| Literature DB >> 29500857 |
Masahide Saito1, Naoki Sano1, Yuki Shibata1, Kengo Kuriyama1, Takafumi Komiyama1, Kan Marino1, Shinichi Aoki1, Kazunari Ashizawa1, Kazuya Yoshizawa1, Hiroshi Onishi1.
Abstract
The purpose of this study was to compare the MLC error sensitivity of various measurement devices for VMAT pre-treatment quality assurance (QA). This study used four QA devices (Scandidos Delta4, PTW 2D-array, iRT systems IQM, and PTW Farmer chamber). Nine retrospective VMAT plans were used and nine MLC error plans were generated for all nine original VMAT plans. The IQM and Farmer chamber were evaluated using the cumulative signal difference between the baseline and error-induced measurements. In addition, to investigate the sensitivity of the Delta4 device and the 2D-array, global gamma analysis (1%/1, 2%/2, and 3%/3 mm), dose difference (1%, 2%, and 3%) were used between the baseline and error-induced measurements. Some deviations of the MLC error sensitivity for the evaluation metrics and MLC error ranges were observed. For the two ionization devices, the sensitivity of the IQM was significantly better than that of the Farmer chamber (P < 0.01) while both devices had good linearly correlation between the cumulative signal difference and the magnitude of MLC errors. The pass rates decreased as the magnitude of the MLC error increased for both Delta4 and 2D-array. However, the small MLC error for small aperture sizes, such as for lung SBRT, could not be detected using the loosest gamma criteria (3%/3 mm). Our results indicate that DD could be more useful than gamma analysis for daily MLC QA, and that a large-area ionization chamber has a greater advantage for detecting systematic MLC error because of the large sensitive volume, while the other devices could not detect this error for some cases with a small range of MLC error.Entities:
Keywords: MLC error; VMAT; patient QA; radiotherapy
Mesh:
Year: 2018 PMID: 29500857 PMCID: PMC5978943 DOI: 10.1002/acm2.12288
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Phantom setup for each device ((a)–(d)). A Farmer chamber was used by inserting to the solid phantom (a). The IQM was used by mounting on the linac gantry head (b). The 2D‐array (c) was used by inserting to an octagon‐shaped phantom (Octavius Phantom) with a central cavity.
Figure 2The MLC error patterns (a) and typical apertures of the original plan for each treatment site (head and neck (b), lung‐SBRT (c), and prostate (d)).
Figure 3Cumulative signal differences between the baseline and systematic open/closed MLC error plans for each treatment site. The results of IQM (a) and Farmer chamber (b) are represented. The results for the IQM and Farmer chamber had a good linear correlation (R >0.99) between the cumulative signal difference and the magnitude of the MLC errors.
Signal differences (S ) between the baseline and nine MLC error plans for the Farmer chamber and IQM. The average values and standard deviations for all nine plans are represented. The P‐values are calculated by the Wilcoxon test
| Error type | Magnitude of MLC error (mm) | Diff (%, mean ± SD) | ||
|---|---|---|---|---|
| Farmer chamber | IQM |
| ||
| Random |
| 0.31 ± 0.99 | 0.14 ± 0.18 | 0.250 |
| Bank A open | 1.00 | 1.58 ± 1.03 | 3.16 ± 0.86 | 0.008 |
| Bank A close | 1.00 | −0.98 ± 1.86 | −2.81 ± 0.90 | 0.020 |
| Systematic open/close | −0.75 | −1.78 ± 2.05 | −4.67 ± 1.19 | 0.008 |
| −0.50 | −1.14 ± 1.32 | −3.09 ± 0.77 | 0.008 | |
| −0.25 | −0.55 ± 0.72 | −1.50 ± 0.37 | 0.008 | |
| 0.25 | 0.63 ± 0.66 | 1.65 ± 0.43 | 0.004 | |
| 0.50 | 1.30 ± 1.33 | 3.30 ± 0.90 | 0.004 | |
| 0.75 | 1.96 ± 1.98 | 4.94 ± 1.37 | 0.004 | |
Figure 4The gamma pass rates between the baseline and MLC error plans for the Delta4 device and the 2D‐array using the three criteria (1%/1, 2%/2, and 3%/3 mm). The results of Delta4 ((a)–(c)) and 2D‐array ((d)–(f)) are represented.
