Evolution of stereoscopic imaging in surgery and recent advances.

Katie Schwab1, Ralph Smith1, Vanessa Brown1, Martin Whyte1, Iain Jourdan1.   

Abstract

In the late 1980s the first laparoscopic cholecystectomies were performed prompting a sudden rise in technological innovations as the benefits and feasibility of minimal access surgery became recognised. Monocular laparoscopes provided only two-dimensional (2D) viewing with reduced depth perception and contributed to an extended learning curve. Attention turned to producing a usable three-dimensional (3D) endoscopic view for surgeons; utilising different technologies for image capture and image projection. These evolving visual systems have been assessed in various research environments with conflicting outcomes of success and usability, and no overall consensus to their benefit. This review article aims to provide an explanation of the different types of technologies, summarise the published literature evaluating 3D vs 2D laparoscopy, to explain the conflicting outcomes, and discuss the current consensus view.

Entities:  

Keywords:  Endoscopy; Minimally invasive surgery; Stereoscopic; Three-dimensional displays; Three-dimensional laparoscopy

Year:  2017        PMID: 28874957      PMCID: PMC5565502          DOI: 10.4253/wjge.v9.i8.368

Source DB:  PubMed          Journal:  World J Gastrointest Endosc


Core tip: Capture of true stereopsis from the operative field is crucial for the subsequent projection of a high quality stereoptic image. The latest three-dimensional (3D) systems using dual channel stereoendoscopes and passive polarizing stereoscopic projection generate high quality 3D images for minimally invasive surgery. There is subjective and objective laboratory based evidence supporting use of 3D vs two-dimensional for surgeons of all experience. However, their clinical application has yet to be addressed with Level 1 evidence.

INTRODUCTION

When Phillipe Bozzini first designed and used his “Lichtleiter” in 1803 to peer into the human body, the medical world unwittingly became reliant on observing the endoscopic view of the human body in only two-dimensions (2D). In 1838 Charles Wheatstone[1] was the first to accurately describe and publish the phenomenon of stereopsis - “… the mind perceives an object of three dimensions by means of the two dissimilar pictures projected by it on the two retinae …”. He described in his paper how the illusion of light projecting outwards from the surface of a metal plate that had been turned on a lathe had brought him to this realisation. He demonstrated the validity of his proposed mechanism of stereopsis by creating the “Wheatsone Stereoscope”. This created an illusion of stereopsis simply by projecting different images to each eye of the viewer. By adjusting each image to give an impression of the perspective that would have been seen by that eye the viewer was left with a sense of a three-dimensional (3D) image. The first endoscopic procedures were performed with single eyepiece rigid scopes which provided a monocular view for the operating surgeon. In the 1970s these images were relayed via a camera to a video monitor. Thus was born the modern era of “off screen” videoscopic operating. In the late 1980s the first laparoscopic cholecystectomies were performed and popularity for laparoscopic surgery began to increase exponentially. This prompted a sudden rise in surgical and technological innovations as the benefits and feasibility of minimal access surgery became more universally recognised. As minimal access surgery became more widely adopted the steepness of the learning curve for surgeons became more apparent. In particular the monocular laparoscopic view providing two-dimensional viewing, and associated reduced depth perception, became the focus of technological advances. Attention therefore turned to producing a usable 3D endoscopic view for surgeons, utilising different technologies for image capture and image projection. These evolving visual systems have been assessed in various research environments with conflicting outcomes of success and usability, and no overall consensus to their benefit. This review article aims to provide an explanation of the different types of technologies, summarise the published literature evaluating 3D vs 2D laparoscopy, to explain the conflicting outcomes, and discuss the current consensus view.

First stereoptic views

Binocular microscopes were first used in 1922 in otolaryngology to overcome the lack of depth perception associated with monocular operating microscopes by surgeon Gunnar Holmgren (1875-1954), Head of the University Clinic of Stockholm[2]. These provided a stereoptic magnified view of the operating field and were quickly adopted by Otolaryngology, Neurosurgery and Orthodontics. In the 1980s, a German surgeon, Dr. Gerhard Buess[3], pioneered Transanal Endoscopic Microsurgery (TEMS) utilising the first “stereoendoscope” with two optical channels, viewed through binocular eye pieces. In 1992, his team trialed the first prototype laparoscopic stereoendoscope in animal studies and clinically during laparoscopic cholecystectomies, and concluded the stereopsis facilitated complex laparoscopy[4].

