Literature DB >> 28638159

In Vitro Evaluation of the Fresenius Kabi CATSmart Autotransfusion System.

Melissa Alberts1, Robert C Groom2, Richard Walczak1, Robert Kramer2, Adrienne Karpiel3, Jeanette Dieter3, Lisa Sheth3, Nathaniel H Greene4, Edmund H Jooste4.   

Abstract

Use of autotransfusion systems to collect, wash, and concentrate shed blood during surgical procedures is a widely used method for reducing postoperative anemia and the need for blood transfusions. The aim of this study was to evaluate the CATSmart Continuous Autotransfusion System wash program performance with small (200 or 700 mL) and large volumes (1,000 mL) of shed blood and to determine non-inferiority of the CATSmart to the C.A.T.S plus system. Human whole blood was collected in citrate phosphate dextrose, diluted, and divided into two aliquots to be processed as a pair using the C.A.T.S plus and CATSmart systems with their corresponding wash programs: low-volume, high quality/smart, or emergency wash. Final packed red cell product was analyzed for red blood cell (RBC), white blood cell, and platelet counts; hemoglobin; hemolysis; RBC recovery rates; and elimination of albumin, total protein, and potassium. The mean hematocrit (HCT) after processing with CATSmart and C.A.T.S plus systems were 59.63% and 57.71%, respectively. The calculated overall RBC recovery rates on the CATSmart and C.A.T.S plus systems were 85.41% and 84.99%, respectively. Elimination of albumin (97.5%, 98.0%), total proteins (97.1%, 97.5%), and potassium (92.1%, 91.9%) were also calculated for the CATSmart and C.A.T.S plus systems. The CATSmart and C.A.T.S plus systems both provided a high-quality product in terms of HCT, protein elimination, and hemolysis rates across the range of tested shed blood volumes and all wash programs. The study was able to confirm the CATSmart is non-inferior to the C.A.T.S plus system.

Entities:  

Keywords:  autotransfusion; blood conservation; cell salvage

Mesh:

Year:  2017        PMID: 28638159      PMCID: PMC5474889     

Source DB:  PubMed          Journal:  J Extra Corpor Technol        ISSN: 0022-1058


  9 in total

1.  Quality of processed blood for autotransfusion.

Authors:  G Shulman
Journal:  J Extra Corpor Technol       Date:  2000-03

2.  Quality of red blood cells using autotransfusion devices: a comparative analysis.

Authors:  Cyril J Serrick; Mary Scholz; Arthur Melo; Onkar Singh; Dionne Noel
Journal:  J Extra Corpor Technol       Date:  2003-03

3.  Processing of small volumes in blood salvage devices.

Authors:  Timo Seyfried; Anita Breu; Michael Gruber; Julia Reipert; Ernil Hansen
Journal:  Transfusion       Date:  2014-06-23       Impact factor: 3.157

Review 4.  2011 update to the Society of Thoracic Surgeons and the Society of Cardiovascular Anesthesiologists blood conservation clinical practice guidelines.

Authors:  Victor A Ferraris; Jeremiah R Brown; George J Despotis; John W Hammon; T Brett Reece; Sibu P Saha; Howard K Song; Ellen R Clough; Linda J Shore-Lesserson; Lawrence T Goodnough; C David Mazer; Aryeh Shander; Mark Stafford-Smith; Jonathan Waters; Robert A Baker; Timothy A Dickinson; Daniel J FitzGerald; Donald S Likosky; Kenneth G Shann
Journal:  Ann Thorac Surg       Date:  2011-03       Impact factor: 4.330

5.  Washing of banked blood by three different blood salvage devices.

Authors:  Michael Gruber; Anita Breu; Melanie Frauendorf; Timo Seyfried; Ernil Hansen
Journal:  Transfusion       Date:  2012-08-15       Impact factor: 3.157

6.  Fat elimination during intraoperative autotransfusion: an in vitro investigation.

Authors:  M Booke; M Fobker; D Fingerhut; M Storm; Y Mortlemans; H Van Aken
Journal:  Anesth Analg       Date:  1997-11       Impact factor: 5.108

Review 7.  Perioperative blood transfusion and blood conservation in cardiac surgery: the Society of Thoracic Surgeons and The Society of Cardiovascular Anesthesiologists clinical practice guideline.

Authors:  Victor A Ferraris; Suellen P Ferraris; Sibu P Saha; Eugene A Hessel; Constance K Haan; B David Royston; Charles R Bridges; Robert S D Higgins; George Despotis; Jeremiah R Brown; Bruce D Spiess; Linda Shore-Lesserson; Mark Stafford-Smith; C David Mazer; Elliott Bennett-Guerrero; Steven E Hill; Simon Body
Journal:  Ann Thorac Surg       Date:  2007-05       Impact factor: 4.330

8.  Predictors of transfusion requirements for cardiac surgical procedures at a blood conservation center.

Authors:  David M Moskowitz; James J Klein; Aryeh Shander; Katherine M Cousineau; Richard S Goldweit; Carol Bodian; Seth I Perelman; Hyun Kang; Daniel A Fink; Howard C Rothman; M Arisan Ergin
Journal:  Ann Thorac Surg       Date:  2004-02       Impact factor: 4.330

9.  Intraoperative autotransfusion. Experience in 725 consecutive cases.

Authors:  M M Keeling; L A Gray; M A Brink; V K Hillerich; K I Bland
Journal:  Ann Surg       Date:  1983-05       Impact factor: 12.969

  9 in total
  2 in total

1.  Bedside Allogeneic Erythrocyte Washing with a Cell Saver to Remove Cytokines, Chemokines, and Cell-derived Microvesicles.

Authors:  Ian J Welsby; Philip J Norris; William J Mauermann; Mihai V Podgoreanu; Chelsea M Conn; Laurie Meade; Tamara Cannon; Sheila M Keating; Christopher C Silliman; Marguerite Kehler; Phillip J Schulte; Daryl J Kor
Journal:  Anesthesiology       Date:  2021-03-01       Impact factor: 7.892

2.  Cell salvage using the continuous autotransfusion device CATSmart - an observational bicenter technical evaluation.

Authors:  Simone Lindau; Madeline Kohlhaas; Michael Nosch; Suma Choorapoikayil; Kai Zacharowski; Patrick Meybohm
Journal:  BMC Anesthesiol       Date:  2018-12-12       Impact factor: 2.217

  2 in total