Endoscopy enables minimally invasive procedures in many medical fields, such as urology. However, current endoscopes are normally cable-driven, which limits their dexterity and makes them hard to miniaturize. Indeed, current urological endoscopes have an outer diameter of about 3 mm and still only possess one bending degree-of-freedom. In this article, we report a novel wireless actuation mechanism that increases the dexterity and that permits the miniaturization of a urological endoscope. The novel actuator consists of thin active surfaces that can be readily attached to any device and are wirelessly powered by ultrasound. The surfaces consist of two-dimensional arrays of microbubbles, which oscillate under ultrasound excitation and thereby generate an acoustic streaming force. Bubbles of different sizes are addressed by their unique resonance frequency, thus multiple degrees-of-freedom can readily be incorporated. Two active miniaturized devices (with a side length of around 1 mm) are demonstrated: a miniaturized mechanical arm that realizes two degrees-of-freedom, and a flexible endoscope prototype equipped with a camera at the tip. With the flexible endoscope, an active endoscopic examination is successfully performed in a rabbit bladder. The results show the potential medical applicability of surface actuators wirelessly powered by ultrasound penetrating through biological tissues.
Endoscopy enables minimally invasive procedures in many medical fields, such as urology. However, current endoscopes are normally cable-driven, which limits their dexterity and makes them hard to miniaturize. Indeed, current urological endoscopes have an outer diameter of about 3 mm and still only possess one bending degree-of-freedom. In this article, we report a novel wireless actuation mechanism that increases the dexterity and that permits the miniaturization of a urological endoscope. The novel actuator consists of thin active surfaces that can be readily attached to any device and are wirelessly powered by ultrasound. The surfaces consist of two-dimensional arrays of microbubbles, which oscillate under ultrasound excitation and thereby generate an acoustic streaming force. Bubbles of different sizes are addressed by their unique resonance frequency, thus multiple degrees-of-freedom can readily be incorporated. Two active miniaturized devices (with a side length of around 1 mm) are demonstrated: a miniaturized mechanical arm that realizes two degrees-of-freedom, and a flexible endoscope prototype equipped with a camera at the tip. With the flexible endoscope, an active endoscopic examination is successfully performed in a rabbit bladder. The results show the potential medical applicability of surface actuators wirelessly powered by ultrasound penetrating through biological tissues.
Entities:
Keywords:
acoustic streaming; active surface; endoscopy; ultrasound; wireless actuator
Flexible endoscopy is an important diagnostic
tool in many medical fields including urology. Its major applications
include the detection of hematuria, the diagnosis of bladder carcinoma
and other pathologies, as well as the visualization of the upper urinary
tract. Bladder cancer is the most common malignancy of the urinary
tract,[1] with annually 74 000 new
cases in the United States and an estimated 16 000 deaths.[2] Cystoscopic controls are necessary not only for
the initial diagnosis of bladder cancer but also for follow-up examinations.
Patients afflicted with bladder cancer undergo mostly repeated endoscopies
of the urinary bladder using flexible devices. Compared to rigid instruments,
the tip of the flexible endoscope benefits the procedure due to larger
coverage areas and reduced trauma. Unlike larger flexible endoscopes,
such as colonoscopes, current flexible thin endoscopes (overall diameter
∼3 mm), as are for instance used in urology, only have one
bending section near the tip. Apart from the overall movements of
the entire endoscope (push/pull and rotate by the operator), only
one bending degree-of-freedom (DoF) can be realized in the tip itself.
