Dosimetric quality assurance (QA) of the new Elekta Unity (MR-linac) will differ from the QA performed of a conventional linac due to the constant magnetic field, which creates an electron return effect (ERE). In this work we aim to validate PRESAGE® dosimetry in a transverse magnetic field, and assess its use to validate the research version of the Monaco TPS of the MR-linac. Cylindrical samples of PRESAGE® 3D dosimeter separated by an air gap were irradiated with a cobalt-60 unit, while placed between the poles of an electromagnet at 0.5 T and 1.5 T. This set-up was simulated in EGSnrc/Cavity Monte Carlo (MC) code and relative dose distributions were compared with measurements using 1D and 2D gamma criteria of 3% and 1.5 mm. The irradiation conditions were adapted for the MR-linac and compared with Monaco TPS simulations. Measured and EGSnrc/Cavity simulated profiles showed good agreement with a gamma passing rate of 99.9% for 0.5 T and 99.8% for 1.5 T. Measurements on the MR-linac also compared well with Monaco TPS simulations, with a gamma passing rate of 98.4% at 1.5 T. Results demonstrated that PRESAGE® can accurately measure dose and detect the ERE, encouraging its use as a QA tool to validate the Monaco TPS of the MR-linac for clinically relevant dose distributions at tissue-air boundaries.
Dosimetric quality assurance (QA) of the new Elekta Unity (MR-linac) will differ from the QA performed of a conventional linac due to the constant magnetic field, which creates an electron return effect (ERE). In this work we aim to validate PRESAGE® dosimetry in a transverse magnetic field, and assess its use to validate the research version of the Monaco TPS of the MR-linac. Cylindrical samples of PRESAGE® 3D dosimeter separated by an air gap were irradiated with a cobalt-60 unit, while placed between the poles of an electromagnet at 0.5 T and 1.5 T. This set-up was simulated in EGSnrc/Cavity Monte Carlo (MC) code and relative dose distributions were compared with measurements using 1D and 2D gamma criteria of 3% and 1.5 mm. The irradiation conditions were adapted for the MR-linac and compared with Monaco TPS simulations. Measured and EGSnrc/Cavity simulated profiles showed good agreement with a gamma passing rate of 99.9% for 0.5 T and 99.8% for 1.5 T. Measurements on the MR-linac also compared well with Monaco TPS simulations, with a gamma passing rate of 98.4% at 1.5 T. Results demonstrated that PRESAGE® can accurately measure dose and detect the ERE, encouraging its use as a QA tool to validate the Monaco TPS of the MR-linac for clinically relevant dose distributions at tissue-air boundaries.
The integration of magnetic resonance imaging (MRI) into a radiotherapy treatment
platform provides images with high soft-tissue contrast that can be used to perform
accurate image-guidance during treatment delivery, giving confidence to apply radiation
with tight planning target volumes margins, and thus potencially reducing the toxicity
risks and leading to higher tumour control. There are currently five separate designs
for integrating an MRI scanner and a radiotherapy treatment device (Fallone et
al
2009, Raaymakers et
al
2009, Keall et al
2014, Mutic and Dempsey 2014, Mutic et al
2016). These designs differ in the
method of radiation delivery (linear accelerator (linac) or cobalt teletherapy), the
orientation of the main (B0) magnetic field (parallel or perpendicular to
the radiation beam) and the magnetic field strength (0.35 T to 1.5 T). One of seven
MR-linac prototypes (Elekta Unity, Elekta AB, Stockholm, Sweden), consisting of a
modified 1.5 T MRI system (Philips, Best, The Netherlands) and a 7 MV linac (Elekta AB,
Stockholm, Sweden) that rotates around a ring outside of the magnet (Lagendijk
et al
2008, Raaymakers et
al
2009), has been installed at our
institute as part of the Elekta MR-linac Consortium (Kerkmeijer et al
2016).Quality assurance (QA) performed in this MR-linac will differ from that required for a
conventional linac due to the presence of the imaging B0 field, which is
perpendicular to the beam direction. The Lorentz force produced by this constant
magnetic field deflects the paths of moving electrons, thus redistributing the absorbed
dose, with some of these secondary electrons that would normally exit the tissue being
directed back into it, leading to an electron return effect (ERE). The impact of the
