Literature DB >> 28702482

Visualization of simulated small vessels on computed tomography using a model-based iterative reconstruction technique.

Toru Higaki1, Fuminari Tatsugami1, Chikako Fujioka2, Hiroaki Sakane1, Yuko Nakamura1, Yasutaka Baba1, Makoto Iida1, Kazuo Awai1.   

Abstract

This article describes a quantitative evaluation of visualizing small vessels using several image reconstruction methods in computed tomography. Simulated vessels with diameters of 1-6 mm made by 3D printer was scanned using 320-row detector computed tomography (CT). Hybrid iterative reconstruction (hybrid IR) and model-based iterative reconstruction (MBIR) were performed for the image reconstruction.

Entities:  

Keywords:  CT angiography; Computed tomography (CT); Hybrid iterative reconstruction; Image reconstruction; Model-based iterative reconstruction

Year:  2017        PMID: 28702482      PMCID: PMC5484979          DOI: 10.1016/j.dib.2017.06.024

Source DB:  PubMed          Journal:  Data Brief        ISSN: 2352-3409


Specifications Table [please fill in right-hand column of the table below] Value of the data The data has described the effect of MBIR for visualizing small vessels in CT images. Researchers can recognize the difference in the appearance of small vessels in various reconstruction method. Also researchers can recognize the difference in the appearance of small vessels in various diameters. This data shows the superiority of MBIR for visualization of small vessels.

Experimental design, materials and methods

Vessel phantom

The vessel phantom (outer diameter 80 mm) was made by a 3D printer (Agilista 3100, KEYENCE, Osaka, Japan) and included cylinders that simulated 1-, 2-, 3-, 4-, 5-, and 6-mm vessels (Fig. 1). The aorta was simulated by a 30-mm diameter cylinder at the center of the phantom. The cylinders were filled with diluted iodine contrast material (Iohexol, Daiichi-Sankyo, Tokyo, Japan, concentration 13 mgI/ml) to simulate the vascular space.
Fig. 1

Design of our vessel phantom. The cylinders measuring 1–6 mm in diameter simulate small vessels. The 30-mm diameter cylinder in the center simulates the ascending aorta.

Design of our vessel phantom. The cylinders measuring 1–6 mm in diameter simulate small vessels. The 30-mm diameter cylinder in the center simulates the ascending aorta.

CT Scanning protocol

CT scans were obtained on a 320-detector scanner (Aquilion ONE ViSION Ed., Toshiba Medical Systems Corp., Tokyo, Japan) with ECG-triggering. Scanning was at a tube voltage of 120 kV in volume-scanning mode. The tube current was set at 60 mA to emulate the routine imaging conditions (noise index 25 SD) for coronary CT angiography (CCTA). The other scanning parameters were 0.275 s rotation time, 0.5 mm slice thickness, 40 mm (0.5 mm×80 slices) z-coverage range, and 100-mm field of view.

CT Image reconstruction

Adaptive iterative dose reduction 3D (AIDR 3D, Toshiba Medical Systems Corp.) was used as the hybrid IR. To obtain AIDR 3D images we used three kinds of standard soft tissue convolution kernels (FC13, FC14 and FC15); they are recommended by the vendor for routine CCTA studies. Forward projected model-based iterative reconstruction solution (FIRST, Toshiba Medical Systems Corp.) with CCTA mode was used as the MBIR.

Data analysis and quantitative evaluation

To evaluate the visibility of simulated vessels we recorded their CT attenuation profile curves (APCs) on CTA images. We used ImageJ software [1] and its particle analysis tool (Plot Profile) to generate APCs and recorded 36 APCs radially around the vessel centers at 5° intervals. As the objective edge response index on APCs we determined the 10 - 90% edge rise distance (ERD) [2] and the 10–90% edge rise slope (ERS) at the vessel boundaries (Fig. 2). The distance and slope values were examined on both sides of the simulated vascular wall, hence, a total of 72 of ERDs and 72 ERSs was obtained and they were averaged in every simulated vessel. We also recorded the peak CT attenuation number in Hounsfield units (HU, PCT-HU) on the APCs and the CT attenuation value and the standard deviation (image noise) for the simulated aorta. The signal-to-noise ratio (SNR) for each vessel was calculated by dividing PCT-HU by the image noise.
Fig. 2

Definition of the edge rise distance (ERD) and the edge rise slope (ERS) on CT attenuation profile curves. ERD is defined as the distance between the site showing 10% and the site showing 90% of the CT number and ERS as the slope of the line connecting the 10% and the 90% HU site.