As they are cable-driven, they are hard to be miniaturized, for example,
with a sheath diameter of 8.7 Fr. (2.9 mm diameter).[3] Furthermore, two cables are required to drive one rotational
DoF and adding more DoFs or more bending sections will result in an
even larger device diameter, as is the case for some colonoscopes,[4] which are then so large that they will no longer
be suitable for urological endoscopic applications.These disadvantages
motivate the following technical developments: (1) Miniaturization,
to reduce the diameter of the endoscope; (2) Multiple DoFs, to achieve
two bending degrees-of-freedom of the endoscope tip and more bending
sections; (3) Wireless actuation, to remove the constraint of the
cable. The overall goal is to achieve more space coverage with minimal
damage to the human tissue and minimal auxiliary medical measures,
e.g., avoid general anesthesia. The ultimate solution is a wireless
microrobot that can actively move through a lumen or swim through
biological fluids[5,6] to perform medical inspection
or treatment. In the quest to realize active endoscopes, different
actuation schemes have been investigated, such as shape memory alloy
actuators,[7] sliding concentric tubes,[8,9] and pneumatic/hydraulic actuators.[10] These
tethered approaches offer large actuation force and precise control
of the position; however, their structures are often complicated and
they are rigid and difficult to be miniaturized. Wireless approaches,
such as magnetics, have been also applied, e.g., to steer magnetic
tips of catheters[11] or untethered capsule
endoscopes.[12−14] However, large magnetic fields and/or gradients might
be required to achieve the forces needed for operation, which results
in the need for cumbersome and expensive dedicated equipment. It is
necessary to develop a new kind of actuator that has a simple structure
and is powered wirelessly and that therefore can be miniaturized and
integrated more efficiently.Ultrasound has been widely used
in urology for imaging, tissue ablation, lithotripsy, and drug delivery,
and it has also been proposed as a means to transfer power wirelessly.[15,16] It has recently been shown that ultrasound can also be shaped to
form complex three-dimensional patterns and this provides additional
opportunities for medical ultrasound.[17] However, the common approaches convert the mechanical vibrations
to electricity using the piezoelectric effect and then utilize the
electricity to power a device. This conversion process adds additional
complexity and suffers from low efficiency. Recently, we developed
an actuator that directly converts ultrasound power into mechanical
work via acoustic streaming from a surface array of microbubbles.[18,19] Acoustic streaming is a steady fluid flow driven by acoustically
driven oscillations, which can be achieved at the transducer surface
or at the air–liquid interface.[20,21] It is well
known that oscillating bubbles in fluids can cause resonant oscillations
and thus acoustic streaming[22] and microfluidic
pumps and mixers have been developed on the basis of this phenomenon.[23] It was shown by Dijkink et al. that oscillating
bubbles in a tube can propel or power a rotary motor on the centimeter
scale,[24] and microswimmers propelled by
one or a few bubbles have been reported by several groups.[25,26] However, the powering of a miniaturized medical instrument based
on acoustic active surfaces has not yet been reported.In this
article, we report the application of our wireless acoustic-surface
actuators to a miniaturized, multi-DoF, ultrasound-powered endoscope
prototype. As illustrated in the schematic of Figure , the device is actuated by acoustic streaming,
which is generated from an active surface consisting of a two-dimensional
(2D) array of microbubbles. Bubbles of different sizes are addressed
with unique ultrasound frequency; thus, multiple DoFs are achieved.
Two miniaturized devices with side length around 1 mm were developed,
i.e., a two DoFs miniaturized mechanical arm (mini arm) and a flexible
endoscope equipped with a camera. The endoscope is inserted into a
rabbit bladder through the tool channel of a rigid cystoscope, and
is successfully used for a cystoscopic examination.
Figure 1
Schematic of the wireless
actuated miniaturized cystoscope. (a) The miniaturized cystoscope
performs bladder imaging with two rotational DoFs and one translational
DoF. (b) An enlarged view of the cystoscope tip. The active surface,
which is powered wirelessly by ultrasound, rotates the first section
relative to the second section. (c) An enlarged view of the active
surface, which consists of an array of microbubbles. The oscillation
of the interface between air and liquid causes liquid streaming, generates
a propulsion force in the opposite direction and actuates the device.
Schematic of the wireless
actuated miniaturized cystoscope. (a) The miniaturized cystoscope
performs bladder imaging with two rotational DoFs and one translational
DoF. (b) An enlarged view of the cystoscope tip. The active surface,
which is powered wirelessly by ultrasound, rotates the first section
relative to the second section. (c) An enlarged view of the active
surface, which consists of an array of microbubbles. The oscillation
of the interface between air and liquid causes liquid streaming, generates
a propulsion force in the opposite direction and actuates the device.
