Literature DB >> 29225424

Parametric analysis of electromechanical and fatigue performance of total knee replacement bearing with embedded piezoelectric transducers.

Mohsen Safaei1, R Michael Meneghini2, Steven R Anton1.   

Abstract

Total knee arthroplasty (TKA) is a common procedure in the United States; it has been estimated that about 4 million people are currently living with primary knee replacement in this country. Despite huge improvements in material properties, implant design, and surgical techniques, some implants fail a few years after surgery. A lack of information about in vivo kinetics of the knee prevents the establishment of a correlated intra- and postoperative loading pattern in knee implants. In this study, a conceptual design of an ultra high molecular weight (UHMW) knee bearing with embedded piezoelectric transducers is proposed, which is able to measure the reaction forces from knee motion as well as harvest energy to power embedded electronics. A simplified geometry consisting of a disk of UHMW with a single embedded piezoelectric ceramic is used in this work to study the general parametric trends of an instrumented knee bearing. A combined finite element and electromechanical modeling framework is employed to investigate the fatigue behavior of the instrumented bearing and the electromechanical performance of the embedded piezoelectric. The model is validated through experimental testing and utilized for further parametric studies. Parametric studies consist of the investigation of the effects of several dimensional and piezoelectric material parameters on the durability of the bearing and electrical output of the transducers. Among all the parameters, it is shown that adding large fillet radii results in noticeable improvement in the fatigue life of the bearing. Additionally, the design is highly sensitive to the depth of piezoelectric pocket. Finally, using PZT-5H piezoceramics, higher voltage and slightly enhanced fatigue life is achieved.

Entities:  

Keywords:  Total knee replacement; UHMW fatigue; embedded sensors; energy harvesting; orthopedic implant; piezoelectric sensing

Year:  2017        PMID: 29225424      PMCID: PMC5718211          DOI: 10.1088/1361-665X/aa814e

Source DB:  PubMed          Journal:  Smart Mater Struct        ISSN: 0964-1726            Impact factor:   3.585


  16 in total

1.  Contact stresses and fatigue life in a knee prosthesis: comparison between in vitro measurements and computational simulations.

Authors:  Tomaso Villa; Francesco Migliavacca; Dario Gastaldi; Maurizio Colombo; Riccardo Pietrabissa
Journal:  J Biomech       Date:  2004-01       Impact factor: 2.712

2.  Design, calibration and pre-clinical testing of an instrumented tibial tray.

Authors:  Bernd Heinlein; Friedmar Graichen; Alwina Bender; Antonius Rohlmann; Georg Bergmann
Journal:  J Biomech       Date:  2007-04-16       Impact factor: 2.712

3.  In vivo contact stresses during activities of daily living after knee arthroplasty.

Authors:  Darryl D D'Lima; Nikolai Steklov; Benjamin J Fregly; Scott A Banks; Clifford W Colwell
Journal:  J Orthop Res       Date:  2008-12       Impact factor: 3.494

4.  The yielding, plastic flow, and fracture behavior of ultra-high molecular weight polyethylene used in total joint replacements.

Authors:  S M Kurtz; L Pruitt; C W Jewett; R P Crawford; D J Crane; A A Edidin
Journal:  Biomaterials       Date:  1998-11       Impact factor: 12.479

5.  Can Intraoperative Sensors Determine the "Target" Ligament Balance? Early Outcomes in Total Knee Arthroplasty.

Authors:  Robert M Meneghini; Mary M Ziemba-Davis; Luke R Lovro; Phillip H Ireland; Brent M Damer
Journal:  J Arthroplasty       Date:  2016-04-04       Impact factor: 4.757

6.  Comparison of ISO standard and TKR patient axial force profiles during the stance phase of gait.

Authors:  Hannah J Lundberg; Valentina Ngai; Markus A Wimmer
Journal:  Proc Inst Mech Eng H       Date:  2012-03       Impact factor: 1.617

7.  Low-power circuits for the bidirectional wireless monitoring system of the orthopedic implants.

Authors: 
Journal:  IEEE Trans Biomed Circuits Syst       Date:  2009-12       Impact factor: 3.833

8.  Deformation, yielding, fracture and fatigue behavior of conventional and highly cross-linked ultra high molecular weight polyethylene.

Authors:  Lisa A Pruitt
Journal:  Biomaterials       Date:  2005-03       Impact factor: 12.479

9.  Estimating the burden of total knee replacement in the United States.

Authors:  Alexander M Weinstein; Benjamin N Rome; William M Reichmann; Jamie E Collins; Sara A Burbine; Thomas S Thornhill; John Wright; Jeffrey N Katz; Elena Losina
Journal:  J Bone Joint Surg Am       Date:  2013-03-06       Impact factor: 5.284

10.  Why are total knee arthroplasties failing today--has anything changed after 10 years?

Authors:  Peter F Sharkey; Paul M Lichstein; Chao Shen; Anthony T Tokarski; Javad Parvizi
Journal:  J Arthroplasty       Date:  2014-07-05       Impact factor: 4.757

View more
  5 in total

1.  Energy Harvesting and Sensing with Embedded Piezoelectric Ceramics in Knee Implants.

Authors:  Mohsen Safaei; R Michael Meneghini; Steven R Anton
Journal:  IEEE ASME Trans Mechatron       Date:  2018-01-15       Impact factor: 5.303

2.  Force detection, center of pressure tracking, and energy harvesting from a piezoelectric knee implant.

Authors:  Mohsen Safaei; R Michael Meneghini; Steven R Anton
Journal:  Smart Mater Struct       Date:  2018-09-25       Impact factor: 3.585

3.  Fabrication and selection of surrogate knee implant bearings for experimental evaluation of embedded in-vivo sensors.

Authors:  Robert I Ponder; Mohsen Safaei; Steven R Anton
Journal:  J Mech Behav Biomed Mater       Date:  2018-12-18

Review 4.  Bioelectronic multifunctional bone implants: recent trends.

Authors:  Marco P Soares Dos Santos; Rodrigo M C Bernardo
Journal:  Bioelectron Med       Date:  2022-09-21

5.  Performance of a Piezoelectric Energy Harvesting System for an Energy-Autonomous Instrumented Total Hip Replacement: Experimental and Numerical Evaluation.

Authors:  Hans-E Lange; Nils Arbeiter; Rainer Bader; Daniel Kluess
Journal:  Materials (Basel)       Date:  2021-09-08       Impact factor: 3.623

  5 in total