B Alper Gultekin1, Ali Sirali2, Pinar Gultekin3, Selim Ersanli1.
Abstract
PURPOSE: The aim of this study was to evaluate the stability during healing and before loading of implants placed at two different supracrestal levels according to their collar texture.Entities:
Keywords: Dental implant; bone–implant interface; implant stability quotient; osseointegration; supracrestal level
Year: 2016 PMID: 28955572 PMCID: PMC5573511 DOI: 10.17096/jiufd.96003
Source DB: PubMed Journal: J Istanb Univ Fac Dent ISSN: 2149-2352
Figure 1.Representative images of the laser-microtextured group (group L) and machined collar group (group M) implants used in this study. The L group implant was placed 1 mm above the crestal bone, while the M group implant was placed 0.3 mm above the crestal bone. The implant stability was measured from the buccal and mesial sections of the transducer in the resonance frequency analysis system.
Figure 2.(a and b) Placement of an implant with a machined collar (M group) 0.3 mm above the crestal bone (c) Implant stability is measured using resonance frequency analysis (d) Healing abutment was attached to the implant for transmucosal healing.
Figure 3.(a and b) Placement of an implant with lasermicrotextured collar (L group) 1 mm above the supracrestal bone and measurement of implant stability using resonance frequency analysis (c and d) Healing abutment was attached to the implant for transmucosl healing.
Comparison of stability between L and M groups according to implant stability quotient (ISQ) values from baseline to 12 weeks after placement. L group: implants with laser-microtextured collar placed 1 mm above the crestal bone M group: implants with a machined collar placed 0.3 mm above the crestal bone.
| L group (n = 47) | M group (n = 56) | ap | |||
|---|---|---|---|---|---|
| Median (Q1, Q3) | Median (Q1, Q3) | ||||
| Baseline (R-0) | 70 (67, 73) | 72 (70.5, 76) | 0.006** | ||
| 1 week (R-1) | 70 (67, 73) | 71.5 (70, 76) | 0.031* | ||
| 4 weeks (R-4) | 68 (65, 72) | 69.5 (67, 73) | 0.212 | ||
| 8 weeks (R-8) | 71 (68, 74) | 72 (70, 75.5) | 0.434 | ||
| 12 weeks (R-12) | 73 (68, 76) | 73 (70.5, 76.5) | 0.516 | ||
| d0.001** | d0.001** | ||||
| Median (Q1, Q3) | cp | Median (Q1, Q3) | cp | ap | |
| 0-1weeks | 0 (-1, 1) | 0.538 | 0 (-1, 0) | 0.001** | 0.028* |
| 0-4 weeks | ?2 (-3, -1) | 0.001** | -3 (-4,-2) | 0.001** | 0.001** |
| 0-8 weeks | 0 (-1, 2) | 0.065 | -1 (-2, 0) | 0.001** | 0.001** |
| 0-12 weeks | 2 (0, 4) | 0.001** | 0 (-1, 2) | 0.480 | 0.001** |
| 1-4 weeks | ?2 (-3, -1) | 0.001** | -3 (-4, -2) | 0.001** | 0.003** |
| 1-8 weeks | 0 (-1, 2) | 0.153 | -1 (-1.5, 0) | 0.010* | 0.003** |
| 1-12 weeks | 1 (0, 4) | 0.001** | 1 (-1, 2) | 0.036* | 0.033* |
| 4-8 weeks | 2 (1, 3) | 0.001** | 2 (1, 3) | 0.001** | 0.544 |
| 4-12 weeks | 3 (2, 5) | 0.001** | 3 (2, 5) | 0.001** | 0.635 |
| 8-12 weeks | 1 (0, 3) | 0.001** | 1 (0, 2) | 0.001** | 0.918 |
aMann–Whitney U Test, cWilcoxon Signed-Rank Test, dFriedman Test *p<0.05, **p<0.01
Figure 4.Comparison of stability between L group and M group implants according to implant stability quotient (ISQ) values from baseline to 12 weeks after placement. L group: implants with laser-microtextured collar placed 1 mm above the crestal bone M group: implants with a machined collar placed 0.3 mm above the crestal bone.