Methods
Fabrication
of the Active Surface
The active surface consists of a regular
array of cylindrical microcavities with a carefully chosen size of
the cavity. Photolithography was used to fabricate the surface as
reported previously.[18] In short, negative
photoresist (SU-8 2000, MicroChem, MA) was spin-coated on a silicon
wafer (WS-650 spin coater, Laurel, PA), exposed on a mask aligner
(MJB4, SUSS MicroTec, Garching, Germany) with a chromium mask (Delta
Mask, Enschede, Netherlands), developed, and hard-baked. Then, the
wafer was diced by an automatic dicing saw (DAD 321, DISCO, Tokyo,
Japan) to a rectangular array.When a dried active surface is
immersed under water, air is trapped in the cavities as bubbles due
to the small size of the cavity and the hydrophobicity of the SU-8
photoresist (∼90° contact angle[27]). The diameter of the bubble is the same as that of the cavity and
the length is normally shorter than that of the cavity, as shown in
the inset of Figure . The size of the bubble determines the resonance frequency of a
single bubble. This corresponds to the frequency at which the maximal
acoustic streaming of the active surface occurs, as reported in our
previous study.[18] Sodium dodecyl sulfate
(SDS, 2.4 g/L) is added to the solution as a surfactant to lower the
surface tension of water from 73 to 38 mN/m.[28] Our previous study shows both theoretically and experimentally that
reducing the surface tension increases the oscillation amplitude of
the air–liquid interface, which substantially increases the
acoustic streaming force (up to ∼3.5 times).[28]
Figure 2
Wirelessly actuated endoscope powered by ultrasonic active surface.
(a) A picture comparing a commercial flexible urological endoscope
2.9 mm in diameter (left), with the mini arm (middle) and flexible
endoscope prototype (right) developed in this work. (b) An enlarged
microscopic image of the red-dashed box in (a) shows the active surface
under water, which consists of an array of microbubbles with controllable
sizes. The inset is a zoom-in of the bubble in the white dashed box,
which is trapped in a cavity 30 μm in diameter and 120 μm
in height.
Wirelessly actuated endoscope powered by ultrasonic active surface.
(a) A picture comparing a commercial flexible urological endoscope
2.9 mm in diameter (left), with the mini arm (middle) and flexible
endoscope prototype (right) developed in this work. (b) An enlarged
microscopic image of the red-dashed box in (a) shows the active surface
under water, which consists of an array of microbubbles with controllable
sizes. The inset is a zoom-in of the bubble in the white dashed box,
which is trapped in a cavity 30 μm in diameter and 120 μm
in height.
Assembly of the Miniaturized
Endoscope
The mini arm has a dimension of 1 mm × 1 mm
in cross section and is 5 mm long. It contains three sections with
two hinges, which are mounted orthogonal to each other (Figure a middle) such that the arm
can achieve two DoF. The sections and the pins were made out of aluminum
and brass, respectively. After assembly of the mini arm, two active
surfaces (1 mm × 4 mm) with different microcavity sizes, i.e.,
actuator 1: an array of 9600 “small” cavities, each
55 μm in depth, 10 μm in diameter, 10 μm in spacing;
actuator 2: an array of 1520 “large” cavities, each
120 μm in depth, 30 μm in diameter, 20 μm in spacing,
were attached to two orthogonal surfaces of the first section using
cyanoacrylate (Superflex Gel, UHU, Bühl, Germany).The
flexible arm was assembled and integrated with a Naneye 2D CMOS sensor
(Awaiba, Yverdon-Les-Bains, Switzerland), which is 1 mm × 1 mm
in cross section and 1.85 mm long. The flat wire connected to the
Naneye is a flexible ribbon of four wires with a total dimension of
0.72 mm width and 0.188 mm height, which also serves as an elastic
sheet to provide the restoring force in the flexible arm design. An
active surface (1 mm × 9 mm, with “large” cavities)
is attached to one side of the flat wire using cyanoacrylate.
Ultrasound
Excitation Setup
A customized ultrasound excitation setup
was made to test the endoscope. It is a glass tank of 50 × 50
× 50 mm3, with one piezoelectric ceramic disk (Ø50
mm × 3 mm, PIC 181, PI Ceramic, Lederhose, Germany) attached
to one side wall by epoxy glue. To increase the propulsive force and
stabilize the bubbles, 2.4 g/L SDS surfactant was added in all fluids.[28] The driving signal was a sinusoidal wave generated
from a function generator (33220A, Agilent technologies, CA), and
amplified 30 times with a customized power amplifier and transformer.