Image capture

In the laparoscopic settling, an image of the operative field may be captured in one of two ways. A traditional rod-lens laparoscope may be used to transmit the light from the image to outside the patient where a video camera then captures the image and sends it as an electrical signal to an image processor. Rod lens technology is now being superseded by “chip on the tip” technology utilizing small camera chips which capture the image at the tip of the laparoscope and then transmit the electrical signal along the laparoscope to an image processor. The technology used to capture the 3D characteristics of the operating field includes the laparoscope, the camera and the image processor. Various systems have been developed and trialed in the literature. Single channel systems attempt to extract two perspectives of the operative field from a single point of view by splitting the image either with a prism or filter. The result is therefore not a true binocular image[5]. Dual channel systems provide two horizontally separated images and thus produce two truly different perspectives of the operative field. “Insect eye” scopes allow for multi images to be captured and processed simultaneously. There is significant variety in the design of the video capture systems, which results in differences in the quality of the perceived image.

Projection systems

Projection systems aim to deliver the 3D view to the observer. Early systems used active shuttering projection, where alternate left and right views are displayed at high frequency on a display. With these systems the operator wears active shuttering glasses so that each eye receives only the corresponding right or left eye image. Robotic systems evolved to use a fixed viewing environment, where, like in a microscope, the observer has a separate image displayed to each eye. This concept was used in Head Mounted Displays (HMDs) where each eye was provided with its own screen to achieve stereopsis. The latest commercial projection systems use passive polarizing technology, which allows for two images to be projected simultaneously in different polarized waveforms. A high definition image is made up of 1080 horizontal pixel lines. For passive polarizing projection the image projected has odd horizontal pixel lines emitting light polarized vertically and even lines emitting light polarized horizontally. The user then wears lightweight polarizing glasses to separate the correct image to each eye. The horizontal resolution of the image is therefore reduced by half to 540 pixels but the vertical resolution remains at 1080 pixels and the resulting image therefore remains high quality. When this technology was transferred from cinema projection systems to home television monitors the opportunity to use this system in the operating theatre became a possibility. More recently there has been the experimental development of complex waveform projection systems (advanced systems based on anaglyph separation), autostereoscopic “glass-free” displays and holographic displays.

LITERATURE REVIEW

We aimed to identify from the literature, all published work evaluating 3D laparoscopic systems compared to 2D standard “classical laparoscopic” systems. PubMed, EMBASE, Ovid and Medline where used as search engines to identify any published full English language papers since 1996 which referenced stereopsis, 3D, vs two-dimensional or 2D, laparoscopy, endoscopic surgery, imaging and 3D. Overall, 361 titles were identified and 275 were discounted on further review of their titles. Of the 86 abstracts reviewed, 45 were further discounted as they didn’t compare 3D with 2D. Review of these 41 papers acknowledged another six papers not identified by the original search. In total, 47 papers reported assessing 3D imaging systems against 2D systems in laparoscopic surgery. A further four titles were discounted on reading the whole paper, leaving 43 to be assessed. Ninety-six percent of the studies describe laboratory based experiments, involving a variety of laparoscopic skills tasks, some from validated curriculum programmes and others designed to mimic advanced laparoscopic skills. The studies also use a variety of subjects from non-surgical participants to those with a variety of experience in laparoscopic surgery. The number of tasks, repetitions, cross over in visual systems, assessment of a learning curve and number of individual subjects involved varied in each study. Universally, the common themes assessed in the majority of studies were the time for task completion and performance, either by clearly defined errors or by other assessment defined scoring systems. There has been speculation for the last 18 years over the benefit of 3D operating visual systems, largely based on conflicting reports in the literature and the ongoing evolution of the system technology. We separated data by the type of optical or projection system in order to clarify the results and explain the conflicting outcomes observed by different researchers.

Single channel endoscope studies

We identified 13 studies which used single channeled scopes to capture the laparoscopic view (Table 1). Seven of these studies[6-12] utilised active shuttering projection systems with only one study[7] identifying a significant improvement in outcomes using the 3D system compared to the 2D standard. All of these studies also reported poor subjective outcomes associated with the 3D systems, including visual strain, headaches and nausea as well as an awareness of flickering of the screen. Four studies[13-16] assessed a second-generation 3D system, which used a single channel scope and projected left and right images to head mounted display systems, allowing individual eye projection without loss of light or image quality. Three of the studies reported significant improvement in performance for novices. The HMDs, although bulky, did not cause any of the cortical disturbances reported by the active shuttering systems. The final two studies[17,18] used single channel scopes and the latest passive polarizing systems. Neither identified a significant difference in respective outcomes with the 3D systems. Both studies reported that a period of adaptation was required to overcome any higher processing symptoms that the 3D visual system induced[17].
Table 1

