Comparison of Z-axis automatic tube current modulation technique with fixed tube current CT scanning of abdomen and pelvis.

Mannudeep K Kalra1, Michael M Maher, Thomas L Toth, Ravi S Kamath, Elkan F Halpern, Sanjay Saini.   

Abstract

PURPOSE: To compare image quality, diagnostic acceptability, and radiation exposure associated with 16-section multi-detector row computed tomographic (CT) examinations of abdomen and pelvis performed with z-axis modulation technique of automatic tube current modulation and with manual selection of fixed tube current.
MATERIALS AND METHODS: Sixty-two consecutive subjects (mean age, 60 years; age range, 19-84 years; male-to-female ratio, 35:27) underwent follow-up CT of abdomen and pelvis with use of a 16-section multi-detector row scanner and z-axis modulation technique (10.5-12.0-HU noise index, 10-380 mA). Scanning parameters included 140 kVp, 0.5-1.0-second gantry rotation time, 0.938:1 beam pitch, and 5-mm reconstructed section thickness. For each subject, images obtained with z-axis modulation were compared with previous images obtained with fixed tube current (200-300 mA) and with other parameters identical. Images were compared for noise and diagnostic acceptability by two subspecialty radiologists using a five-point scale (1, unacceptable; 3, acceptable; 5, excellent) at five levels: upper liver at diaphragm, porta hepatis, right kidney hilum, iliac crest, and upper margin of acetabulum. Tube current and gantry rotation time used for acquisitions at these levels were recorded. Data were analyzed with parametric and nonparametric statistical tests.
RESULTS: Although no significant differences were found (P =.34), images acquired with z-axis modulation at the levels of the upper liver (diaphragm) and acetabulum had a higher noise and lower diagnostic quality, compared with images acquired with fixed tube current. Compared with fixed tube current, z-axis modulation resulted in tube current-time product reduction in 54 (87%) of 62 examinations (mean reduction, 71.2 mAs) and increase in eight (13%) (mean increase, 17.0 mAs).
CONCLUSION: Compared with manually selected fixed tube current, z-axis automatic tube current modulation resulted in reduced tube current-time product and similar image noise and diagnostic acceptability at CT of abdomen and pelvis. Copyright RSNA, 2004

Entities:  

Mesh:

Year:  2004        PMID: 15286306     DOI: 10.1148/radiol.2322031304

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  35 in total

1.  Current status of low dose multi-detector CT in the urinary tract.

Authors:  Mi Kim Sung; Sarabjeet Singh; Mannudeep K Kalra
Journal:  World J Radiol       Date:  2011-11-28

Review 2.  Quantitation of stone burden: imaging advances.

Authors:  Terri J Vrtiska
Journal:  Urol Res       Date:  2005-11-13

Review 3.  Reducing radiation dose in emergency computed tomography with automatic exposure control techniques.

Authors:  Mannudeep K Kalra; Stefania M R Rizzo; Robert A Novelline
Journal:  Emerg Radiol       Date:  2005-05-12

4.  Optimization of Z-axis automatic exposure control for multidetector row CT evaluation of neck and comparison with fixed tube current technique for image quality and radiation dose.

Authors:  S Namasivayam; M K Kalra; K M Pottala; S M Waldrop; P A Hudgins
Journal:  AJNR Am J Neuroradiol       Date:  2006 Nov-Dec       Impact factor: 3.825

5.  Radial differential interior tomography and its image reconstruction with differentiated backprojection and projection onto convex sets.

Authors:  Shaojie Tang; Xiangyang Tang
Journal:  Med Phys       Date:  2013-09       Impact factor: 4.071

6.  Radiation dose in CT colonography--trends in time and differences between daily practice and screening protocols.

Authors:  M H Liedenbaum; H W Venema; J Stoker
Journal:  Eur Radiol       Date:  2008-05-20       Impact factor: 5.315

7.  Automatic exposure control in CT: the effect of patient size, anatomical region and prescribed modulation strength on tube current and image quality.

Authors:  Antonios E Papadakis; Kostas Perisinakis; John Damilakis
Journal:  Eur Radiol       Date:  2014-07-17       Impact factor: 5.315

8.  CT dose index and patient dose: they are not the same thing.

Authors:  Cynthia H McCollough; Shuai Leng; Lifeng Yu; Dianna D Cody; John M Boone; Michael F McNitt-Gray
Journal:  Radiology       Date:  2011-05       Impact factor: 11.105

9.  Practical interior tomography with radial Hilbert filtering and a priori knowledge in a small round area.

Authors:  Shaojie Tang; Yi Yang; Xiangyang Tang
Journal:  J Xray Sci Technol       Date:  2012       Impact factor: 1.535

10.  Computed tomography colonography (virtual colonoscopy): climax of a new era of validation and transition into community practice.

Authors:  Jacob Thomas; Jeffrey Carenza; Elizabeth McFarland
Journal:  Clin Colon Rectal Surg       Date:  2008-08
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.