Simone Mastrogiacomo1, Weiqiang Dou2, Olga Koshkina3, Otto C Boerman2, John A Jansen1, Arend Heerschap2, Mangala Srinivas3, X Frank Walboomers1. 1. Department of Biomaterials, Radboud University Medical Center , P.O. Box 9101, 6500 HB Nijmegen (309), The Netherlands. 2. Department of Radiology and Nuclear Medicine, Radboud University Medical Center , Geert Grooteplein Zuid 10, 6525 GA Nijmegen, The Netherlands. 3. Department of Tumor Immunology, Radboud Institute for Molecular Life Sciences (RIMLS) , Geert Grooteplein Zuid 28, 6525 GA Nijmegen, The Netherlands.
Abstract
Calcium phosphate cement (CPC) is used in bone repair because of its biocompatibility. However, high similarity between CPC and the natural osseous phase results in poor image contrast in most of the available in vivo imaging techniques such as computed tomography (CT) and magnetic resonance imaging (MRI). For accurate identification and localization during and after implantation in vivo, a composition with enhanced image contrast is needed. In this study, we labeled CPC with perfluoro-15-crown-5-ether-loaded (PFCE) poly(latic-co-glycolic acid) nanoparticles (hydrodynamic radius 100 nm) and gold nanoparticles (diameter 40 nm), as 19F MRI and CT contrast agents, respectively. The resulting CPC/PFCE/gold composite is implanted in a rat model for in vivo longitudinal imaging. Our findings show that the incorporation of the two types of different nanoparticles did result in adequate handling properties of the cement. Qualitative and quantitative long-term assessment of CPC/PFCE/gold degradation was achieved in vivo and correlated to the new bone formation. Finally, no adverse biological effects on the bone tissue are observed via histology. In conclusion, an easy and efficient strategy for following CPC implantation and degradation in vivo is developed. As all materials used are biocompatible, this CPC/PFCE/gold composite is clinically applicable.
Calcium phosphate cement (CPC) is used in bone repair because of its biocompatibility. However, high similarity between CPC and the natural osseous phase results in poor image contrast in most of the available in vivo imaging techniques such as computed tomography (CT) and magnetic resonance imaging (MRI). For accurate identification and localization during and after implantation in vivo, a composition with enhanced image contrast is needed. In this study, we labeled CPC with perfluoro-15-crown-5-ether-loaded (PFCE) poly(latic-co-glycolic acid) nanoparticles (hydrodynamic radius 100 nm) and gold nanoparticles (diameter 40 nm), as 19F MRI and CT contrast agents, respectively. The resulting CPC/PFCE/gold composite is implanted in a rat model for in vivo longitudinal imaging. Our findings show that the incorporation of the two types of different nanoparticles did result in adequate handling properties of the cement. Qualitative and quantitative long-term assessment of CPC/PFCE/gold degradation was achieved in vivo and correlated to the new bone formation. Finally, no adverse biological effects on the bone tissue are observed via histology. In conclusion, an easy and efficient strategy for following CPC implantation and degradation in vivo is developed. As all materials used are biocompatible, this CPC/PFCE/gold composite is clinically applicable.
Bone grafting is the most
common transplantation procedure, after
blood, with more than 2 million patients worldwide receiving a bone
transplant every year.[1,2] To date, autografting of autologous
bone is still the gold standard procedure. However,
drawbacks such as the need for a second surgery, high donor-site morbidity,
and shortage of donor bone are increasing the demand for artificial
bone substitutes. Since 1920, calcium phosphate-based cement (CPC)
has been extensively studied and used for orthopedic and dental applications
due to its high biocompatibility, biodegradability, and osteoconductivity.[3−5] Moreover, the ability to be injected and to set at body temperature
in vivo makes CPC suitable for minimally invasive surgeries.[6] Two criteria are important in the application
of CPC in bone repair: mechanical properties that are tailored to
the specific application and an adequate imaging contrast that allows
for monitoring cement after injection, particularly to monitor its
degradation.[6,7] However, only a few strategies
have been successful in the synthesis of CPC that meet both criteria,
as labeling the cement typically affects its mechanical properties.