Single channelled scopes

Ref.YearProjection system for 3DWho and what assessedObjective outcomesSubjective outcomes
McDougall et al[6]1996Active shuttering screen and glasses22 urological and gynaecological surgeons, non-novice Pig-lab, laparoscopic vessel dissection and securing, suturing and knot tyingTime for completion. No significant difference found3D not felt to enhance image quality or enhance performance. Blurred vision and eye fatigue with 3D
Dion et al[7]1997Active shuttering screen and glassesSurgeons and non-surgeons. Lab visual (n = 8) and motor skills (n = 9)Time and errors. Improvement in both with 3DGlasses bothersome and dizziness reported
Chan et al[8]1997Active shuttering screen and glasses32 surgeons, 11 with and 21 without laparoscopic experience 1 × lab based skills taskTime for completion in 2D and 3D (1 repetition). No significant difference50% felt no improved performance although 66% felt depth perception improved 40% felt reduced image quality and dimmer; 10% reported dizziness and eyestrain
Hanna et al[9]1998Active shuttering screen and glasses (A/S)4 surgical SpRs performing 60 laparoscopic cholecystectomiesTime for completion and errors No significant differenceVisual strain, headache and facial discomfort with 3D system
Mueller et al[10]1999Active shuttering screen and glasses30 subjects (10 with and 20 without laparoscopic experience) 4 × lab based skills tasks for all, then experienced did suturing tasksTime for attempts, and success/failure of attempt No significant differenceReported loss of concentration, headaches and distraction with 3D system
Herron et al[11]19993D (active shuttering screen and glasses) and 3D HMD50 laparoscopic novices 3 × lab based skills tasksTime to completion of 3 skills tasks in each visual system (2 × repitions) No significant differenceAlthough 48% preferred 3D A/S screen over all, 7% and 25% respectively reported headaches with 3D screen and 3D HMD. 82% found HMD uncomfortable
Mueller-Richter et al[12]20033D (active shuttering screen and polarising glasses) and 3D Autostereoscopic screen59 laparoscopic novices 3 × lab based skills tasksNumber of completions in time limit and subjective difficulty No significant differenceFlickering reported with both 3D systems
Bhayani et al[13]2005HMD24 surgical residents, minimal laparoscopic experience. 1 × lab based skills taskTime for completion in 2D and 3D (1 repetition) Significant reduction in time> 50% preferred the 3D system and found task easier in 3D No subjective assessment on physical symptoms
Patel et al[14]2007HMD15 novices and 2 experts 5 × lab based skills tasksTime and accuracy in 2D and 3D (1 repetition) of the novices compared to the experts Significant difference in both for novices only in 3DNA
Bittner et al[15]2008HMD2 novices, 2 intermediate and 2 experts 2 × lab based suturing tasks (based on handedness, visual system and articulating needle holder)Time and accuracy in 2D and 3D (multi repetitions with each variable) No significant difference83% felt improved depth perception. No reported physical symptoms
Votanopoulos et al[16]2008HMD36 surgical residents and medical students (11 with and 25 without laparoscopic experience) 6 × lab based skills tasks (rpt 3/12 later)Time and errors in 2D and 3D (1 repetition) Significant improvement in time and errors in novice group onlyNA
Kong et al[17]2009Passive polarising screen and glasses21 novices and 6 experienced surgeons 2 × lab based skills tasksTime and errors in 2D and 3D (4 repetitions of each over 4 d) Significant reduction in errors in 3D novices, no other significant difference notedDizziness and eye fatigue in novice with 3D system which improved with time
Mistry et al[18]2013Passive polarising screen and glasses31 medical students (novices) 4 × lab based skills tasks (MISTELS)Task Performance in 2D and 3D as per MISTELS scoring system No significant differenceNo detrimental symptoms with 3D

NA: Not available; 3D: Three-dimensional; 2D: Two-dimensional; HMD: Head mounted display.

Single channelled scopes NA: Not available; 3D: Three-dimensional; 2D: Two-dimensional; HMD: Head mounted display.

Dual channel endoscope studies

Robotic “fixed screen” studies: Nine studies investigated the effect of stereopsis in laparoscopic surgery utilising the Da Vinci robotic system (Intuitive, California United States) (Table 2)[19-27]. Stereopsis is achieved with a binocular endoscope and two camera heads for separate left and right image capture. Each image is received by the respective eye, simultaneously using a fixed console, alleviating the need for shuttering, polarizing or head mounted projection. All studies reported significant improvement in performance with the Da Vinci system in 3D mode over 2D mode. Notably, performance advantages were independent of participant experience[27].
Table 2