Lorentz force on the secondary electrons creates a reduced build-up distance and a
strong dose increase at the proximal side of an air cavity and a reduction at its distal
side (Raaijmakers et al
2004, 2005, Raaijmakers et al
2007a, 2007b).Being able to have high resolution 3D dosimetry is of great interest for the
increasingly sophisticated modern radiotherapy treatments, which often include high
spatial dose modulation. Evidence is steadily accumulating to show the improved survival
benefit of tight compliance radiation oncology delivery protocols of the actual delivery
doses (Weber et al
2012, Dyk et al
2013).The PRESAGE® 3D dosimeter consists of a radiochromic plastic that shows an
optical density change when irradiated and can thus be read out using an optical-CT
scanner (Guo et al
2006). Contrary to other 3D
dosimeters such as Fricke gels and polymer gels (Schreiner 2015), PRESAGE® does not suffer from
diffusion, or need an external container, making it a promising candidate to detect dose
values at dosimeter-air interfaces. Previous work using large PRESAGE®
samples in an MR-guided radiotherapy context has shown good agreement with MC
simulations, providing that spatial and temporal corrections are applied (Mein
et al
2017, Rankine et al
2017). However, the critical issue of
irradiating near tissue-air interfaces, where ERE can still occur even when opposing
beams are used (Bol et al
2015), was not investigated.This work is a demonstration of the use of PRESAGE® for accurate relative
dosimetry under the effect of a magnetic field to detect the ERE when air-gaps are
present. We define a methodology by comparing PRESAGE® measurements with
EGSnrc MC simulations of an experimental set-up using an in-house phantom placed in an
electromagnet and irradiated by a cobalt-60 (60Co) source. This methodology
is of great interest to validate the research version of the Monaco (Elekta AB,
Stockholm, Sweden) TPS which has been specially developed to model the magnetic field
effect and dosimetry in the MR-linac (Bol et al
2012). We present the results of
initial investigations in this area.
Material and methods
Experimental set-up
At the UK’s National Physical Laboratory (NPL, Teddington, UK) an electromagnet
(250MM Electromagnet, GMW, USA) was installed adjacent to a (60Co) unit
(Theratron 780C, Theratronics, Ontario, Canada), to study the effects of magnetic
field (B) on dose distribution prior to the MR-linac coming into service.A Perspex phantom was developed, consisting of a number of separate 5 5 1 slabs that attach to each other and to the 5 7 surface of a black Acetal base, with the help of
cylindrical rods. The phantom accommodates cylindrical samples of PRESAGE®
(Heuris Pharma, Skillman, NJ) of 2 cm diameter and is placed between the poles of the
electromagnet as shown in (figure 1(a)).
Figure 1.
(a) Perspex phantom attached to the black Acetal base and positioned between
the poles of the electromagnet (5 cm gap allow varying the magnetic field
strength from 0 to 2 T). (b) PRESAGE® sample configuration within
the phantom to be placed between the electromagnet poles, at a distance of
162 cm from the 60Co source. (c) The same PRESAGE® sample
configuration within the phantom before its placement at the isocentre of the
MR-linac. (d) EGSnrc MC simulation diagram of the 60Co unit set-up.
Measurements were performed at 127.45 cm from the source, for convenience, and
compared with BEAMnrc simulations. The output from BEAMnrc was saved as a phase
space file and used as source input for EGSnrc/Cavity, where the electromagnet
and the phantom geometry were created.
(a) Perspex phantom attached to the black Acetal base and positioned between
the poles of the electromagnet (5 cm gap allow varying the magnetic field
strength from 0 to 2 T). (b) PRESAGE® sample configuration within
the phantom to be placed between the electromagnet poles, at a distance of
162 cm from the 60Co source. (c) The same PRESAGE® sample
configuration within the phantom before its placement at the isocentre of the
MR-linac. (d) EGSnrc MC simulation diagram of the 60Co unit set-up.
Measurements were performed at 127.45 cm from the source, for convenience, and
compared with BEAMnrc simulations. The output from BEAMnrc was saved as a phase
space file and used as source input for EGSnrc/Cavity, where the electromagnet
and the phantom geometry were created.
