Comparing subjective and objective image quality at two different radiation exposure ranges of the paranasal sinus CT examinations using a volumetric 320-row detector CT system.

Berhan Pirimoglu1, Recep Sade1, Muhammed Sedat Sakat2, Hayri Ogul1, Suat Eren1, Mecit Kantarci1.   

Abstract

OBJECTIVES: To evaluate image quality of protocol including 135 kVp, 5 mAs and 80 kVp, 5 mAs in the paranasal sinus CT examinations using single volumetric 320-row multidetector CT technique.
METHODS: From September to December 2016, both of our control group including 135 kVp and 5 mAs and our study group including 80 kVp and 5 mAs paranasal CT protocols were simultaneously performed on 40 patients using single volumetric 320-row multidetector CT device. Image quality for bony structures, air-filled structures and soft tissues were independently assessed for each group by three blinded observers using a 3-point grading scale (0 = not diagnostic, 1 = partially diagnostic, 2 = diagnostic). Objective image quality was also performed by region of interests were placed on axial soft tissue reconstructions in both eye bulbs, retrobulbar fat and maxillary sinuses to obtain different values for aqueous tissue, soft tissue and air, respectively. Effective dose was calculated from the dose-length product.
RESULTS: The effective radiation dose which calculated for the control group scans was 0.037 ± 0.003 mSv. But, it was 0.0099 ± 0.001 mSv for the study group scans. The effective radiation dose of study group was statistically significant lower than control group (p < 0.001). Despite significant lowering of the radiation doses, image qualities were sufficient for evaluating all the bony structures, air-filled structures and soft tissues except for eye muscle, retrobulbar fat and eye bulb.
CONCLUSIONS: Our results present that our protocols for study and control group provide significant dose reduction without the loss of diagnostic image quality for paranasal sinus CT. Paranasal sinus CT imaging can be performed at very low radiation exposure maintaining high image quality using a single volume 320-row detector CT device using 135 kVp and 5 mAs.

Entities:  

Keywords:  320-row detector CT; low radiation dose; paranasal sinus; volumetric

Mesh:

Year:  2017        PMID: 28937286      PMCID: PMC5965943          DOI: 10.1259/dmfr.20170212

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  18 in total

1.  Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation.

Authors:  Tom H Mulkens; Patrick Bellinck; Michel Baeyaert; Dirk Ghysen; Xavier Van Dijck; Elvier Mussen; Caroline Venstermans; Jean-Luc Termote
Journal:  Radiology       Date:  2005-08-26       Impact factor: 11.105

Review 2.  [Imaging of paranasal sinuses today].

Authors:  F Dammann
Journal:  Radiologe       Date:  2007-07       Impact factor: 0.635

3.  Comparison of radiation dose estimates and scan performance in pediatric high-resolution thoracic CT for volumetric 320-detector row, helical 64-detector row, and noncontiguous axial scan acquisitions.

Authors:  Daniel J Podberesky; Erin Angel; Terry T Yoshizumi; Greta Toncheva; Shelia R Salisbury; Alan S Brody; Christopher Alsip; Alessandra Barelli; John C Egelhoff; Colin Anderson-Evans; Giao B Nguyen; David Dow; Donald P Frush
Journal:  Acad Radiol       Date:  2013-09       Impact factor: 3.173

4.  Reducing the dose of CT of the paranasal sinuses: potential of an iterative reconstruction algorithm.

Authors:  Lars-Arne Schaafs; Julian Lenk; Bernd Hamm; Stefan Markus Niehues
Journal:  Dentomaxillofac Radiol       Date:  2016-07-15       Impact factor: 2.419

5.  Low-Dose CT of the Paranasal Sinuses: Minimizing X-Ray Exposure with Spectral Shaping.

Authors:  Wolfgang Wuest; Matthias May; Marc Saake; Michael Brand; Michael Uder; Michael Lell
Journal:  Eur Radiol       Date:  2016-02-24       Impact factor: 5.315

6.  Reducing the radiation dose for low-dose CT of the paranasal sinuses using iterative reconstruction: feasibility and image quality.

Authors:  Stefan Bulla; Philipp Blanke; Frederike Hassepass; Tobias Krauss; Jan Thorsten Winterer; Christine Breunig; Mathias Langer; Gregor Pache
Journal:  Eur J Radiol       Date:  2011-06-12       Impact factor: 3.528

Review 7.  How Can New Imaging Modalities Help in the Practice of Radiology?

Authors:  Berhan Pirimoglu; Recep Sade; Hayri Ogul; Mecit Kantarci; Suat Eren; Akın Levent
Journal:  Eurasian J Med       Date:  2016-10

8.  CT image quality improvement using Adaptive Iterative Dose Reduction with wide-volume acquisition on 320-detector CT.

Authors:  Alban Gervaise; Benoît Osemont; Sophie Lecocq; Alain Noel; Emilien Micard; Jacques Felblinger; Alain Blum
Journal:  Eur Radiol       Date:  2011-09-17       Impact factor: 5.315

Review 9.  CT coronary angiography: 256-slice and 320-detector row scanners.

Authors:  Edward M Hsiao; Frank J Rybicki; Michael Steigner
Journal:  Curr Cardiol Rep       Date:  2010-01       Impact factor: 2.931

10.  Comparison between low-dose and standard-dose multidetector CT in patients with suspected chronic sinusitis.

Authors:  Denis Tack; Jacques Widelec; Viviane De Maertelaer; Jean-Marie Bailly; Christian Delcour; Pierre Alain Gevenois
Journal:  AJR Am J Roentgenol       Date:  2003-10       Impact factor: 3.959

View more
  1 in total

1.  Volumetric growth analysis of maxillary sinus using computed tomography scan segmentation: a pilot study of Indonesian population.

Authors:  Erli Sarilita; Yurika Ambar Lita; Harry Galuh Nugraha; Nani Murniati; Harmas Yazid Yusuf
Journal:  Anat Cell Biol       Date:  2021-12-31
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.