Two frequencies, i.e., 51.3 and 115.3 kHz, were used to address the
arrays of “large” and “small” bubbles,
respectively. For the wireless actuation of the endoscope in the rabbit
bladder, an ultrasonic bath (2510E-DTH, Bransonic, CT) was used for
higher power ultrasound of 100 W at 42 kHz ± 6%.
Endoscope Actuation
in the Rabbit Bladder
The cystoscopy with the ultrasound-powered
endoscope is demonstrated in an isolated rabbit bladder (Oryctolagus cuniculusforma domestica). The organs were obtained from a local slaughterhouse. The careful
representation of the urinary bladder in the retroperitonal convolute
was followed by the preparation including ligation of both ureters
and the removal of the bladder including the proximal urethra. To
observe the movement of the endoscope inside the bladder, a rigid
8 Fr. cystoscope (Richard Wolf, Knittlingen, Germany) was used. The
bladder was fixed to the cystoscope by a 4-0 Vicryl suture (Ethicon,
Somerville) and filled with ∼100 mL SDS aqueous solution. The
developed flexible endoscope was introduced by a 3 Fr. flexible gripper
(Richard Wolf) through the tool channel of the cystoscope into the
bladder, as shown in detail in Figure S-1. The illumination and visualization were achieved by a 5132 Auto
LP HLight Module and a 5520 1CCD Endocam Module (Richard Wolf) connected
to the rigid cystoscope. The video from the cystoscope and the flexible
miniaturized endoscope are recorded and shown as Videos in the Supporting Information.
Results and Discussion
Wireless
Actuation of the Mini Arm
As shown in Figure , the mini arm consists of three sections
connected in series and can achieve two rotational DoF. Two active
surfaces with different microcavity sizes are attached to two orthogonal
surfaces on the first section. The mini arm is mechanically held under
water by tweezers. The mini arm contains no electrical components
or connections. Instead, the mini arm is powered completely wirelessly
by ultrasound. The ultrasound that is used to excite the arm is generated
by a piezoelectric transducer, which can also be seen in Figure , behind the arm
at a distance of ∼30 mm.
Figure 3
Wireless actuation of the mini arm. (a)
Sketch of the miniaturized arm made of aluminum and assembled through
brass hinges. Two actuators 1 and 2 are attached orthogonally to the
first section, which actuates the arm with two rotational DoF. (b)
Sequential snapshots of a video. At 9–11 s, the first section
is actuated with ultrasound at 115.3 kHz, and at 20–24 s, the
second section is actuated with ultrasound at 51.3 kHz. See also Video S-1 in the Supporting Information.
Wireless actuation of the mini arm. (a)
Sketch of the miniaturized arm made of aluminum and assembled through
brass hinges. Two actuators 1 and 2 are attached orthogonally to the
first section, which actuates the arm with two rotational DoF. (b)
Sequential snapshots of a video. At 9–11 s, the first section
is actuated with ultrasound at 115.3 kHz, and at 20–24 s, the
second section is actuated with ultrasound at 51.3 kHz. See also Video S-1 in the Supporting Information.Once the mini arm is put under
water, air is trapped in the microcavities of the active surfaces
as individual bubbles due to the surface hydrophobicity and small
diameter of the cavities (see Methods). The
size of the microbubbles is determined by the cavity size, which has
a unique different acoustic resonant frequency. A detailed physical
study of the frequency response of the bubbles is reported in a previous
work.[18] In short, the resonant frequencies
of micron-sized bubbles lie in the range of 50–200 kHz and
the smaller the bubble size the higher is its resonant frequency.
