.7 | 3.2 | 2.0 | 1.5 | 1.5 | 2.5 | 1.7 | 1.2 | 1.0 |
| 1.0 | 3.0 | 2.2 | 2.0 | 1.7 | 3.3 | 2.5 | 2.3 | 2.0 | 3.2 | 2.5 | 2.0 | 2.0 | 3.0 | 2.2 | 1.7 | 1.5 |
Figure 4Increased drivewire travel is needed to traverse the hypotenuse from the extractor hook to the template. For the system used in this study, the template center (D/4.0) was the calibration position.
FIRST conformance to applicable AAPM Task Group report recommendations
| Item | Recommendation description | Task Group Report (TG‐XX) | Conformance (yes/no) |
|---|---|---|---|
| 1 | 2D dosimetry formalism | TG‐43 | yes |
| 2 | 1D dosimetry formalism | TG‐43 | yes |
| 3 | input data acquisition preceding clinical implementation | TG‐43 | yes |
| 4 | method used to derive consensus dataset | TG‐43 | yes |
| 5 | source strength specification format | TG‐43 | yes |
| 6 | customer acceptance testing procedure | TG‐53 | no |
| 7 | image acquisition and integrity | TG‐53 | yes |
| 8 | ability to contour, set position points of interest, and calculate volumes | TG‐53 | yes |
| 9 | hard copy printout | TG‐53 | yes |
| 10 | visual display verification and visibility of echogenic‐tipped needles | TG‐53 | yes |
| 11 | global dose scaling with Λ, | TG‐53 | yes |
| 12 | accuracy and consistency of units | TG‐53 | yes |
| 13 | DVH and DHI calculations | TG‐53 | yes |
| 14 | ability to perform routine QA testing of input data | TG‐53 | yes |
| 15 | functioning and accuracy of peripheral devices: printer | TG‐53 | yes |
| 16 | Daily and clinical use QA procedures | TG‐53 | no |
| 17 | routine QA log of hardware and software changes/updates | TG‐53 | yes |
| 18 | computer systems electronic data management: storage & security | TG‐53 | yes |
| 19 | vendor responsibilities: training, support, and education | TG‐53 | yes |
| 20 | user responsibilities: feedback mechanisms | TG‐53 | yes |
| 21 | comparison of source strength with manufacturer calibration | TG‐56 | yes |
| 22 | temporal variations and needle/seed attenuation | TG‐56 | yes |
| 23 | comparison of preplan with clinical plan | TG‐56 | yes |
| 24 | minimization of radiation exposure | TG‐56 | yes |
| 25 | delivery cable calibration procedure | TG‐59 | yes |
| 26 | generation of reports | TG‐59 | yes |
| 27 | amendment of institution radioactive materials license | TG‐59 | yes |
| 28 | need for understanding error codes/status messages | TG‐59 | yes |
| 29 | positioning accuracy and reproducibility of ultrasound probe vs. SPOT | TG‐64 | yes |
| 30 | capability to cope with pubic arch collisions by skewing needles | TG‐64 | yes |
| 31 | intraoperative seed localization and treatment planning | TG‐64 | yes |
The system uses the same (r) and (r) data, which produce errors that increase as r decreases.
did not include TLD results, and methods to evaluate g(0), F(0,θ), and were not described. However, AAPM TG‐43U1 recommendations were not available at the time of FIRST system release.
No acceptance testing procedure form was present (see Appendix C); only an installation engineer checklist was available.
Terminology for positioning within the seedselectron build element appeared inconsistent and misleading.
The manufacturer did not provide a recommended QA procedure form. Therefore, a daily QA form is included in Appendix A. However, it is often the user's regulatory responsibility to prepare an institutional quality management program for new treatment modalities.
Needle attenuation of radiation dose distributions was not accounted for since they are removed within a relatively short timeframe. However, seeds are “permanently” implanted, and seed attenuation might be clinically significant. To our knowledge, no commercially available RTP systems account for seed attenuation.
Figure 5Radiograph of the selectSeed. For this seed, the capsule length, capsule diameter, and active length are 4.64 mm, 0.82 mm, and 3.45 mm, respectively. For the seeds examined in this investigation, the average dimensions were 4.53 mm, 0.79 mm, and 3.39 mm, respectively.
