Literature DB >> 19963490

Wireless steering mechanism with magnetic actuation for an endoscopic capsule.

A Menciassi1, P Valdastri, C Quaglia, E Buselli, P Dario.   

Abstract

This paper illustrates the design, development and testing of a miniature mechanism to be integrated in endoscopic capsules for precise steering capabilities (Magnetic Internal Mechanism, MIM). The mechanism consists of an electromagnetic motor connected to a couple of small permanent magnets and immersed in a static magnetic field produced by an external permanent magnet or a by an electromagnetic coil. The overall steering capsule, integrating the magnetic steering mechanism and the vision system is 15.6 mm in diameter, 48 mm in length, 14.4 g in weight and can be oriented with an accuracy of 0.01 degrees . As regards system scalability, the capsule size could be reduced down to 11 mm in diameter by optimizing some mechanical components. On the other hand, the magnets size cannot be reduced because the magnetic link between internal and external magnets at typical operation distances (about 15 mm) would be weak.

Entities:  

Mesh:

Year:  2009        PMID: 19963490     DOI: 10.1109/IEMBS.2009.5332426

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X


  4 in total

Review 1.  New vision in video capsule endoscopy: current status and future directions.

Authors:  Laurel R Fisher; William L Hasler
Journal:  Nat Rev Gastroenterol Hepatol       Date:  2012-05-08       Impact factor: 46.802

Review 2.  Capsule endoscopy in patients refusing conventional endoscopy.

Authors:  Javier Romero-Vázquez; Federico Argüelles-Arias; Josefa Maria García-Montes; Ángel Caunedo-Álvarez; Francisco Javier Pellicer-Bautista; Juan Manuel Herrerías-Gutiérrez
Journal:  World J Gastroenterol       Date:  2014-06-21       Impact factor: 5.742

3.  Wireless Acoustic-Surface Actuators for Miniaturized Endoscopes.

Authors:  Tian Qiu; Fabian Adams; Stefano Palagi; Kai Melde; Andrew Mark; Ulrich Wetterauer; Arkadiusz Miernik; Peer Fischer
Journal:  ACS Appl Mater Interfaces       Date:  2017-11-28       Impact factor: 9.229

4.  Evaluation of the tip-bending response in clinically used endoscopes.

Authors:  Esther D Rozeboom; Rob Reilink; Matthijs P Schwartz; Paul Fockens; Ivo A M J Broeders
Journal:  Endosc Int Open       Date:  2016-03-30
  4 in total