For the actuation of the mini arm, an ultrasonic field of 115.3 kHz
was first applied and it only excited the “small” bubbles
on the actuator 1, which caused ultrasonic streaming and thus propelled
the first section into rotating to the left (image at 11 s in Figure b). When the ultrasound
was switched off, the arm returned to the original position due to
gravity. Similarly, when an ultrasound of 51.3 kHz was applied, only
actuator 2 was activated; thus, the second section of the mini arm
rotated out of the plane (image at 24 s in Figure b). See also Video S-1 in the Supporting Information for the real-time movement of
the mini arm. The reproducibility is demonstrated by switching each
frequency on and off two times (see Video).Increasing the DoF in the tendon-driven endoscope requires
adding at least two more driving wires; thus, the endoscope structure
is more rigid and complicated and the diameter is larger. The wireless-driven
active surface reported here solves this problem. The bandwidth response
of each kind of actuator is about 10 kHz, thus in principle 18 independent
actuators can be achieved within the 20–200 kHz frequency range.
Whereas for cystoscopy three DoFs (two rotational DoFs ω1, ω2, and one translational DoF, see Figure ) are normally enough;
for more complicated intraluminal anatomy, such as the upper urinary
tract, more DoFs will be beneficial. If the hinges are arranged in
orthogonal directions, three arm sections will realize three independent
rotational DoFs. Because the back and forth motions of one DoF requires
two actuators, six actuators in total will be needed. The redundant
DoFs could be useful for applications other than bending the endoscope,
for example, to open and close a gripper. More importantly, as all
actuators are driven wirelessly, no additional wire is required for
more DoFs, and the overall diameter of the endoscope can remain very
small. It is noteworthy that each active surface can be 100–200
μm thin; therefore, equipping the endoscope tip with two DoFs
(four active surfaces) will increase its diameter by much less than
half a millimeter. Moreover, even more active surfaces can be added
at different positions along the articulated tip without making the
diameter any larger. Notably, the streaming actuation force is independent
of the direction of the incoming ultrasound, which can be applied
from the most convenient position. It is also interesting to point
out that the adoption of active surfaces is a scalable approach. The
mini arm is designed to be 1 mm × 1 mm in cross section, because
the smallest commercial camera has a cross section of 1 mm ×
1 mm. In other words, 1 mm is far from the size limit for the active
surface and the principle illustrated here is still valid for even
smaller devices in the micrometer range.
Wireless Actuation of the
Flexible Endoscope
Although the actuation scheme is wireless,
there is currently no wireless camera small enough to fit through
the tool channel of a cystoscope. Therefore a tethered camera such
as the Naneye 2D sensor was used. The design of the flexible arm takes
advantage of the connecting wires for the camera and adopts it as
an elastic sheet to provide the restoring force when the ultrasonic
transducer is turned off. Figure a,b show the movement of the flexible endoscope, which
consists of three parts: a miniaturized camera at the tip pointing
forward, a flat wire connecting the camera and also acting as a recoil
spring, and an active surface. The effective bending length of the
wire is 22.5 mm, and the streaming active surface is attached next
to the camera causing the deflection in one direction. The deflection
is driven by an ultrasound transducer ∼30 mm away, and is controlled
by the driving voltage which shows a linear relationship to the output
power in this range (Figure c). The total length of the flexible arm is about 33 mm, which
is realistic for cystoscopy in a human bladder. A normal human bladder
capacity is ∼400 mL,[29] so considered
as a sphere, a filled human bladder has an average radius of approximately
46 mm. This also means the miniaturized camera can be quite close
to the bladder inner surface, thus providing a clearer image of the
bladder wall. When the wire is thinner or softer, or the ultrasound
power is increased, it is possible to realize even larger deflections.
Figure 4
Wireless
actuation of the flexible endoscope prototype. (a) Sketch of the flexible
arm. The active surface is attached to the flat wire, which serves
as an elastic sheet to provide the restoring force. (b) A superimposed
image showing the deflection of the flexible arm is controlled by
the driving voltage of the ultrasonic transducer. See also Video S-2 in the Supporting Information. (c)
The deflection exhibits a linear relationship with the driving voltage,
and reaches a maximum of 60° in the current setup.