PRESAGE® 3D dosimeter / cuvettes irradiation
PRESAGE® is a commercially available dosimeter which consists of a
solid-state polyurethane matrix with a free radical initiator and a leuco-dye. When
irradiated, PRESAGE® undergoes an optical density (OD) change as the
leuco-dye is oxidised from leuco malachite green to malachite green with a peak
absorption maximum around 633 nm (Guo et al
2006). Two batches of custom-made
cylindrical samples of PRESAGE® (density 1.07 g⋅ ) with 2 cm diameter and approximately 6 cm length
were created with a plastic mould and cut to have 2.5 cm length (5 cm samples cut in
half). For batch 1, the meniscus end of the samples was removed (1 cm), while for batch 2, both ends of the samples
were removed (5 mm each end).The reproducibility of our PRESAGE® samples was determined via two
irradiations, one using the cobalt source and one with an Elekta Synergy Linac
(Elekta, Crawley, UK) with a 6 MV beam. On each platform, three sets of three
cuvettes were irradiated in a water tank, with doses of 2, 6 and 10 Gy. Each sample
was measured on four separate occasions (1, 2, 4 and 6 days) after irradiation, and
the absorbance at 633 nm was recorded (an average of three consecutive readings)
using a spectrophotometer (6705 Jenyway, Staffordshire, UK). For each radiation
source, time point and each dose value, the mean and sample standard deviation for
the three cuvettes were obtained and used to calculate a coefficient of variance.The effect of magnetic field on the PRESAGE® dose sensitivity was examined
by irradiating cuvettes on the MR-linac under similar conditions except for the
magnetic field. In the first case, the magnet was ramped down and in the second it
was at field (1.5 T). Doses of 2.67, 6.23 and 9.79 Gy were given (300, 700 and 1100
MU). Two samples were irradiated at each combination of dose and magnetic field and
the results read out on a Cary 50 Bio UV-VIS spectrophotometer (Agilent Technologies,
California, USA), by recording a spectrum three times consecutively and taking the
average of the points at 633 nm (the difference in readout equipment was dictated by
equipment availability). Data were analysed by performing a linear fit on the
absorbance values for the three dose points and using the LINEST function of Excel to
estimate the sensitivity and its corresponding error from the gradient.
EGSnrc MC simulations in the presence of magnetic field
A source model for the 60Co unit has been previously implemented and
validated at the NPL in BEAMnrc, an MC code based on EGSnrc (EGSnrc v4-r2-4-0)
(Rogers et al
2005), for reference conditions
(source to surface distance (SSD) = 95 cm, 10 10 cm2 field). The 60Co
unit comprises the 60Co source shielded by stainless steel and lead
encapsulation, a tungsten collimator and 5 pairs of lead jaws that move together. The
available source model was modified and a phase space file (ph-sp) was generated for
non-reference conditions of SSD = 127.45 cm, 10 10 cm2 field (Costa et
al
2017).To validate the BEAMnrc simulations, photon energy fluence spectra using the BEAMnrc
utility code BEAMDP (Ma and Rogers 2016) were determined based on the ph-sp data and compared against
measurements of air kerma. These profiles were obtained between −15 cm and 15 cm in
the and directions perpendicular to the beam, with an
ionization chamber (Semiflex 31010, PTW, Freiburg, Germany) placed on an empty water
tank. A schematic of the arangement for the measurements is shown on the diagram in
figure 1(d). Gamma index analysis
(Depuydt et al
2002), with 1.5% as dose
difference (DD) and 1.5 mm as distance to agreement (DTA), was applied to assess the
agreement between the measurements and the simulated photon energy fluence.A second EGSnrc user code called Cavity, a C++ based MC system (Kawrakow et
al
2009) was used to simulate the
experimental set-up described in section 2.1. Both electromagnetic coils and the phantom were simulated and the
ph-sp of the 60Co head was used as the input source (figure 1(d)). Simulations of two
PRESAGE® cylinders with their long axes parallel to the beam in the
presence of different magnetic field strengths (0.5 T to 2 T) and different sizes air
gaps (0.5 to 2 cm) were performed to study the influence of these parameters on the
dose distribution. These simulations allowed us to identify the measurement
conditions that would give us easily visible changes in the dose distribution, in
particular in the ERE region, by only changing one parameter (air gap size or
magnetic field strength). The lengths of the simulated PRESAGE® samples
were chosen based on the phantom length limitation (i.e. total for two halves of the
phantom plus air gap) of 7 cm. One-dimensional profiles were obtained along the
central region with variable sampling density, corresponding to the dose gradient
along the beam direction (i.e. finer sampling where the dose changed more rapidly
with distance). 0.5 1 1 mm3 and 1 1 1 mm3 voxels were used in the build-up
region (first 4.5 mm of the sample); 2.5 1 1 mm3 voxels were used in the more
slowly varying region distal to the depth of dose maximum; and 1 1 1 m3, followed by 0.5 1 1 mm3, in the steepening part of the
curve in the edge region affected by the ERE (last 3 mm). Data were normalized to the
centre of each simulated profile. The transport cut-offs were set to
AE = ECUT = 0.521 MeV for electrons, which corresponds to the rest mass plus the
kinetic energy. The photons cut-offs were set to AP = PCUT = 0.01 MeV. The macro
provided by EGSnrc (emf_macros.mortran with EM ESTEPE = 0.02) was included in the
code to simulate magnetic fields (Kirkby et al
2008, Kawrakow et
al
2016, Malkov and Rogers 2016), which were homogeneous and
present through the simulation volume.