Dual channel laparoscopes - Robotic fixed screen

Ref.YearProjection system for 3DWho and what assessedObjective outcomesSubjective outcomes
Falk et al[19]2001Da Vinci15 experienced laparoscopic surgeons 6 × lab based skills tasks (increasing difficulty)Time and errors in 2D and 3D and 2DHD (I repetition in each view) Significant differences in time and errors in 3DOnly 33% felt 3D better view No detrimental symptoms reported
Munz et al[20]2004Da Vinci11 experienced laparoscopic surgeons 4 × lab based skills tasksErrors and performance (ICSAD assessment - time, no. movements and distance moved) Significant difference in both in 3DNA
Moorthy et al[21]2004Da Vinci10 surgeons of varying experience Lab based suturing taskTime and distance travelled of instruments in 2D and 3D Significant difference in both in 3DNA
Badani et al[22]2005Da Vinci7 surgeons (3 experienced with Da Vinci, 4 not) 2 × lab based suturing tasksTime and errors Significant difference in 3D in all areasNA
Blavier et al[23]2007Da Vinci40 medical students Lab based skills taskErrors, performance and learning curve Significant difference in 3DNo detrimental symptoms reported
Byrn et al[24]2007Da Vinci12 surgeons of varying experience 4 × lab based skills tasksTime and errors in 2D and 3D Significant difference in 3DNo detrimental symptoms reported
Blavier et al[25]2007Da Vinci60 medical students 4 × lab based skills task (increasing difficulty)Specific performance metric score Significant difference in 3D in all tasksNo detrimental symptoms reported
Fishman et al[27]2008Da Vinci and prototype Ames stereoscopic camera12 subjects of varying exposure to stereoptic systemsTime for completion while altering binocular disparity of stereoptic camera until 0% (matching 2D vision)NA
Blavier et al[28]2009Da VinciLab based skills task using Da Vinci manipulator 80 subjects (60 novice individuals and 20 expert laparoscopic surgeons) Lab based taskSignificant difference with 3D from binocular disparity Time for task completion and estimation of time in 2D or 3D not both Significant difference in 3D for novices, similar results for expertsNA

NA: Not available; 3D: Three-dimensional; 2D: Two-dimensional.

Dual channel laparoscopes - Robotic fixed screen NA: Not available; 3D: Three-dimensional; 2D: Two-dimensional.

Studies using screen projection and eye-glass technology

Five studies reported outcomes with binocular stereoendoscopes (Table 3), alternating screen image and active shuttering glasses[28-32]. Four of the five studies reported significant improvements in performance with 3D systems[28-32]. In the one study (Wentink et al[30], 2002) the screen was placed very close to the surgeon while the working environment from the stereoendoscope was 12 cm. This produces conflict between convergence and focus for the operating surgeon, and it is therefore unsurprising that the 3D system showed poorer performance.
Table 3

Dual channel laparoscopes - Screen projection and glasses

Ref.YearProjection system for 3DWho and what assessedObjective outcomesSubjective outcomes
Birkett et al[26]1994Active shuttering screen and Active glasses then polarised glasses vs 2D10 Subjects? experience 2 × lab based skills tasksTime take for repetitive cycles; No difference in simples task, reduced time in complex taskNA
Peitgen et al[29]1996Active shuttering screen and glasses60 subjects (20 novices, 20 beginners, 20 advanced laparoscopic surgeons) 2 × lab based skills tasksTime and accuracy of tasks Both significantly improved in 3D, independent of experienceNA
Wentink et al[30]2002Active shuttering screen and polarised glasses vs TFT display vs projection vs standard (2D)8 surgeons with laparoscopic experience Lab based skills taskTime for task completion, 10 repetitions but only 2 surgeons per visual system No improvement with 3DFelt image quality poorer with 3D
Jourdan et al[31]2004Active shuttering screen and glasses8 experienced laparoscopic surgeons 5 × lab based skills tasksTime and errors, 10 repetitions each, in each visual system Significant improvement in both in 3DNA
Feng et al[32]2010Active shuttering screen and polarised glasses (SD vs 2D SD vs 2D HD)27 subjects (16 novices, 11 with varying laparoscopic experience) Lab based skills taskTime and economy of movement Time significantly improved over both 2D systems in 3D, economy of movement improved in 3D vs HD, not SD 2DFelt improved depth perception in 3D
Hubber et al[33]2003Prototype passive polarising screen and glasses16 Medical Students (novices) Lab based skills tasksTime and performance (ICSAD) Improvements in 3D significant over 2DNA
Honeck et al[34]2012Passive polarising screen and glassed10 novices and 10 experienced laparoscopic surgeons 5 × lab based skills tasksTime and errors (1 × repetition, in only 1 of the visual systems) No significant improvement in time, reduction in errors significant in both groups in 3DNo impairment felt in subjective feedback when using the 3D system
Smith et al[35]2012Passive polarising screen and glassed20 novices 4 × lab based skills tasksTime and errors (10 repetitions of each task in each visual condition) Significant improvement in time and errors in 3DNA
Bilgen et al[36]2013Passive polarising screen and glassed3 surgeons Clinical - 11 laparoscopic cholecystectomies performed in 3D (compared to 11 performed retrospectively in 2D)Time Significant reduction in time when performed in 3D, compared to case matched lap choles performed previously in 2DNA
Sinha et al[37]2013Passive polarising screen and glassedRetrospective analysis of 451 clinical gynaecological surgery performed in 3D Case matched assessment of 200 hysterectomies performed in 3D vs 2DTime Significant reduction in operating time and morcellation time when performed in 3DNA
Cicione et al[38]2013Passive polarising screen and glassed33 subjects (10 experts and 23 novices) 5 × lab based skills tasks (Basic Laparoscopic Urological Skills)Time and errors Overall, significant improvement in time and errors (although experts only improved time in 1 task in 3D)Subjective Questionnaire - felt tasks were easier in 3D universally
Lusch et al[39]2014Passive polarising screen and glassed24 subjects (10 medical students, 7 residents, 7 expert surgeons) 6 × lab based skills tasksTime and errors 4 out of 5 skills tasks had significantly improved time and errors when done in 3D, independent on experienceOptical resolution and depth perception improved in 3D
Smith et al[40]2014Passive polarising screen and glassed20 experienced surgeons 4 × lab based skills tasksTime and errors (10 repetitions of each task in each visual condition) Significant improvement in time and errors in 3DSubjective assessments using NASA Task Load Index - improvements with 3D all sections

