Quantitative imaging of peripheral trabecular bone microarchitecture using MDCT.

Cheng Chen1, Xiaoliu Zhang1, Junfeng Guo2,3, Dakai Jin1, Elena M Letuchy4, Trudy L Burns4, Steven M Levy4,5, Eric A Hoffman2,3, Punam K Saha1,2.   

Abstract

PURPOSE: Osteoporosis associated with reduced bone mineral density (BMD) and microarchitectural changes puts patients at an elevated risk of fracture. Modern multidetector row CT (MDCT) technology, producing high spatial resolution at increasingly lower dose radiation, is emerging as a viable modality for trabecular bone (Tb) imaging. Wide variation in CT scanners raises concerns of data uniformity in multisite and longitudinal studies. A comprehensive cadaveric study was performed to evaluate MDCT-derived Tb microarchitectural measures. A human pilot study was performed comparing continuity of Tb measures estimated from two MDCT scanners with significantly different image resolution features.
METHOD: Micro-CT imaging of cadaveric ankle specimens (n=25) was used to examine the validity of MDCT-derived Tb microarchitectural measures. Repeat scan reproducibility of MDCT-based Tb measures and their ability to predict mechanical properties were examined. To assess multiscanner data continuity of Tb measures, the distal tibias of 20 volunteers (age:26.2±4.5Y,10F) were scanned using the Siemens SOMATOM Definition Flash and the higher resolution Siemens SOMATOM Force scanners with an average 45-day time gap between scans. The correlation of Tb measures derived from the two scanners over 30% and 60% peel regions at the 4% to 8% of distal tibia was analyzed.
RESULTS: MDCT-based Tb measures characterizing bone network area density, plate-rod microarchitecture, and transverse trabeculae showed good correlations (r∈0.85,0.92) with the gold standard micro-CT-derived values of matching Tb measures. However, other MDCT-derived Tb measures characterizing trabecular thickness and separation, erosion index, and structure model index produced weak correlation (r<0.8) with their micro-CT-derived values. Most MDCT Tb measures were found repeatable (ICC∈0.94,0.98). The Tb plate-width measure showed a strong correlation (r = 0.89) with experimental yield stress, while the transverse trabecular measure produced the highest correlation (r = 0.81) with Young's modulus. The data continuity experiment showed that, despite significant differences in image resolution between two scanners (10% MTF along xy-plane and z-direction - Flash: 16.2 and 17.9 lp/cm; Force: 24.8 and 21.0 lp/cm), most Tb measures had high Pearson correlations (r > 0.95) between values estimated from the two scanners. Relatively lower correlation coefficients were observed for the bone network area density (r = 0.91) and Tb separation (r = 0.93) measures.
CONCLUSION: Most MDCT-derived Tb microarchitectural measures are reproducible and their values derived from two scanners strongly correlate with each other as well as with bone strength. This study has highlighted those MDCT-derived measures which show the greatest promise for characterization of bone network area density, plate-rod and transverse trabecular distributions with a good correlation (r ≥ 0.85) compared with their micro-CT-derived values. At the same time, other measures representing trabecular thickness and separation, erosion index, and structure model index produced weak correlations (r < 0.8) with their micro-CT-derived values, failing to accurately portray the projected trabecular microarchitectural features. Strong correlations of Tb measures estimated from two scanners suggest that image data from different scanners can be used successfully in multisite and longitudinal studies with linear calibration required for some measures. In summary, modern MDCT scanners are suitable for effective quantitative imaging of peripheral Tb microarchitecture if care is taken to focus on appropriate quantitative metrics.
© 2017 American Association of Physicists in Medicine.

Entities:  

Keywords:  bone mechanics; microarchitecture; modulation transfer function; multidetector row CT; trabecular bone

Mesh:

Year:  2017        PMID: 29064579      PMCID: PMC6345384          DOI: 10.1002/mp.12632

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  16 in total

Review 1.  Update on Imaging-Based Measurement of Bone Mineral Density and Quality.

Authors:  Thomas M Link; Galateia Kazakia
Journal:  Curr Rheumatol Rep       Date:  2020-04-09       Impact factor: 4.592

Review 2.  A Review of Experimentally Informed Micromechanical Modeling of Nanoporous Metals: From Structural Descriptors to Predictive Structure-Property Relationships.

Authors:  Claudia Richert; Norbert Huber
Journal:  Materials (Basel)       Date:  2020-07-24       Impact factor: 3.623

Review 3.  X-ray-based quantitative osteoporosis imaging at the spine.

Authors:  M T Löffler; N Sollmann; K Mei; A Valentinitsch; P B Noël; J S Kirschke; T Baum
Journal:  Osteoporos Int       Date:  2019-11-14       Impact factor: 4.507

4.  Generation of Vertebra Micro-CT-like Image from MDCT: A Deep-Learning-Based Image Enhancement Approach.

Authors:  Dan Jin; Han Zheng; Qingqing Zhao; Chunjie Wang; Mengze Zhang; Huishu Yuan
Journal:  Tomography       Date:  2021-11-12

5.  3-D X-Ray-Induced Acoustic Computed Tomography With a Spherical Array: A Simulation Study on Bone Imaging.

Authors:  Yang Li; Pratik Samant; Siqi Wang; A Behrooz; Dengwang Li; Liangzhong Xiang
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2020-04-06       Impact factor: 2.725

6.  CT-Based Characterization of Transverse and Longitudinal Trabeculae and Its Applications.

Authors:  Xiaoliu Zhang; Elena M Letuchy; Steven M Levy; James C Torner; Punam K Saha
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2020-02-28

7.  Deep Learning Based High-Resolution Reconstruction of Trabecular Bone Microstructures from Low-Resolution CT Scans using GAN-CIRCLE.

Authors:  Indranil Guha; Syed Ahmed Nadeem; Chenyu You; Xiaoliu Zhang; Steven M Levy; Ge Wang; James C Torner; Punam K Saha
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2020-02-28

8.  Application of a Novel Ultra-High Resolution Multi-Detector CT in Quantitative Imaging of Trabecular Microstructure.

Authors:  G Shi; S Subramanian; Q Cao; S Demehri; J H Siewerdsen; W Zbijewski
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2020-03-05

9.  Effects of fluoride intake on cortical and trabecular bone microstructure at early adulthood using multi-row detector computed tomography (MDCT).

Authors:  Punam K Saha; Reem Reda Oweis; Xiaoliu Zhang; Elena Letuchy; Julie M Eichenberger-Gilmore; Trudy L Burns; John J Warren; Kathleen F Janz; James C Torner; Linda G Snetselaar; Steven M Levy
Journal:  Bone       Date:  2021-02-10       Impact factor: 4.398

10.  The reproducibility of measuring trabecular bone parameters using a commercially available high-resolution magnetic resonance imaging approach: A pilot study.

Authors:  Sarah L West; Chamith S Rajapakse; Tammy Rayner; Rhiannon Miller; Michelle A Slinger; Greg D Wells
Journal:  Bone Rep       Date:  2018-04-26
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.