PRESAGE® dose measurements in the presence of magnetic field
PRESAGE® irradiations versus EGSnrc MC simulations
On the basis of the simulation results, two sets of measurements were identified
for subsequent experiments. Two sets of two PRESAGE® samples from batch
1, separated by 0.5 cm air gap, were irradiated (3 Gy at the centre of the
electromagnet poles, measured with an ionization chamber), at 0.5 T and 1.5 T
magnetic field strengths (figure 1(b)). The choice of these two irradiations was based on the
simulations performed in section 2.3, where the effect of the air gap size and magnetic field strengths
was tested. The 60Co output value was determined with a NE2611(UK)
ionization chamber based on Lillicrap et al (1990) guidelines.Axial and sagittal 2D dose distributions were also simulated for comparison with
measurements. The beam exit end of the PRESAGE samples was simulated to compare
with measured data. The axial 2D distribution obtained with PRESAGE®
was averaged over the last 1 mm of the sample. Percentage depth dose (PDD)
profiles were simulated independently with variable sampling from 0.5 1 1 mm3 to 2.5 1 1 mm3, performed in a similar
manner as described in section 2.3. The statistical uncertainty of all simulations was within 2% for 2D
data and 1% for 1D profiles. Measured and simulated profiles were normalized to
the center of each simulated profile and compared using 1D gamma criterion
analysis of 3% DD and 1.5 mm DTA (3%, 1.5 mm) (Depuydt et al
2002). 2D dose distributions
were compared using 2D gamma analysis with the same criteria.
PRESAGE®/optical-CT scanner
PRESAGE® samples were imaged using an improved version of an in-house
optical-CT microscopy scanner, which has previously been described and
characterized for microbeam radiotherapy applications (Doran et
al
2013). Improvements to the
camera, computer and positioning system allowed faster and more accurate scans
(McErlean et al
2016).Optical-CT scanning was performed following the guidelines of Doran (2013a). Each sample was attached
to the rotary sample holder and placed inside a glass tank with a matching liquid
that has the same refractive index as PRESAGE®. A custom designed cap,
developed in-house, was used to position PRESAGE® cylinders within the
sample holder. A small mark was made with a scalpel on both sample edges and the
cap for reproducible positioning in pre and post-irradiation scans. As the
positioning caps clamp onto the top 5 mm of the sample, it is not currently
possible to image both distal and proximal ends of the sample simultaneously. Thus
a pre-scan was performed for two separate regions of length 2.5 cm at opposite
ends of the sample (i.e. the sample was inverted and then repositioned), allowing
both end surfaces to be imaged. Each scan takes approximately 1 minute to obtain
1000 projections, each of 512 512 pixels, over 180 rotation for a field of view (FOV) of
approximately (2.6 cm)2 to obtain a reconstructed image with a voxel
size of (0.052 mm)3. Projections were reconstructed by filtered
back-projection and the reconstructed images were post processed using
Matlab®. The scans were assembled together and the high intensity
values caused by artefacts within the sample were identified and replaced by the
median of the surrounding pixel values. A median filter was also applied to smooth
the images.
