Investigation of signal thresholding to reduce the effects of instrument noise of an EMCCD based micro-CT system.

Alexander R Podgorsak1,2, Sumukh Bysani Krishnakumar1,2, Sv Setlur Nagesh2, Daniel R Bednarek2, Stephen Rudin1,3,2, Ciprian N Ionita1,3,2.   

Abstract

This project investigated the signal thresholding effectiveness at reducing the instrument noise of an electron multiplying charged coupled device (EMCCD) based micro-CT system at low x-ray exposure levels. Scans of a mouse spine and an iodine phantom were taken using an EMCCD detector coupled with a micro-CT system. An iodine filter of 4 mg/cm2 area density was placed in the beam. The output signal was thresholded using some multiple of the inherent background noise. For each threshold, 100, 200, and 300 frames were summed for each projection to evaluate the effect on the reconstructed image. The projection images from the scans were compared using line profiles and their SNR. Our results indicate that, as the threshold was increased, the line profiles of the projection images showed less statistical variation, but also lower signal levels, so that the SNR of the projection images decreased as the threshold increased. When the line profile of a projection image obtained using a signal threshold is compared with one obtained using energy integrating mode, the profile obtained using thresholding had less variation than that obtained using energy integration, which indicates less instrument noise. The SNR at the edges of the scan object is higher in the thresholded images when compared with the energy integrated projection images. We conclude that thresholding the output signal from an EMCCD detector at low x-ray exposure levels is an effective method to reduce the instrument noise of an EMCCD detector.

Entities:  

Keywords:  EMCCD; Signal thresholding; instrument noise; micro-CT

Year:  2016        PMID: 28649164      PMCID: PMC5482266          DOI: 10.1117/12.2216272

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  4 in total

1.  Photon-counting versus an integrating CCD-based gamma camera: important consequences for spatial resolution.

Authors:  Freek J Beekman; Gerralt A de Vree
Journal:  Phys Med Biol       Date:  2005-05-25       Impact factor: 3.609

2.  Cone-beam micro-CT system based on LabVIEW software.

Authors:  Ciprian N Ionita; Keneth R Hoffmann; Daniel R Bednarek; Ravishankar Chityala; Stephen Rudin
Journal:  J Digit Imaging       Date:  2007-02-28       Impact factor: 4.056

3.  Progress in electron-multiplying CCD (EMCCD) based, high-resolution, high-sensitivity x-ray detector for fluoroscopy and radiography.

Authors:  Andrew T Kuhls; Girijesh Yadava; Vikas Patel; Daniel R Bednarek; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2007

4.  Self-calibration of a cone-beam micro-CT system.

Authors:  V Patel; R N Chityala; K R Hoffmann; C N Ionita; D R Bednarek; S Rudin
Journal:  Med Phys       Date:  2009-01       Impact factor: 4.071

  4 in total
  1 in total

1.  A comprehensive non-uniformity correction method for EMCCD.

Authors:  Li Qiao; Mingfu Wang; Zheng Jin; Danbo Mao
Journal:  Sci Rep       Date:  2021-12-14       Impact factor: 4.379

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.