X-ray and CT are the most common methods to monitor bone defects and
the formation of new bone.[8,9] Nonlabeled CPCs have
a slightly higher radiodensity than the natural bone, resulting in
a brighter signal on conventional X-ray radiographs. However, the
geometrical conformation of the cement is barely recognizable after
surgery. Longitudinal monitoring becomes even more challenging when
degradation of the material and the in-growth of the new bone begins.[10−12] Addition of radiopaque salts, such as barium sulfate or tantalum
oxide, can enhance the radiocontrast of CPC without affecting its
biological behavior.[13−16] However, the incorporation of such salts affects the setting and
mechanical properties of the cement, making it less suitable for clinical
application. To avoid changes in cement properties after labeling,
different strategies, such as the encapsulation of probes into a silica
carrier, have been described.[17,18] Also, gold nanoparticles
(AuNPs) are promising for CT because gold has a high atomic number
(Z = 79) and high X-ray attenuation per mass (5.16
cm2 g–1 at 100 keV). Moreover, biocompatibility,
low toxicity, and high affinity of gold nanoparticles to different
functional groups, for example thiol or phosphine, make AuNPs suitable
for labeling biomaterials.[19−21] However, no successful strategy
for long-term labeling of CPC with gold has been described thus far.MR imaging of hard tissues (i.e., bone and teeth) requires acquisition
sequences sensitive to their very short transversal relaxation such
as ultrashort echo time (UTE) and zero echo time (ZTE) sequences.[22,23] However, in MR images with these sequences, bone transplants do
not provide enough contrast, making labeling essential.[24] Multimodal imaging with MRI and CT is advantageous,
as it allows the acquisition of independent information using two
different imaging techniques. The use of two independent labels with
different physicochemical properties and different affinity to CPC
could reduce problems such as a decrease or complete loss of the signal
due to, for example, diffusion of noncovalently attached label from
the CPC matrix. The first CPC mixture that was labeled with multiple
imaging agents contained superparamagnetic iron oxide particles (SPIO,
200 nm in diameter) and colloidal gold (4 nm in diameter) respectively
as MRI and CT contrast agents.[17,18] However, the strong
susceptibility artifacts caused by SPIO particles resulted in a gross
overestimation of the implant shape.[25,26] Together with
the fast decrease of the CT contrast over time, the use of this strategy
became very limited. The labeling of CPC with heteronuclei, such as 19F, 31P, or 23Na, natural isotopes with
spin of 1/2 that can be directly detected with MRI, could be an excellent
alternative to SPIO. Fluorine is especially attractive as it is almost
completely absent from biological tissues, which results in a high
MR contrast-to-noise ratio and specificity, and it has a similar intrinsic
high sensitivity as that of protons.[27−32] Typical 19F MRI labels contain organofluorine compounds,
in particular liquid perfluorocarbons (PFCs). Perfluorocarbons are
biologically inert and exhibit low toxicity. Therefore, they have
been used in the clinic as blood substitutes and more recently as
MR imaging agents.[33−36]In this study, we combined 19F MRI and CT for imaging
the CPC in bone for the first time, using perfluoro-15-crown-5-ether
(PFCE)-loaded poly(d,l-lactide-co-glycolide) (PLGA) nanoparticles and commercially available gold
nanoparticles (AuNPs). PFCE nanoparticles have previously only been
applied to image labeled cells, i.e., soft tissues.[35] Here we use these nanoparticles for imaging of bone substitutes.
We decided to use AuNPs because they are also effective for both promoting
osteoblast differentiation and bone formation.[37,38] This newly developed CPC/PFCE/gold composite was first characterized
in vitro and then implanted in vivo in a rat femoral condyle model.
We followed material degradation by in vivo 1H/19F MRI and CT up to 8 weeks postsurgery and correlated the images
to the biological tissue response.
Materials and Methods
CPC Composite
Calcium phosphate cement
(CPC) was prepared as a w/w mixture of 78.5% α-tricalcium phosphate
(α-TCP; CAM Bioceramics BV, Leiden, The Netherlands), 1.5% carboxymethylcellulose
(CMC; CAM Bioceramics), and 20% of cryo-grinded poly(d,l-lactide-co-glycolide) microparticles (<200
μm) with a 50:50 ratio of lactic to glycolic acid (PURASOB 5002A,
Purac, Gorinchem, The Netherlands).
Contrast
Agents
PFCE-loaded PLGA
nanoparticles were prepared as described elsewhere.[35] Briefly, PLGA (100 mg) was dissolved in dichloromethane
(3 mL) and mixed with PFCE (0.9 mL) and Prohance (1.78 mL). The resulting
emulsion was added quickly under sonication to 25 mL of 1.96% (w/w)
solution of poly(vinyl alcohol) (9–10 kDa, 80% hydrolyzed)
and sonicated for 3 min at 40% amplitude (Branson Digital Sonifier
250; Branson Sonic power, Danbury, CT). After evaporation of the solvent
overnight the particles were washed four times with water at 16 000g, resuspended in water, and freeze-dried, yielding of approximately
100–150 mg of particles as white powder. PFCE was purchased
from Exfluor, Rond Rock, TX; PLGA Resomer RG 502H, lactide:glycolide
molar ratio 48:52–52:48, from Evonic, Germany; dichloromethane
from Merck, Darmstadt, Germany; and PVA from Sigma-Aldrich, St. Louis,
MO.Gold nanoparticles (diameter 40 nm) stabilized by citrate
in 0.1 mM phosphate buffered saline solution (PBS) were purchased
from Sigma and used without any additional modification.
Material Preparation
The CPC control
(i.e., without contrast agents) was prepared by adding the mixed powders
(i.e., α-TCP, CMC, and PLGA) into an exit-closed 2 mL syringe
(Terumo Europe N.V., Leuven, Belgium). Afterward, 50 μL of sterile-filtered
(0.2 μm filter) sodium dihydrogen phosphate solution (4% NaH2PO4·2H2O) was added into the syringe
and shaken for 30 s by means of a dental shaker machine (Silamat mixing
apparatus, Vivadent, Schaan, Liechtenstein).In order to add
the contrast agents to the starting CPC composition, additional steps
were requested. Briefly, 100 mg of initial powders of CPC was first
mixed with 1 mL of AuNPs solution (corresponding to ∼7.2 ×
109 particles), homogenized by vortex, and freeze-dried
overnight. Thereafter, 20 mg of PLGA/PFCE nanoparticles was first
resuspended in 1 mL of Milli-Q water and then added to the CPC/gold
powder. The solution was then vortexed and freeze-dried again providing
the final composition, which is indicated in this study as CPC/PFCE/gold.An additional CPC composite was prepared by performing the same
freeze-dry steps as described before, but using Milli-Q water alone,
without any contrast agents. This CPC/freeze-dry composite was used
as an internal control for the characterization of the handling and
mechanical properties. Samples were sterilized by using 25 kGy of
γ radiation (Synergy Health Ede B.V., Ede, The Netherlands).