Calculations using brachytherapy dosimetry parameters for the 2D and 1D AAPM TG‐43 formalisms
|
|
| φ( | A | B | C | Ratio C/A |
|---|---|---|---|---|---|---|
| 0.3 | 1.087 | 1.051 | 11.093 | 12.078 | 12.694 | 1.144 |
| 0.5 | 1.078 | 0.959 | 4.196 | 4.312 | 4.135 | 0.985 |
| 0.7 | 1.052 | 0.938 | 2.126 | 2.147 | 2.014 | 0.947 |
| 1.0 | 1.000 | 0.933 | 1.000 | 1.000 | 0.933 | 0.933 |
| 1.5 | 0.907 | 0.932 | 0.405 | 0.403 | 0.376 | 0.927 |
| 2.0 | 0.808 | 0.936 | 0.203 | 0.202 | 0.189 | 0.929 |
| 2.5 | 0.713 | 0.939 | 0.115 | 0.114 | 0.107 | 0.932 |
| 3.0 | 0.627 | 0.941 | 0.070 | 0.070 | 0.066 | 0.933 |
| 3.5 | 0.548 | 0.942 | 0.045 | 0.045 | 0.042 | 0.934 |
| 4.0 | 0.477 | 0.943 | 0.030 | 0.030 | 0.028 | 0.935 |
| 4.5 | 0.414 | 0.944 | 0.021 | 0.020 | 0.019 | 0.936 |
| 5.0 | 0.357 | 0.945 | 0.014 | 0.014 | 0.013 | 0.936 |
| 6.0 | 0.265 | 0.946 | 0.007 | 0.007 | 0.007 | 0.937 |
| 7.0 | 0.196 | 0.948 | 0.004 | 0.004 | 0.004 | 0.939 |
| 8.0 | 0.144 | 0.950 | 0.002 | 0.002 | 0.002 | 0.941 |
| 9.0 | 0.106 | 0.951 | 0.001 | 0.001 | 0.001 | 0.942 |
| 10.0 | 0.078 | 0.951 | 0.001 | 0.001 | 0.001 | 0.942 |
| Equipment | Tolerance | Complete |
|---|---|---|
| Inspect probe, ECRM, all cables, plugs, screens, accessories for damage | No damage | |
| Assemble and connect ECRM, cradle, probe, stepper stabilizer | ||
| Power on equipment | No errors | |
| Sterile components present: | All present | |
| Template, needles, compose, deliver, drive wire, seeds, spacers, long obturator, extractor knob, emergency kit | ||
|
| ||
| Battery Check | ok | |
| Range | 10 uSv/h | |
| Fluoro ‐ Power on, Patient name, check printer | Ready |
| SPOT‐ECRM‐US communications (before patient in room) | ||
|---|---|---|
| Template Display on US screen spacing and labels | Functional and correct | |
| US transducers | No signal dead spots | |
| SPOT field of view set to capture entire US screen | Functional and correct | |
| Depth (5, 6, 8 cm) |
| |
| frame rate | 30 fps | |
| Scanning angle‐maximum | 140° | |
| Probe moves through full range and returns to home position | Smooth motion | |
| 3D cube reconstruction | Functional | |
| Prepare Probe (OR Nurse), gel and brachyballoon | ||
| Ensure acoustic contact between probe and balloon | No signal dead spots |
|
| ||
|---|---|---|
| Power‐on and self‐test seedSelectron | Self‐test okay | |
| Sterile person assemble compose, delivery drive, cartridges & compose/delivery element, extract or knob | n/a | |
| Enter certificate source information | n/a | |
| “Seed QA” ‐ follow on screen instructions | No errors | |
| Verify seed strength in well chamber |
| |
| Enter measured seed strength in seedSelectron | n/a |
Tolerances are listed only where relevant If a test fails to meet tolerances specified, action should be taken to identify the source of the discrepancy and a strategy should be developed to return the system to specified tolerances. If there is mechanical or safety test failure, no treatment should proceed until the failure is repaired and shown to be in proper working order.
