| Literature DB >> 29432459 |
Tomohiro Kaneta1, Hiromitsu Daisaki2, Matsuyoshi Ogawa1, En-Tao Liu1, Hitoshi Iizuka1, Tetsu Arisawa1, Ayako Hino-Shishikura1, Keisuke Yoshida1, Tomio Inoue1.
Abstract
Standardized uptake values (SUVs) are the most widely used quantitative imaging biomarkers in PET. It is important to evaluate the variability and repeatability of measured SUVs. Phantom studies seem to be essential for this purpose; however, repetitive phantom scanning is not recommended due to the decay of radioactivity. In this study, we performed count-based image reconstruction to avoid the influence of decay using two different PET/CT scanners. By adjusting the ratio of 18F-fluorodeoxyglucose solution to tap water, a NEMA IEC body phantom was set for SUVs of 4.0 inside six hot spheres. The PET data were obtained using two scanners (Aquiduo and Celesteion; Toshiba Medical Systems, Tochigi, Japan). We set the start time for image reconstruction when the total radioactivity in the phantom was 2.53 kBq/cc, and employed the counts of the first 2-min acquisition as the standard. To maintain the number of counts for each image, we set the acquisition time for image reconstruction depending on the decay of radioactivity. We obtained 50 images, and calculated the SUVmax and SUVpeak of all six spheres in each image. The average values of the SUVmax were used to calculate the recovery coefficients to compare those measured by the two different scanners. Bland-Altman analyses of the SUVs measured by the two scanners were also performed. The measured SUVs using the two scanners exhibited a 10-30% difference, and the standard deviation (SD) of the measured SUVs was between 0.1-0.2. The Celesteion always exhibited higher values than the Aquiduo. The smaller sphere exhibited a larger SD, and the SUVpeak had a smaller SD than the SUVmax. The Bland-Altman analyses showed poor agreement between the SUVs measured by the two scanners. The recovery coefficient curves obtained from the two scanners were considerably different. The Celesteion exhibited higher recovery coefficients than the Aquiduo, especially at approximately 20-mm-diameter. Additionally, the curves were lower than those calculated from the standard 30-min acquisition images. We propound count-based image reconstruction to evaluate the variability and repeatability of measured SUVs. These results are also applicable for the standardization and harmonization of SUVs in multi-institutional studies.Entities:
Mesh:
Substances:
Year: 2018 PMID: 29432459 PMCID: PMC5809066 DOI: 10.1371/journal.pone.0192549
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Time calculations for image reconstruction.
| Reconstruction No. | Time (min:sec) | Duration (min:sec) | Decay factor |
|---|---|---|---|
| 1 | (A) | 2:00 | 1 |
| 2 | (A) + 2:00 | 2:02 | 0.987 |
| 3 | (A) + 4:02 | 2:03 | 0.975 |
| 4 | (A) + 6:05 | 2:05 | 0.962 |
| 5 | (A) + 8:10 | 2:06 | 0.950 |
| 6 | (A) + 10:16 | 2:08 | 0.937 |
At the time (A), the total radioactivity in the phantom corresponds to 2.53 kBq/cc
Decay factor =
Duration = 2 min / Decay factor, e.g., Duration for reconstruction no. 2 = 2 min / 0.987
Fig 1The measured SUVmax and SUVpeak of a 37-mm-diameter sphere with a 2-min acquisition over time.
Standardized uptake values (SUVs) obtained from constant 2-min acquisition periods using the Celesteion scanner are shown over time to compare the resulting SUVs. The scatter plot depicts SUVmax (solid, black circle) and SUVpeak (empty/open circle) values. The solid and dotted trend lines correspond to SUVmax and SUVpeak, respectively.
Fig 2The distributions of count-based measured SUVmax and SUVpeak.
The distribution of the SUVmax (A-C) and SUVpeak (D-F) of the representative three spheres. The 50 count-based measurements were performed using the Celesteion (solid, black circle) and Aquiduo (empty/open circle) scanners. The solid and dotted trend lines correspond to the Celesteion and Aquiduo, respectively.
Mean and standard deviation (SD) of measured SUVs.
| Diameter of the sphere | 37mm | 22mm | 13mm |
|---|---|---|---|
| SUVmax Mean | 5.077 | 4.707 | 3.150 |
| SD | 0.146 | 0.189 | 0.194 |
| SUVpeak Mean | 4.657 | 4.194 | 2.523 |
| SD | 0.112 | 0.136 | 0.125 |
| SUVmax Mean | 4.534 (89.3%) | 3.848 (81.8%) | 2.291 (72.7%) |
| SD | 0.110 | 0.135 | 0.132 |
| SUVpeak Mean | 4.303 (92.3%) | 3.560 (84.9%) | 1.994 (79.0%) |
| SD | 0.095 | 0.109 | 0.090 |
The percentage in parentheses represents the ratio to the SUVs measured by Celesteion.
Fig 3The Bland-Altman plots of count-based measured SUVmax and SUVpeak.
The Bland-Altman plots of SUVmax (A-C) and SUVpeak (D-F) measured by the Celesteion and Aquiduo.
The results of Bland-Altman analyses of the SUVs measured by the Celesteion and Aquiduo scanners.
| Diameter of the sphere (mm) | 37 | 28 | 22 | 17 | 13 | 10 |
|---|---|---|---|---|---|---|
| SUVmax | ||||||
| Difference Mean | 0.543 | 0.587 | 0.859 | 0.969 | 0.859 | 0.590 |
| SD | 0.177 | 0.196 | 0.237 | 0.253 | 0.251 | 0.215 |
| 95% CI | 0.049 | 0.054 | 0.066 | 0.070 | 0.070 | 0.060 |
| SUVpeak | ||||||
| Difference Mean | 0.354 | 0.391 | 0.634 | 0.734 | 0.530 | 0.315 |
| SD | 0.132 | 0.167 | 0.194 | 0.164 | 0.163 | 0.130 |
| 95% CI | 0.037 | 0.046 | 0.054 | 0.045 | 0.045 | 0.036 |
SD = standard deviation, CI = confidence interval
Fig 4The curves of the recovery coefficients.
The curves of the recovery coefficients measured by the Celesteion (a) and Aquiduo (b) scanners. The solid line with black circles corresponds to the average of the count-based measurements. The dashed line with black squares corresponds to the standard 30-min acquisition measurements.