| Literature DB >> 28835607 |
Markandey M Tripathi1, Satoru Egawa1,2, Alexandra G Wirth1, Diane M Tshikudi1, Elizabeth M Van Cott3, Seemantini K Nadkarni4.
Abstract
Prothrombin time (PT) and the associated international normalized ratio (INR) are routinely tested to assess the risk of bleeding or thrombosis and to monitor response to anticoagulant therapy in patients. To measure PT/INR, conventional coagulation testing (CCT) is performed, which is time-consuming and requires the separation of cellular components from whole blood. Here, we report on a portable and battery-operated optical sensor that can rapidly quantify PT/INR within seconds by measuring alterations in the viscoelastic properties of a drop of whole blood following activation of coagulation with thromboplastin. In this study, PT/INR values were measured in 60 patients using the optical sensor and compared with the corresponding CCT values. Our results report a close correlation and high concordance between PT/INR measured using the two approaches. These findings confirm the accuracy of our optical sensing approach for rapid PT/INR testing in whole blood and highlight the potential for use at the point-of-care or for patient self-testing.Entities:
Mesh:
Year: 2017 PMID: 28835607 PMCID: PMC5569083 DOI: 10.1038/s41598-017-08693-5
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1LSR-based coagulation sensor: (A) Photograph of the hand-held sensor and compatible Window surface tablet interface, (B) Computer aided diagram of the optical and mechanical device configuration. Light from a 690 nm diode laser (DL) was focused (spot size 100 μm) by a lens on the disposable test cartridge (IC) containing 40 μL of thromboplastin-activated whole blood. Cross-polarized laser speckle patterns were acquired at 180° back-scattering geometry via a beam-splitter (BS) using a USB CMOS camera (CM) equipped with an imaging optics consist of a linear polarizer, a 500 µm aperture and an f 9 mm focusing lens. Beam dump (BD) dumps the laser beam transmitted through the beam splitter (BS). A miniature heating element (HP), temperature controller (TM) and a custom cartridge tray was incorporated within the hand-held sensor. The captured speckle patterns were transferred to a Microsoft SurfaceTM tablet computer for further processing. (C) Inexpensive test cartridges were fabricated by laser cutting a silicone base with a transparent polycarbonate film overlay to sample a drop of whole blood (40 μL).
Figure 2Viscoelastic modulus, G, at 5 Hz measured by the LSR sensor is plotted as a function of coagulation time from blood samples obtained from a normal patient (with reported normal laboratory PT/INR value) and a patient on Coumadin medication. The G trace measured in real-time during coagulation was analyzed to estimate the PT and INR values. The blood sample from normal patient showed shorter PTLSR value (12 s) in comparison to the blood sample from the patient on Coumadin therapy (51 s). The corresponding PTLab values obtained from standard laboratory testing were 14 s and 43 s for the normal and Coumadin treated patient respectively.
Quality assessment test for LSR sensor.
| Reference PT Range | PTLSR | |
|---|---|---|
| Level 1 | 10–13 Sec | (12.1 ± 0.6) Sec |
| Level 2 | 15–20 Sec | (17.3 ± 0.9) Sec |
| Level 3 | 22–28 Sec | (26.9 ± 1.3) Sec |
Repeatability testing of PT/INR measurement with LSR sensor.
| Normal PT/INR | High PT/INR | |
|---|---|---|
| PTLAB | 13.4 | 27.2 |
| INRLAB | 1.08 | 2.52 |
| PTLSR, mean ± SD (%CV) | 13.7 ± 0.6 (4.4%) | 28.0 ± 1.4 (4.9%) |
| INRLSR, mean ± SD (%CV) | 1.08 ± 0.05 (5.2%) | 2.52 ± 0.14 (5.7%) |
(n = 10 measurements per blood sample, %CV was calculated by dividing the SD by mean).
Figure 3The plots show (A) PT and (B) INR measured using the LSR sensor against corresponding values obtained from standard laboratory tests measured from 60 patient whole blood samples. A strong statistically significant correlation between the two approaches (R = 0.94, p < 0.001) is observed. Bland-Altman plots comparing (C) PT and (D) INR measured with the LSR sensor and standard laboratory results show high concordance between both test measurements. Upper and lower dotted lines represent 95% limit of agreement (LOA) while the middle line represents bias that shows the average difference in the values measured with LSR sensor and laboratory test.