| Literature DB >> 28946680 |
Ross D Peterson1, Kenneth R Wilund2,3, Brian T Cunningham4,5, Juan E Andrade6,7.
Abstract
The total analytical error of a photonic crystal (PC) biosensor in the determination of ferritin and soluble transferrin receptor (sTfR) as biomarkers of iron deficiency anemia in chronic kidney disease (CKD) patients was evaluated against certified ELISAs. Antigens were extracted from sera of CKD patients using functionalized iron-oxide nanoparticles (fAb-IONs) followed by magnetic separation. Immuno-complexes were recognized by complementary detection Ab affixed to the PC biosensor surface, and their signals were followed using the BIND instrument. Quantification was conducted against actual protein standards. Total calculated error (TEcalc) was estimated based on systematic (SE) and random error (RE) and compared against total allowed error (TEa) based on established quality specifications. Both detection platforms showed adequate linearity, specificity, and sensitivity for biomarkers. Means, SD, and CV were similar between biomarkers for both detection platforms. Compared to ELISA, inherent imprecision was higher on the PC biosensor for ferritin, but not for sTfR. High SE or RE in the PC biosensor when measuring either biomarker resulted in TEcalc higher than the TEa. This did not influence the diagnostic ability of the PC biosensor to discriminate CKD patients with low iron stores. The performance of the PC biosensor is similar to certified ELISAs; however, optimization is required to reduce TEcalc.Entities:
Keywords: analytical quality specification; iron deficiency biomarkers; method validation; photonic crystal biosensor; total allowable error
Mesh:
Substances:
Year: 2017 PMID: 28946680 PMCID: PMC5677296 DOI: 10.3390/s17102203
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1Standard curves for ferritin (left) or soluble transferrin receptor (sTfR; right) on (A,B) the photonic crystal (PC) biosensor or (C,D) certified enzyme-linked immunosorbent assays (ELISAs). PWV: peak wavelength value.
Performance, means, SD, coefficient of variations, and range of measurements for all biosensing platforms and analytes.
| Sensing Platform | Mean | SD | CV (%) | Range | LOD c | Dynamic Range | |
|---|---|---|---|---|---|---|---|
| sTfR a | PC Biosensor | 1.21 | 0.16 | 12.8 | 0.62–1.93 | 21 | 20–500 |
| R&D ELISA | 1.30 | 0.15 | 11.2 | 0.82–2.47 | 42.5 | 255–6800 | |
| Ferritin b | PC Biosensor | 279.5 | 23.7 | 8.5 | 56–414 | 26 | 26–2000 |
| ALPCO ELISA | 276.8 | 11.3 | 4.1 | 0–504 | 5 | 15–1000 | |
a sTfR measurements are in µg/mL for both platforms; b Ferritin measurements are in ng/mL for both platforms. For ELISA, according to vendor specifications; c LOD and dynamic range in ng/mL. For PC biosensor, LOD was calculated as in [42]. CV: coefficient of variation; LOD: limit of detection; SD: standard deviation.
Figure 2Difference plots comparing serum ferritin and sTfR concentrations from hemodialysis patients using the PC biosensor against the certified ELISAs.
Figure 3Comparison plots showing the analytical range and Pearson correlation coefficient of ferritin and sTfR measured in sera of hemodialysis patients using the PC biosensor and ELISAs.
Total calculated analytical error (TEcalc) for iron deficiency anemia (IDA) biomarkers measured on the PC biosensor against established quality specifications for total allowed error (TEa).
| Biomarker | n 1 | TEcalc 2 | TEcalc 3 (%) | TEa (%) | TEcalc 4 (%) |
|---|---|---|---|---|---|
| sTfR (µg/mL) | 27 | 0.52 | 43.0 | 17.6 5 | 9.8 |
| Ferritin (ng/mL) | 23 | 94 | 33.6 | 16.9 6 | n/a |
1 Number of patients included in analysis; 2 TEcalc as absolute concentration; 3 TEcalc as a percentage; 4 TEcalc for clinical cutoff at 5.3 µg/mL; 5 TEa cited by Bailey et al., 2014; 6 TEa cited on Westgard’s online database [48].