PRESAGE® irradiation on the MR-linac versus Monaco TPS
simulations
PRESAGE® irradiations were also performed at the MR-linac. The perspex
phantom was used again to accommodate two PRESAGE® samples with 2.5 cm
length separated by a 0.5 cm air gap. The phantom was placed so that the MR-linac
isocentre was 1 cm downstream with respect to the incident radiation from the
proximal face of the phantom, as shown in figure 1(c). Two irradiations were performed, using samples
from batch 2, one at 0 T and the other at 1.5 T, with the 7MV beam of the MR-linac,
gantry 0, a 20 20 cm2 field and 500MU.
PRESAGE® samples, from batch 2, were scanned as described in section
2.4.1. The experimental set-up
was recreated in silico with CARPE DICOM (Elekta AB, Stockholm, Sweden) and simulated
with 1 mm resolution and 1% uncertainty using the GPU-based Monte Carlo dose
calculation algorithm (GPUMCD) in Monaco TPS. Again, measured and simulated profiles
were normalized to the centre of each simulated profile.
Results
PRESAGE® cuvettes irradiation
For the reproducibility experiment, the cobalt source samples gave a coefficient of
variance with a mean over all times and doses of 1.8% and range of [0.9%, 4.1%],
whilst the equivalent result for samples irradiated on the linac was 2.9% [0.4%,
5.0%]. For the experiment testing the effect of magnetic field on the
PRESAGE®, the measured sensitivities with and without the magnetic
field were (0.0425 0.0017) ⋅ and (0.0434 0.0001) ⋅, a difference of 2.1%. It should be noted that
from the three data points measured, the error in the slope given by the LINEST
algorithm will not be accurately determined.
EGSnrc model validation and EGSnrc MC simulation in the presence of magnetic
field
The phase space file validation of the source model is shown in figure 2(a) and (b). Profiles were normalized to the mean value of the
central region, and showed good agreement with 100% points passing the gamma
criterion of 1.5%, 1.5 mm.
Figure 2.
Comparison between air kerma measurements (Measurement) and simulated energy
fluence profile (BEAM simulation) at 127.45 cm from the source, in (a) direction (horizontal) and (b) direction (vertical). Gamma values for
1.5%, 1.5 mm criterion are also shown. Simulated 1D profiles obtained by taking
a central profile along the length of the two PRESAGE® samples with
different (c) air gap sizes (at 1.5 T) and (d) magnetic field strengths (with a
0.5 cm air gap). The simulated uncertainty is given by the error bars and was
below 1% for each simulated data.
Comparison between air kerma measurements (Measurement) and simulated energy
fluence profile (BEAM simulation) at 127.45 cm from the source, in (a) direction (horizontal) and (b) direction (vertical). Gamma values for
1.5%, 1.5 mm criterion are also shown. Simulated 1D profiles obtained by taking
a central profile along the length of the two PRESAGE® samples with
different (c) air gap sizes (at 1.5 T) and (d) magnetic field strengths (with a
0.5 cm air gap). The simulated uncertainty is given by the error bars and was
below 1% for each simulated data.Initial MC simulations of the experimental set-up showed that keeping an air gap of
0.5 cm and changing the magnetic field from 0.5 T to 1.5 T maximizes the differences
between each PRESAGE® sample (20% at the end of the first and second
samples shown in figure 2(d)). This
occurs as a 0.5 cm air gap is large enough for the electrons to curl back to the
sample when a perpendicular 1.5 T magnetic field is present (radius = 3.7 mm in
vacuum), but less so when a 0.5 T (radius = 11.2 mm in vacuum) is present instead.
This justifies the selection of this set-up for the measurements described in section
2.4.1. The same does not happen
for an air gap larger than 1 cm, as electrons will curve back for magnetic field
strengths between 0.5 T and 2 T. At 1.5 T, changing the air gap size from 0.5 cm to
2 cm or more did not produce significant differences on the ERE values (2–3% on the
first PRESAGE® and 5–7% on the second) as shown in figure 2(c).Representative central slices from a sagittal view of the optical CT data are
shown for two of the PRESAGE® samples separated by an air gap of 0.5 cm
and irradiated at 0.5 T and 1.5 T magnetic field strengths in figures 3(a) and (c) respectively. A number of image artefacts are
present. These include microscopic inclusions in the PRESAGE® samples
and schlieren effects which are attributed to localised refractive index
inhomogeneities (Doran 2009).