Wireless
actuation of the flexible endoscope prototype. (a) Sketch of the flexible
arm. The active surface is attached to the flat wire, which serves
as an elastic sheet to provide the restoring force. (b) A superimposed
image showing the deflection of the flexible arm is controlled by
the driving voltage of the ultrasonic transducer. See also Video S-2 in the Supporting Information. (c)
The deflection exhibits a linear relationship with the driving voltage,
and reaches a maximum of 60° in the current setup.The flexible arm has a simpler design compared
with the above mentioned mini arm, i.e., no mechanical hinges are
required and it can be fabricated more easily. There are ∼3400
bubbles on an area of 1 mm × 9 mm surface, each generating a
force up to ∼120 nN (bubbles 120 μm in depth, 30 μm
in diameter).[30] Therefore, the estimated
output force generated by the active surface is ∼0.4 mN. The
force is smaller than those typically generated by tethered approaches,
but large enough to effectively steer a miniaturized camera, as demonstrated
in the experiments. As the force is small and the flexible endoscope
prototype is compliant, the actuator–camera setup should be
safe for the urinary tract wall inspection and cause no damage to
the tissue.[31] Furthermore, such systems
might be potentially introduced via urologic catheters such as ureteric
catheters replacing conventional endoscopic inspection of the urinary
tract by rigid, semirigid and flexible devices.
Safety and
Stability of the Ultrasonic Streaming Actuation
Safety is
one of the most important concerns of medical devices, especially
for invasive procedures. The safety of the ultrasound-powered acoustic
streaming is discussed here. It is difficult to compare the values
directly with current ultrasonic medical devices because in diagnostic
ultrasound, a higher frequency (2 MHz) coupled with pulse mode operation
is used. The ultrasound used here is continuous wave with a frequency
between 50 and 200 kHz. Two parameters are often used to represent
the ultrasound intensity, i.e., the spatial peak-temporal average
intensity ISPTA and the spatial peak-pulse
average intensity ISPPA. A typical pressure
amplitude in a B-scan is 1.68 MPa,[32] and
the recommended intensity limits set by the U.S. Food and Drug Administration
(FDA) for use in a peripheral vessel are 720 mW/cm2 for ISPTA and 190 W/cm2 for ISPPA.[33] A continuous wave means
the duty cycle is 100%; therefore, ISPTA = ISPPA = ISP in this case. Currently, there is no safety guideline for continuous
ultrasound, however, a previous study shows that the tissue injury
intensity threshold of animal tissues for continuous ultrasound is
470 W/cm2.[34]Measured
by a needle hydrophone, the pressure amplitudes of the ultrasound
used to power the mini arm and the flexible endoscope are both around
100–200 kPa, which is much lower than 1.68 MPa in the B-scan.
The corresponding ultrasound intensity can be calculated by the following
equation[35]where ISP is the spatial-peak ultrasound intensity, P is the pressure amplitude, ρ is the density of water
(1000 kg/m3), and c is the speed of sound
in water, which is ∼1520 m/s at 37 °C. If we take the
highest pressure P = 200 kPa, then ISP = 2.6 W/cm2. This value is higher than the
FDA recommended value for ISPTA 720 mW/cm2, which may lead to some thermal effects;[32] but 2 orders of magnitude lower than the FDA recommended
value for ISPPA 190 W/cm2 and
the reported tissue injury intensity threshold 470 W/cm2, so it is low enough that it is unlikely that the ultrasound used
here will cause lasting mechanical damage or induce cavitation.[32] However, further studies on animal experiments
are required to establish more carefully if and when one may expect
thermal effects of continuous wave ultrasound on tissue.Another
issue is the stability of the air bubbles. It is observed in the experiments
that the air bubbles trapped in the active surface are stable under
ultrasonic excitation for at least 30 min, and the streaming actuation
can be excited repeatedly during this period. Under the microscope,
we observed that the air bubbles will gradually dissolve in water
and the bubble size will become smaller, thus the driving frequency
needs to be adjusted and be increased accordingly. But the dissolution
process is relatively slow compared to the operating time of a cystoscope
(depending on the examiner normally 5–15 min for an examination).
The dimension of the bubble only changes by about 0.8%/min (see Figure S-3 in the Supporting Information). The
dimension change is slow enough that no frequency adjustments are
needed during the typical operation time of the active surface (∼5–15
min for a cystoscopy examination). Changing air to other inert gases
that are insoluble in water, such as SF6, a commonly used
contrast agent for ultrasound imaging,[36] will result in even longer operating time. SDS is added to the solution
as a surfactant to lower the surface tension, increase the oscillation
amplitude of the air–liquid interface, and thus increase the
acoustic streaming force (up to ∼3.5 times).[28] SDS was chosen as it is readily available in the laboratory,
but it is not supposed to be used in clinical applications. Many biocompatible
surfactants, such as amino acid-based surfactants,[37] can instead be used as safe substitutes with comparable
effects on the acoustic actuation. Particular attention to safety
and biocompatibility will be given in the future translational studies.
