| Literature DB >> 30541447 |
Simone Lindau1, Madeline Kohlhaas1, Michael Nosch2, Suma Choorapoikayil1, Kai Zacharowski1, Patrick Meybohm3.
Abstract
BACKGROUND: The use of cell salvage and autologous blood transfusion has become an important method of blood conservation. So far, there are no clinical data about the performance of the continuous autotransfusion device CATSmart.Entities:
Keywords: Auto transfusion; Cell salvage; Hematocrit value
Mesh:
Substances:
Year: 2018 PMID: 30541447 PMCID: PMC6292025 DOI: 10.1186/s12871-018-0651-0
Source DB: PubMed Journal: BMC Anesthesiol ISSN: 1471-2253 Impact factor: 2.217
Fig. 1Flowchart of participants
Fig. 2Red cell separation performance defined by hematocrit values. Data are presented as median [25%;75%]. Hematocrit (Hct) values are shown before (IN, shed blood) and after blood processing (OUT, packed red cells) by central laboratory (LAB) and CATSmart.
Fig. 3Hematocrit values in RBC concentrates. Hct values of the packed red cell concentrate measured by CATSmart and the laboratory (LAB) significantly correlated (r = 0.51; p ≤ 0.05).
Washout quality defined by removal ratio of blood parameters
| Parameters | RRmedian [25%;75%] (%) | RRmin (%) | RRmax (%) |
|---|---|---|---|
| K+ (mmol/l) | 92.5 [90.8;95.0] | 73.3 | 99.1 |
| Alb (g/dl) | 97.9 [96.6;98.5] | 90.7 | 100.0 |
| aXa (U/ml) | 99.9 [99.9;100.0] | 90.0 | 100.0 |
| fHb (g/dl) | 89.3 [85.2;94.9] | 45.8 | 98.5 |
| WBC (/nl) | 72.4 [57.9;87.3] | 9.0 | 95.9 |
| Plt (/nl) | 94.7 [88.2;96.7] | 63.4 | 99.6 |
Data are presented as median [25%;75%], minimum (min) and maximum (max). RR Removal ratio, K potassium, Alb albumin, aXa heparin, fHb free hemoglobin, WBC white blood cells, Plt platelets
Fig. 4Washout quality defined by removal ratio (%) of multiple blood parameters for each of the 50 analyzed patients. WBC = white blood cells; Plt = platelets; Alb = albumin; fHb = free hemoglobin; aXa = heparin; K+ = potassium.