In vitro, contrast agent-based evaluation of the influence of flow diverter size and position on intra-aneurysmal flow dynamics using syngo iFlow.

Giorgio Franco Maria Cattaneo1, Andreas Ding2, Tobias Jost2, Désirée Ley3, Ruben Mühl-Bennighaus3, Umut Yilmaz3, Heiko Körner3, Wolfgang Reith3, Andreas Simgen3.   

Abstract

PURPOSE: Treatment of intracranial aneurysm with flow-diverting devices has become widespread in recent years. Despite that, intra-aneurysmal flow changes are yet not fully understood and can lead to different complications. Our aim was an in vitro contrast-based evaluation of the influence of flow diverter size and position on intra-aneurysmal flow dynamics.
METHODS: Flow-diverting devices with different sizes (diameters 4.0, 4.5, and 6.0 mm) were deployed in seven silicone aneurysm models at different positions relative to the aneurysm neck (proximal, central, distal). Using syngo iFlow, we defined quantitative evaluation criteria based on contrast medium intensity and performed a flow evaluation.
RESULTS: Intra-aneurysmal flows were heavily dependent on both size and position of flow-diverting devices at the aneurysm neck. We observed a higher peak intensity delay and intra-aneurysmal washout delay with the centrally placed 4.0- and 4.5-mm device, respectively, compared to the proximal and distal positions. Especially distally placed 4.0-mm devices led to an earlier filling of the aneurysm and increased intra-aneurysmal contrast agent intensity compared to the parent vessel, due to a potential endoleak.
CONCLUSIONS: Not only size but also position of flow-diverting devices have a considerable impact on the intra-aneurysmal flow dynamics. The suggested evaluation criteria allowed a quantitative comparison of flow-diverting effect using syngo iFlow and could represent an efficient tool for predicting flow diversion pre-procedurally.

Entities:  

Keywords:  Flow diverter; Intracranial aneurysms; Stenting; iFlow

Mesh:

Substances:

Year:  2017        PMID: 28842742     DOI: 10.1007/s00234-017-1903-2

Source DB:  PubMed          Journal:  Neuroradiology        ISSN: 0028-3940            Impact factor:   2.804


  28 in total

1.  An in vitro study of silk stent morphology.

Authors:  Thaweesak Aurboonyawat; Raphaël Blanc; Paul Schmidt; Michel Piotin; Laurent Spelle; Amir Nakib; Jacques Moret
Journal:  Neuroradiology       Date:  2010-10-28       Impact factor: 2.804

2.  Aneurysm rupture following treatment with flow-diverting stents: computational hemodynamics analysis of treatment.

Authors:  J R Cebral; F Mut; M Raschi; E Scrivano; R Ceratto; P Lylyk; C M Putman
Journal:  AJNR Am J Neuroradiol       Date:  2010-11-11       Impact factor: 3.825

3.  In-room assessment of cerebral blood volume for guidance during intra-arterial thrombolytic therapy.

Authors:  C-J Lin; M Yu; S-C Hung; M M H Teng; W-Y Guo; F-C Chang; C-B Luo; W-F Chu; C-Y Chang
Journal:  Interv Neuroradiol       Date:  2012-12-03       Impact factor: 1.610

4.  Distant and delayed (>7 days) hemorrhage after treatment by flow-diverter stents in intracranial aneurysms: a rare but potentially serious complication.

Authors:  F Clarençon; F Di Maria; A Biondi; J Chiras; N-A Sourour
Journal:  AJNR Am J Neuroradiol       Date:  2013-04-25       Impact factor: 3.825

5.  Parametric color coding of digital subtraction angiography in the evaluation of carotid cavernous fistulas.

Authors:  P Gölitz; T Struffert; H Lücking; J Rösch; F Knossalla; O Ganslandt; Y Deuerling-Zheng; A Doerfler
Journal:  Clin Neuroradiol       Date:  2012-12-14       Impact factor: 3.649

6.  Flow diversion treatment: intra-aneurismal blood flow velocity and WSS reduction are parameters to predict aneurysm thrombosis.