Material Characterization
Morphological
assessment of the PFCE-loaded PLGA nanoparticles and of preset cylinders
of CPC/PFCE/gold were performed by electron scanning microscopy (SEM,
JEOL 6310, Jeol Corp., Tokyo, Japan). Images were acquired at 5 kV
with ×5000 magnification.The size and the zeta potential
of the PLGA/PFCE nanoparticles were measured by dynamic light scattering
(DLS) using a Malvern Zetaziser Nano ZS (Malvern Instrument, Worcestershire,
UK) at sample concentration of 0.1 mg/mL, using ultrapure water as
a solvent for DLS and 5 mM sodium chloride solution for zeta potential
measurements. The content of PFCE was determined by 19F
NMR (Bruker Avance 400) using trifluoroacetic acid as an internal
reference and deuterium oxide as solvent (both from Sigma-Aldrich).
Cement Handling and Mechanical Properties
Setting times, elasticity, compression strength, injectability,
and cohesion properties of the CPC with and without contrast agents
were investigated. As internal control the CPC/freeze-dried composite
was also used.For the setting times, a Gillmore apparatus according
to ASTM C266 was used. The cement pastes were injected in a cylindrical
bronze mold of 6 mm in diameter and 12 mm in height and immersed in
a water bath at 37 °C after which initial and final setting times
were recorded. The elasticity and the compressive strength were calculated
by using a testing bench machine (858 miniBionixII, MTS, Eden Prairie,
MN). Cylinders of 4.5 mm in diameter and 9 mm in height were prepared
from the CPC mixtures and compressed with a loading force of 2.5 kN
at a constant speed of 0.5 mm min–1. The injectability
was assessed by using the same testing bench machine (858 miniBionixII)
set in a compression mode and adapted with a custom-made fixture metallic
cage. A total weight of 500 mg for each cement composition was mixed
into an exit-closed 2 mL syringe (Terumo Europe N.V.) with 250 μL
of setting solution (4% NaH2PO4·2H2O). After mixing the components for 30 s, the syringe was
placed in the metallic cage, the exit was opened, and a compression
force of 100 N (i.e., estimated maximum force applicable by human
operator) was applied with a constant speed of 20 mm min–1 until all the material was extruded from the syringe (33 ±
2 s). From the raw data the extrusion curve was obtained as reported
in the literature as applied force (N) by the time (s).[39] Finally, all the cements extruded from the syringes
were accumulated in 10 mL of PBS at 37 °C, and the cohesive properties
were qualitatively assessed (i.e., by counting the number of fragments).
In Vitro Assay
A cylindrical hole
(3 × 3 mm) was drilled into bone blocks (∼1 cm3) obtained from pig cadaver jaw. The hole was filled either with
CPC or CPC with contrast agents and cured overnight. All the samples
were prepared in triplicate (n = 3).For MRI
scan, the samples were first embedded in 5% gelatin type A (Sigma-Aldrich)
in order to simulate a water environment. The gelatin was poured in
a 15 mL plastic tube and solidified at +4 °C. For the microcomputed
tomography (μCT) measurements the samples were wrapped in Parafilm
(SERVA Electrophoresis GmbH, Heidelberg, Germany) to avoid drying
artifacts during the scan.
In Vitro μCT
The bone blocks
were scanned horizontally along the X-ray beam by using a μCT
imaging system (Skyscan 1072, Kontich, Belgium). Samples were recorded
using ×15 magnification (i.e., pixel resolution = 18.88 μm),
X-ray source of 100 kV/98 μA, exposure time 3.9 s, and 1 mm
aluminum filter. The obtained projected files were reconstructed with
NRecon software (Skyscan) and analyzed with CtAnalyser software (version
1.10.1.0; Skyscan). The volume of interest (VOI) was defined manually
by selecting a total of 135 slices (in height; i.e., 2.55 mm) and
a circle of 3 mm (in diameter). Two-dimensional (2D) reconstructions
were finally obtained by DataViewer software (Version 1.5.2.4; Skyscan).
In Vitro MRI
All the samples embedded
in gelatin were scanned on a 11.7 T MRI system (Biospec, Bruker, Germany)
equipped with a 1H/19F volume coil. 1H/19F MR images of the bone phantom were acquired by a
zero echo time (ZTE) sequence with the following parameters: repetition
time (TR) = 2 ms/4 ms, image resolution = 1.56 × 1.56 ×
1.56 mm, 1 average/16 averages, flip angle (FA) = 2°/4°,
and acquisition time (TA) = 27 s/13.54 min. The obtained 1H and 19F MR images were processed with Matlab R2014b
(MathWorks Inc., Natick, MA) and overlaid by using MRIcro software
(Smith Micro software, Aliso Vijeo, CA).[40]
In Vivo Assay
The animal study was
performed in agreement with the standards and the protocols of the
Radboud University Medical Center, Nijmegen, The Netherlands. All
the surgeries were performed after the approval of the Animal Ethics
Committee (RU-DEC number 2015-0035) for the care and the use of laboratory
animals. Sixteen healthy male Wistar rats (body weight: 250–300
g) were used as experimental animals. A ratfemoral condyle defect
model was used as described elsewhere.[42] In order to reduce the number of animals for this experiment, both
posterior legs of the animals were used. First, the animal was anesthetized
by inhalation of a mix of Isoflurane (Rhodia Organique Fine, Avonmouth,
Bristol, UK) and oxygen, and then the legs were shaved and disinfected
using povidone–iodine solution. The animal was located in the
supine position on a heating mat in order to prevent hypothermia.