NA: Not available; 3D: Three-dimensional; 2D: Two-dimensional; HMD: Head mounted display.

Dual channel laparoscopes - Screen projection and glasses NA: Not available; 3D: Three-dimensional; 2D: Two-dimensional; HMD: Head mounted display. Eight studies evaluated passive polarizing screen and glass technology (Table 3)[33-40]. Two of these studies retrospectively compared a series of operations (laparoscopic cholecystectomies and laparoscopic gynaecological operations) with case matched procedures in standard 2DHD systems[36,37]. Both reported a significant reduction in operating times for case matched procedures. Six laboratory based studies identified significant improvements in most of the tested parameters when tasks were performed in 3D[33-35,38-40]. Two other studies (Honeck et al[34], 2012, and Cicione et al[38], 2013) found varied performance improvements in 3D. Honeck found reduced errors but no significant time improvements, while Cicione et al[38] (2013) found an overall significant improvement with 3D over 2D. These advantages were only observed in the expert subgroup when performing one task, the “Peg Transfer”. However both studies only allowed for a single repetition of tasks in 3D and 2D before comparison. In studies which allowed for repetitions and plateauing of the learning curve in both visual environments before comparison, there was a universal improvement when comparing 3D over 2D, independent of experience[33,35,39,40].

Comparing different scopes and projection systems

Four papers described using more than one type of 3D system in their comparison of 3D vs 2D (Table 4)[41-44]. Hanna et al[42] (2000) assessed single-channel scope and dual-channel scope systems, both using active shuttering screen/glasses systems compared to a standard 2D system when performing laboratory based bowel anastomosis. The 3D systems were evaluated together, rather than separately and showed no significant difference in time or precision compared to 2D. However, closer analysis of the data implies the dual channel scope demonstrated a trend of improved time and precision compared to its single channel counterpart. Visual strain was reported using both stereoendoscopes. Wilhelm et al[43] (2014) reported all performance parameters were superior in 3D over 2D using a variety of experimental and commercially available systems, although visual disturbance related to the autostereoscopic screen only. Finally, Wagner et al[44] (2012), compared single channel scope with HMD technology (in 3D and 2D settings) with robotic dual channel fixed screen technology (2D and 3D settings) and demonstrated significant time reductions with robotic 3D across all other laparoscopic outcomes.
Table 4

Comparing multisystems

Ref.YearProjection system for 3DWho and what assessedObjective outcomesSubjective outcomes
van Bergen et al[41]19982 × single channelled and 2 × dual channelled scopes + active shuttering screen vs 2D40 subjects - novices Variety of different models and skills tasksTimes and errors Objectively - significant improvement in 3D throughoutSubjectively - all tasks judged easier in 3D
Hanna et al[42]2000Single-channel scope + active shuttering screen and glasses; double-channel scope + active10 experienced surgeons Lab based endoscopic anastomotic suturingTime, precision of suture placement and pressure leakage score of anastomosis (2 × repetitions in each visual system) 3D systems evaluated together, no significant difference noted in 3DVisual strain reported with 3D systems
Wilhelm et al[43]2014Dual channel scope + passive polarising screen and glasses vs 2D vs autostereoscopic screen48 subjects, varying experience Lab based suturing taskTime, economy of movement (electromagnetic tracking) and workload assessments (using NASA Task Index Score All performance parameters were superior in 3DNo symptoms in 3D PP system, visual disturbance reported with autostereoscopic display
Wagner et al[44]2012Single-channel scope + HMD vs robotic dual channel scope + fixed head view34 subjects (18 novices) 3 × lab based skills tasksTime 3D robotic performance faster than all others, significantlyNA

NA: Not available; 3D: Three-dimensional; 2D: Two-dimensional; HMD: Head mounted display.

