Emulsification action of longitudinal and torsional ultrasound tips and the effect on treatment of the nucleus during phacoemulsification.

Teruyuki Miyoshi1, Hironori Yoshida.   

Abstract

PURPOSE: To compare the results of emulsification of test materials when applying torsional and longitudinal emulsification tips as well as to define emulsification and determine how it occurs by examining the effect of phacoemulsification.
SETTING: Miyoshi Eye Hospital, Fukuyama, Japan.
METHOD: An ultra-high-speed digital video camera was used to record emulsification of chestnuts and human nuclei in a test chamber, simulating real phacoemulsification. The treatment of the nucleus as a result of phacoemulsification was analyzed visually.
RESULTS: Visual comparison of video data showed that the test chamber materials were completely desiccated by the torsional phaco device. Longitudinal phacoemulsification resulted in cutting and coring of the test chamber materials, but not total disruption.
CONCLUSIONS: Visual comparisons showed that complete obliteration and desiccation of the nucleus, or emulsification, occurs only with torsional phacoemulsification. These observations open the way to understanding the best approach to new tip design and surgical strategy development. FINANCIAL DISCLOSURE: Neither author has a financial or proprietary interest in any material or method mentioned. Copyright 2010 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20610101     DOI: 10.1016/j.jcrs.2010.01.028

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  Comparative clinical study of Whitestar Signature phacoemulsification system with standard and Ellips FX handpieces.

Authors:  Daniele Tognetto; Rossella D'Aloisio; Paolo Cecchini; Marta Di Nicola; Giuseppe Di Martino
Journal:  Int Ophthalmol       Date:  2017-07-13       Impact factor: 2.031

2.  Comparison of 45-degree Kelman and 45-degree balanced phaco tip designs in torsional microcoaxial phacoemulsification.

Authors:  Süleyman Demircan; Mustafa Ataş; Emre Göktaş; Burhan Başkan
Journal:  Int J Ophthalmol       Date:  2015-12-18       Impact factor: 1.779

3.  Effect of IOP based infusion system with and without balanced phaco tip on cumulative dissipated energy and estimated fluid usage in comparison to gravity fed infusion in torsional phacoemulsification.

Authors:  Praveen K Malik; Taru Dewan; Arun Kr Patidar; Ekta Sain
Journal:  Eye Vis (Lond)       Date:  2017-09-17
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.