| Literature DB >> 29222607 |
Arnaud Diffo Kaze1,2,3, Stefan Maas4,5, James Belsey6, Alexander Hoffmann7,8,5, Dietrich Pape7,8,5.
Abstract
BACKGROUND: The purpose of the present study was to compare the mechanical static and fatigue strength of the size 2 osteotomy plate "Activmotion" with the following five other common implants for the treatment of medial knee joint osteoarthritis: the TomoFix small stature, the TomoFix standard, the Contour Lock, the iBalance and the second generation PEEKPower.Entities:
Keywords: Activmotion- TomoFix; Biomechanics; ContourLock; Correction angle; Fatigue strength; High tibial osteotomy (HTO); Mechanical stiffness; Osteoarthritis; PEEKPower; Permanent deformation; Static strength; iBalance
Year: 2017 PMID: 29222607 PMCID: PMC5722784 DOI: 10.1186/s40634-017-0115-3
Source DB: PubMed Journal: J Exp Orthop ISSN: 2197-1153
Different HTO implants considered in the present study. The plates are precontoured to fit the proximal tibia head. The ContourLock is wider than the other plates
Specimen grouping and assignment, depending on used implants and the performed test
| Performed test | Group I; n = 5 | Group II; | Group III; | Group IV; n = 5 | Group V; n = 5 | Group VI; n = 6 |
|---|---|---|---|---|---|---|
| Static: | TomoFix 1 | PEEKPower 1 | iBalance 1 | TomoFix sm 1 | Contour Lock 1 | Activmotion 1 |
| TomoFix 2 | PEEKPower 2 | iBalance 2 | TomoFix sm 2 | Contour Lock 2 | Activmotion 2 | |
| Dynamic: | TomoFix 3 | PEEKPower 3 | iBalance 3 | TomoFix sm 3 | Contour Lock 3 | Activmotion 3 |
| TomoFix 4 | PEEKPower 4 | iBalance 4 | TomoFix sm 4 | Contour Lock 4 | Activmotion 4 | |
| TomoFix 5 | PEEKPower 5 | iBalance 5 | TomoFix sm 5 | Contour Lock 5 | Activmotion 5 | |
| iBalance 6 | Activmotion 6 |
Fig. 1Specimen and sensors’ locations: (a) Specimen before mounting to hydraulic press. b Specimen under test: The lateral and the medial sensor (LS and MS) register the relative lateral and medial vertical displacements from the tibial head, while VS measured its vertical displacement. The sensors DX, DY1 and DY2 register the horizontal displacements of the tibial head; along the transverse axis for the first and the sagittal axis for the latter
Used failure types and their defining criteria (Diffo Kaze, 2016; Diffo Kaze et al., 2015; Maas et al., 2013)
| Failure type | Criteria |
|---|---|
| 1 | Medial or lateral displacements of the tibial head in relation to the tibial shaft of more than 2 mm equivalent to a rotation of more than 1.4 °. A counter-clockwise rotation corresponds to a valgus malrotation of the tibia head. This criterion can only be checked in the unloaded condition. |
| 2 | Visible collapse of lateral cortex. Small hairline cracks are not considered as failure. |
| 3 | Maximal displacement range of more than 0.5 mm within one hysteresis loop in the case of cyclic testing only. |
| 4 | Cracks of the screws of more than 1 mm |
Fig. 2Maximal displacement range within a hysteresis loop of an ideal spring-damper element: The hysteresis loop of an ideal spring-damper element is an inclined ellipsis
Fig. 3Fracture of the lateral cortical bone in specimens: (a) Activmotion 1 and (b) Activmotion 2. The opposite cortex was the weak point of the specimens
Fig. 4Static test results: (a) Activmotion 1: the rupture of the lateral cortex occurred without observable cracks formation. b Activmotion 2: Cracks formations preceded the final rupture of lateral cortex
Static tests summary: Displacements, valgus-malrotation of the tibial head and their corresponding crack and ultimate loads, including mean values and standard deviations (SD)
| Specimen | Crack / Ultimate load [kN] | Medial displ. at crack/ ultimate load [mm] | Lateral displ. at crack/ ultimate load [mm] | valgus-malrotation at crack/ ultimate load (°) | Lateral stiffness at crack/ ultimate load [kN/mm] | Failure types |
|---|---|---|---|---|---|---|
| Activmotion 1 | - / 8.9 | - / 1.3 | - / 2.5 | - / 0.6 | - / 3.6 | 2 |
| Activmotion 2 | 3.7 / 7.5 | 0.7 / 2.1 | 2.6 / 5.1 | 0.9 / 1.4 | 1.4 / 1.5 | 1 and 2 |
| Mean: | - / 8.2 | - / 1.7 | - / 3.8 | - / 1.0 | - / 2.6 | |
