Literature DB >> 16142805

Transcatheter atrial septal defect closure: modified balloon sizing technique to avoid overstretching the defect and oversizing the Amplatzer septal occluder.

Karina M Carlson1, Henri Justino, Richard E O'Brien, V Vivian Dimas, Glenn T Leonard, Ricardo H Pignatelli, Charles E Mullins, E O'Brian Smith, Ronald G Grifka.   

Abstract

The objective of this study was to evaluate a new technique of sizing atrial septal defects (ASDs) for transcatheter device closure. ASD closure using the Amplatzer septal occluder (ASO) device is commonly performed. Complications, including arrhythmias, pericardial effusions, and perforations, may be related to oversizing ASDs and choosing larger devices. Two methods were used to size ASDs using a compliant balloon. In some patients, the balloon was inflated until a waist was visible [(+)waist]; in others, only until no shunting was demonstrable by echocardiogram [echo; (-)waist]. The device was selected and implanted using standard procedure and echo guidance. One hundred seventeen patients underwent secundum ASD closure with an ASO device. There were 43 patients in the (-)waist group and 74 in the (+)waist group. All devices were implanted successfully. The initial echo ASD diameter was larger in the (-)waist group compared to the (+)waist group (P = 0.01). There was a smaller difference between the initial echo and balloon-sized ASD diameters in the (-)waist group (P < 0.02). ASO device size implanted (in mm greater than echo ASD diameter) was smaller in the (-)waist group (P < 0.01). There were 0/43 complications in the (-)waist group and 5/74 in the (+)waist group. The complete closure rate was the same in both groups. Sizing an ASD by inflating a compliant balloon just until shunting is eliminated, and not until a waist is visible, results in less overstretching of the ASD and selection of a smaller ASO device, achieving similar closure rates and potentially fewer complications. Copyright 2005 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2005        PMID: 16142805     DOI: 10.1002/ccd.20443

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  17 in total

Review 1.  Atrial septum defect closure device in a beating heart, from the perspective of a researcher in artificial organs.

Authors:  Yasuko Tomizawa
Journal:  J Artif Organs       Date:  2012-06-23       Impact factor: 1.731

2.  Risk factors and prognosis of atrioventricular block after atrial septum defect closure using the Amplatzer device.

Authors:  Yibin Wang; Yimin Hua; Li Li; Xiaoqin Wang; Lina Qiao; Xiaoqing Shi; Jiping Hua; Yi Qu; Dezhi Mu
Journal:  Pediatr Cardiol       Date:  2013-10-30       Impact factor: 1.655

3.  How to Size ASDs for Percutaneous Closure.

Authors:  Isabelle Boon; Katrien Vertongen; Bernard P Paelinck; Laurent Demulier; An Van Berendoncks; Catherine De Maeyer; Fabienne Marchau; Joseph Panzer; Kristof Vandekerckhove; Daniel De Wolf
Journal:  Pediatr Cardiol       Date:  2017-09-27       Impact factor: 1.655

4.  Transcatheter closure of secundum atrial septal defects in adults with the Amplatzer septal occluder: intermediate and long-term results.

Authors:  Christian Spies; Ines Timmermanns; Rainer Schräder
Journal:  Clin Res Cardiol       Date:  2007-02-26       Impact factor: 5.460

5.  Appropriate selection of echocardiographic guidance for transcatheter atrial septal defect closure.

Authors:  Michiyo Yamano; Tetsuhiro Yamano; Takeshi Nakamura; Kan Zen; Hirokazu Shiraishi; Takeshi Shirayama; Satoaki Matoba
Journal:  Int J Cardiovasc Imaging       Date:  2020-02-10       Impact factor: 2.357

Review 6.  Echocardiography guidance of atrial septal defect closure.

Authors:  Bushra S Rana
Journal:  J Thorac Dis       Date:  2018-09       Impact factor: 2.895

7.  Accuracy of Transthoracic Echocardiography in Assessing Retro-aortic Rim prior to Device Closure of Atrial Septal Defects.

Authors:  Michael L O'Byrne; Andrew C Glatz; David J Goldberg; Russell Shinohara; Yoav Dori; Jonathan J Rome; Matthew J Gillespie
Journal:  Congenit Heart Dis       Date:  2014-09-16       Impact factor: 2.007

8.  Transcatheter Atrial Septal Defect Closure under Transthorasic Echocardiography in Children.

Authors:  Keyhan Sayadpour Zanjani; Aliakbar Zeinaloo; Elaheh Malekan-Rad; Abdolrazagh Kiani; Mohammad-Mehdi Bagheri
Journal:  Iran J Pediatr       Date:  2011-12       Impact factor: 0.364

9.  Comparison of the response of the right ventricle with endovascular occlusion and surgical closure in adults with atrial septal defect one year after intervention.

Authors:  Niloufar Samiei; Fariba Bayat; Masoud Moradi; Mozhgan Parsaei; Seyedeh Zahra Ojaghi Haghighi; Ahmad Mohebbi; Negar Hamzepour; Fereidoon Noohi
Journal:  Clin Med Insights Cardiol       Date:  2010-12-20

10.  Trans-catheter closure of atrial septal defect: Balloon sizing or no Balloon sizing - single centre experience.

Authors:  Saurabh Kumar Gupta; S Sivasankaran; S Bijulal; Jagan Mohan Tharakan; S Harikrishnan; Kvk Ajit
Journal:  Ann Pediatr Cardiol       Date:  2011-01
View more