Comparing multisystems NA: Not available; 3D: Three-dimensional; 2D: Two-dimensional; HMD: Head mounted display.

Other prototype projection systems

Four publications assessed prototype projection systems (Table 5)[45-48]. Three used autostereoscopic screen technology with binocular scopes thus negating the need for eyewear[45,46,48]. Improvements in all outcomes were seen with the 3D group. Storz et al[47] (2011) used a novel projection system with a wavelength multiplex camera and monitor with wavelength polarizing eyewear (a technology based on original anaglyph systems). This again returned a true sense of stereopsis and improvements in outcomes were significant in 3D over 2D.
Table 5

Other prototype projection systems

Ref.YearProjection system for 3DWho and what assessedObjective outcomesSubjective outcomes
Taffinder et al[45]1999Dual channel scope with autostereoscopic/glass free screen28 subjects (16 novices and 12 experienced laparoscopic surgeons) Novices = basic grasping and cutting lab based skills Experienced = suturing and complex cutting lab based skillsTime and performance score (ICSAD assessment tool) Significant improvement in 3D over 2D laparoscopyNo side effects reported with 3D
Ohuchida et al[46]2009Dual channel scope with “Cyberdome” projection system23 novices 6 × lab based skills tasksTime, errors and performance Significant improvement in all parameters in 3D with cyberdome over 2DNA
Storz et al[47]2011Dual-channel scope + wavelength multiplex camera and monitor with polarising glasses30 subjects (20 medical students and 10 experienced laparoscopic surgeons) 5 × lab based skills tasksTime and errors In 4 out 5 tasks, significant reduction in time in 3D, in 4out of 5 tasks, significant reduction in errorsNA
Khoshabeh et al[48]2012Dual-channel scope + Multiview autostereoscopic display/glass free screen3 experienced laparoscopic surgeons 2 × lab based skills tasksTime and errors Reduced time and errors using 3DNA

NA: Not available; 3D: Three-dimensional; 2D: Two-dimensional.

Other prototype projection systems NA: Not available; 3D: Three-dimensional; 2D: Two-dimensional.