A longitudinal incision was performed through the skin and the muscle
on the medial surface of the knee (Supporting Information Figure S1). After the exposition of the medial
side of the distal femoral condyle, the patella was laterally dislocated
in order to have a clear view of the knee joint. Three different dental
burs with increasing diameter up to 2.5 mm were used in order to perform
a cylindrical defect (2.5 mm in diameter and 3 mm in depth) along
the same direction of the femur. A sterile 0.9% saline solution (Fresenius
Kabi B.V., Emmer-Compascuum, The Netherlands) was used to cool down
the dental bur and to clean the drilled cavity. In the meantime, the
CPC powders were mixed with the setting solution as described before.
The defect was filled with CPC with or without contrast agents, while
it was left empty in the case of the positive control group. As negative
control group, the leg was kept untreated (Table S1). After cement implantation, the patella was moved back
to the original position, and the muscle and the skin were closed
with absorbable sutures (Vicryl 4.0 Ethicon, Somerville, NY). A subcutaneous
injection of pain killer (5 mg/mL Rimadyl, Pfizer animal health, NY)
was performed postsurgery in order to decrease postoperative discomfort.
After the MRI and CT scans the skin was further fastened by using
metallic wound clips (Becton and Dickinson, Franklin Lakes, NJ). In
vivo MRI and CT were performed right directly postoperative and at
4 and 8 weeks postsurgery. After 8 weeks the animals were sacrificed
by CO2/O2 inhalation, and the femora were harvested
for histological assessments.
In Vivo
CT
For the in vivo CT a
small animal CT scanner was used (Inveon Micro-CT/PET, Siemens Medical
Solution, Knoxville, TN). The animals were located in the supine position
on a heating mat and always assessed under general anesthesia (Isoflurane/O2). Images were recorded with an acquisition time of 6 min,
spatial resolution of 30 μm, 80 kV tungsten anode source, and
exposure time of 1000 ms. Inveon Research Workplace (IRW, Siemens)
software was used for 3D reconstruction of the projected files and
in order to define the VOI corresponding to the implanted material.
As the shape of the implant was heterogeneous from leg to leg, the
outline of the implant was carefully drawn. For each VOI, the total
volume in mm3 and the mean attenuation intensity in Hounsfield
units (HU) were calculated. For each implant, the level of attenuation
intensity was adjusted by the corresponding volume. Finally, the mean
value of the signal intensity at each time point was computed. Signal
decrease over time was also investigated. Based on a constant VOI
(of 15 mm3), the relative signal intensity at each time
point was expressed as percentage and calculated as total signal intensity
from the implants to the mean of the signal intensity from normal
bone (i.e., no defect group) (Figure S2).
In Vivo MRI
Animal MR experiments
were performed with the same 11.7 T MRI system and 1H/19F volume coil as for in vitro experiments. Animals were anesthetized
by Isoflurane/O2 and placed in the supine position. One
by one each leg was immobilized inside the coil while body temperature
and breathing were constantly monitored. For fluorine content quantification
a 200 mL Eppendorf tube filled with 20 mg of PLGA/PFCE nanoparticles
dispersed in Milli-Q water was used as reference. The Eppendorf tube
was placed adjacent to the medial side of the leg. 1H/19F images for each rat leg were acquired by ZTE sequence with
TR = 2 ms, image resolution = 0.31 mm3/1.25 mm3, 1 average/32 averages, and FA = 2°/4°. The scan times
were less than 7 and 14 min for 1H and 19F,
respectively. The 19F signals in the rat legs as well as
in the control sample were quantified in MRIcro software.[40] Regions of interest (ROIs) were manually outlined
based on the detected 19F signal on 19F ZTE
images. 19F signal per ROI was calculated through multiplying
the mean of the pixel intensity of the ROI by its volume (i.e., area
per slice thickness). The total 19F amount was then summed
over slices and expressed in arbitrary units (au).[41] Based on the quantified 19F signals of the control
sample scanned together with each leg at day 0, 4, and 8 weeks, the
normalization procedure for 19F signals has also been performed
within each leg at three time points and between legs.
MRI, CT, and Implant Volume Linear Correlation
Correlation
analysis was performed by comparing 19F
MRI signal (expressed in au), CT signal (expressed in HU/mm3), and volume of the implant (expressed in mm3) at each
time point. The volume of the implant at each time point was calculated
by IRW (Siemens) based on the in vivo CT acquisitions. For each time
point, the two-tailed Pearson correlation coefficient (ρ) between
MRI signal and volume of the implant, CT signal and volume of the
implant, and CT signal and MRI signal was calculated. Afterward, Pearson
coefficients were converted to Fisher’s coefficient (z), and the interval of confidence at 95% was computed by
backconverting z to ρ. Pearson coefficients
were ranked according to the rule of thumb.[43]
Histology
Samples were decalcified
in 10% ethylenediaminetetraacetic acid (EDTA) for 2 weeks. The
decalcified bones were then dehydrated in a gradual ethanol series
(from 70% to 100%) and embedded in paraffin. Sections of 4 μm
in thickness were cut along the axial direction of the femur by using
a microtome (RM2165, Leica Microsystems, Rijswijk, The Netherlands).