Definition of the edge rise distance (ERD) and the edge rise slope (ERS) on CT attenuation profile curves. ERD is defined as the distance between the site showing 10% and the site showing 90% of the CT number and ERS as the slope of the line connecting the 10% and the 90% HU site.

Data

The ERDs and ERSs measured on vessels with different diameters are shown in Figs. 3a and 3b. With FIRST the ERD was smaller and the ERS was larger than with AIDR 3D. The PCT-HU value of the vessels is shown in Fig. 3c. For 6-mm vessels the PCT-HU obtained with the different reconstruction methods was similar. On images reconstructed with AIDR 3D and any of the convolution kernels-, the smaller the vessel diameter, the lower was the PCT-HU value. On FIRST images, the PCT-HU value were constantly stable and highest at the vessel diameter of 2–5 mm. The aortic PCT-HU of images acquired with AIDR 3D-FC13, AIDR 3D-FC14, and AIDR 3D-FC15 and with FIRST was 412.3, 410.6, 410.8, and 411.1, respectively. Representative APCs are shown in Fig. 4. The image noise on those images was 21.5, 25.8, 30.6, and 24.2, respectively (Fig. 5); it was comparable on AIDR 3D-FC14- and FIRST images (24.2 and 25.8 HU, respectively).
Fig. 3

Results of attenuation profile curve analysis. On images reconstructed with FIRST, the ERD was smaller and the ERS was larger than on vessel images reconstructed with the other methods.

Fig. 4

Representative attenuation profile curves (APCs). The peak CT number of 2-mm vessels was strongly affected by the applied reconstruction algorithm; with FIRST we recorded the highest HU value. The HU value for 2-mm vessels was similar with all reconstruction algorithms we investigated.

Fig. 5

Representative images of our vessel phantom. The effect of the convolution kernels changed gradually from FC13 to FC15. With FC13 the image noise and image sharpness were the lowest; with FC15 they were the highest. The image noise on FIRST- and AIDR 3D-FC14 images was comparable.

Results of attenuation profile curve analysis. On images reconstructed with FIRST, the ERD was smaller and the ERS was larger than on vessel images reconstructed with the other methods. Representative attenuation profile curves (APCs). The peak CT number of 2-mm vessels was strongly affected by the applied reconstruction algorithm; with FIRST we recorded the highest HU value. The HU value for 2-mm vessels was similar with all reconstruction algorithms we investigated. Representative images of our vessel phantom. The effect of the convolution kernels changed gradually from FC13 to FC15. With FC13 the image noise and image sharpness were the lowest; with FC15 they were the highest. The image noise on FIRST- and AIDR 3D-FC14 images was comparable. As shown in Fig. 6, the SNR on AIDR 3D-FC14 and AIDR 3D-FC15 images of vessels was lower than on FIRST scans for vessels of all diameters examined. We obtained higher SNR values for vessels whose diameter exceeded 4 mm when AIDR 3D-FC13 rather than FIRST was applied.
Fig. 6

Signal-to-noise ratios (SNRs). The SNR of simulated vessels with a diameter of up to 3 mm was higher on FIRST- than on AIDR 3D-FC14 and AIDR 3D-FC15 images. For vessels with a diameter greater than 4 mm the SNR value was higher on AIDR 3D-FC13- than FIRST images.

