Speech audiometry and data logging in CI patients : Implications for adequate test levels.

M Hey1, T Hocke2, P Ambrosch3.   

Abstract

BACKGROUND: As part of postoperative cochlear implant (CI) diagnostics, speech comprehension tests are performed to monitor audiological outcome. In recent years, a trend toward improved suprathreshold speech intelligibility in quiet and an extension of intelligibility to softer sounds has been observed. Parallel to audiometric data, analysis of the patients' acoustic environment can take place by means of data logging in modern CI systems.
OBJECTIVES: Which speech test levels reflect the individual listening environment in a relevant manner and how can these be reflected in a clinical audiometric setting? PATIENTS AND METHODS: In a retrospective analysis, data logs of 263 adult CI patients were evaluated for sound level and the listening situation (quiet, speech in quiet, noise, speech in noise, music, and wind). Additionally, monosyllabic word comprehension in quiet was analyzed in experienced CI users at presentation levels of 40-80 dB.
RESULTS: For the sound level in the acoustic environment of postlingually deafened adult CI users, data logging shows a maximum occurrence of speech signals in the range of 50-59 dB. This demonstrates the relevance of everyday speech comprehension at levels below 60 dB.
CONCLUSION: Individual optimization of speech intelligibility with a CI speech processor should not be performed in the range of 65-70 dB only, but also at lower levels. Measurements at 50 dB currently seem to be a useful addition.

Entities:  

Keywords:  Cochlear implant; Data logging; Presentation level; Speech audiometry; Speech discrimination test

Mesh:

Year:  2018        PMID: 29119199     DOI: 10.1007/s00106-017-0419-8

Source DB:  PubMed          Journal:  HNO        ISSN: 0017-6192            Impact factor:   1.284


  27 in total

Review 1.  [Device-based treatment of mixed hearing loss: An audiological comparison of current hearing systems].

Authors:  T Rahne; S K Plontke
Journal:  HNO       Date:  2016-02       Impact factor: 1.284

Review 2.  The multi-channel cochlear implant: multi-disciplinary development of electrical stimulation of the cochlea and the resulting clinical benefit.

Authors:  Graeme M Clark
Journal:  Hear Res       Date:  2014-08-24       Impact factor: 3.208

3.  [Improving speech comprehension using a new cochlear implant speech processor].

Authors:  J Müller-Deile; T Kortmann; U Hoppe; H Hessel; A Morsnowski
Journal:  HNO       Date:  2009-06       Impact factor: 1.284

Review 4.  Electric-acoustic stimulation of the auditory system: a review of the first decade.

Authors:  Christoph A von Ilberg; Uwe Baumann; Jan Kiefer; Jochen Tillein; Oliver F Adunka
Journal:  Audiol Neurootol       Date:  2011-05-24       Impact factor: 1.854

5.  [Speech perception with hearing aids in comparison to pure-tone hearing loss].

Authors:  U Hoppe; A Hast; T Hocke
Journal:  HNO       Date:  2014-06       Impact factor: 1.284

6.  [Evaluation of the perceptual equivalence of test lists in the Freiburg monosyllabic speech test].

Authors:  I Baljić; A Winkler; T Schmidt; I Holube
Journal:  HNO       Date:  2016-08       Impact factor: 1.284

7.  Hearing aids in the real world: use of multimemory and volume controls.

Authors:  Shilpi Banerjee
Journal:  J Am Acad Audiol       Date:  2011-06       Impact factor: 1.664

Review 8.  The Hybrid cochlear implant: a review.

Authors:  Erika A Woodson; Lina A J Reiss; Christopher W Turner; Kate Gfeller; Bruce J Gantz
Journal:  Adv Otorhinolaryngol       Date:  2009-11-25

9.  [Evaluation of the Freiburg monosyllabic speech test in background noise].

Authors:  J Löhler; B Akcicek; M Pilnik; K Saager-Post; S Dazert; S Biedron; J Oeken; D Mürbe; J Löbert; R Laszig; T Wesarg; C Langer; S Plontke; T Rahne; U Machate; R Noppeney; K Schultz; P Plinkert; S Hoth; M Praetorius; P Schlattmann; E F Meister; H W Pau; K Ehrt; R Hagen; W Shehata-Dieler; M Cebulla; L E Walther; A Ernst
Journal:  HNO       Date:  2013-07       Impact factor: 1.284

10.  Clinical evaluation of the Nucleus 6 cochlear implant system: performance improvements with SmartSound iQ.

Authors:  Stefan J Mauger; Chris D Warren; Michelle R Knight; Michael Goorevich; Esti Nel
Journal:  Int J Audiol       Date:  2014-08       Impact factor: 2.117

View more
  2 in total

1.  Cochlear implantation performance outcomes in patients over 80 years old.

Authors:  Helena Wichova; Dawna Mills; Sarah Beatty; Kevin Peng; Mia Miller
Journal:  Laryngoscope Investig Otolaryngol       Date:  2022-05-27

2.  Speech comprehension across multiple CI processor generations: Scene dependent signal processing.

Authors:  Matthias Hey; Britta Böhnke; Alexander Mewes; Patrick Munder; Stefan J Mauger; Thomas Hocke
Journal:  Laryngoscope Investig Otolaryngol       Date:  2021-06-15
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.