| Literature DB >> 29089820 |
Yan Xing1, Yuan Zhao1, Ning Guo2, Cun-Xue Pan1, Gulina Azati1, Yan-Wei Wang1, Wen-Ya Liu1.
Abstract
OBJECTIVE: Using a pulsating coronary artery phantom at high heart rate settings, we investigated the efficacy of a motion correction algorithm (MCA) to improve the image quality in dual-energy spectral coronary CT angiography (CCTA).Entities:
Keywords: Coronary arteries; Motion correction algorithm; Spectral imaging; Tomography, X-ray computed
Mesh:
Year: 2017 PMID: 29089820 PMCID: PMC5639153 DOI: 10.3348/kjr.2017.18.6.881
Source DB: PubMed Journal: Korean J Radiol ISSN: 1229-6929 Impact factor: 3.500
Fig. 1Pulsating coronary artery phantom and segmentations of vessels.
A. Pulsating coronary artery phantom. B. Volume rending images demonstrate segmentations of vessels.
Comparison of Image Interpretability and Image Score between CA and MCA at Monochromatic Images of 70 keV
| 60 bpm | 70 bp | 80 bpm | 90 bpm | 100 bpm | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| CA | MCA | CA | MCA | CA | MCA | CA | MCA | CA | MCA | ||||||
| Overall interpretability (%) | 97.2 | 100 | > 0.05 | 83.3 | 94.4 | > 0.05 | 77.8 | 86.1 | > 0.05 | 38.9 | 72.2 | < 0.05 | 33.3 | 63.9 | < 0.05 |
| Likert Scale | < 0.05 | < 0.05 | < 0.05 | < 0.001 | < 0.001 | ||||||||||
| 1. Non-interpretable | 1 | 0 | 6 | 2 | 8 | 5 | 22 | 10 | 24 | 13 | |||||
| 2. Sub-optimal | 4 | 2 | 7 | 6 | 10 | 5 | 5 | 6 | 10 | 11 | |||||
| 3. Mild motion | 28 | 29 | 19 | 23 | 17 | 25 | 9 | 20 | 2 | 12 | |||||
| 4. Excellent | 3 | 5 | 4 | 5 | 1 | 1 | 0 | 0 | 0 | 0 | |||||
bpm = beats/minute, CA = conventional algorithm, MCA = motion correction algorithm
Fig. 2Comparison of image score and 95% confidence interval (CI) between MCA and CA at heart rates of 70, 80, 90, and 100 bpm, at different monochromatic levels.
A. Comparison of image quality score (95% CI) between MCA and CA at heart rate of 70 bpm. B. Comparison of image quality score (95% CI) between MCA and CA at heart rate of 80 bpm. C. Comparison of image quality score (95% CI) between MCA and CA at heart rate of 90 bpm. D. Comparison of image quality score (95% CI) between MCA and CA at heart rate of 100 bpm. *0.05, †p < 0.01. bpm = beats/minute, CA = conventional algorithm, MCA = motion correction algorithm
Fig. 3Comparison of phantom axial images after CA and MCA at different keV levels at heart rate of 90 bpm.
Comparision of Image Interpretability between CA and MCA at Different Heart Rates and on Different Monochromatic Levels
| Heart Rates (bpm) | Monochromatic Level (keV) | CA (% [n/n]) | MCA (% [n/n]) | χ2 | |
|---|---|---|---|---|---|
| 70 | 50 | 91.7 (33/36) | 100 (36/36) | 0.158 | 0.691 |
| 60 | 88.9 (32/36) | 97.2 (35/36) | 1.934 | 0.164 | |
| 70 | 83.3 (30/36) | 94.4 (34/36) | 2.250 | 0.134 | |
| 80 | 69.4 (25/36) | 75.0 (27/36) | 0.277 | 0.599 | |
| 90 | 61.1 (22/36) | 66.7 (24/36) | 0.241 | 0.624 | |
| 80 | 50 | 88.9 (32/36) | 97.2 (35/36) | 2.022 | 0.155 |
| 60 | 83.3 (30/36) | 91.7 (33/36) | 1.143 | 0.285 | |
| 70 | 77.8 (28/36) | 86.1 (31/36) | 0.845 | 0.358 | |
| 80 | 72.2 (26/36) | 75.0 (27/36) | 0.071 | 0.789 | |
| 90 | 58.3 (21/36) | 52.8 (19/36) | 0.225 | 0.635 | |
| 90 | 50 | 77.8 (28/36) | 88.9 (32/36) | 1.600 | 0.206 |
| 60 | 55.6 (20/36) | 86.1 (31/36) | 8.134 | 0.004 | |
| 70 | 38.9 (14/36) | 72.2 (26/36) | 8.100 | 0.004 | |
| 80 | 38.9 (14/36) | 44.4 (16/36) | 0.229 | 0.633 | |
| 90 | 36.1 (13/36) | 41.7 (15/36) | 0.234 | 0.629 | |
| 100 | 50 | 69.4 (25/36) | 91.7 (33/36) | 5.675 | 0.017 |
| 60 | 55.6 (20/36) | 83.3 (30/36) | 6.545 | 0.011 | |
| 70 | 33.3 (12/36) | 63.9 (23/36) | 6.727 | 0.009 | |
| 80 | 8.3 (3/36) | 11.1 (4/36) | 0.158 | 0.691 | |
| 90 | 11.1 (4/36) | 8.3 (3/36) | 0.158 | 0.691 |