Miniaturized Endoscope for Cystoscopy in a Rabbit Bladder
The flexible endoscope was tested in a rabbit bladder ex vivo. The
rigid cystoscope is used to hold and refill the bladder (Figure ), and visualize
the movement of the flexible endoscope inside the bladder. However,
in a real application the cystoscope is not necessary, as the flexible
endoscope can be inserted into the bladder through the urethra alone
(as illustrated in Figure ) or through a passive catheter. In clinical practice, cystoscopy
often starts with an empty bladder without urine; then, to obtain
a clear view of the bladder wall, the operator fills the bladder with
saline solution through the irrigation channel of the cystoscope.
As there is no irrigation channel in our miniaturized endoscope prototype
due to its small size, another thin catheter could be used to fill
the bladder with saline solution before the examination by the acoustically
actuated endoscope, which requires an aqueous environment to operate.
Figure 5
Setup
for cystoscopy in rabbit bladder. (a) A schematic shows the flexible
arm entering the bladder through the tool channel of a rigid cystoscope,
which is fixed to urethra. The bending of the flexible arm is powered
by the streaming of the active surface, and its movement is viewed
via the cystoscope. (b) A picture of the bladder fixed with the endoscope.
(c) An enlarged picture of the flexible arm coming out of the tool
channel of a cystoscope.
Setup
for cystoscopy in rabbit bladder. (a) A schematic shows the flexible
arm entering the bladder through the tool channel of a rigid cystoscope,
which is fixed to urethra. The bending of the flexible arm is powered
by the streaming of the active surface, and its movement is viewed
via the cystoscope. (b) A picture of the bladder fixed with the endoscope.
(c) An enlarged picture of the flexible arm coming out of the tool
channel of a cystoscope.The miniaturized endoscope entered the bladder through the
working channel of the endoscope; at first it was stationary pointing
downward (image at 0 s in Figure a). When the ultrasound was applied, streaming from
the active surface occurred and the flexible arm or the endoscope
is bent to the right (3 s in Figure a). The streamline is weakly visualized in the image,
but clearer in the Video S-3 in the Supporting Information. When the ultrasound was turned off, the flexible
arm returned to the original position due to the restoring force of
the flat wire (6 s in Figure a). We found that the bubbles on the active surface were stable
during the whole period of examination (∼10 min) and contact
with the bladder wall did not release the bubbles. From the miniaturized
camera on the tip of the endoscope, another video was recorded (snapshots
in Figure b). When
the flexible arm was pointing downwards, the structure of the fat
tissue was seen as a marker and can be directly compared to the cystoscopic
images in Figure a.
When the flexible arm is bent, a blood vessel on the bladder side
wall can be visualized. This vessel can be seen from the outside of
the bladder in Figure b, but because of its position it could not be visualized with the
cystoscope in Figure a. In the Video S-3 in the Supporting Information, the two videos from the cystoscope (left) and miniaturized camera
(right) are synchronized and to demonstrate the reproducibility of
the imaging procedure the ultrasound is switched on and off two times.
In the video, a certain level of shake of the camera is observed,
which can be a problem in endoscopy. The reason is that the ultrasound
power is not very stable in the experiment and the device is operated
in an open-loop control. The image stability can be improved by correcting
for fluctuations in ultrasound power, e.g., with feedback from an
in situ pressure sensor, or by sensing the bending angle of the endoscope.
The miniaturized camera has a high frame rate over 60 frames/s, which
generates redundant frames; thus, the video can be digitally processed
with antishaking algorithms to minimize the shaking effect. The current
image quality is still not comparable to the current endoscope, but
with the fast development of microelectronic devices, we believe,
the imaging quality of the miniaturized camera can be further improved
and the miniaturized wireless device will be the trend in the future
endoscopy.