The sections were stained with hematoxylin/eosin (H/E) and with trichrome
Elastic van Gieson (EVG). For each specimen, at least two images from
three sections (at 100 μm of interdistance) were analyzed. Images
were acquired by using a light microscope (Axio Imager Z1, Carl Zeiss
AG Light Microscopy, Göttingen, Germany) equipped with a digital
camera (AxioCam MRc5, Carl Zeiss AG Light Microscopy). Histomorphometrical
analysis was performed based on the EVG-stained slices by using a
computer image analysis technique based on ImageJ software (Wayne
Rasband, Research Services Branch, National Institute of Mental Health,
Bethesda, MD). The amount of bone was calculated as area percentage
inside a defined ROI (circle of 2.5 mm diameter, Figure S3).
Statistical Analysis
GraphPad Instat
(GraphPad Software, San Diego, CA) was used for all statistical measurements;
data were reported as mean ± standard deviation. For comparison
of data one-way analysis of variance (ANOVA) with a Tukey’s
post hoc test was used. All differences were considered significant
at p-values <0.05.
Results
and Discussion
Simultaneous Labeling of
CPC with PFCE-Loaded
PLGA and Gold Nanoparticles
To label the cement for 19F MRI, we used PFCE-loaded PLGA nanoparticles. These nanoparticles
were developed at the Radboud University Medical Center (Nijmegen,
The Netherlands) and are now being prepared for use in clinical trial
for labeling dendritic cell-based vaccines (clinicaltrial.gov identifier
NCT02574377).[35] These particles typically
have a hydrodynamic radius of 100 nm, as determined by DLS, and zeta-potential
values in a range of −1 to −5 mV, due to steric stabilization
by poly(vinyl alcohol). A scanning electron microscope (SEM) image
of the nanoparticles is shown in Figure A. The amount of PFCE in the particles was
determined by nuclear magnetic resonance (NMR) using an internal reference
and was around 20 wt %. As CT contrast agent, we used commercially
available AuNPs that are stabilized with citric acid in PBS.
Figure 1
Scanning electron
micrographs of the PFCE-loaded PLGA nanoparticles
(A) and of the surface of a preset cylinder of CPC/PFCE/gold (B).
Yellow arrows indicate the agglomerates of PLGA/PFCE nanoparticles
that appeared after the material preparation. Scale bar 1 μm.
In (C) a 2 mL syringe containing CPC/PFCE/gold paste is shown.
Scanning electron
micrographs of the PFCE-loaded PLGA nanoparticles
(A) and of the surface of a preset cylinder of CPC/PFCE/gold (B).
Yellow arrows indicate the agglomerates of PLGA/PFCE nanoparticles
that appeared after the material preparation. Scale bar 1 μm.
In (C) a 2 mL syringe containing CPC/PFCE/gold paste is shown.The largest initial challenge
was the incorporation of both nanoparticle
labels into the CPC matrix. Earlier combinations of CPCs and several
contrast agents were produced by mixing the dried powders and therefore
had only one easy production step.[13−18] However, our PFCE-loaded PLGA nanoparticles formed clusters in the
freeze-dried state, making homogeneous incorporation into the CPC
difficult. Moreover, AuNPs were provided dispersed in PBS solution.
Therefore, we developed an easy two-step method that allowed for the
simultaneous incorporation of both nanoparticles into the CPC phase.
All components were independently dispersed in water and mixed at
high shear stress followed by fast freezing and freeze-drying the
mixture to avoid the setting of CPC. In the second step, the obtained
dual labeled-CPC powder could be mixed with the setting solution and
handled further as the nonlabeled CPC. This method was more efficient
for incorporation of both labels than adding the imaging agents, mixing
them in the setting solution, or mixing the cement directly with freeze-dried
powder of nanoparticles. The color of the cement composite changed
from white to pink after AuNPs were added (Figure C). The uniform distribution of color in
CPC indicated a homogeneous distribution of AuNPs. Morphological analysis
based on SEM (Figure B and Figure S4) proved that PFCE-loaded
PLGA nanoparticles were incorporated into the CPC matrix. After incorporation
and setting of the CPC, the initial PFCE-loaded PLGA nanoparticles
formed clusters of approximately 1–5 μm in diameter.
This cluster formation could be due to reduced colloidal stability
of nanoparticles in the cement mixture and to the relatively high
concentration of nanoparticles at 17% w/w.[44] However, MR imaging of labeled CPC revealed that
these agglomerates were homogeneously distributed in the cement matrix
(Figure C). Thus,
the agglomeration did not influence the imaging properties of the
final material. Likely, this agglomeration could even be an advantage
for our purpose due to possibly slower degradation of aggregated nanoparticles
that allowed the PFCE signal to last longer in the CPC matrix. Unfortunately,
the AuNPs were too small to be identified by SEM in the morphologically
complex CPC matrix. However, the change of the color and the enhanced
CT visibility indicated that the incorporation was successful (Figure E–G).
Figure 2
In vitro MRI
and CT imaging of a bone block obtained from pig jaw.
In (A) the morphological representation of the bone block. In (B)
and (C) 1H and 19F ZTE MR images of CPC/PFCE/gold
composite obtained at 11.7 T. Note that the 19F signal
corresponds to the cylindrical defect in the 1H ZTE image
as highlighted by the yellow lines. (D) ZTE MR images of 19F overlaid on those of 1H ZTE. In (E) and (F) the gray
scale 2D images from the μCT of the pig block filled with the
CPC composite are shown without (E) and with (F) contrast agents.
The yellow arrows indicate the implanted material. In (G) the graph
from the gray values distribution calculated for a defined volume
of interest (n = 3). Note that the signal intensity
is elevated and much more defined.
In vitro MRI
and CT imaging of a bone block obtained from pig jaw.
