| Literature DB >> 29236918 |
Nicolau Silveira-Neto1, Mateus Ericson Flores1, João Paulo De Carli1, Max Dória Costa2, Felipe de Souza Matos3, Luiz Renato Paranhos4, Maria Salete Sandini Linden1.
Abstract
OBJECTIVES: This research evaluated detail registration in peri-implant bone using two different cone beam computer tomography systems and a digital periapical radiograph.Entities:
Mesh:
Substances:
Year: 2017 PMID: 29236918 PMCID: PMC5707201 DOI: 10.6061/clinics/2017(11)10
Source DB: PubMed Journal: Clinics (Sao Paulo) ISSN: 1807-5932 Impact factor: 2.365
Figure 1(a) Upper view of the planned clinical situations: BBD, PBD, and BC. (b) and (c) Buccal and upper views of the installed implants showing the three clinical situations.
Specifications for the Kodak 9000 3D CBCT.
| Parameters | kp1 | kp2 | kp3 |
|---|---|---|---|
| kVp | 68 | 60 | 80 |
| mAs | 8.0 | 6.3 | 12.0 |
| Voxel | 0.076 | 0.1 | 0.2 |
| FOV (mm) | 4.7 x 5.9 | 4.7 x 5.9 | 4.7 x 5.9 |
| Exam time (s) | 10.8 | 10.8 | 10.8 |
| Software | Kodak Dental Imaging Software 3D | Kodak Dental Imaging Software 3D | Kodak Dental Imaging Software 3D |
Specifications for the i-CAT CBCT.
| Parameters | ip1 | ip2 | ip3 |
|---|---|---|---|
| kVp | 120 | 120 | 120 |
| mAs | 37.07 | 20.27 | 18.54 |
| Voxel | 0.2 | 0.125 | 0.4 |
| FOV (mm) | 8.0 x 8.0 | 8.0 x 8.0 | 8.0 x 8.0 |
| Exam time (s) | 26.09 | 26.09 | 8.9 |
| Software | i-CAT Vision Xoran | i-CAT Vision Xoran | i-CAT Vision Xoran |
Figure 2Image results from the BBD, PBD, and BC cases for the different image acquisition protocols from sagittal CBCT images.
Figure 3Measurements of a real object made with a digital caliper in regions A, B, and E.
Figure 4Model of presentation for evaluators for the PBD situation.
Mean values for the BBD, PBD, and BC cases using the Kodak 9000 3D CBCT.
| Protocol | Region A (mm) | Region B (mm) | Region E (mm) |
|---|---|---|---|
| BBD - buccal bone defect | |||
| Control | 4.622b | 0.688a | 2.750a |
| kp1 | 4.730a | 0.704a | 2.740a |
| kp2 | 4.686a | 0.528b | 2.640b |
| kp3 | 4.690a | 0.534b | 2.604b |
| PBD - peri-implant bone defect | |||
| Control | 9.914a | 3.216a | 2.858a |
| kp1 | 8.980c | 2.856c | 2.784b |
| kp2 | 9.676b | 3.040b | 2.846a |
| kp3 | 9.560b | 2.914c | 2.810ab |
| BC - bone contact | |||
| Control | 0.878a | 0.862a | 2.518a |
| kp1 | 0.914a | 0.700b | 2.450b |
| kp2 | 0.918a | 0.574c | 2.408bc |
| kp3 | 0.872a | 0.588c | 2.370c |
Means followed by the same letter are not significantly different (Tukey’s test, α=0.05).
Mean values for the BBD, PBD, and BC cases using the i-CAT CBCT.
| Protocol | Region A (mm) | Region B (mm) | Region E (mm) |
|---|---|---|---|
| BBD - buccal bone defect | |||
| Control | 4.622a | 0.688a | 2.750a |
| ip1 | 4.234d | 0.620ab | 2.260c |
| ip2 | 4.500b | 0.570b | 2.510b |
| ip3 | 4.384c | 0.394c | 2.020d |
| PBD - peri-implant bone defect | |||
| Control | 9.914a | 3.216a | 2.858a |
| ip1 | 8.712c | 2.592b | 2.632b |
| ip2 | 8.980b | 2.586b | 2.510c |
| ip3 | 8.888bc | 2.460c | 2.010d |
| BC - bone contact | |||
| Control | 0.878a | 0.862a | 2.518a |
| ip1 | 0.648c | 0.534c | 2.388b |
| ip2 | 0.856a | 0.564c | 2.250c |
| ip3 | 0.764b | 0.724b | 1.736d |
Means followed by the same letter are not significantly different (Tukey’s test, α=0.05).
Score interpretations for tomographic images and digital periapical radiographs in each situation.
| Situation | Kodak | i-CAT | Digital Periapical Radiograph |
|---|---|---|---|
| BBD | p1 - Very Good | p1- Good | Very Poor |
| p2 - Very Good | p2 - Good | ||
| p3 - Very Good | p3 - Good | ||
| PBD | p1 - Very Poor | p1 - Poor | Very Good |
| p2 - Very Poor | p2 - Very Poor | ||
| p3- Very Poor | p3 - Very Poor | ||
| BC | p1- Very Good | p1 - Poor | Very Good |
| p2 - Very Good | p2 - Good | ||
| p3- Poor | p3 - Poor |