Signal-to-noise ratios (SNRs). The SNR of simulated vessels with a diameter of up to 3 mm was higher on FIRST- than on AIDR 3D-FC14 and AIDR 3D-FC15 images. For vessels with a diameter greater than 4 mm the SNR value was higher on AIDR 3D-FC13- than FIRST images.
Subject areaRadiology
More specific subject areaEffect of Image reconstruction methods for small blood vessels in CT.
Type of dataImage, graph, text
How data was acquiredPhantom with simulated small vessels was scanned with CT, and it was reconstructed by hybrid IR and MBIR.
Data formatRaw, Analyzed
Experimental factorsThe sharpness of the blood vessel boundary was measured with a quantitative index.
Experimental featuresRadiation dose was determined by the routinely used noise level in coronary CT angiography. Adaptive Iterative Dose Reduction 3D (AIDR 3D) was used as the hybrid IR, Forward-projected model-based Iterative Reconstruction SoluTion (FIRST) was used as the MBIR.
Data source location1-2-3 Kasumi, Minami-ku, Hiroshima, 34° 22′ 44.4′′ N; 132° 28′ 38.26′′ E
Data accessibilityThe data are available with this article
  1 in total

1.  Measurement of vascular wall attenuation: comparison of CT angiography using model-based iterative reconstruction with standard filtered back-projection algorithm CT in vitro.

Authors:  Shigeru Suzuki; Haruhiko Machida; Isao Tanaka; Eiko Ueno
Journal:  Eur J Radiol       Date:  2012-03-19       Impact factor: 3.528

  1 in total
  7 in total

1.  Potential value of the PixelShine deep learning algorithm for increasing quality of 70 kVp+ASiR-V reconstruction pelvic arterial phase CT images.

Authors:  Shi-Feng Tian; Ai-Lian Liu; Jing-Hong Liu; Yi-Jun Liu; Ju-Dong Pan
Journal:  Jpn J Radiol       Date:  2018-12-06       Impact factor: 2.374

2.  Quantitative measurements of emphysema in ultra-high resolution computed tomography using model-based iterative reconstruction in comparison to that using hybrid iterative reconstruction.

Authors:  Shun Muramatsu; Kazuhiro Sato; Tsuneo Yamashiro; Kunio Doi
Journal:  Phys Eng Sci Med       Date:  2022-01-13

3.  Quantitative 3D dynamic contrast-enhanced (DCE) MR imaging of carotid vessel wall by fast T1 mapping using Multitasking.

Authors:  Nan Wang; Anthony G Christodoulou; Yibin Xie; Zhenjia Wang; Zixin Deng; Bill Zhou; Sangeun Lee; Zhaoyang Fan; Hyukjae Chang; Wei Yu; Debiao Li
Journal:  Magn Reson Med       Date:  2018-10-28       Impact factor: 4.668

4.  Deep Learning-based CT Image Reconstruction: Initial Evaluation Targeting Hypovascular Hepatic Metastases.

Authors:  Yuko Nakamura; Toru Higaki; Fuminari Tatsugami; Jian Zhou; Zhou Yu; Naruomi Akino; Yuya Ito; Makoto Iida; Kazuo Awai
Journal:  Radiol Artif Intell       Date:  2019-10-09

5.  The effect of sirolimus on angiomyolipoma is determined by decrease of fat-poor compartments and includes striking reduction of vascular structures.

Authors:  Elieser Hitoshi Watanabe; Fernando Morbeck Almeida Coelho; Hilton Leão Filho; Bruno Eduardo Pedroso Balbo; Precil Diego Miranda de Menezes Neves; Fernanda Maria Franzin; Fernando Ide Yamauchi; Luiz Fernando Onuchic
Journal:  Sci Rep       Date:  2021-04-19       Impact factor: 4.379

6.  Six-dimensional quantitative DCE MR Multitasking of the entire abdomen: Method and application to pancreatic ductal adenocarcinoma.

Authors:  Nan Wang; Srinivas Gaddam; Lixia Wang; Yibin Xie; Zhaoyang Fan; Wensha Yang; Richard Tuli; Simon Lo; Andrew Hendifar; Stephen Pandol; Anthony G Christodoulou; Debiao Li
Journal:  Magn Reson Med       Date:  2020-01-21       Impact factor: 4.668

7.  Image quality improvement with deep learning-based reconstruction on abdominal ultrahigh-resolution CT: A phantom study.

Authors:  Takashi Shirasaka; Tsukasa Kojima; Yoshinori Funama; Yuki Sakai; Masatoshi Kondo; Ryoji Mikayama; Hiroshi Hamasaki; Toyoyuki Kato; Yasuhiro Ushijima; Yoshiki Asayama; Akihiro Nishie
Journal:  J Appl Clin Med Phys       Date:  2021-06-23       Impact factor: 2.102

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.