In (A) the morphological representation of the bone block. In (B)
and (C) 1H and 19F ZTE MR images of CPC/PFCE/gold
composite obtained at 11.7 T. Note that the 19F signal
corresponds to the cylindrical defect in the 1H ZTE image
as highlighted by the yellow lines. (D) ZTE MR images of 19F overlaid on those of 1H ZTE. In (E) and (F) the gray
scale 2D images from the μCT of the pig block filled with the
CPC composite are shown without (E) and with (F) contrast agents.
The yellow arrows indicate the implanted material. In (G) the graph
from the gray values distribution calculated for a defined volume
of interest (n = 3). Note that the signal intensity
is elevated and much more defined.
Imaging of Cement Injected in Vitro in Pig
Jaw Blocks
To find out the optimal ratio between the CPC
powders and the contrast agents, we performed μCT and 19F MR imaging of the final material. On the basis of these results,
we determined that a concentration of 20 mg of freeze-dried PFCE-loaded
PLGA nanoparticles per 100 mg of CPC powder provided sufficient labeling
with PFCE. For CT contrast, 1 mL of AuNPs suspension was necessary
for significant enhancement (data not shown).Next, we imaged
CPC composites injected in pig bone blocks (Figure A) using 1H and 19F
MRI. On the 1H MR image (Figure B), CPC could not be distinguished from pig
bone. However, a strong fluorine signal was detected from the CPC/PFCE/gold
composite in the 19F ZTE MR image (Figure C). Thus, overlaying the 19F on
the 1H image allowed identification of the labeled CPC
(Figure D).Two-dimensional gray scale images of the CPC injected in pig blocks
were obtained by μCT acquisition. As shown in Figure , CPC with contrast agents
(Figure F) appeared
visually darker than nonlabeled CPC (Figure E). A quantitative estimation of the gray
value distribution confirmed this observation. CPC/PFCE/gold reported
significantly higher pixel frequency compared to the CPC control as
well as a rightward shift of the gray values curve (Figure G). Furthermore, the narrowing
of the peak indicated an even distribution of the particles in the
CPC phase.
Characterization of the
Handling and Mechanical
Properties
The mechanical and handling properties of the
final cement mixture are very important for the application as a bone
filler. These mechanical properties should match those of the treated
bone. Furthermore, the in situ setting should be fast but also provide
enough time for the surgeon to implant and model the paste. Further,
hydraulic features should allow easy injection of the material, enabling
minimally invasive operations.[4−6] To investigate whether the incorporation
of the two types of nanoparticles affected the material properties
of the CPC, we performed handling and mechanical testing. As a control
we used CPC that was treated in the usual way (CPC-control) and another
CPC that we prepared in the same way as the labeled CPC but without
nanoparticles (CPC/freeze-dried). The last control is important as
possible changes in material properties can be caused either by incorporation
of nanoparticles or by the extra steps involving resuspension in water
and freeze-drying. Setting times of the CPC/PFCE/gold composite that
we obtained from the Gillmore tests (Figure A,B) showed an increase in the initial and
final times by 1.7 ± 0.3 min and 3.2 ± 0.3 min, respectively,
compared to CPC-control. However, this small increase did not affect
injection during surgery. Setting times of both nonlabeled CPC controls—the
freeze-dried and the non-freeze-dried—were similar to each
other. This behavior indicates that the incorporation of the particles
did cause an increase in setting time. Ideally, the initial setting
time should be between 6 and 10 min, and the final hardening should
be reached within 20 min.[5,6] Thus, the setting time
of our labeled composite is within this range confirming that it can
be used as a bone implant.
Figure 3
Mechanical and handling properties of the CPC
composite with and
without contrast agents, respectively; the initial (A) and the final
(B) setting time from Gilmore tests (n = 3); the E-modulus (C) and the compressive strength (D) from the
compression tests (n = 5); the extrusion curves from
the injectability tests (E) (n = 3); CPC/PFCE/gold
composite (F) and CPC (G) representative samples from cohesion tests
(n = 3). # and ∗ mean p <
0.05.
Mechanical and handling properties of the CPC
composite with and
without contrast agents, respectively; the initial (A) and the final
(B) setting time from Gilmore tests (n = 3); the E-modulus (C) and the compressive strength (D) from the
compression tests (n = 5); the extrusion curves from
the injectability tests (E) (n = 3); CPC/PFCE/gold
composite (F) and CPC (G) representative samples from cohesion tests
(n = 3). # and ∗ mean p <
0.05.Comparing the compressive strength
and the E-modulus
of the CPC/PFCE/gold composite to the nonlabeled CPC showed that labeling
does not affect the mechanical properties of CPC (Figure C,D). Particularly, both values
are similar compared to the CPC that was only freeze-dried (i.e.,
CPC/freeze-dried), indicating that the freeze-drying rather than the
incorporation of the nanoparticles affected the properties of the
final material. Importantly, these differences were not significantly
different when compared to standard CPC. Thus, compressive strength
and E-modulus are comparable with the values of the
most common α-TCP composites reported in the literature that
are typically between 1 and 70 MPa and 0.5 and 9 GPa, respectively,
depending on the exact composition of the material.[5] In particular, a commercially available injectable bone
cement, chronOSTM Inject, has a compressive strength of 3.0 ±
0.6 MPa.[44] This value is very close to
the results of this study (i.e., CPC/PFCE/gold = 4.6 ± 2.1 MPa)
meaning that our composite could be used for similar applications.