Suspended particles in the matching liquid also degrade the image quality but are
hard to remove completely in practice.
Figure 3.
Normalized dose maps of the central sagittal slice along the two
PRESAGE® samples at (a) 0.5 T and (c) 1.5 T and respective (b)
and (d) dose profile at the central region of interests (ROI) compared with
simulated profiles. Error bars are representative of 1 standard deviation of
the measured values within the ROI and account for readout artefacts and
sample inhomogeneities. Gamma values for 3%, 1.5 mm gamma criterion are also
shown.
Normalized dose maps of the central sagittal slice along the two
PRESAGE® samples at (a) 0.5 T and (c) 1.5 T and respective (b)
and (d) dose profile at the central region of interests (ROI) compared with
simulated profiles. Error bars are representative of 1 standard deviation of
the measured values within the ROI and account for readout artefacts and
sample inhomogeneities. Gamma values for 3%, 1.5 mm gamma criterion are also
shown.The normalized measured profiles in figures 3(b) and (d) were obtained from the regions of interest (ROI) shown in figures
3(a) and (c) respectively. Comparison with simulated profiles
showed good agreement with a gamma passing rate (3%, 1.5 mm) of 99.9% for 0.5T and
99.8% for 1.5 T.In figures 3(a) and (c) one can see that the dose increase
due to the ERE is not uniform laterally within the last 2 mm of the beam exit area
of each PRESAGE® sample. The red region is not central but displaced
upwards. This asymmetry is also evident in figures 4(a) and (b), where the simulated and measured last 1 mm axial slice of the
first PRESAGE® sample irradiated at 1.5 T are shown. This vertical
shift is perpendicular to both beam and magnetic field, as the electron paths
curve and deposit their energy at the top of the sample. This leads to a
corresponding dose deficit at the bottom of the sample, and this is visualized
quantitatively in the vertical profile of figure 4(d) for both simulated and experimental data. By
contrast the horizontal profile in figure 4(e) shows no variation. Comparison between measured and simulated
vertical and a horizontal profiles are shown in figures 4(d) and (e) respectively and these are discussed in further detail below. For a
visual 2D dose comparison, both simulated and measured averaged 2D slices were
resized to a 0.25 cm pixel size and their relative difference shown in figure
4(c). A 2D gamma analysis for
3%, 1.5 mm was also performed and the gamma map can be seen in figure 4(f).
Figure 4.
Normalized (a) simulated and (b) measured axial slice dose distribution of
the last 1 mm of the first PRESAGE® cylinder shown in figure
3(c). (c) Relative
difference between measurements and EGSnrc MC simulations. (d) Vertical and
(e) horizontal measured and simulated profiles excluding the last 1 mm edge.
Error bars are representative of 1 standard deviation of the measured values
within the ROI. (f) Gamma map for 3%, 1.5 mm of measured and simulated axial
slice.
Normalized (a) simulated and (b) measured axial slice dose distribution of
the last 1 mm of the first PRESAGE® cylinder shown in figure
3(c). (c) Relative
difference between measurements and EGSnrc MC simulations. (d) Vertical and
(e) horizontal measured and simulated profiles excluding the last 1 mm edge.
Error bars are representative of 1 standard deviation of the measured values
within the ROI. (f) Gamma map for 3%, 1.5 mm of measured and simulated axial
slice.
PRESAGE® irradiation on the MR-linac versus Monaco TPS
simulations
Comparison between PRESAGE® results and Monaco TPS simulation at the
MR-linac are shown in figures 5(b)
and (d), based on profiles along
the length of both samples. A representative normalized dose distribution and
central slices from a sagittal view are shown for both simulations (figure 5(a)) and Optical-CT data results with
PRESAGE® (figure 5(c)).
Figure 5.
(a) Normalized dose distribution in the central sagittal plane simulated
with Monaco TPS at 1.5 T. (c) Dose distribution obtained with
PRESAGE® samples irradiated at the same conditions. (b) Monaco
and PRESAGE® normalized dose profile at the central ROI at 1.5 T
and (d) 0 T. Error bars are representative of 1 standard deviation of the
measured values within the ROI. 1D gamma values for 3%, 1.5 mm gamma
criterion are also shown.
