| Literature DB >> 29165913 |
Abstract
The Canadian Organization of Medical Physicists (COMP) in close partnership with the Canadian Partnership for Quality Radiotherapy (CPQR) has developed a series of Technical Quality Control (TQC) guidelines for radiation treatment equipment. These guidelines outline the performance objectives that equipment should meet in order to ensure an acceptable level of radiation treatment quality. The TQC guidelines have been rigorously reviewed and field tested at various Canadian radiation treatment facilities. The development process enables rapid review and update to keep the guidelines current with changes in technology (the most updated version of this guideline can be found on the CPQR website). This particular TQC details recommended quality control for kilovoltage X Ray radiotherapy machines.Entities:
Keywords: kilovoltage; quality control; radiotherapy
Mesh:
Year: 2017 PMID: 29165913 PMCID: PMC5849838 DOI: 10.1002/acm2.12228
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Daily quality control tests
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
| Daily | |||
| DK1 | kVp and mA indicators | Functional | |
| DK2 | Beam‐off at key‐off test | Functional | |
| DK3 | Beam interrupt | Functional | |
| DK4 | Backup timer/monitor unit channel check | 1% | 2% |
| DK5 | Dosimetric test: output check | 2% | 4% |
Notes on daily tests
| DK1 | Functional check of kVp and mA indicators |
| DK2 | Functional check of beam‐off at key‐off |
| DK3 | Functional check of the beam interrupt button (radiation stop and restart when the radiation on button is pressed again) |
| DK4 | Quantitative verification of correct operation of backup timer or backup monitor unit |
| DK5 | Quantitative dosimetric test: output reproducibility test at the chosen energies and filter combinations. If the output is stable with time, this test can be done weekly (so replaced by WK4 according to Table |
Weekly quality control tests
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
| Weekly | |||
| WK1 | Couch movement and brakes | Functional | |
| WK2 | Unit motions and motion stops | Functional | |
| WK3 | Interlocks for added filters/kVp‐filter choice | Functional | |
| WK4 | Dosimetric test: output check | 2% | 4% |
Notes on weekly tests
| WK1 | Functional check of couch motion and brakes (where applicable) |
| WK2 | Functional check of unit motions and motion stops |
| WK3 | Functional check of interlocks for added filters, correct placement of filters, and the matching of filters with kVp value |
| WK4 | Quantitative dosimetric test: output reproducibility test for all energies. This test can be limited to the energies used clinically on the condition that any additional energies cannot be chosen at any time (by removing the corresponding filters from the treatment room, for example) |
Monthly quality control tests
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
| Monthly | |||
| MK1 | Mechanical stability and safety | Functional | |
| MK2 | Cone/filters integrity and cone indicators | Functional | |
| MK3 | Physical distance indicators | 2 mm | 3 mm |
| MK4 | Accuracy of head tilt and rotation readouts | 1° | 1.5° |
| MK5 | Light/x ray field coincidence | 2 mm | 3 mm |
| MK6 | Light field size | 2 mm | 3 mm |
| MK7 | X ray field size indicator | 2 mm | 3 mm |
| MK8 | X ray field uniformity | 5% | 8% |
| MK9 | Output verification and reproducibility with head tilt and rotation | 2% | 4% |
| MK10 | Timer accuracy verification | 1% | 2% |
| MK11 | Dose rate output constancy | 2% | 4% |
| MK12 | Records | Complete | |
Notes on monthly tests
| MK1 | Verification that the unit and accessories are firmly anchored and may be used without endangering patients or staff |
| MK2 | Verification of the integrity of the filters and cones and cone indicators |
| MK3 | Verification of the optical and/or mechanical distance indicator if the unit is equipped with one |
| MK4 | Verification of the angle readouts |
| MK5 | Performance parameters refer to agreement at each edge. This test does not apply to all machine designs |
| MK6 | Geometric test to verify the light field sizes (where applicable) |
| MK7 | Confirmation of radiation field size when a variable collimation system is provided. At least two field sizes must be checked |
| MK8 | Using a film, the flatness and symmetry of the X ray beam must be assessed for the largest cone |
| MK9 | Quantitative dosimetric test: output reproducibility test at all energies with varying head tilt and rotation |
| MK10 | If the unit is equipped with a timer, its accuracy must be checked against a stop watch over a range of doses of 10−1000 cGy |
| MK11 | Should be checked for all beam qualities for MU‐based systems |
| MK12 | Documentation relating to the daily quality control checks, preventive maintenance, service calls and subsequent checks must be complete, legible, and the operator identified |
Annual quality control tests
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
| Annual | |||
| AK1 | Reference dosimetry | 1% | 2% |
| AK2 | Timer and end‐effect error | Characterize | ±0.05 min |
| AK3 | Output linearity | n/a | 1% |
| AK4 | Output reproducibility | 2% | 3% |
| AK5 | Output error associated with beam interrupt | 2% | 4% |
| AK6 | Beam quality | 10% | 15% |
| AK7 | Alignment of focal spots | 0.5 mm | 1 mm |
| AK8 | Focal spot size | Reproducible | |
| AK9 | Percentage depth dose verification | Characterize and document | |
| AK10 | Profiles verification | Characterize and document | |
| AK11 | Independent quality control review | Complete | |
Notes on annual tests
| AK1 | Using a high‐quality dosimetry system calibrated against the local secondary standard, all beams and cones in use are recalibrated |
| AK2 | Timer and end‐effect error measurement may be performed in conjunction with AK3 |
| AK3 | Output linearity measurement for a clinically used filter/cone combination at a standard SSD and a dose range of 10−1000 cGy |
| AK4 | Output reproducibility verification for a clinically used filter/cone combination. These measurements should be repeated at typical tilt and head rotation used for treatments |
| AK5 | Output error when the beam is interrupted during the irradiation for a clinically used filter/cone combination |
| AK6 | The HVL of any clinically used beams is measured. The HVLs measured in millimeters of Al or millimeters of Cu as appropriate are compared with the values obtained at commissioning. These tolerances acknowledge measurement uncertainty |
| AK7 | Focal spot—quantitative measurement, assessed relative to acceptance test value where applicable |
| AK8 | Using a pin hole or resolution tool |
| AK9 | Verification of percentage depth dose measurements for all used filter/cone combinations against baseline |
| AK10 | Verification of inplane and crossplane profiles at different depths for all used filter/cone combinations against baseline |
| AK11 | To ensure redundancy and adequate monitoring, a second qualified medical physicist must independently verify the implementation, analysis, and interpretation of the quality control tests at least annually |