Surely, CPC/PFCE/gold is indicated for nonload-bearing bone defects,
specifically in cancellous bone, which shows similar compressive strength
and elasticity (i.e., 4–12 MPa and 0.1–0.5 GPa, respectively).[46]Finally, CPC/PFCE/gold composite showed
excellent hydraulic and
cohesive features. All samples could be completely extruded from the
syringe through a 1.7 mm orifice and thereafter set in an aqueous
solution (Figure E–G
and Figure S5). Even if we used no needles
in this study, our results indicate that CPC/PFCE/gold composite is
suitable for use with needles up to 15 gauge. Such needles are used,
for example, in several back and face surgeries.[6,39]
Imaging of Cement Injected in Vivo in Rat
Bone Model
Having promising in vitro results on both imaging
and materials properties, we then investigated the performance of
our labeled composite in vivo for longitudinal monitoring up to 8
weeks from the injection in a ratfemoral condyle defect. This defect
involves relatively simple surgery and is a well-established nonload-bearing
model to study bone biomaterials.[42] The
time of 8 weeks was chosen, as we expected from our in vitro study
(data not shown) and from previous works on CPC degradation in the
same model, that the CPC would stay in the bone for at least 4 weeks.[17,18]For in vivo testing, we used non freeze-dried CPC as a control
as well as another group of animals with an empty defect. None of
the animals showed any sign of discomfort, swelling, or restriction
in movement after surgery. To monitor the cement after injection,
we performed MRI and CT image acquisitions immediately after surgery
(i.e., day 0), and at 4 and 8 weeks postsurgery. High-resolution anatomical
MR images of the rat leg were acquired at day 0, 4 weeks and 8 weeks
after injection (Figure S6). As expected
from in vitro imaging results, it was not possible to distinguish
the nonlabeled CPC composite in rat legs from the surrounding bone
on 1H ZTE images. In contrast, after injection, the labeled
composite could be imaged with both MRI and CT for the whole duration
of 8 weeks.The shape of the labeled composite was clearly recognizable
on 19F ZTE MR images, shown in false color, providing precise
geometrical information (Figure ). Moreover, comparing the images at different time-points
after injection showed that the 19F signal area was shrinking.
Indeed, further quantitative analysis of 19F signal confirmed
this observation and showed a gradual decrease of the 19F-containing region over 8 weeks (Figure ). Moreover, this reduction in signal area
correlates to the decrease in the volume of the implant as a consequence
of cement degradation. This suggests that dissolution of the cement
causes clearance of the PFCE-loaded PLGA nanoparticles, resulting
in a signal decrease.
Figure 4
Representative in vivo ZTE MR images of 19F
overlaid
on those of 1H of a rat leg at day 0, 4 weeks and 8 weeks
in axial and coronal directions. The presence of 19F is
shown in false color. Yellow arrows indicate the implanted material,
while the bright yellowish spot next to each leg corresponds to the
reference (i.e., Eppendorf tube containing PFCE-loaded nanoparticles
dispersed in water).
Figure 5
19F MRI and CT signal quantification of the CPC/PFCE/gold
composite at day 0, 4 weeks and 8 weeks. 19F signal (red)
in 19F ZTE MR images and relative signal (blue) in CT images
were quantified and expressed as arbitrary unit and in percentage
of relative attenuation intensity, respectively (n = 8). Lines indicate the trend of the data points.
Representative in vivo ZTE MR images of 19F
overlaid
on those of 1H of a rat leg at day 0, 4 weeks and 8 weeks
in axial and coronal directions. The presence of 19F is
shown in false color. Yellow arrows indicate the implanted material,
while the bright yellowish spot next to each leg corresponds to the
reference (i.e., Eppendorf tube containing PFCE-loaded nanoparticles
dispersed in water).19F MRI and CT signal quantification of the CPC/PFCE/gold
composite at day 0, 4 weeks and 8 weeks. 19F signal (red)
in 19F ZTE MR images and relative signal (blue) in CT images
were quantified and expressed as arbitrary unit and in percentage
of relative attenuation intensity, respectively (n = 8). Lines indicate the trend of the data points.Because of the high calcium salts content, CPCs
generally show
higher radiographic attenuation when compared to the natural bone
phase.[10] However, there are many available
calcium phosphate-based composites which may consist of different
mineral components (i.e., α-TCP, hydroxyapatite) often combined
with different additives, polymers, or fibers.[4−6] Depending on
the specific mixture of components, CPCs may show different X-ray
attenuation, with better or worst contrast when compared to the natural
bone. Thus, a “natural” contrast between CPC and bone
is not always a given yet is dependent on parameters, such as the
CPC composition, the type/density of the bone, and the biological
phase of bone remodeling. We aimed to achieve a labeling methodology
in which, irrespective of these circumstances, a reliable contrast
can always be achieved. Composites with improved contrast are demanded
especially when need to be injected in the vertebral spine. In such
clinical circumstances, leakage of CPC from the vertebral body can
occur leading to several symptomatic complications (i.e., neurological
complication, collapse of the adjacent vertebrae).[11,12] A clear view of the implanted CPC can be also hindered by the presence
of extensive soft tissue around the implant that can obscure its precise
shape identification.[11] Furthermore, CPC
monitoring over time is changeling and less precise. It has been proven
that only after disappearance of the 50% of the implanted material
and subsequent replacement with new bone is it possible to recognize
a visible change in bone remodeling.[10,11] CPC/PFCE/gold
composite showed not only visual enhancement of CT attenuation right
after the surgery (i.e., day 0), but the shape of the implant was
clearly visible during the all duration of the experiment (Figure ). By contrast, the
nonlabeled CPC used as control shows similar attenuation as the natural
bone phase. Although it is possible to identify the implanted nonlabeled
CPC after surgery, the border between the bone and the implanted cement
was not always recognizable. The identification of the implanted control
cement became more challenging at 4 weeks postsurgery after the CPC
degradation took place (Figure ). Evaluation of the mean attenuation coefficients expressed
in Hounsfield unit (HU) per mm3 confirmed these results.