(a) Normalized dose distribution in the central sagittal plane simulated
with Monaco TPS at 1.5 T. (c) Dose distribution obtained with
PRESAGE® samples irradiated at the same conditions. (b) Monaco
and PRESAGE® normalized dose profile at the central ROI at 1.5 T
and (d) 0 T. Error bars are representative of 1 standard deviation of the
measured values within the ROI. 1D gamma values for 3%, 1.5 mm gamma
criterion are also shown.
Discussion
The aim of this work was to perform a perliminary study of the effects of magnetic field
on dose distributions as part of our methodological development prior to the MR-linac
coming into service. PRESAGE® has been used previously to measure highly
conformal dose distributions in radiotherapy applications, showing accurate and reliable
results to validate patient treatment plans before its delivery (Oldham et
al
2008, Sakhalkar et
al
2009, Brady et al
2010, Jackson et al
2015). PRESAGE® also
showed good potential to detect the ERE in the presence of cylindrical air cavities (Lee
et al
2017a), but has not yet been assessed
in the presence of air gaps and compared directly with MC simulations.Previous published literature has reported a PRESAGE® reproducibility of
better than 2% (Guo et al
2006). It was not the original
intention of this study to make definitive measures of absolute absorption, since many
prior studies have demonstrated the difficulty of this task and the level of precautions
necessary (Olding and Schreiner 2011,
Skyt et al
2011). For our ERE measurements, it
was the relative dosimetry within a single scan that was of interest. Thus, we did not
attempt to control for day-to-day variations in temperature and other factors. Nor did
we perform prior studies to establish the performance of the spectrophotometers as a
gold standard. Instead, the results from the reproducibility experiment were used as a
guide to determine whether the radiation sensitivities with and without magnetic field
were significantly different. We observed that this difference was 2.1% and this is of
the same order as the mean coefficient of variance in the reproducibility study. This
suggests that the magnetic field does not affect the sensitivity, but, in view of other
research in this area (Mathis et al
2014, Lee et al
2017b), further dedicated experiments
would be needed to establish this conclusively.Similar results were reported by others for doses up to 5 Gy (Lee et al
2017b), but for higher doses,
sensitivity differences above 9% were obtained (Mathis et al
2014, Choi 2016). This difference might be explained by variations
in the fromulation of the samples. Measured dose profiles in the presence of magnetic
field showed generally very good agreement near dosimeter-air interfaces. In particular,
the 60Co irradiations represent an extremely challenging test case, with the
ERE that we wish to detect occurring over the last 3 mm of the sample (figures 3(b) and (d)). On the MR-linac itself, the range of the effect is
closer to 1 cm and correspondingly less affected by optical artefacts (figure 5(b)).These findings are very encouraging, because optical measurements have historically been
extremely challenging at interfaces between media. There are two main reasons:The chemical manufacturing process can lead to sample inhomogeneities near
surfaces, for example, through the effect of temperature during curation and,
in the case of PRESAGE® because of a loss of solvent from the
dosimeter edges, leading to a concentration of various reactants. This can lead
to a varying radiation sensitivity profile. Different radiation sensitivities
within a PRESAGE® sample have been reported before for large
PRESAGE® cylinders, showing 5% to 20% differences between the
centre and the periphery, depending on the sample (Dekker et
al
2016). We found similar
non-uniformities in our samples, as exemplified by figures 3(a) and 5(c). The effect might be expected to be particularly pronounced in
our small-diameter samples, where the surface-to-volume ratio is large. These
effects may be responsible for the slight disagreement in the build-up region
shown in the left-hand parts of the profiles in figures 3(c) and (d). The inhomogenities are also visible in figure 5, but the build-up shows better
agreement, which could be due to the use of a different batch (batch 2), or the
physical removal of both sample edges, whereas for batch 1, only the beam exit
side of the samples was removed. A further batch-specific difficulty that we
encountered during this work was the formation of surface nubs/bubbles on the
PRESAGE® due to low levels of moisture interacting at the
interface between the mould surface and the curing polyurethane.The difficulty of obtaining an exact match in optical index between a sample
and the liquid in the optical CT scanner matching tank leads to a well known
ring artefact (Doran 2013a).
