Literature DB >> 27795606

Evaluation of a photon counting Medipix3RX CZT spectral x-ray detector.

Steven M Jorgensen1, Andrew J Vercnocke1, David S Rundle2, Philip H Butler3, Cynthia H McCollough4, Erik L Ritman1.   

Abstract

We assessed the performance of a cadmium zinc telluride (CZT)-based Medipix3RX x-ray detector as a candidate for micro-computed tomography (micro-CT) imaging. This technology was developed at CERN for the Large Hadron Collider. It features an array of 128 by 128, 110 micrometer square pixels, each with eight simultaneous threshold counters, five of which utilize real-time charge summing, significantly reducing the charge sharing between contiguous pixels. Pixel response curves were created by imaging a range of x-ray intensities by varying x-ray tube current and by varying the exposure time with fixed x-ray current. Photon energy-related assessments were made by flooding the detector with the tin foil filtered emission of an I-125 radioisotope brachytherapy seed and sweeping the energy threshold of each of the four charge-summed counters of each pixel in 1 keV steps. Long term stability assessments were made by repeating exposures over the course of one hour. The high properly-functioning pixel yield (99%), long term stability (linear regression of whole-chip response over one hour of acquisitions: y = -0.0038x + 2284; standard deviation: 3.7 counts) and energy resolution (2.5 keV FWHM (single pixel), 3.7 keV FWHM across the full image) make this device suitable for spectral micro-CT. The charge summing performance effectively reduced the measurement corruption caused by charge sharing which, when unaccounted for, shifts the photon energy assignment to lower energies, degrading both count and energy accuracy. Effective charge summing greatly improves the potential for calibrated, energy-specific material decomposition and K edge difference imaging approaches.

Entities:  

Keywords:  CZT-based spectral micro-CT; Medipix; charge sharing; photon counting; x-ray microtomography

Year:  2016        PMID: 27795606      PMCID: PMC5080655          DOI: 10.1117/12.2236501

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  1 in total

1.  Performance of a Medipix3RX spectroscopic pixel detector with a high resistivity gallium arsenide sensor.

Authors:  Elias Hamann; Thomas Koenig; Marcus Zuber; Angelica Cecilia; Anton Tyazhev; Oleg Tolbanov; Simon Procz; Alex Fauler; Tilo Baumbach; Michael Fiederle
Journal:  IEEE Trans Med Imaging       Date:  2014-04-14       Impact factor: 10.048

  1 in total
  2 in total

1.  Evaluation of a photon counting Medipix3RX cadmium zinc telluride spectral x-ray detector.

Authors:  Jeffrey F Marsh; Steven M Jorgensen; David S Rundle; Andrew J Vercnocke; Shuai Leng; Philip H Butler; Cynthia H McCollough; Erik L Ritman
Journal:  J Med Imaging (Bellingham)       Date:  2018-12-04

2.  Beam profile assessment in spectral CT scanners.

Authors:  Marzieh Anjomrouz; Muhammad Shamshad; Raj K Panta; Lieza Vanden Broeke; Nanette Schleich; Ali Atharifard; Raja Aamir; Srinidhi Bheesette; Michael F Walsh; Brian P Goulter; Stephen T Bell; Christopher J Bateman; Anthony P H Butler; Philip H Butler
Journal:  J Appl Clin Med Phys       Date:  2018-02-07       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.