Literature DB >> 29226198

Restoration of the Mechanical Axis in Total Knee Artrhoplasty Using Patient-Matched Technology Cutting Blocks. A Retrospective Study of 132 Cases.

Dimitrios N Lyras1, Robert Greenhow1, Craig Loucks1.   

Abstract

BACKGROUND: The aim of this study is to evaluate the accuracy of bone cuts and the resultant alignment, using the MyKnee patient specific cutting blocks.
METHODS: We retrospectively reviewed 132 patients undergoing primary TKR for osteoarthritis by one single surgeon. The operative time, the preoperative Hip-Knee-Ankle (HKA) axis based on the CT-scan, the postoperative HKA axis based on long axis standing x-rays, the planned and the actual size of the femoral and the tibial components, and the number of the recuts which has been made intraoperative were measured.
RESULTS: The average preoperative HKA axis was 177.50 (range 163.50 to 1940), whereas the average postoperative HKA axis was 179.40 (range 177.10 to 182.70). No outliers were reported in the study (0%). Intraoperatively, 4 femoral components (3.03%), and 7 tibial components (5.30%) applied to the patients were different than the planned size. There was no need of recuts in any of our cases intraoperatively.
CONCLUSION: The MyKnee system evaluated in this study was shown to be remarkable reliable in the coronal plane alignment, and the prediction of the component size. However, further studies are needed to determine whether there are any clinically important improvements in outcomes or patient satisfaction when using patient-specific cutting blocks for TKA.

Entities:  

Keywords:  CT scan; HKA ankle; Knee alignment; Patient specific instruments; Total knee arthroplasty

Year:  2017        PMID: 29226198      PMCID: PMC5712393     

Source DB:  PubMed          Journal:  Arch Bone Jt Surg        ISSN: 2345-461X


  26 in total

1.  Alignment in total knee arthroplasty. A comparison of computer-assisted surgery with the conventional technique.

Authors:  H Bäthis; L Perlick; M Tingart; C Lüring; D Zurakowski; J Grifka
Journal:  J Bone Joint Surg Br       Date:  2004-07

2.  Imageless computer assisted versus conventional total knee replacement. A Bayesian meta-analysis of 23 comparative studies.

Authors:  Yaron S Brin; Vassilios S Nikolaou; Lawrence Joseph; David J Zukor; John Antoniou
Journal:  Int Orthop       Date:  2010-04-08       Impact factor: 3.075

3.  Press-fit condylar design total knee arthroplasty. Fourteen to seventeen-year follow-up.

Authors:  David J Rodricks; Shantanu Patil; Pamela Pulido; Clifford W Colwell
Journal:  J Bone Joint Surg Am       Date:  2007-01       Impact factor: 5.284

4.  Coronal alignment after total knee replacement.

Authors:  R S Jeffery; R W Morris; R A Denham
Journal:  J Bone Joint Surg Br       Date:  1991-09

5.  The accuracy of acquisition of an imageless computer-assisted system and its implication for knee arthroplasty.

Authors:  S Lustig; C Fleury; D Goy; P Neyret; S T Donell
Journal:  Knee       Date:  2010-01-12       Impact factor: 2.199

6.  Accuracy of 3D cartilage models generated from MR images is dependent on cartilage thickness: laser scanner based validation of in vivo cartilage.

Authors:  Seungbum Koo; Nicholas J Giori; Garry E Gold; Chris O Dyrby; Thomas P Andriacchi
Journal:  J Biomech Eng       Date:  2009-12       Impact factor: 2.097

7.  The Chitranjan Ranawat award: is neutral mechanical alignment normal for all patients? The concept of constitutional varus.

Authors:  Johan Bellemans; William Colyn; Hilde Vandenneucker; Jan Victor
Journal:  Clin Orthop Relat Res       Date:  2012-01       Impact factor: 4.176

8.  Improved accuracy of alignment with patient-specific positioning guides compared with manual instrumentation in TKA.

Authors:  Vincent Y Ng; Jeffrey H DeClaire; Keith R Berend; Bethany C Gulick; Adolph V Lombardi
Journal:  Clin Orthop Relat Res       Date:  2012-01       Impact factor: 4.176

9.  The value of patient-matched instrumentation in total knee arthroplasty.

Authors:  John W Noble; Chris A Moore; Ning Liu
Journal:  J Arthroplasty       Date:  2011-09-09       Impact factor: 4.757

10.  Patient-specific instrumentation improved mechanical alignment, while early clinical outcome was comparable to conventional instrumentation in TKA.

Authors:  Werner Anderl; Leo Pauzenberger; Roman Kölblinger; Gabriele Kiesselbach; Georg Brandl; Brenda Laky; Bernhard Kriegleder; Philipp Heuberer; Eva Schwameis
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-10-19       Impact factor: 4.342

View more
  1 in total

1.  Accurate implant fit and leg alignment after cruciate-retaining patient-specific total knee arthroplasty.

Authors:  Jörg Arnholdt; Yama Kamawal; Konstantin Horas; Boris M Holzapfel; Fabian Gilbert; Axel Ripp; Maximilian Rudert; Andre F Steinert
Journal:  BMC Musculoskelet Disord       Date:  2020-10-22       Impact factor: 2.362

  1 in total