DISCUSSION

There is subjective and objective laboratory based evidence supporting use of 3D vs 2D for surgeons of all experiences as it provides the most realistic view of the operating field. It is also evident that stereoscopic imaging technology is continuing to evolve to generate higher quality 3D images. Capture of true stereopsis from the operative field is crucial for the subsequent projection of a true stereoptic image. However, with such focus on producing an effective projection system, the acquisition and true stereopsis of the image has sometimes been overlooked. It is clear from this review that in systems that compromised on the capture of two truly separate images of the operative field, they yielded no advantage for the participants using 3D over 2D. In studies using dual channel stereoendoscopes, the separate lenses within the laparoscope provided a greater spatial impression of stereopsis[49-51]. Consequently, for the operator, there is a more accurate appreciation of depth. Fishman et al[27] (2008) concluded there was deterioration in laparoscopic performance by reducing horizontal lens separation in an experimental dual channel scope (thereby reducing stereopsis impression). However single channel systems produce images of greater clarity and resolution due to the greater size of the single optic channel for light transfer[52]. Single channel optics can produce convincing stereopsis only at close operating distances, whereas dual channel systems provide significant stereopsis in larger cavities, where there is greater distance from the end of the stereoendoscope to the operating site[51]. Close operating or near field objects with dual channel systems can cause visual discomfort due to the fixed focal point of the two lenses and our natural convergence conflicting. Therefore it is not surprising that the majority of studies which utilised single channel laparoscopes did not show a benefit of 3D laparoscopy as all used target operating points distant to the scopes key stereoptic capabilities, irrespective of the projection system employed. Modern projection systems attempt to provide as true a representation of the natural 3D view as possible, whilst balancing comfort and visual ease for the observer(s) and maintaining the brightness and resolution quality of the image. Active systems caused visual disturbances, headaches and symptoms of nausea due to the conflict of convergence and accommodation, as well as flickering and discomfort for the viewer due to the cumbersome battery powered glasses. Early 3D images had poor resolution and luminosity as early cameras could not cope with low light levels or capture at high resolution. Projection systems were equally constrained by low refresh rates, low resolution and brightness. This added to discomfort and degraded the early 3D view[51]. Falk et al[19], 2001, demonstrated that image quality is vital for precision and surgical performance, as 2DHD systems produced better results when compared with standard view 2D and 3D. The use of polarizing glasses and filters over the shuttering screen provides a more comfortable wear experience for the observer but this is at the expense of image brightness. Head-mounted displays provide good quality images with no degradation in quality or light and preserve the normal hand-eye axis[53]. However open sided head units, which do not block surrounding visual stimuli, can cause headaches and dizziness due to conflicting information from visual input and body position whilst with sealed units the surgeons are isolated from their surroundings and unable to react to unforeseen environmental incidents[42]. The Da Vinci robotic system (intuitive, United States) allows for fixed console viewing and so provides an unparalleled quality of stereopsis for the surgeon. All the studies which assessed binocular and biocular (same view through each eye, therefore 2D view)[51], showed statistically significant advantages with 3D performance for time and errors, reduced motion, and all other comparative markers for surgical performance. There can be no doubt that the advantages noted were purely due to the improvement in view provided by reintroduction of natural stereoptic depth cues. However use of the robot is limited to a relatively small number of procedures where advantage of the robotic platform over standard laparoscopic techniques has been established. Later studies (Table 3), which used binocular endo-scopes and the latest passive polarizing projection systems, identified no subjective impairment or “side effects” to using the 3D systems. The majority identified significant differences in their respective markers of surgical performance when comparing classical laparoscopy to 3D systems. Whilst surgeon experience does affect outcomes, it must be appreciated that experience in classical laparoscopy leads to the development of techniques to overcome the lack of stereopsis. This therefore favours poorer outcomes with the 3D system in studies where the assessment was made after short exposure times and single repetition of skills[34,38,39]. Studies which accounted for learning curves by allowing familiarisation with the system with multiple repetitions and well powered sample sizes demonstrate clearly the benefits in performance achievable with 3D laparoscopy[31,33,35,40]. High quality experimental studies have shown that the latest 3D systems using dual channel stereo-endoscopes and passive polarizing technology provide a “near natural” view, almost comparable to that observed by the Da Vinci. However, their clinical application has yet to be addressed with Level 1 evidence. The only randomised clinical trial assessing 3D systems[9], and addressed by Cochrane review[54], showed no discernible difference for laparoscopic cholecystectomy performance. However, this study is over ten years old and the system assessed used a single channel scope and active shuttering projection, which was unlikely to have provided a true spatial impression of the operating field throughout. Studies that investigated the clinical application of the latest 3D systems identify performance advantages but are underpowered[36,37]. Establishing the benefits of these systems can only truly be addressed within randomised clinical trials, using appropriately powered sample sizes.
  52 in total

1.  The effect of high-definition imaging on surgical task efficiency in minimally invasive surgery: an experimental comparison between three-dimensional imaging and direct vision through a stereoscopic TEM rectoscope.

Authors:  P van Bergen; W Kunert; G F Buess
Journal:  Surg Endosc       Date:  2000-01       Impact factor: 4.584

2.  Three-dimensional laparoscopy. Gadget or progress? A randomized trial on the efficacy of three-dimensional laparoscopy.

Authors:  M D Mueller; C Camartin; E Dreher; W Hänggi
Journal:  Surg Endosc       Date:  1999-05       Impact factor: 4.584

Review 3.  Possibilities and limitations of current stereo-endoscopy.

Authors:  U D A Mueller-Richter; A Limberger; P Weber; K W Ruprecht; W Spitzer; M Schilling
Journal:  Surg Endosc       Date:  2004-04-27       Impact factor: 4.584

4.  Advanced stereoscopic projection technology significantly improves novice performance of minimally invasive surgical skills.

Authors:  R Smith; A Day; T Rockall; K Ballard; M Bailey; I Jourdan
Journal:  Surg Endosc       Date:  2012-01-11       Impact factor: 4.584

5.  The effect of a second-generation 3D endoscope on the laparoscopic precision of novices and experienced surgeons.

Authors:  N Taffinder; S G Smith; J Huber; R C Russell; A Darzi
Journal:  Surg Endosc       Date:  1999-11       Impact factor: 4.584

6.  Three-dimensional vision enhances task performance independently of the surgical method.

Authors:  O J Wagner; M Hagen; A Kurmann; S Horgan; D Candinas; S A Vorburger
Journal:  Surg Endosc       Date:  2012-05-12       Impact factor: 4.584

7.  Multiview glasses-free 3-D laparoscopy.

Authors:  Ramsin Khoshabeh; Jason Juang; Mark A Talamini; Truong Q Nguyen
Journal:  IEEE Trans Biomed Eng       Date:  2012-08-07       Impact factor: 4.538

8.  Comparative evaluation of HD 2D/3D laparoscopic monitors and benchmarking to a theoretically ideal 3D pseudodisplay: even well-experienced laparoscopists perform better with 3D.