Gamma pass rate and dose difference between a baseline and the nine error plans for a 2D‐array and Delta4 for all nine plans. Three criteria of gamma pass rate (1%/1, 2%/2, and 3%/3 mm) are represented. The P‐values are calculated by the Wilcoxon test
| Error type | Magnitude of MLC error (mm) | Pass rate (%, mean ± SD) | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| 1%/1 mm | 2%/2 mm | 3%/3 mm | ||||||||
| Delta4 | 2D‐array |
| Delta4 | 2D‐array |
| Delta4 | 2D‐array |
| ||
| Random | σ = 0.50 | 95.8 ± 9.1 | 97.3 ± 3.3 | 1.000 | 98.5 ± 4.2 | 99.8 ± 0.6 | 1.000 | 99.5 ± 1.5 | 100.0 ± 0.1 | 1.000 |
| Bank A open | 1.00 | 68.1 ± 17.7 | 79.0 ± 16.6 | 0.203 | 91.4 ± 11.0 | 96.6 ± 3.9 | 0.297 | 97.5 ± 5.6 | 99.1 ± 1.4 | 0.438 |
| Bank A close | 1.00 | 81.8 ± 16.4 | 80.0 ± 15.0 | 0.820 | 98.7 ± 1.7 | 97.0 ± 4.2 | 0.578 | 99.8 ± 0.3 | 99.1 ± 1.6 | 0.438 |
| Systematic open/close | −0.75 | 54.3 ± 33.1 | 64.9 ± 27.3 | 0.652 | 89.0 ± 15.4 | 96.8 ± 6.2 | 0.156 | 97.7 ± 3.4 | 99.6 ± 1.0 | 0.250 |
| −0.50 | 79.1 ± 24.6 | 89.6 ± 15.3 | 0.301 | 97.9 ± 3.0 | 99.6 ± 1.0 | 0.250 | 99.9 ± 0.1 | 100 ± 0.0 | 0.500 | |
| −0.25 | 98.0 ± 2.8 | 99.7 ± 0.8 | 0.250 | 100 ± 0.0 | 100 ± 0.0 | 1.000 | 100 ± 0.0 | 100 ± 0.0 | 1.000 | |
| 0.25 | 99.4 ± 8.0 | 89.5 ± 12.5 | 0.438 | 99.7 ± 0.5 | 99.0 ± 1.4 | 0.313 | 100 ± 0.1 | 100 ± 0.0 | 1.000 | |
| 0.50 | 75.4 ± 28.8 | 62.5 ± 27.5 | 0.301 | 94.6 ± 7.7 | 91.8 ± 10.9 | 0.563 | 99.0 ± 1.6 | 98.3 ± 2.7 | 0.438 | |
| 0.75 | 48.8 ± 30.0 | 28.3 ± 20.7 | 0.203 | 85.2 ± 19.4 | 85.3 ± 16.3 | 0.844 | 94.9 ± 7.3 | 92.2 ± 10.8 | 0.469 | |
Dose difference between a baseline and the nine error plans for a 2D‐array and Delta4 for all nine plans. Three criteria of dose difference (1%, 2%, and 3%) are represented. The P‐values are calculated by the Wilcoxon test
| Error type | Magnitude of MLC error (mm) | Dose difference (%, mean ± SD) | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| 1% | 2% | 3% | ||||||||
| Delta4 | 2D‐array |
| Delta4 | 2D‐array |
| Delta4 | 2D‐array |
| ||
| Random |
| 93.5 ± 11.0 | 98.3 ± 1.9 | 0.219 | 98.2 ± 4.9 | 99.9 ± 0.2 | 0.625 | 99.4 ± 1.7 | 100.0 ± 0.1 | 1.000 |
| Bank A open | 1.00 | 44.9 ± 20.4 | 85.8 ± 7.6 | 0.004 | 76.2 ± 14.0 | 94.5 ± 5.1 | 0.004 | 90.4 ± 8.4 | 97.8 ± 2.8 | 0.020 |
| Bank A close | 1.00 | 51.8 ± 16.4 | 87.7 ± 7.7 | 0.004 | 83.6 ± 8.7 | 96.1 ± 3.9 | 0.008 | 93.2 ± 4.9 | 98.6 ± 1.9 | 0.020 |
| Systematic open/close | −0.75 | 19.6 ± 20.1 | 30.5 ± 23.7 | 0.734 | 55.2 ± 26.5 | 73.5 ± 20.7 | 0.074 | 78.2 ± 19.0 | 90.2 ± 16.0 | 0.203 |
| −0.50 | 39.7 ± 26.5 | 53.1 ± 30.3 | 0.496 | 78.4 ± 17.9 | 91.9 ± 12.2 | 0.098 | 92.2 ± 9.1 | 98.5 ± 3.0 | 0.129 | |
| −0.25 | 78.4 ± 16.7 | 97.3 ± 4.1 | 0.055 | 96.6 ± 4.5 | 100 ± 0.0 | 0.031 | 99.7 ± 0.9 | 100 ± 0.0 | 1.000 | |
| 0.25 | 78.6 ± 18.1 | 73.1 ± 23.5 | 0.820 | 97.9 ± 3.1 | 97.2 ± 4.0 | 0.887 | 99.5 ± 1.0 | 99.9 ± 0.3 | 0.500 | |
| 0.50 | 42.7 ± 24.9 | 35.2 ± 21.4 | 0.496 | 78.3 ± 19.0 | 77.2 ± 22.2 | 1.000 | 92.8 ± 9.0 | 92.7 ± 11.1 | 1.000 | |
| 0.75 | 19.7 ± 17.8 | 12.5 ± 12.6 | 0.359 | 57.2 ± 25.7 | 57.6 ± 23.7 | 1.000 | 79.2 ± 18.2 | 79.7 ± 21.2 | 1.000 | |