| Literature DB >> 28955394 |
Semi Ozturk1, Ahmed Seyfettin Gurbuz2, Suleyman Cagan Efe3, Mehmet Fatih Yılmaz4, Cevat Kırma4.
Abstract
BACKGROUND AND OBJECTIVES: Balloon sizing remains the main technique for determining occluder device size for atrial septal defects (ASDs). New evidence has proposed that accurate estimation of device size could be possible without using the balloon technique. Operators have predicted the amount of possible enlargement depending on their experiences. Thus, selection criteria have mostly relied on personal observations and experiences. The objective of this study was to determine the relationship between age, sex, defect size, and deployed device size based on the balloon technique. SUBJECTS AND METHODS: Sixty-six patients who underwent percutaneous ASD closure with a Cardi-O-Fix occluder between 2011 and 2012 were retrospectively evaluated. Patients whose maximum defect size and device size were available were included. Enlargement amount (EA) (device size-defect size) and enlargement ratio (ER) (EA/defect size) were calculated. The relationship between these 2 calculations and age, sex, and defect size were analyzed.Entities:
Keywords: Echocardiography; Heart defects, congenital; Heart septal defects, atrial
Year: 2017 PMID: 28955394 PMCID: PMC5614952 DOI: 10.4070/kcj.2017.0073
Source DB: PubMed Journal: Korean Circ J ISSN: 1738-5520 Impact factor: 3.243
Baseline characteristics
| Characteristics | Male (n=19) | Female (n=47) | p value |
|---|---|---|---|
| Age (years) | 37.4±15.0 | 34.5±14.1 | 0.480 |
| BMI (kg/m2) | 25.5±5.5 | 25.1±4.2 | 0.850 |
| Defect size (mm) | 14 (7, 30) | 13 (6, 25) | 0.280 |
| EA (mm) | 4 (0, 15) | 5 (0, 13) | 0.800 |
| ER (%) | 30.0 (0, 115) | 33.3 (0, 125) | 0.430 |
Values are presented as mean±SD or median (min, max).
BMI = body mass index; EA = enlargement amount; ER = enlargement ratio; SD = standard deviation.
Figure 1ER-age distribution.
ER = enlargement ratio.
ASD characteristics of young and older patients
| Characteristics | Age >40 (n=25) | Age ≤40 (n=41) | p value |
|---|---|---|---|
| Defect size (mm) | 15 (8, 25) | 13 (6, 30) | 0.120 |
| EA (mm) | 4 (0, 9) | 5 (0, 15) | 0.110 |
| ER (%) | 27.3 (0, 88) | 38.5 (0, 125) | 0.030 |
Values are presented as median (min, max).
ASD = atrial septal defect; EA = enlargement amount; ER = enlargement ratio.
Figure 2Device and ASD size* correlation.
ASD = atrial septal defect; TEE = transesophageal echocardiography.
*TEE measurement.