| Literature DB >> 20717089 |
Mehran Goharian1, Ian Nygren, Mauro Tambasco, Jose Eduardo Villarreal Barajas.
Abstract
During the recommissioning of a Philips RT-250 kilovoltage X-ray unit, unexpected output variations with tube head rotation (cross-plane) and tube head tilt (in-plane) were observed. The measured output showed an increase of up to 7.3% relative to the neutral position (0? in-plane and 0? cross-plane) over the possible range of angles of in-plane rotation for 75 kVp (half-value layer, HVL = 1.84 mm Al). A less pronounced but noticeable output change (with respect to the neutral position) was observed for cross-plane rotation reaching 2% for the 225 kVp beam (HVL = 0.90 mm Cu). This output variation was observed while manually adjusting the current to maintain constancy according to the current meter gauge. In order to address the observed output dependence with head orientation, the dose rate monitor chamber of the kilovoltage unit was calibrated to monitor the beam output in real time. The dose rate was manually adjusted to maintain a constant dose rate (in r/min) as displayed on the r/min gauge. This approach resulted in maintaining beam output for the 75 kVp and the 225 kVp beams within +/- 2% for the in-plane angle variation and +/- 0.5% for the cross-plane angle variation. A daily output check that includes ion chamber-based measurements at the neutral position and an in-plane angle of 45? has been implemented using the constant dose rate approach to monitor the stability of the X-ray beams. As a result of the output variations with in/cross-plane rotation, the quality control (QC) procedures that are typically used for clinical setup have been modified to test the stability of the beams under the non-neutral positioning of the X-ray tube.Entities:
Mesh:
Year: 2010 PMID: 20717089 PMCID: PMC5720435 DOI: 10.1120/jacmp.v11i3.3216
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1X‐ray tube head tilt (in‐plane) (a) and head rotation (cross‐plane) (b) orientations.
Figure 2The ion chamber and holder were used for output consistency check
Figure 3X‐ray beam output variation with head tilt angles for 75 kVp (a) and 225 kVp (b) using current meter to adjust output.
Figure 4X‐ray beam output variation with head rotation angles at zero head tilt for 75 kVp (a) and 225 kVp (b) using current meter to adjust output.
Figure 5X‐ray beam output variation with head tilt angles for 75 kVp (a) and 225 kVp (b) using dose rate meter to adjust output.
Figure 6X‐ray beam output variation with head rotation angles for three different head tilt angles for 75 kVp (a) and 225 kVp (b) using dose rate meter to adjust output (dose rate maintained at 40 r/min and 100 r/min for 75 kVp and 225 kVp, respectively).