| SD ±: | - / 1.0 | - / 0.4 | - / 1.3 | - / 0.4 | - / 1.1 |
Results of the t-test comparing the previous tested implants to the Activmotion implant. Mean values were compared. All p values were greater than 0.05
| Groups | Ultimate load [kN] | Medial displ. at ultimate load [mm] | Lateral displ. at ultimate load [mm] | valgus-malrotation at ultimate load (°) | Lateral stiffness at ultimate load [kN/mm] | |
|---|---|---|---|---|---|---|
| TomoFix std. | Mean: | 5.3 | 1.2 | 4.7 | 2.8 | 1.1 |
| SD ±: | 0.1 | 0.1 | 0.4 | 0.2 | 0.1 | |
|
| > 0.05 | > 0.05 | > 0.05 | > 0.05 | > 0.05 | |
| PEEKPower | Mean: | 4.4 | 0.3 | 3.1 | 1.6 | 1.4 |
| SD ±: | 0.1 | 0.3 | 0.3 | 0.1 | 0.1 | |
|
| > 0.05 | > 0.05 | > 0.05 | > 0.05 | > 0.05 | |
| iBalance | Mean: | 5.5 | 0.3 | 1.9 | 0.9 | 3.1 |
| SD ±: | 0.2 | 0 | 0.4 | 0.4 | 0.7 | |
|
| > 0.05 | > 0.05 | > 0.05 | > 0.05 | > 0.05 | |
| TomoFix sm | Mean: | 3.4 | 0.8 | 2.1 | 1.4 | 1.7 |
| SD ±: | 0.3 | 0.2 | 0.4 | 0.1 | 0.1 | |
|
| > 0.05 | > 0.05 | > 0.05 | > 0.05 | > 0.05 | |
| Contour Lock | Mean: | 3.6 | 0.5 | 4.1 | 2.2 | 0.9 |
| SD ±: | 0.5 | 0 | 0.2 | 0.1 | 0.1 | |
|
| > 0.05 | > 0.05 | > 0.05 | > 0.05 | > 0.05 | |
Fig. 5Fracture of the contralateral cortical bone during the fatigue testing: The specimens failed by fracture of the contralateral cortical bone, similar to the static tests
Fig. 6Cracking of the contralateral cortex during the cyclical testing: Unlike the case of Activmotion 4 showed on the picture, the cracking was generally not observable
Fig. 7Examples of hysteresis loops (Activmotion 4): Curves force versus lateral displacement. The maximal displacement range, which increase with the failure, is 0,07 mm
Fig. 8Comparison of the deflection angle or valgus-malrotation of the tibial head before and after the failure for groups 1, 2, 3 and 6: The failure type 1 was observed in the case of the specimen iBalance 6 after the collapse of the opposite cortex. LS “n” means the failure occurred at load step “n”. The values of the first 3 groups are retrieved from our previous studies
Summary of fatigue failure tests: maximal load, vertical & lateral stiffness, number of cycles (all values prior to failure) and failure types. The values of the first 5 groups have been retrieved from our previous studies and reported here for comparison purposes. All the differences were not statistically significant
| Specimen | Maximal load [N] | Vertical | Lateral | Number of cycles | Failure types |
|---|---|---|---|---|---|
| TomoFix std. 3 | 1280 | 1350 | 2000 | > 60000 | 2 |
| TomoFix std. 4 | 1440 | 2000 | 2500 | > 80000 | 2 |
| TomoFix std. 5 | 1760 | 2500 | 2200 | > 120000 | 2 |
| Mean: | 1.5 | 1950 | 2233 | > 86000 | |
| SD ±: | 0.2 | 577 | 252 | 30550 | |
| PEEKPower 3 | 1440 | 2000 | 2500 | > 80000 | 2 |
| PEEKPower 4 | 1280 | 1950 | 2140 | > 60000 | 2 |
| PEEKPower 5 | 1440 | 2785 | 2250 | > 80000 | 2 |
| Mean: | 1.4 | 2245 | 2297 | > 73000 | |
| SD ±: | 0.1 | 468 | 184 | 11500 | |
| iBalance 3 | 1760 | 4000 | 3600 | > 120000 | 2,4 |
| iBalance 4 | 1760 | 3000 | 3400 | > 120000 | 2 |
| iBalance 5 | 1920 | 3000 | 2952 | > 140000 | 2 |
| iBalance 6 | 1760 | 3500 | 2500 | > 120000 | 1,2 |
| Mean: | 1.8 | 3375 | 3113 | 125000 | |
| SD ±: | 0.1 | 479 | 490 | 10000 | |
| TomoFix sm 3 | 1280 | 2200 | 2000 | > 60000 | 2,3 |
| TomoFix sm 4 | 1280 | 1750 | 1500 | > 60000 | 2,3 |
| TomoFix sm 5 | 1760 | 2000 | 2300 | > 120000 | 1,2 |
| Mean: | 1.4 | 1983 | 1933 | > 80000 | |
| SD ±: | 0.3 | 184 | 330 | 28,300 | |
| Contour Lock 3 | 2400 | 2100 | 4400 | > 200000 | 2 |
| Contour Lock 4 | 1760 | 2300 | 2400 | > 120000 | 2 |
| Contour Lock 5 | 2400 | 2700 | 2600 | > 200000 | 1,2,3 |
| Mean: | 2.2 | 2367 | 3133 | 173000 | |
| SD ±: | 0.4 | 250 | 900 | 37700 | |
| Activmotion 3 | 2240 | 2500 | 6300 | > 180000 | 2 |
| Activmotion 4 | 2240 | 2500 | 2900 | > 180000 | 2 |
| Activmotion 5 | 1600 | 2500 | 4750 | > 100000 | 2 |
| Activmotion 6 | 1600 | 3100 | 5100 | > 100000 | 2 |
| Mean: | 1.9 | 2650 | 4763 | 140000 | |
| SD ±: | 0.3 | 260 | 1219 | 40000 |
Fig. 9Average relative strength values: The TomoFix std. group has been taken as reference