| Patient Setup | Tolerance | Complete |
|---|---|---|
| Insert probe | Image ok | |
| Lock stepper‐stabilizer base | Locked | |
| Template QA | Fits in check jig | |
| Mount template onto stepper stabilizer | Locked | |
| Locker needles into prostate | Note position | |
| Doc sets base plane in transverse US and stepper | Stepper set at 0.0 | |
| Check stepper stabilizer détente | At zero position | |
| Lock down all stepper‐stabilizer knobs | Locked | |
| Confirm patient name & number | Correct |
| Contouring | Tolerance | Complete |
|---|---|---|
| Slice Spacing | 2.5mm | |
| Contour prostate (Red) | ||
| Contour Urethra (Yellow) | ||
| Determine prostate volume | ||
| Physician reviews contours |
| Planning | ||
|---|---|---|
| Live Planning | ||
| Enter source information | Certificate source strength | |
| Select Template – serial number should match mounted template | Identical | |
| Set baseplane in SPOT baseplane to be the identical to baseplane in transverse US when stepper at 0.0 | Identical | |
| Add needles – AutoLoad | ||
| Follow OR Needle Record for needle placement | ||
| Activate All |
| Review Isodoses and DVHs | Tolerance | Complete |
|---|---|---|
| Distance outside prostate contour for 100% isodose coverage | Approx. 3mm | |
| Volume of target getting 100% of prescription dose |
| |
| Volume of Urethra getting |
| |
|
| ||
| Print needle & seed configurations | ||
| Export to selectSeed |
| Treatment | Tolerance | Complete |
|---|---|---|
| Mount seedSelectron on rails | ||
| Ensure no more than 10mm between extractor flange and hook when connected to deepest needle | 10mm | |
| Secure/Lock seedSelectron to rails | Locked | |
| Calibrate baseplane delivery distance – | ||
| Confirm needle coordinates and retractions distance with Physician | Confirmed | |
| Build (indicator tolerances | Indicators | |
| Re‐Confirm needle coordinates | ||
| Deliver | ||
| Physician to survey each needle after removal from patient | Signal | |
|
| ||
| “CLOSE” cartridges after last needle | ||
| Radiation survey of physician, laundry, garbage, and room | ||
|
| ||
| Fluoro patient and count seeds | 2 counts match delivery record | |
| Fluoro seeds cartridge and count remaining seeds | Count matches expected | |
|
| ||
| Label storage containers and return to hot lab | ||
| Complete inventory control records | Accurate and complete |
These surveys represent only part of the Radiation Safety practices which adhere to local and national standards
| Tolerance | Complete | |
|---|---|---|
| Inspect probe, ECRM, all cables, plugs, screens, accessories for damage | No damage | |
| Assemble and connect ECRM, cradle, probe, stepper stabilizer, seedSelectron | ||
| Power on equipment | No errors | |
| Transducer check ‐with a little gel on finger, run finger along transverse and sagittal transducer to ensure signal, check for dead spots | No dead spots | |
| Geometric fidelity (US & SPOT systems) | Functional, 2mm or 3% | |
| ‐measurement tools available and perform as expected | ||
| ‐Distance, Area, Volume | 2mm or 3% | |
| ‐Remove test QA jig | ||
| ‐Acquire transverse and sagittal images (US) and 3‐D Scan (SPOT) | ||
| ‐Take linear distance measurements in all 3 planes on SPOT system | ||
| ‐Take linear distance measurements in Trans and Sag planes on US system | ||
| ‐Record results or print images, compare with “true” values of the jig | ||
| ‐In Spot system contour a triangle on a few slices | ||
| ‐Have the system determine the area and volume ‐ compare with true values | ||
| Image quality | Satisfactory image; no unusual artifacts | |
| In transverse and sagittal mode acquire image on US screen of one of the prostate phantoms and examine the image quality | ||
| Note anything unusual, such as streaks, artifacts, dead spots, etc. | ||
| Monitor grey levels – compare US and SPOT screens | Min. 8 levels; US & SPOT match | |
| Template ‐displayed spacing and labels (US & SPOT) | Matches physical template spacing and labels | |
| Template – Calibration Check (Complete the Template Alignment Record for SPOT and US) | Collar settings: | |
| ‐Measure distance from base of template to collars should be within tolerance of calibrated settings | ||
| ‐Mount template on stepper stabilizer stand – numbers and letters facing the ECRM | ||
| ‐Carefully place needles in the template calibration jig | ||
| ‐Fasten the template calibration jig to the stepper stabilizer stand | ||
| ‐Acquire a scan | ||
