Comparative clinical study of Whitestar Signature phacoemulsification system with standard and Ellips FX handpieces.

Daniele Tognetto1,2, Rossella D'Aloisio3, Paolo Cecchini3, Marta Di Nicola4, Giuseppe Di Martino5.   

Abstract

PURPOSE: To compare the intraoperative parameters and postoperative outcomes using two different phaco handpieces with different energy delivery systems of the same phacoemulsification machine.
METHODS: One hundred and sixty eyes of 160 cataract patients were randomly assigned to 2.40 mm phacoemulsification using the AMO WhiteStar Signature® phaco system with the standard phaco handpiece (Group 1) or the Ellips FX phaco handpiece (Group 2). According to the lens opacities classification system III, each group was divided into four subgroups of 20 patients each. U/S total time (UST), total phaco time (TPT), mean effective phaco time, cumulative dissipated energy (CDE) and balanced salt solution (BSS) consumption were studied. Visual acuity, corneal endothelium cell count and corneal pachymetry were evaluated before surgery and 1, 7 and 30 days after surgery.
RESULTS: No statistically significant difference in terms of visual acuity was observed between the two groups. The postoperative endothelial cell loss and corneal pachymetry were similar between the two groups. In higher density cataracts, Ellips FX showed the lower CDE and the lowest UST. TPT and BSS consumption were significantly lower in the Ellips FX handpiece group.
CONCLUSIONS: The two phaco handpieces appeared to be safe. AMO Ellips FX handpiece showed better performances and may be safer in lens removal, especially in hard cataract patients.

Entities:  

Keywords:  Cataract surgery; Corneal endothelium; Phaco handpieces; Phacoemulsification; Visual acuity

Mesh:

Year:  2017        PMID: 28707255     DOI: 10.1007/s10792-017-0649-5

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


  22 in total

1.  Phacoemulsification using the burst mode.

Authors:  Daniel Badoza; Julio Fernández Mendy; Marcela Ganly
Journal:  J Cataract Refract Surg       Date:  2003-06       Impact factor: 3.351

Review 2.  New phacoemulsification technology.

Authors:  Richard S Hoffman; I Howard Fine; Mark Packer
Journal:  Curr Opin Ophthalmol       Date:  2005-02       Impact factor: 3.761

3.  Microincisional cataract surgery (MICS) with pulse and burst modes.

Authors:  M Baykara; I Ercan; H Ozcetin
Journal:  Eur J Ophthalmol       Date:  2006 Nov-Dec       Impact factor: 2.597

4.  Torsional mode versus conventional ultrasound mode phacoemulsification: randomized comparative clinical study.

Authors:  Yizhi Liu; Mingbing Zeng; Xialin Liu; Lixia Luo; Zhaohui Yuan; Yuanlin Xia; Yangfa Zeng
Journal:  J Cataract Refract Surg       Date:  2007-02       Impact factor: 3.351

5.  Comparison of torsional and longitudinal modes using phacoemulsification parameters.

Authors:  Marek Rekas; Robert Montés-Micó; Karolina Krix-Jachym; Adam Kluś; Andrzej Stankiewicz; Teresa Ferrer-Blasco
Journal:  J Cataract Refract Surg       Date:  2009-10       Impact factor: 3.351

Review 6.  The phaco machine: analysing new technology.

Authors:  William J Fishkind
Journal:  Curr Opin Ophthalmol       Date:  2013-01       Impact factor: 3.761

7.  Emulsification action of longitudinal and torsional ultrasound tips and the effect on treatment of the nucleus during phacoemulsification.

Authors:  Teruyuki Miyoshi; Hironori Yoshida
Journal:  J Cataract Refract Surg       Date:  2010-07       Impact factor: 3.351

8.  Comparison of two different ultrasound methods of phacoemulsification.

Authors:  Firat Helvacioglu; Celal Yeter; Sadik Sencan; Zeki Tunc; O Murat Uyar
Journal:  Am J Ophthalmol       Date:  2014-04-30       Impact factor: 5.258

9.  Comparison of burst, pulse, and linear modes used in phacoemulsification surgery.

Authors:  Yelda B Ozkurt; Tufan Evciman; Tomris Sengor; Melih Haboglu; Gokcen Bas; Sevda Kurna; Suat Aki
Journal:  Eur J Ophthalmol       Date:  2010 Mar-Apr       Impact factor: 2.597

10.  Comparison of endothelial changes and power settings between torsional and longitudinal phacoemulsification.

Authors:  Anna Reuschel; Holger Bogatsch; Thomas Barth; Renate Wiedemann
Journal:  J Cataract Refract Surg       Date:  2010-11       Impact factor: 3.351

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.