Maged Alnawaiseh1, Lisann Hömberg1, Nicole Eter1, Verena Prokosch2.
Abstract
PURPOSE: To compare the structure-function relationships between retinal nerve fiber layer thickness (RNFLT) and visual field defects measured either by standard automated perimetry (SAP) or by Pulsar perimetry (PP).Entities:
Year: 2017 PMID: 29119021 PMCID: PMC5651118 DOI: 10.1155/2017/8014294
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Figure 1Retinal nerve fiber layer sectors and Octopus visual field clusters. IN: inferonasal; IT: inferotemporal; SN: superonasal; ST: superotemporal; T: temporal; L: nasal. The map is based on the structure-function map developed by Garway-Heath and colleagues [15].
Characteristics of the study groups.
| Entire group ( | Group A ( | Group B ( |
| |
|---|---|---|---|---|
| Age | 44.92 ± 20.65 | 40.06 ± 20.22 | 58.30 ± 15.33 |
|
| Sex (m : w) | 102 : 161 | 73 : 120 | 29 : 41 | |
| RNFLT | 88.37 ± 15.39 | 95.83 ± 8.10 | 67.79 ± 11.45 |
|
| MD (SAP) | −2.26 ± 4.03 | −1.04 ± 2.18 | −5.65 ± 5.7 |
|
| PSD/sLV (SAP) | 2.89 ± 2.78 | 1.97 ± 0.97 | 5.42 ± 4.23 |
|
| MD (PP)∗ | 1.59 ± 3.58 | 0.51 ± 2.29 | 4.56 ± 4.67 |
|
| PSD/sLV (PP) | 2.22 ± 1.24 | 1.82 ± 0.72 | 3.33 ± 1.63 |
|
Bold: statistically significant differences between group A and group B. ∗Octopus instruments show defects with a positive MD sign.
Structure-function relationship between RNFL thickness, measured with SD-OCT, and Pulsar perimetry or standard automated perimetry for the entire study group.
| Entire group ( | Global thickness∗ | Temporal inferior thickness∗ | Temporal thickness | Temporal superior thickness∗ | Nasal superior thickness∗ | Nasal thickness∗ | Nasal inferior thickness∗ | |
|---|---|---|---|---|---|---|---|---|
| RNFL and Pulsar |
| 0.59 | 0.57 | 0.26 | 0.55 | 0.46 | 0.33 | 0.33 |
|
|
|
|
|
|
|
|
| |
| RNFL and SAP |
| 0.38 | 0.40 | 0.27 | 0.40 | 0.24 | 0.18 | 0.21 |
|
|
|
|
|
|
|
|
|
r Spe: Spearman's correlation coefficient. ∗Statistically significant difference between correlations for PP and SAP.
Structure-function relationship between RNFL thickness, measured with SD-OCT, and Pulsar perimetry or standard automated perimetry for the glaucoma group (A).
| Group A (70) | Global thickness∗ | Temporal inferior thickness∗ | Temporal thickness | Temporal superior thickness∗ | Nasal superior thickness∗ | Nasal thickness∗ | Nasal inferior thickness∗ | |
|---|---|---|---|---|---|---|---|---|
| RNFL and Pulsar |
| 0.56 | 0.73 | 0.43 | 0.72 | 0.29 | 0.08 | 0.26 |
|
|
|
|
|
|
| 0.506 |
| |
| RNFL and SAP |
| 0.60 | 0.75 | 0.49 | 0.74 | 0.39 | 0.08 | 0.34 |
|
|
|
|
|
|
| 0.515 |
|
r Spe: Spearman's correlation coefficient. ∗Statistically significant difference between correlations for PP and SAP.
Structure-function relationship between RNFL thickness, measured with SD-OCT, and Pulsar perimetry or standard automated perimetry for group B.
| Group B ( | Global thickness∗ | Temporal inferior thickness∗ | Temporal thickness | Temporal superior thickness∗ | Nasal superior thickness∗ | Nasal thickness∗ | Nasal inferior thickness∗ | |
|---|---|---|---|---|---|---|---|---|
| RNFL and Pulsar |
| 0.31 | 0.31 | −0.03 | 0.30 | 0.28 | 0.05 | −0.02 |
|
|
|
| 0.722 |
|
| 0.537 | 0.794 | |
| RNFL and SAP |
| 0.06 | 0.13 | 0.04 | 0.12 | 0.05 | −0.01 | −0.10 |
|
| 0.436 | 0.067 | 0.57 | 0.106 | 0.514 | 0.929 | 0.178 |
r Spe: Spearman's correlation coefficient. ∗Statistically significant difference between correlations for PP and SAP.