| ‐Go to Pre‐ Planning or Live Planning | ||
| ‐Place the template on an appropriate “baseplane” | ||
| ‐Compare the needle positions as seen with the transverse transducer on the US unit, displaying the appropriate template | ||
| ‐Compare the needle position as seen with the sagittal transducer on the SPOT system | Template alignment: Average for 5 calibration points | |
| ‐Compare current images with images from original calibration to determine if template calibrated okay | than | |
| ‐Print copies of alignment on US and SPOT systems for reference | ||
| ‐Complete the Template Alignment Record for SPOT and US | ||
| ECRM rotation alignment for needle navigator | Needle visible on screen/angle of ECRM correct | |
| ‐place needle at known position in template, plan needle, navigate to that position in software, determine if the needle is visible at the angle selected by the probe | ||
| Dosimetry constancy | Identical to commissioning values | |
| ‐“standard” plan in pre, live and post‐planning gives isodoses, DVHs, point dose values identical to values obtained at commissioning | ||
|
| ||
| Communication between SPOT and seedSelectron | Correct; hard and soft copy agree | |
| ‐plan information transfer fidelity (number of seeds, needles, locations) | ||
| ‐hard copy and soft copy consistency | ||
| seedSelectron quarterly QA (complete seedSelectron Quarterly QA log sheet) | ||
| seed placement accuracy: the first seed in the train should be within tolerance of expected position |
| |
| Build sequences: identical to the planned sequences | Identical | |
| Build indicators: functional and accurately reflect the planned and built sequence | Functional and accurate | |
| seedSelectron measured seed activity: within tolerance of the activity measured in the well chamber |
| |
| Indicators: functional and accurate | Functional and accurate | |
| Safety, interrupt and dispose features | Functional |
If a test fails to meet tolerances specified, action should be taken to identify the source of the discrepancy and a strategy should be developed to return the system to specified tolerances. If there is mechanical or safety test failure, no treatment should proceed until the failure is repaired and shown to be in proper working order.
more detailed tests follow the ABR and AAPM TG‐1 recommendations for ultrasound QA
TG‐53 provides detailed guidelines on planning system testing
| Alignment in SPOT as measured from centre of dot to centre of first echo Template ID | ||||
|---|---|---|---|---|
| Position | Distance on image (mm) | Direction | Scaling factor | Distance scaled to actual (mm) |
| A 2 | ||||
| M 2 | ||||
| G 7 | ||||
| A 11 | ||||
| M 11 | ||||
| Average STDEV | ||||
| Alignment in US as measured from centre of dot to centre of first echo Template ID | ||||
|---|---|---|---|---|
| Position | Distance on image (mm) | Direction | Scaling factor | Distance scaled to actual (mm) |
| A 2 | ||||
| M 2 | ||||
| G 7 | ||||
| A 11 | ||||
| M 11 | ||||
| Average STDEV | ||||
| Delivery Order | Needle | Retraction (mm) | Pattern | Delivery Method | Feature Tested |
|---|---|---|---|---|---|
| 1 | D3.0 | 0 | OxOxOxOxOxOxOxO | Auto | build sequence accuracy, ctivity accuracy, diode detector operation, seed placement |
| 2 | C3.0 | 0 | OOxOxOxOxOxOxOxO | Auto | build sequence accuracy, activity accuracy, diode detector operation, seed placement |
| 3 | A3.0 | 35 | Ox | Auto | build sequence accuracy, activity accuracy, diode detector operation, retraction distance, extra spacer, seed placement |
| 4 | G3.0 | 50 | Ox | Auto | build sequence accuracy, activity accuracy, diode detector operation, retraction distance, extra spacer, seed placement |
| 5 | E3.0 | 5 | OOxOOOxOOx | Manual | build sequence accuracy, activity accuracy, diode detector operation, double and triple seed compensation, retraction distance, extra spacer, seed placement, |
| 6 | F3.0 | 0 | OxOxOx | Manual; Interrupt | error detection in build, software & hardware, extra spacer feature, interrupt button |
| Test | Diode Position | |||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 0 | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 | 13 | 14 | 15 | |
| Planned pattern | ||||||||||||||||
| Select seed activity | ||||||||||||||||
| SS indicator corresponds to SS measured activity | ||||||||||||||||
| Well chamber activity | ||||||||||||||||
| Built as planned? If not, diagram | ||||||||||||||||
| First seed position (mm) | ||||||||||||||||