| Literature DB >> 29411524 |
Marzieh Anjomrouz1,2, Muhammad Shamshad1,2, Raj K Panta1,2, Lieza Vanden Broeke3, Nanette Schleich4, Ali Atharifard1,2, Raja Aamir1,2, Srinidhi Bheesette1,5, Michael F Walsh2, Brian P Goulter2, Stephen T Bell2,5, Christopher J Bateman1,2,3, Anthony P H Butler1,2,3,5, Philip H Butler2,3,5.
Abstract
In this paper, we present a method that uses a combination of experimental and modeled data to assess properties of x-ray beam measured using a small-animal spectral scanner. The spatial properties of the beam profile are characterized by beam profile shape, the angular offset along the rotational axis, and the photon count difference between experimental and modeled data at the central beam axis. Temporal stability of the beam profile is assessed by measuring intra- and interscan count variations. The beam profile assessment method was evaluated on several spectral CT scanners equipped with Medipix3RX-based detectors. On a well-calibrated spectral CT scanner, we measured an integral count error of 0.5%, intrascan count variation of 0.1%, and an interscan count variation of less than 1%. The angular offset of the beam center ranged from 0.8° to 1.6° for the studied spectral CT scanners. We also demonstrate the capability of this method to identify poor performance of the system through analyzing the deviation of the experimental beam profile from the model. This technique can, therefore, aid in monitoring the system performance to obtain a robust spectral CT; providing the reliable quantitative images. Furthermore, the accurate offset parameters of a spectral scanner provided by this method allow us to incorporate a more realistic form of the photon distribution in the polychromatic-based image reconstruction models. Both improvements of the reliability of the system and accuracy of the volume reconstruction result in a better discrimination and quantification of the imaged materials.Entities:
Keywords: Medipix detector; beam profile; geometric calibration; spectral CT; x-ray source model
Mesh:
Year: 2018 PMID: 29411524 PMCID: PMC5849855 DOI: 10.1002/acm2.12260
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1An overview of beam profile assessment method.
Figure 2A schematic diagram of the components of MARS spectral CT with a single‐chip camera alongside the frame sequence acquired at each camera position (CPOS). (a) S shows the position of x‐ray source with respect to the detector plane that both rotate simultaneously around the object volume. The single‐chip camera is translated along the vertical axis. In this diagram, photon distribution along a solid angle in both θ and φ directions are also demonstrated. (b) A series of flat‐field images were acquired at each CPOS using a CZT‐Medipix3RX in a typical scan.
Figure 3Normalized integrated count distribution of the source model. The magnified region identifies the portion of the beam targeted for the experiment.
Figure 4A series of flat‐field images in a single camera position taken by a CdTe‐Medipix3RX detector. The red dividers indicate how row pixels are categorized into several groups to increase the resolution of the beam profile along θ. These count groups are averaged across all frames within each time interval, labeled T1, T2, and T3.
Figure 5Classified counts against time and position. The measured counts classified against time and position to produce a beam profile from a scan with five camera positions (CPOS). In a MARS scanner with the single‐chip camera, the dataset of each camera position is collected sequentially. The flat‐field dataset at each camera position is divided by three, representing three time intervals (T1 − T3). This classification is shown in the graph by three horizontal dotted lines at each camera position. Each temporal beam profile is constructed by stitching data of the respective time interval from each camera position. Each dot along the horizontal lines represents the pixel classification based on the position that is calculated in the same way as illustrated by θ1 − θ5 in Fig. 3.
Figure 6An example of quantitative assessment of the beam profile. The measured and molded beam profiles have the same concavity and small difference of the latus rectums. The arrows show how the measured beam profile parabolas deviate from the model in both properties of angular offset and intrascan variations.
Figure 7(a) The comparison of the measured counts and fitted curve plotted against the modeled beam profile. (b) Histogram of the measured to the expected noise ratio following the Poisson distribution.
Figure 8(a) Comparing the modeled beam profile shape with MARS experiments before and after angular offset adjustments. (b) Variation of the temporal beam profiles in three time intervals plotted against the modeled beam profile.
Figure 9Comparison of the integrated counts between the MARS source model and three different experimental beam profiles.
Figure 10A beam profile with a minor defect due to intrascan variation at the end of the scan (i.e., most positive θ value). Although this dataset has relatively large angular offset, the variation in the angular offsets between all temporal beam profiles is within the acceptable range.
Figure 11A beam profile with a major defect behaving chaotically in different time intervals. Both beam properties of intrascan variation and angular offset show large variation.
Summary of beam profile assessment results
| Properties | Status | ||
|---|---|---|---|
| Stable | Minor defect | Major defect | |
| Concavity | Concave down | Concave down | Concave down |
| Average latus rectum | 2.26°± 0.07° | 1.02°± 0.12° | 3.6°± 3.57° |
| Latus rectum diff. from model | 0.03° | 1.5° | 1.2° |
| Angular offset along (ξθ) | 0.8 ± 0.07° | 1.6 ± 0.14° | 3.6 ± 1.6° |
| Intrascan count variation | 0.1% | 0.6% | 1.4% |
| Integral count diff. at the beam center | 0.3–0.5% | 25% | 30% |
| Interscan count variation | 0.3% | – | – |