Design and Development of a Miniaturized Percutaneously Deployable Wireless Left Ventricular Assist Device: Early Prototypes and Feasibility Testing.

Brian Letzen, Jiheum Park, Zeynep Tuzun, Pramod Bonde.   

Abstract

The current left ventricular assist devices (LVADs) are limited by a highly invasive implantation procedure in a severely unstable group of advanced heart failure patients. Additionally, the current transcutaneous power drive line acts as a nidus for infection resulting in significant morbidity and mortality. In an effort to decrease this invasiveness and eliminate drive line complications, we have conceived a wireless miniaturized percutaneous LVAD, capable of being delivered endovascularly with a tether-free operation. The system obviates the need for a transcutaneous fluid purge line required in existing temporary devices by utilizing an incorporated magnetically coupled impeller for a complete seal. The objective of this article was to demonstrate early development and proof-of-concept feasibility testing to serve as the groundwork for future formalized device development. Five early prototypes were designed and constructed to iteratively minimize the pump size and improve fluid dynamic performance. Various magnetic coupling configurations were tested. Using SolidWorks and ANSYS software for modeling and simulation, several geometric parameters were varied. HQ curves were constructed from preliminary in vitro testing to characterize the pump performance. Bench top tests showed no-slip magnetic coupling of the impeller to the driveshaft up to the current limit of the motor. The pump power requirements were tested in vitro and were within the appropriate range for powering via a wireless energy transfer system. Our results demonstrate the proof-of-concept feasibility of a novel endovascular cardiac assist device with the potential to eventually offer patients an untethered, minimally invasive support.

Entities:  

Mesh:

Year:  2018        PMID: 28938307      PMCID: PMC5823723          DOI: 10.1097/MAT.0000000000000669

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  12 in total

1.  Toward total implantability using free-range resonant electrical energy delivery system: achieving untethered ventricular assist device operation over large distances.

Authors:  Benjamin Waters; Alanson Sample; Joshua Smith; Pramod Bonde
Journal:  Cardiol Clin       Date:  2011-10-02       Impact factor: 2.213

2.  Innovative Free-range Resonant Electrical Energy Delivery system (FREE-D System) for a ventricular assist device using wireless power.

Authors:  Benjamin H Waters; Joshua R Smith; Pramod Bonde
Journal:  ASAIO J       Date:  2014 Jan-Feb       Impact factor: 2.872

3.  Sixth INTERMACS annual report: a 10,000-patient database.

Authors:  James K Kirklin; David C Naftel; Francis D Pagani; Robert L Kormos; Lynne W Stevenson; Elizabeth D Blume; Marissa A Miller; J T Baldwin; J Timothy Baldwin; James B Young
Journal:  J Heart Lung Transplant       Date:  2014-04-21       Impact factor: 10.247

Review 4.  Energy transmission and power sources for mechanical circulatory support devices to achieve total implantability.

Authors:  Jake X Wang; Joshua R Smith; Pramod Bonde
Journal:  Ann Thorac Surg       Date:  2014-02-12       Impact factor: 4.330

5.  Impella 5.0 as a Second-Line Mechanical Circulatory Support Strategy After Extracorporeal Life Support.

Authors:  David Schibilsky; Tobias Kruger; Henning F Lausberg; Christoph Eisenlohr; Christoph Haller; Attila Nemeth; Barbara Schibilsky; Helene Haeberle; Peter Rosenberger; Tobias Walker; Christian Schlensak
Journal:  Artif Organs       Date:  2016-09       Impact factor: 3.094

6.  Percutaneous Mechanical Circulatory Support Versus Intra-Aortic Balloon Pump in Cardiogenic Shock After Acute Myocardial Infarction.

Authors:  Dagmar M Ouweneel; Erlend Eriksen; Krischan D Sjauw; Ivo M van Dongen; Alexander Hirsch; Erik J S Packer; M Marije Vis; Joanna J Wykrzykowska; Karel T Koch; Jan Baan; Robbert J de Winter; Jan J Piek; Wim K Lagrand; Bas A J M de Mol; Jan G P Tijssen; José P S Henriques
Journal:  J Am Coll Cardiol       Date:  2016-10-31       Impact factor: 24.094

7.  Effect of the hemopump left ventricular assist device on regional myocardial perfusion and function. Reduction of ischemia during coronary occlusion.

Authors:  M E Merhige; R W Smalling; D Cassidy; R Barrett; G Wise; J Short; R K Wampler
Journal:  Circulation       Date:  1989-11       Impact factor: 29.690

8.  The diameter of the common femoral artery in healthy human: influence of sex, age, and body size.

Authors:  T Sandgren; B Sonesson; R Ahlgren; T Länne
Journal:  J Vasc Surg       Date:  1999-03       Impact factor: 4.268

9.  The Impella CP device for acute mechanical circulatory support in refractory cardiac arrest.

Authors:  Henrik Vase; Steffen Christensen; Aage Christiansen; Christian Juhl Therkelsen; Evald Høj Christiansen; Hans Eiskjær; Steen Hvidtfeldt Poulsen
Journal:  Resuscitation       Date:  2016-10-14       Impact factor: 5.262

10.  Percutaneous left ventricular assist in ischemic cardiac arrest.

Authors:  Vegard Tuseth; Mohamed Salem; Reidar Pettersen; Ketil Grong; Svein Rotevatn; Tore Wentzel-Larsen; Jan Erik Nordrehaug
Journal:  Crit Care Med       Date:  2009-04       Impact factor: 7.598

View more
  3 in total

1.  Chugging to silent machines: development of mechanical cardiac support.

Authors:  Cumaraswamy Sivathasan
Journal:  Indian J Thorac Cardiovasc Surg       Date:  2020-08-25

2.  Electrical power to run ventricular assist devices using the Free-range Resonant Electrical Energy Delivery system.

Authors:  Benjamin H Waters; Jiheum Park; J Christopher Bouwmeester; John Valdovinos; Arnar Geirsson; Alanson P Sample; Joshua R Smith; Pramod Bonde
Journal:  J Heart Lung Transplant       Date:  2018-08-11       Impact factor: 10.247

Review 3.  Systems of conductive skin for power transfer in clinical applications.

Authors:  Andreas P Kourouklis; Julius Kaemmel; Xi Wu; Evgenij Potapov; Nikola Cesarovic; Aldo Ferrari; Christoph Starck; Volkmar Falk; Edoardo Mazza
Journal:  Eur Biophys J       Date:  2021-09-03       Impact factor: 1.733

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.