Measurement of vascular wall attenuation: comparison of CT angiography using model-based iterative reconstruction with standard filtered back-projection algorithm CT in vitro.

Shigeru Suzuki1, Haruhiko Machida, Isao Tanaka, Eiko Ueno.   

Abstract

OBJECTIVES: To compare the performance of model-based iterative reconstruction (MBIR) with that of standard filtered back projection (FBP) for measuring vascular wall attenuation. STUDY
DESIGN: After subjecting 9 vascular models (actual attenuation value of wall, 89 HU) with wall thickness of 0.5, 1.0, or 1.5 mm that we filled with contrast material of 275, 396, or 542 HU to scanning using 64-detector computed tomography (CT), we reconstructed images using MBIR and FBP (Bone, Detail kernels) and measured wall attenuation at the center of the wall for each model. We performed attenuation measurements for each model and additional supportive measurements by a differentiation curve. We analyzed statistics using analyzes of variance with repeated measures.
RESULTS: Using the Bone kernel, standard deviation of the measurement exceeded 30 HU in most conditions. In measurements at the wall center, the attenuation values obtained using MBIR were comparable to or significantly closer to the actual wall attenuation than those acquired using Detail kernel. Using differentiation curves, we could measure attenuation for models with walls of 1.0- or 1.5-mm thickness using MBIR but only those of 1.5-mm thickness using Detail kernel. We detected no significant differences among the attenuation values of the vascular walls of either thickness (MBIR, P=0.1606) or among the 3 densities of intravascular contrast material (MBIR, P=0.8185; Detail kernel, P=0.0802).
CONCLUSIONS: Compared with FBP, MBIR reduces both reconstruction blur and image noise simultaneously, facilitates recognition of vascular wall boundaries, and can improve accuracy in measuring wall attenuation.
Copyright © 2012 Elsevier Ireland Ltd. All rights reserved.

Mesh:

Year:  2012        PMID: 22436433     DOI: 10.1016/j.ejrad.2012.02.009

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  4 in total

1.  Effects of adaptive statistical iterative reconstruction on radiation dose reduction and diagnostic accuracy of pediatric abdominal CT.

Authors:  Sohi Bae; Myung-Joon Kim; Choon-Sik Yoon; Dong Wook Kim; Jung Hwa Hong; Mi-Jung Lee
Journal:  Pediatr Radiol       Date:  2014-07-08

2.  Cerebral aneurysm pulsation: do iterative reconstruction methods improve measurement accuracy in vivo?

Authors:  T Illies; D Säring; M Kinoshita; T Fujinaka; M Bester; J Fiehler; N Tomiyama; Y Watanabe
Journal:  AJNR Am J Neuroradiol       Date:  2014-06-26       Impact factor: 3.825

3.  Computed tomography depiction of small pediatric vessels with model-based iterative reconstruction.

Authors:  Gonca Koc; Jesse L Courtier; Andrew Phelps; Peter A Marcovici; John D MacKenzie
Journal:  Pediatr Radiol       Date:  2014-02-16

4.  Visualization of simulated small vessels on computed tomography using a model-based iterative reconstruction technique.

Authors:  Toru Higaki; Fuminari Tatsugami; Chikako Fujioka; Hiroaki Sakane; Yuko Nakamura; Yasutaka Baba; Makoto Iida; Kazuo Awai
Journal:  Data Brief       Date:  2017-06-16
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.