| Literature DB >> 29308077 |
Roya Sharifpour1,2, Pardis Ghafarian3,4, Mehrdad Bakhshayesh-Karam5, Hamidreza Jamaati3, Mohammad Reza Ay1,2.
Abstract
BACKGROUND: The most important advantage of positron emission tomography/computed tomography (PET/CT) imaging is its capability of quantitative analysis. The aim of the current study was to choose the proper standardized uptake value (SUV) threshold, when the time-of-flight (TOF) and point spread function (PSF) were used for respiratory artifact reduction in the liver dome in a new-generation PET/CT scanner.Entities:
Keywords: PET/CT imaging; Point-Spread-Function; Respiratory artifact; Standardized Uptake Value; Threshold; Time-of-Flight
Year: 2017 PMID: 29308077 PMCID: PMC5749325
Source DB: PubMed Journal: Tanaffos ISSN: 1735-0344
Figure 1.Illustration of measured (a) SUVmax, (b) SUV75%, (c) SUV50% and (d) SUV45% as a function of various sphere sizes in NEMA IEC Body phantom with 4:1 (upper) and 2:1(lower)activity ratio and background concentration of 5.31kBq/mL. Routine Protocol (HD+PSF with 3it, 18sub,6.4mm filter), PSF+TOF1(2it, 18sub, 4.4mm filter), PSF+TOF2(2it, 18sub, 5.4mm filter), PSF+TOF3(2it, 18sub, 6.4mm filter), PSF+TOF4(2it, 24sub, 4.4mm filter), PSF+TOF5(2it, 24sub, 5.4mm filter), PSF+TOF6(2it, 24sub, 6.4mm filter).
Figure 2.Correlation between measured SUV of 27 patients data. SUV50% with regard to PSF + TOF3 method is compared with the other reconstruction methods. The solid line shows the identity line. Routine Protocol (HD+PSF with 3it, 18sub,6.4mm filter), PSF+TOF1(2it, 18sub, 4.4mm filter), PSF+TOF2(2it, 18sub, 5.4mm filter), PSF+TOF3(2it, 18sub, 6.4mm filter), PSF+TOF4(2it, 24sub, 4.4mm filter), PSF+TOF5(2it, 24sub, 5.4mm filter), PSF+TOF6(2it, 24sub, 6.4mm filter).
Lin concordance correlation coefficients Between the SUV Measured with Different Protocols
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | |||||
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | |||||
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | |||||
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | ||||||
| 0.91 | --- | --- | --- | --- | --- | --- | --- | --- | --- | ||||||
| --- | --- | --- | --- | --- | --- | --- | --- | --- | |||||||
| --- | --- | --- | --- | --- | --- | --- | --- | ||||||||
| 0.84 | 0.70 | 0.74 | 0.79 | 0.65 | 0.70 | 0.75 | --- | --- | --- | --- | --- | --- | --- | ||
| 0.92 | 0.82 | 0.86 | 0.90 | 0.77 | 0.82 | 0.87 | --- | --- | --- | --- | --- | --- | |||
| 0.89 | 0.77 | 0.82 | 0.86 | 0.73 | 0.78 | 0.83 | --- | --- | --- | --- | |||||
| 0.86 | 0.73 | 0.78 | 0.83 | 0.68 | 0.73 | 0.79 | --- | --- | --- | ||||||
| 0.92 | 0.84 | 0.87 | 0.91 | 0.80 | 0.84 | 0.89 | 0.92 | --- | --- | ||||||
| 0.90 | 0.79 | 0.84 | 0.88 | 0.75 | 0.80 | 0.85 | --- | --- | |||||||
| 0.87 | 0.75 | 0.80 | 0.84 | 0.71 | 0.76 | 0.81 | --- | ||||||||
Lin Concordance Correlation Coefficients (CCC) greater than 0.95 are in bold.
HD+PSF with 3it, 18sub and 6.4 mm filter
2it, 18sub, 4.4 mm filter
2it, 18sub, 5.4 mm filter
2it, 18sub, 6.4 mm filter
2it, 24sub, 4.4 mm filter
2it, 24sub, 5.4 mm filter
2it, 24sub, 6.4 mm filter
The Mean of Difference Between the SUV Measured with Various Protocols and Routine Protocol
| PSF+TOF1 | 1.47±1.06 | 1.31±0.97 | 1.01±0.73 | 0.93±0.67 |
| PSF+TOF2 | 1.00±0.77 | 0.87±0.66 | 0.68±0.54 | 0.64±0.48 |
| PSF+TOF3 | 0.52±0.57 | 0.45±0.49 | 0.38±0.39 | 0.36±0.35 |
| PSF+TOF4 | 1.95±1.43 | 1.74±1.33 | 1.28±0.93 | 1.21±0.89 |
| PSF+TOF5 | 1.41±1.09 | 1.23±0.92 | 0.93±0.73 | 0.88±0.69 |
| PSF+TOF6 | 0.88±0.89 | 0.73±0.68 | 0.58±0.56 | 0.55±0.51 |
2it, 18 sub, 4.4 mm filter.
2it, 18 sub, 5.4 mm filter.
2it, 18 sub, 6.4 mm filter.
2it, 24 sub, 4.4 mm filter.
2it, 24 sub, 5.4 mm filter.
2it, 24 sub, 6.4 mm filter.
Figure 3.Top (a,b) and bottom (c,d) pulmonary nodulesof a patient with colon cancer.Transverseandcoronal view forlung window of CT images and PET images with various reconstruction methods are illustrated. The impact of TOF on improvement in contrast and lesion detectability is clear in all methods. (Body weight: 105 kg and injected dose: 492MBq of 18F-FDG)