Literature DB >> 29124339

Accuracy of screw fixation using the O-arm® and StealthStation® navigation system for unstable pelvic ring fractures.

Jun Takeba1, Kensuke Umakoshi2, Satoshi Kikuchi3, Hironori Matsumoto2, Suguru Annen3, Naoki Moriyama3, Yuki Nakabayashi2, Norio Sato3, Mayuki Aibiki2.   

Abstract

PURPOSE: Screw fixation for unstable pelvic ring fractures is generally performed using the C-arm. However, some studies reported erroneous piercing with screws, nerve injuries, and vessel injuries. Recent studies have reported the efficacy of screw fixations using navigation systems. The purpose of this retrospective study was to investigate the accuracy of screw fixation using the O-arm® imaging system and StealthStation® navigation system for unstable pelvic ring fractures.
METHODS: The participants were 10 patients with unstable pelvic ring fractures, who underwent screw fixations using the O-arm StealthStation navigation system (nine cases with iliosacral screw and one case with lateral compression screw). We investigated operation duration, bleeding during operation, the presence of complications during operation, and the presence of cortical bone perforation by the screws based on postoperative CT scan images. We also measured the difference in screw tip positions between intraoperative navigation screen shot images and postoperative CT scan images.
RESULTS: The average operation duration was 71 min, average bleeding was 12 ml, and there were no nerve or vessel injuries during the operation. There was no cortical bone perforation by the screws. The average difference between intraoperative navigation images and postoperative CT images was 2.5 ± 0.9 mm, for all 18 screws used in this study.
CONCLUSION: Our results suggest that the O-arm StealthStation navigation system provides accurate screw fixation for unstable pelvic ring fractures.

Entities:  

Keywords:  Navigation; O-arm; Osteosynthesis; Pelvic ring fracture; Screw

Mesh:

Year:  2017        PMID: 29124339     DOI: 10.1007/s00590-017-2075-9

Source DB:  PubMed          Journal:  Eur J Orthop Surg Traumatol        ISSN: 1633-8065


  30 in total

1.  Accuracy of computer-assisted iliosacral screw placement using a hybrid operating room.

Authors:  P H Richter; F Gebhard; C Dehner; A Scola
Journal:  Injury       Date:  2015-12-12       Impact factor: 2.586

2.  Transpedicular wedge osteotomy for treatment of kyphosis after L1 fracture using intraoperative, full rotation, three-dimensional image (O-arm)-based navigation: a case report.

Authors:  Yijie Liu; Xuefeng Li; Han Sun; Huilin Yang; Weimin Jiang
Journal:  Int J Clin Exp Med       Date:  2015-10-15

3.  Removal of an iliosacral screw entrapping the L5 nerve root after failed posterior pelvic ring fixation: a case report.

Authors:  Yoram A Weil; Markku T Nousiainen; David L Helfet
Journal:  J Orthop Trauma       Date:  2007-07       Impact factor: 2.512

4.  Quantitative anatomic basis of antegrade lag screw placement in posterior column of acetabulum.

Authors:  Wei-Dong Mu; Xian-Quan Wang; Tang-Hong Jia; Dong-Sheng Zhou; Ai-Xin Cheng
Journal:  Arch Orthop Trauma Surg       Date:  2009-02-17       Impact factor: 3.067

5.  Three-dimensional image-guided placement of S2 alar screws to adjunct or salvage lumbosacral fixation.

Authors:  Eric W Nottmeier; Stephen M Pirris; Sarah Balseiro; Douglas Fenton
Journal:  Spine J       Date:  2010-07       Impact factor: 4.166

Review 6.  Malposition and revision rates of different imaging modalities for percutaneous iliosacral screw fixation following pelvic fractures: a systematic review and meta-analysis.

Authors:  Jörn Zwingmann; Oliver Hauschild; Gerrit Bode; Norbert P Südkamp; Hagen Schmal
Journal:  Arch Orthop Trauma Surg       Date:  2013-06-08       Impact factor: 3.067

7.  Surgical resection of osteoid osteoma and osteoblastoma of the spine.

Authors:  Muayad Kadhim; Odion Binitie; Patrick O'Toole; Emmanouil Grigoriou; Camila B De Mattos; John P Dormans
Journal:  J Pediatr Orthop B       Date:  2017-07       Impact factor: 1.041

8.  Placement of iliosacral screws using 3D image-guided (O-Arm) technology and Stealth Navigation: comparison with traditional fluoroscopy.

Authors:  A A Theologis; S Burch; M Pekmezci
Journal:  Bone Joint J       Date:  2016-05       Impact factor: 5.082

9.  Percutaneous iliosacral screw fixation in unstable pelvic ring lesions: the interest of O-ARM CT-guided navigation.

Authors:  C Coste; Y Asloum; P S Marcheix; P Dijoux; J L Charissoux; C Mabit
Journal:  Orthop Traumatol Surg Res       Date:  2013-04-29       Impact factor: 2.256

10.  Pedicle violation and Navigational errors in pedicle screw insertion using the intraoperative O-arm: A preliminary report.

Authors:  Jacob E Mathew; Kelvin Mok; Benoit Goulet
Journal:  Int J Spine Surg       Date:  2013-12-01
View more
  6 in total

1.  A novel patient-specific three-dimensional-printed external template to guide iliosacral screw insertion: a retrospective study.

Authors:  Fan Yang; Sheng Yao; Kai-Fang Chen; Feng-Zhao Zhu; Ze-Kang Xiong; Yan-Hui Ji; Ting-Fang Sun; Xiao-Dong Guo
Journal:  BMC Musculoskelet Disord       Date:  2018-11-13       Impact factor: 2.362

2.  Combined 3D Printed Template to Guide Iliosacral Screw Insertion for Sacral Fracture and Dislocation: A Retrospective Analysis.

Authors:  Chao Wu; Jia-Yan Deng; Tao Li; Lun Tan; De-Chao Yuan
Journal:  Orthop Surg       Date:  2020-02       Impact factor: 2.071

3.  Preoperative marking of a submillimeter metastatic pulmonary tumor using a mobile computed tomography scan with a navigation system: A case report.

Authors:  Asuka Uebayashi; Ryo Fujikawa; Yoshifumi Arai; Toru Nakamura; Kazuhito Funai
Journal:  Int J Surg Case Rep       Date:  2021-01-18

4.  Clinical Analysis of Single and Double Sacroiliac Screws in the Treatment of Tile C1 Pelvic Fracture.

Authors:  Hong-Li Deng; Dong-Yang Li; Yu-Xuan Cong; Bin-Fei Zhang; Jin-Lai Lei; Hu Wang; Peng-Fei Wang; Yan Zhuang
Journal:  Biomed Res Int       Date:  2022-01-04       Impact factor: 3.411

5.  Volar Locking Plate Fixation for Distal Radius Fractures by Intraoperative Computed Tomographic-Guided Navigation.

Authors:  Akira Kawabata; Yusuke Sogabe; Yukiko Morimoto; Kiyohito Takamatsu
Journal:  J Hand Surg Glob Online       Date:  2020-09-08

6.  Radiological and Clinical Differences between Tinavi Orthopedic Robot and O-Arm Navigation System in Thoracolumbar Screw Implantation for Reconstruction of Spinal Stability.

Authors:  Jinpeng Du; Lin Gao; Dageng Huang; Lequn Shan; Wentao Wang; Yong Fan; Dingjun Hao; Liang Yan
Journal:  Med Sci Monit       Date:  2020-09-12
  6 in total