Authors:  Zsolt Kulcsár; Luca Augsburger; Philippe Reymond; Vitor M Pereira; Sven Hirsch; Ajit S Mallik; John Millar; Stephan G Wetzel; Isabel Wanke; Daniel A Rüfenacht
Journal:  Acta Neurochir (Wien)       Date:  2012-08-29       Impact factor: 2.216

7.  Analysis of hemodynamics and aneurysm occlusion after flow-diverting treatment in rabbit models.

Authors:  J R Cebral; F Mut; M Raschi; S Hodis; Y-H Ding; B J Erickson; R Kadirvel; D F Kallmes
Journal:  AJNR Am J Neuroradiol       Date:  2014-04-10       Impact factor: 3.825

8.  Treatment of 14 intracranial aneurysms with the FRED system.

Authors:  Orlando Diaz; Taylor L Gist; Ginna Manjarez; Fernando Orozco; Rafael Almeida
Journal:  J Neurointerv Surg       Date:  2013-09-23       Impact factor: 5.836

9.  The DERIVO Embolization Device in the Treatment of Intracranial Aneurysms: Short- and Midterm Results.

Authors:  Erol Akgul; Hasan Bilen Onan; Suha Akpinar; Huseyin Tugsan Balli; Erol Huseyin Aksungur
Journal:  World Neurosurg       Date:  2016-08-08       Impact factor: 2.104

Review 10.  Delayed hemorrhagic complications after flow diversion for intracranial aneurysms: a literature overview.

Authors:  Aymeric Rouchaud; Waleed Brinjikji; Giuseppe Lanzino; Harry J Cloft; Ramanathan Kadirvel; David F Kallmes
Journal:  Neuroradiology       Date:  2015-11-09       Impact factor: 2.804

View more
  6 in total

1.  2D parametric contrast time-density analysis for the prediction of complete aneurysm occlusion at six months' post-flow diversion stent.

Authors:  Ahmed E Hussein; Meghana Shownkeen; Andre Thomas; Christopher Stapleton; Denise Brunozzi; Jessica Nelson; John Naumgart; Andreas Linninger; Gursant Atwal; Ali Alaraj
Journal:  Interv Neuroradiol       Date:  2020-02-26       Impact factor: 1.610

2.  Quantitative color-coded digital subtraction neuroangiography for pediatric arteriovenous shunting lesions.

Authors:  Grace M Y Ma; Adam A Dmytriw; Premal A Patel; Nicholas Shkumat; Timo Krings; Manohar M Shroff; Prakash Muthusami
Journal:  Childs Nerv Syst       Date:  2019-07-06       Impact factor: 1.475

3.  Angiographic Characteristics of Cerebral Perfusion and Hemodynamics of the Bridging Artery After Surgical Treatment of Unilateral Moyamoya Disease.

Authors:  Kun Zhang; Wei Ren; Yu-Xue Sun; Xin-Jun Wang; Chao-Yue Li; Zi-Liang Wang; Tian-Xiao Li; Bu-Lang Gao
Journal:  Front Neurosci       Date:  2022-06-14       Impact factor: 5.152

4.  Retrospective analysis of intracranial aneurysms after flow diverter treatment including color-coded imaging (syngo iFlow) as a predictor of aneurysm occlusion.

Authors:  Andreas Simgen; Christine Mayer; Michael Kettner; Ruben Mühl-Benninghaus; Wolfgang Reith; Umut Yilmaz
Journal:  Interv Neuroradiol       Date:  2021-06-09       Impact factor: 1.764

5.  Reliability and Accuracy of Peri-Interventional Stenosis Grading in Peripheral Artery Disease Using Color-Coded Quantitative Fluoroscopy: A Phantom Study Comparing a Clinical and Scientific Postprocessing Software.

Authors:  Patrick Ghibes; Sasan Partovi; Gerd Grözinger; Petros Martirosian; Fritz Schick; Konstantin Nikolaou; Dominik Ketelsen; Roland Syha; Ulrich Grosse
Journal:  Biomed Res Int       Date:  2018-07-24       Impact factor: 3.411

6.  Quantitative Evaluation of Peripheral Arterial Blood Flow Using Peri-Interventional Fluoroscopic Parameters: An In Vivo Study Evaluating Feasibility and Clinical Utility.

Authors:  Patrick Ghibes; Gerald Hefferman; Konstantin Nikolaou; Roland Syha; Christoph Artzner; Ulrich Grosse; Rüdiger Hoffmann; Gerd Grözinger
Journal:  Biomed Res Int       Date:  2020-01-16       Impact factor: 3.411

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.