Replacing pretreatment verification with in vivo EPID dosimetry for prostate IMRT.

Leah N McDermott1, Markus Wendling, Jan-Jakob Sonke, Marcel van Herk, Ben J Mijnheer.   

Abstract

PURPOSE: To investigate the feasibility of replacing pretreatment verification with in vivo electronic portal imaging device (EPID) dosimetry for prostate intensity-modulated radiotherapy (IMRT). METHODS AND MATERIALS: Dose distributions were reconstructed from EPID images, inside a phantom (pretreatment) or the patient (five fractions in vivo) for 75 IMRT prostate plans. Planned and EPID dose values were compared at the isocenter and in two dimensions using the gamma index (3%/3 mm). The number of measured in vivo fractions required to achieve similar levels of agreement with the plan as pretreatment verification was determined. The time required to perform both methods was compared.
RESULTS: Planned and EPID isocenter dose values agreed, on average, within +/-1% (1 SD) of the total plan for both pretreatment and in vivo verification. For two-dimensional field-by-field verification, an alert was raised for 10 pretreatment checks with clear but clinically irrelevant discrepancies. Multiple in vivo fractions were combined by assessing gamma images consisting of median, minimum and low (intermediate) pixel values of one to five fractions. The "low" gamma values of three fractions rendered similar results as pretreatment verification. Additional time for verification was approximately 2.5 h per plan for pretreatment verification, and 15 min +/- 10 min/fraction using in vivo dosimetry.
CONCLUSIONS: In vivo EPID dosimetry is a viable alternative to pretreatment verification for prostate IMRT. For our patients, combining information from three fractions in vivo is the best way to distinguish systematic errors from non-clinically relevant discrepancies, save hours of quality assurance time per patient plan, and enable verification of the actual patient treatment.

Entities:  

Mesh:

Year:  2007        PMID: 17394951     DOI: 10.1016/j.ijrobp.2006.11.047

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  20 in total

1.  Patient setup verification procedure for a portal image in a computed radiography system with a high-resolution liquid-crystal display monitor.

Authors:  Hideki Fujita; Michihiro Yamaguchi; Yuichi Bessho; Tomio Fujioka; Haruyuki Fukuda; Kenya Murase
Journal:  Radiol Phys Technol       Date:  2009-11-20

2.  A generalized calibration procedure for in vivo transit dosimetry using siemens electronic portal imaging devices.

Authors:  Andrea Fidanzio; Francesca Greco; Laura Gargiulo; Savino Cilla; Domenico Sabatino; Massimo Cappiello; Cinzia Di Felice; Elisabetta Di Castro; Luigi Azario; Mariateresa Russo; Luciano Pompei; Guido D'Onofrio; Angelo Piermattei
Journal:  Med Biol Eng Comput       Date:  2010-11-04       Impact factor: 2.602

3.  A novel approach to SBRT patient quality assurance using EPID-based real-time transit dosimetry : A step to QA with in vivo EPID dosimetry.

Authors:  Christos Moustakis; Fatemeh Ebrahimi Tazehmahalleh; Khaled Elsayad; Francis Fezeu; Sergiu Scobioala
Journal:  Strahlenther Onkol       Date:  2020-01-10       Impact factor: 3.621

4.  Transit dosimetry in dynamic IMRT with an a-Si EPID.

Authors:  Mahsheed Sabet; Pejman Rowshanfarzad; Fred W Menk; Peter B Greer
Journal:  Med Biol Eng Comput       Date:  2014-05-31       Impact factor: 2.602

5.  Accurate IMRT fluence verification for prostate cancer patients using 'in-vivo' measured EPID images and in-room acquired kilovoltage cone-beam CT scans.

Authors:  Ali Sam Ali; Maarten L P Dirkx; Ruud M Cools; Ben J M Heijmen
Journal:  Radiat Oncol       Date:  2013-09-10       Impact factor: 3.481

6.  Assessment of a 2D electronic portal imaging devices-based dosimetry algorithm for pretreatment and in-vivo midplane dose verification.

Authors:  Ali Jomehzadeh; Parvaneh Shokrani; Mohammad Mohammadi; Alireza Amouheidari
Journal:  Adv Biomed Res       Date:  2016-11-28

7.  A Simple Method for 2-D In Vivo Dosimetry by Portal Imaging.

Authors:  Stefano Peca; Derek Wilson Brown; Wendy Lani Smith
Journal:  Technol Cancer Res Treat       Date:  2017-06-06

8.  Feasibility of a remote, automated daily delivery verification of volumetric-modulated arc therapy treatments using a commercial record and verify system.

Authors:  Jonas D Fontenot
Journal:  J Appl Clin Med Phys       Date:  2012-03-08       Impact factor: 2.102

9.  Feasibility of using two-dimensional array dosimeter for in vivo dose reconstruction via transit dosimetry.

Authors:  Heeteak Chung; Jonathan Li; Sanjiv Samant
Journal:  J Appl Clin Med Phys       Date:  2011-04-08       Impact factor: 2.102

10.  Investigation of a real-time EPID-based patient dose monitoring safety system using site-specific control limits.

Authors:  Todsaporn Fuangrod; Peter B Greer; Henry C Woodruff; John Simpson; Shashank Bhatia; Benjamin Zwan; Timothy A vanBeek; Boyd M C McCurdy; Richard H Middleton
Journal:  Radiat Oncol       Date:  2016-08-12       Impact factor: 3.481

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.