| Literature DB >> 8309449 |
Abstract
A commercial megavoltage constancy meter was tested for routine checks of the x-ray output of an orthovoltage therapy unit. A practical method of using the meter for both output and energy checks was developed and is presented as part of an orthovoltage quality assurance program.Mesh:
Year: 1993 PMID: 8309449 DOI: 10.1118/1.596962
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071