Authors:  D Wilhelm; S Reiser; N Kohn; M Witte; U Leiner; L Mühlbach; D Ruschin; W Reiner; H Feussner
Journal:  Surg Endosc       Date:  2014-03-21       Impact factor: 4.584

9.  Effect of passive polarizing three-dimensional displays on surgical performance for experienced laparoscopic surgeons.

Authors:  R Smith; K Schwab; A Day; T Rockall; K Ballard; M Bailey; I Jourdan
Journal:  Br J Surg       Date:  2014-08-18       Impact factor: 6.939

10.  Technique of transanal endoscopic microsurgery.

Authors:  G Buess; K Kipfmüller; D Hack; R Grüssner; A Heintz; T Junginger
Journal:  Surg Endosc       Date:  1988       Impact factor: 4.584

View more
  13 in total

1.  3D presentation in surgery: a review of technology and adverse effects.

Authors:  Tianqi Wang; Bin Zheng
Journal:  J Robot Surg       Date:  2018-12-15

2.  The use of 3D laparoscopic imaging systems in surgery: EAES consensus development conference 2018.

Authors:  Alberto Arezzo; Nereo Vettoretto; Nader K Francis; Marco Augusto Bonino; Nathan J Curtis; Daniele Amparore; Simone Arolfo; Manuel Barberio; Luigi Boni; Ronit Brodie; Nicole Bouvy; Elisa Cassinotti; Thomas Carus; Enrico Checcucci; Petra Custers; Michele Diana; Marilou Jansen; Joris Jaspers; Gadi Marom; Kota Momose; Beat P Müller-Stich; Kyokazu Nakajima; Felix Nickel; Silvana Perretta; Francesco Porpiglia; Francisco Sánchez-Margallo; Juan A Sánchez-Margallo; Marlies Schijven; Gianfranco Silecchia; Roberto Passera; Yoav Mintz
Journal:  Surg Endosc       Date:  2018-12-04       Impact factor: 4.584

3.  Comprehensive review of surgical microscopes: technology development and medical applications.

Authors:  Ling Ma; Baowei Fei
Journal:  J Biomed Opt       Date:  2021-01       Impact factor: 3.170

4.  Learning curves, potential and speed in training of laparoscopic skills: a randomised comparative study in a box trainer.

Authors:  Wolfgang Kunert; Pirmin Storz; Nicolaus Dietz; Steffen Axt; Claudius Falch; Andreas Kirschniak; Peter Wilhelm
Journal:  Surg Endosc       Date:  2020-07-08       Impact factor: 4.584

5.  Comparison of 2D 4K vs. 3D HD laparoscopic imaging systems using a pelvitrainer model: a randomized controlled study.

Authors:  Tibor A Zwimpfer; Claudine Wismer; Bernhard Fellmann-Fischer; James Geiger; Andreas Schötzau; Viola Heinzelmann-Schwarz
Journal:  Updates Surg       Date:  2021-10-26

Review 6.  Endoscopic vitreoretinal surgery: principles, applications and new directions.

Authors:  Radwan S Ajlan; Aarsh A Desai; Martin A Mainster
Journal:  Int J Retina Vitreous       Date:  2019-06-18

7.  3D laparoscopy does not reduce operative duration or errors in day-case laparoscopic cholecystectomy: a randomised controlled trial.

Authors:  Katie E Schwab; Nathan J Curtis; Martin B Whyte; Ralph V Smith; Timothy A Rockall; Karen Ballard; Iain C Jourdan
Journal:  Surg Endosc       Date:  2019-07-16       Impact factor: 4.584

8.  Is 3D faster and safer than 4K laparoscopic cholecystectomy? A randomised-controlled trial.

Authors:  Matt Dunstan; Ralph Smith; Katie Schwab; Andrea Scala; Piers Gatenby; Martin Whyte; Tim Rockall; Iain Jourdan
Journal:  Surg Endosc       Date:  2019-07-18       Impact factor: 4.584

9.  A pilot study comparing ergonomics in laparoscopy and robotics: beyond anecdotes, and subjective claims.

Authors:  Li-Jen Kuo; James Chi-Yong Ngu; Yen-Kuang Lin; Chia-Che Chen; Yue-Her Tang
Journal:  J Surg Case Rep       Date:  2020-02-21

10.  A laparoscopic study investigating 3D vs 2D imaging systems using a pelvitrainer model with experts, non-experts, and students.

Authors:  Tibor Andrea Zwimpfer; Dominik Lacher; Bernhard Fellmann-Fischer; Michael Mueller
Journal:  BMC Surg       Date:  2020-11-09       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.