Relationship between implant stability quotient (ISQ) values and the site of placement.
| Mandible (n=65) | Maxilla (n=38) | ap | |
|---|---|---|---|
| Median (Q1. Q3) | Median (Q1. Q3) | ||
| Baseline (R-0) | 73 (71.76) | 68 (67.72) | 0.001** |
| 1 week (R-1) | 72 (70.76) | 69 (67.71) | 0.001** |
| 4 weeks (R-4) | 71 (68.74) | 66 (63.69) | 0.001** |
| 8 weeks (R-8) | 73 (70.76) | 69 (65.72) | 0.001** |
| 12 weeks (R-12) | 73 (71.77) | 71.5 (66.75) | 0.012* |
aMann–Whitney U Test, *p<0.05, **p<0.01
Relationship between implant stability quotient (ISQ) values and implant diameter.
| Diameter | ||||
|---|---|---|---|---|
| 3.8 mm (n= 77) | 4.6 mm (n = 26) | |||
| Median (Q1. Q3) | Median (Q1. Q3) | ap | ||
| Baseline (R-0) | 72 (68. 72) | 79 (75. 82) | 0.001** | |
| 1 week (R-1) | 71 (67. 72) | 79 (75. 82) | 0.001** | |
| 4 weeks (R-4) | 69 (65. 70) | 75.5 (73. 79) | 0.001** | |
| 8 weeks (R-8) | 71 (67. 73) | 79.5 (75. 82) | 0.001** | |
| 12 weeks (R-12) | 72 (68. 74) | 79.5 (75. 82) | 0.001** | |
| 3.8 mm (n = 50) | 4.6 mm (n = 22) | |||
| Length | Median (Q1. Q3) | Median (Q1. Q3) | ap | |
| 10.5 mm | Baseline (R-0) | 71 (68. 72) | 79 (76. 82) | 0.001** |
| 1 week (R-1) | 70 (67. 72) | 80 (75. 82) | 0.001** | |
| 4 weeks (R-4) | 68 (64. 70) | 77 (73. 79) | 0.001** | |
| 8 weeks (R-8) | 70 (66. 73) | 79.5 (75. 81) | 0.001** | |
| 12 weeks (R-12) | 71 (68. 73) | 79.5 (75. 82) | 0.001** |
aMann–Whitney U Test, **p<0.01
Relationship between implant stability quotient (ISQ) values and implant length.
| Length | 10.5 mm (n= 72) | 12 mm (n = 31) | |
|---|---|---|---|
| Median (Q1. Q3) | Median (Q1. Q3) | ap | |
| Baseline (R-0) | 72 (69.76) | 70 (68.73) | 0.109* |
| 1 week (R-1) | 72 (69.75.5) | 70 (69.73) | 0.192* |
| 4 weeks (R-4) | 69 (66.5.73.5) | 68 (66.71) | 0.256* |
| 8 weeks (R-8) | 72 (68.5.75) | 71 (68.74) | 0.542* |
| 12 weeks (R-12) | 73 (69.5.76) | 74 (70.76) | 0.719* |
| 10.5 mm (n= 50) | 12 mm (n = 27) | ||
| Diameter 3.8 mm | Median (Q1. Q3) | Median (Q1. Q3) | ap |
| Baseline (R-0) | 71 (68. 72) | 70 (68. 72) | 0.859* |
| 1 week (R-1) | 70 (67. 72) | 70 (68. 72) | 0.629* |
| 4 weeks (R-4) | 68 (64. 70) | 68 (66. 70) | 0.476* |
| 8 weeks (R-8) | 70 (66. 73) | 71 (68. 74) | 0.264* |
| 12 weeks (R-12) | 71 (68. 73) | 73 (70. 75) | 0.047** |
aMann–Whitney U Test, *p>0.05 **p<0.01