Comparison of plasma with whole blood prothrombin time and fibrinogen on the same instrument.

Timothy K Amukele1, Chris Ferrell, Wayne L Chandler.   

Abstract

We compared plasma with whole blood (WB) international normalized ratio (INR) and fibrinogen using the same instrument and reagents. WBINRs were 50% higher than plasma INRs. After increasing the WB sample volume 40% and adjusting the International Sensitivity Index, WBINRs were similar to plasma INRs [adjusted WBINR = 0.99(plasma INR) - 0.02; r(2) = 0.98; n = 155], but the average difference in WB vs plasma INR was 4-fold higher than duplicate plasma INRs. Variation in hematocrit was a major determinant of the accuracy of the WBINR, with increased error at high INRs. The WB fibrinogen assay was highly dependent on the sample hematocrit (r(2) = 0.83), even after the sample volume was adjusted. Accurate WB fibrinogen measurements required a mathematical hematocrit correction. We conclude that WBINR and fibrinogen assays can be performed on point-of-care or automated analyzers, but sample volume must be adjusted to account for hematocrit. Accuracy is limited by variations in hematocrit with worsening accuracy for samples with high INRs or low fibrinogen levels.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20231607     DOI: 10.1309/AJCPLDT9OVX1TDGT

Source DB:  PubMed          Journal:  Am J Clin Pathol        ISSN: 0002-9173            Impact factor:   2.493


  4 in total

Review 1.  Clinical review: Canadian National Advisory Committee on Blood and Blood Products--Massive transfusion consensus conference 2011: report of the panel.

Authors:  Walter H Dzik; Morris A Blajchman; Dean Fergusson; Morad Hameed; Blair Henry; Andrew W Kirkpatrick; Teresa Korogyi; Sarvesh Logsetty; Robert C Skeate; Simon Stanworth; Charles MacAdams; Brian Muirhead
Journal:  Crit Care       Date:  2011-12-08       Impact factor: 9.097

2.  Rapid measurement of fibrinogen concentration in whole blood using a steel ball coagulometer.

Authors:  Christoph J Schlimp; Anna Khadem; Anton Klotz; Cristina Solomon; Gerald Hochleitner; Martin Ponschab; Heinz Redl; Herbert Schöchl
Journal:  J Trauma Acute Care Surg       Date:  2015-04       Impact factor: 3.313

3.  Clinical evaluation of whole blood prothrombin time (PT) and international normalized ratio (INR) using a Laser Speckle Rheology sensor.

Authors:  Markandey M Tripathi; Satoru Egawa; Alexandra G Wirth; Diane M Tshikudi; Elizabeth M Van Cott; Seemantini K Nadkarni
Journal:  Sci Rep       Date:  2017-08-23       Impact factor: 4.379

4.  Micro-mechanical blood clot testing using smartphones.

Authors:  Justin Chan; Kelly Michaelsen; Joanne K Estergreen; Daniel E Sabath; Shyamnath Gollakota
Journal:  Nat Commun       Date:  2022-02-11       Impact factor: 14.919

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.