Most authors using larger samples have ignored this effect, but it occurs at
exactly the position of interest in the profiles of figures 4(d) and (e). The effect can be modelled (Doran et
al
2001), but since the shape
of the roll off at the sample edge depends on both the refractive index and the
dose-dependent sample absorption, correction requires detailed work to
characterize each dosimeter batch. Although this is not an issue for this work,
this effect will be investigated in detail before the use of these size
dosimeters to validate more clinically relevant irradiations scenarios.
Arranging the sample in such a way that the desired interface to measure is the
one at the flat face of the dosimeter makes the task much easier. While we can
still get surface artefacts within 1 mm of the samples surface, excellent
results are obtained throughout the profiles as shown in figures 3 and 5.For the above reasons, the comparison between EGSnrc MC simulation and experiment in
figure 4(d) are not satisfactory over the
edge region. However, we suggest that figure 4(e) provides early evidence of the ERE that matches simulation. Also
important to note is the misalignment that might also have occurred between the pre and
post-irradiation scan. As the OD is fairly homogeneous throughout the sample prior to
irradiation, when the pre scan is subtracted from the post scan, small misalignments
(less than 1 rotation about an axis perpendicular to the plane of
the images in figures 3(a) and (b)) are not expected to influence the dose
results. However, the existence of small imperfections inside and also at the surface of
the PRESAGE® samples results in more noise in the data if the pre and the
post irradiation images are not well aligned. This effect, which was not quantified in
this work, highlights the need for a more reproducible way to place the samples.Measured and simulated PDD profiles on Monaco showed good agreement, with a gamma (3%,
1.5 mm) passing rate of 98.4% at 1.5 T and 99.6% at 0 T. This initial assessment based
on profiles shows the potentialities of PRESAGE® to detect the ERE accurately
and to validate more complex dose distributions simulated with Monaco TPS.
Conclusion
While PRESAGE® readout data can be improved by reducing edge and schliere
artefacts, and a better understanding of the sample inhomogenities is still needed, good
agreement between measured and computed results for both 1D profiles and 2D planes in
the ERE region showed the great capabilities of PRESAGE® to detect the ERE.
This study gives encouragement for the use of PRESAGE® on the MR-linac to
validate Monaco TPS for more complex dose distributions near tissue-air interfaces.
Authors: A J E Raaijmakers; B Hårdemark; B W Raaymakers; C P J Raaijmakers; J J W Lagendijk Journal: Phys Med Biol Date: 2007-11-15 Impact factor: 3.609
Authors: Jan J W Lagendijk; Bas W Raaymakers; Alexander J E Raaijmakers; Johan Overweg; Kevin J Brown; Ellen M Kerkhof; Richard W van der Put; Björn Hårdemark; Marco van Vulpen; Uulke A van der Heide Journal: Radiother Oncol Date: 2007-11-26 Impact factor: 6.280
Authors: B W Raaymakers; J J W Lagendijk; J Overweg; J G M Kok; A J E Raaijmakers; E M Kerkhof; R W van der Put; I Meijsing; S P M Crijns; F Benedosso; M van Vulpen; C H W de Graaff; J Allen; K J Brown Journal: Phys Med Biol Date: 2009-05-19 Impact factor: 3.609
Authors: Simon J Doran; A T Abdul Rahman; Elke Bräuer-Krisch; Thierry Brochard; John Adamovics; Andrew Nisbet; David Bradley Journal: Phys Med Biol Date: 2013-08-22 Impact factor: 3.609
Authors: Linda G W Kerkmeijer; Clifton D Fuller; Helena M Verkooijen; Marcel Verheij; Ananya Choudhury; Kevin J Harrington; Chris Schultz; Arjun Sahgal; Steven J Frank; Joel Goldwein; Kevin J Brown; Bruce D Minsky; Marco van Vulpen Journal: Front Oncol Date: 2016-10-13 Impact factor: 6.244
Authors: Jin Dong Cho; Jaeman Son; Chang Heon Choi; Jin Sung Kim; Hong-Gyun Wu; Jong Min Park; Jung-In Kim Journal: PLoS One Date: 2020-03-16 Impact factor: 3.240