Hannele Niiniviita1,2, Jarmo Kulmala1,3, Tuukka Pölönen4, Heli Määttänen1, Hannu Järvinen5, Eeva Salminen3,5.
Abstract
BACKGROUND: The aim of the study was to assess patient dose from whole-body computed tomography (CT) in association with patient size, automatic exposure control (AEC) and intravenous (IV) contrast agent. PATIENTS AND METHODS: Sixty-five testicular cancer patients (mean age 28 years) underwent altogether 279 whole-body CT scans from April 2000 to April 2011. The mean number of repeated examinations was 4.3. The GE LightSpeed 16 equipped with AEC and the Siemens Plus 4 CT scanners were used for imaging. Whole-body scans were performed with (216) and without (63) IV contrast. The ImPACT software was used to determine the effective and organ doses.Entities:
Keywords: automatic exposure control; computed tomography; contrast agent; radiation exposure; waist circumference
Year: 2017 PMID: 28740460 PMCID: PMC5514665 DOI: 10.1515/raon-2017-0012
Source DB: PubMed Journal: Radiol Oncol ISSN: 1318-2099 Impact factor: 2.991
Patient characteristics at the time of the first whole-body scan
| Characteristic | Value |
|---|---|
| Number of patients | 65 |
| Median age (years) | 28 (range 16.5-39.4) |
| Number of patients with seminoma/non-seminoma | 26/39 |
| Median waist circumference (cm) | 92.2 (range 66.9-126.3) |
| Total number of whole-body CTEs | 279 |
| Mean number of whole-body CTEs/patient | 4.3 (SD 2.6) |
| Number of whole body CTEs
- LightSpeed 16 - Plus 4 | 137 (49.1%) 142 (50.2%) |
| Number of whole-body CTE with IV contrast -LightSpeed 16 -Plus 4 | 104 (75.9) 112 (78.9) |
CTE = computed tomography examinations
The waist circumference was measured midway between the lowest rib and the iliac crest from one exial CT image.
Figure 1Boxplot of effective dose by device. The difference between devices is highly significant (p < 0.0001, Wilcoxon’s).
Figure 2Effective dose versus waist circumference by scanner type in whole-body examinations. A wider variation of effective doses is observed for the LightSpeed 16 device.
Mean effective dose and SD (mSv) by device and waist circumference with and without IV contrast
| LightSpeed 16 | Plus 4 | |||
|---|---|---|---|---|
| Waist circumference | < 100 cm | ≥ 100 cm | < 100 cm | ≥ 100 cm |
| With IV | 13.6 (3.2) | 17.5 (6.1) | 7.4 (1.2) | 8.5 (3.3) |
| Without IV | 10.7 (2.0) | 12.8 (1.9) | 6.2 (0.5) | 7.7 (2.5) |
| All | 12.8 (3.2) | 16.7 (5.9) | 7.2 (1.2) | 8.2 (3.0) |
The p-value (Wilcoxon two-sample test) between waist circumference over and under 100 cm p=0.003
The p-value (Wilcoxon two-sample test) between waist circumference over and under 100 cm p=0.003