Figure 6
Cystoscopy in a rabbit bladder. (a) Sequential snapshots of a video
recorded by the rigid cystoscope. At 0 s, the flexible arm is stationary
and pointing downward. At 3 s, ultrasound is turned on and the arm
is bent to the right due to the streaming of the active surface. At
6 s, ultrasound is turned off, the flexible arm returns to the original
position. (b) Sequential snapshots of a video recorded by the Naneye
camera on the flexible arm. At 0 s, the flexible arm is pointing downwards,
and the shape of the fat tissue is seen as a marker comparing with
the cystoscopic images above. At 3 s, when the flexible arm bends,
a blood vessel on the bladder side wall is visualized, which can be
seen from outside the bladder in Figure b but cannot be seen from the cystoscopic
images in (a). See also Video S-3 in the Supporting Information.
Cystoscopy in a rabbit bladder. (a) Sequential snapshots of a video
recorded by the rigid cystoscope. At 0 s, the flexible arm is stationary
and pointing downward. At 3 s, ultrasound is turned on and the arm
is bent to the right due to the streaming of the active surface. At
6 s, ultrasound is turned off, the flexible arm returns to the original
position. (b) Sequential snapshots of a video recorded by the Naneye
camera on the flexible arm. At 0 s, the flexible arm is pointing downwards,
and the shape of the fat tissue is seen as a marker comparing with
the cystoscopic images above. At 3 s, when the flexible arm bends,
a blood vessel on the bladder side wall is visualized, which can be
seen from outside the bladder in Figure b but cannot be seen from the cystoscopic
images in (a). See also Video S-3 in the Supporting Information.Although because of smaller
dimensions, a rabbit’s bladder was studied as a proof-of-concept,
this work reports a general method to actuate a miniaturized endoscope
in low viscosity bodily fluids. The miniaturization of the instrument
leads to a significant reduction of surgical trauma and thus reducing
unwanted complications for the patient. It not only can be used as
a cystoscope, but may also be applied to the upper urinary tract and
the kidney. Alternatively, the flexible endoscope can be tested in
a realistic phantom of the kidney and urinary tract.[38] It can potentially simplify preparatory steps of surgery
procedures; for example, a double-J catheter might be required to
dilate the ureter for a period of time prior to the surgical manipulation
on the upper urinary tract via an endoscope. With the actuation scheme
developed herein, an instrument with less than 1 mm diameter could
potentially avoid unnecessary interventions. Consequently the severeness
of urological endoscopic surgeries could be decreased, resulting in
a shortened hospital stay of the patients and thereby contribute to
a cost reduction.
Conclusions
In this article, we
report the first thin-layer surface actuators that permit the wireless
power and control of an endoscope. The wireless actuation is based
on two-dimensional arrays of microbubbles that resonate with an external
ultrasound, giving rise to frequency-selective, directional acoustic
streaming. The endoscope has an overall side dimension of 1 mm and
fits through the tool channel of a cystoscope. A complete cystoscopy
was successfully performed by this endoscope in a rabbit bladder.
The novel actuation scheme is general and can therefore be further
miniaturized. The method demonstrated herein thus holds potential
as a powerful actuator to steer miniaturized devices for minimally
invasive in vivo inspections.
Authors: Marko Babjuk; Andreas Böhle; Maximilian Burger; Otakar Capoun; Daniel Cohen; Eva M Compérat; Virginia Hernández; Eero Kaasinen; Joan Palou; Morgan Rouprêt; Bas W G van Rhijn; Shahrokh F Shariat; Viktor Soukup; Richard J Sylvester; Richard Zigeuner Journal: Eur Urol Date: 2016-06-17 Impact factor: 20.096
Authors: Yak-Nam Wang; Julianna C Simon; Bryan W Cunitz; Frank L Starr; Marla Paun; Denny H Liggitt; Andrew P Evan; James A McAteer; Ziyue Liu; Barbrina Dunmire; Michael R Bailey Journal: J Ther Ultrasound Date: 2014-03-31
Authors: Tian Qiu; Tung-Chun Lee; Andrew G Mark; Konstantin I Morozov; Raphael Münster; Otto Mierka; Stefan Turek; Alexander M Leshansky; Peer Fischer Journal: Nat Commun Date: 2014-11-04 Impact factor: 14.919