At all the time points, the CPC/PFCE/gold composite showed higher
values than both empty defect group and nonlabeled CPC composite (Figure ). However, this
difference was statistically relevant only at 4 and 8 weeks. The higher
signal intensity from the CPC/PFCE/gold composite allowed for a better
definition of implant shape and volume, in comparison to the nonlabeled
CPC. A narrowing of the shape of the labeled implant was clearly visible
for all the experimental times. For this reason, CT acquisitions enabled
a more precise quantification of the volume of the labeled implant
at each time point. Moreover, by comparing the relative signal intensities
over time, it was possible to see a gradual decrease of the signal
intensity, indicating clearance of the label with cement degradation
(Figure ). Only few
studies on the long-term imaging of radiopaque bone substitutes are
available. However, most of these studies concern in vitro tests or
focus on in vivo biological response without imaging its degradation.[13−16] In a previous study, Ventura et al. used smaller AuNPs with 4 nm
diameter that were embedded in silica matrix to enhance CT contrast.[17] However, these nanoparticles diffused from the
silica particles 4 weeks after the implantation in vivo, resulting
in loss of CT contrast. In this study, larger nanoparticles with diameter
of 40 nm provided long-term enhancement CT contrast, at least up to
8 weeks. We assume the bigger size of AuNPs hindered the diffusion
of nanoparticles out of CPC, resulting in prolonged enhancement of
CT.
Figure 6
Qualitative and quantitative in vivo CT results. The CT images
of the coronal direction of the rat femoral condyle are shown for
defects left empty or/and filled with CPC alone or/and with CPC/PFCE/gold.
From the top to the bottom, the pictures of the same specimen are
shown directly after the surgery (day 0) and at 4 and 8 weeks postsurgery.
The bar graph presents the signal quantification estimated by IRW,
based on a defined region of interest. Values are reported as mean
of attenuation coefficients in HU per the total volume in mm3 (n = 8).
Qualitative and quantitative in vivo CT results. The CT images
of the coronal direction of the rat femoral condyle are shown for
defects left empty or/and filled with CPC alone or/and with CPC/PFCE/gold.
From the top to the bottom, the pictures of the same specimen are
shown directly after the surgery (day 0) and at 4 and 8 weeks postsurgery.
The bar graph presents the signal quantification estimated by IRW,
based on a defined region of interest. Values are reported as mean
of attenuation coefficients in HU per the total volume in mm3 (n = 8).Finally, the combination of both imaging modalities enabled
to
monitor degradation of the cement. To compare quantitative MRI and
CT signals, we investigated if there is a correlation between 19F MRI signal, CT attenuation coefficient, and volume of the
implant at each time point. The Pearson correlation coefficients and
confidence intervals are summarized in Table . Pearson coefficients were ranked according
to the rule of thumb.[43] Very strong (i.e.,
0.9 < r < 1) and strong (i.e., 0.7 < r < 0.9) statistically significant linear relationships
(i.e., p < 0.05) between the two variables were
found at day 0 and at 4 weeks, respectively, for correlation of MRI
signal to implant volume, CT signal to implant volume, and MRI signal
to CT signal. In contrast, at 8 weeks, a strong linear correlation
was found between CT signal and volume of the implant, while the relationships
between MRI signal and implant volume or MRI signal and CT signal
were low and moderate respectively (i.e., moderate = 0.5 < r < 0.7, low = 0.3 < r < 0.5).
This lower correlation of MRI signal could be due to limited MRI resolution
for such a small implant volume (i.e., <4.41 ± 1.96 mm3), resulting in relatively few voxels over the implant and
thus larger error due to the partial volume effect.[29] However, both MRI and CT demonstrated that the degradation
of the cement mainly occurred as a gradual process of dissolution
that starts from the edges of the implant. Thus, in total, it was
possible to follow CPC degradation qualitatively and quantitatively
up to 8 weeks postsurgery.
Table 1
Correlation Analysis
Comparing Mean
Intensities from 19F MRI Signal, CT Attenuation Coefficient,
and Implant Volume
Pearson (ρ)
p < 0.05
95% interval of confidence
day 0
CT vs volume
0.88
yes
0.39 < ρ < 0.96
MRI vs volume
0.79
yes
0.13 < ρ < 0.96
CT vs MRI
0.88
yes
0.43 < ρ < 0.98
4 weeks
CT vs volume
0.93
yes
0.62 < ρ < 0.99
MRI vs volume
0.95
yes
0.70 < ρ < 0.96
CT vs MRI
0.79
yes
–0.01 < ρ < 0.95
8 weeks
CT vs volume
0.82
yes
0.22 < ρ < 0.97
MRI vs volume
0.41
no
0.46 < ρ < 0.88
CT vs MRI
0.56
no
–0.29 < ρ < 0.92
The versatility of a dual-labeling
approach allows to produce patient-tuned
composites that could include either both or only one of the two contrast
agents depending on the specific bone applications. In imaging-guided
vertebroplasty a CPC composite with initial enhanced CT contrast is
preferred by the surgeons as it can help for the identification of
any cement leakage.[11] By contrast, when
the monitoring of the material degradation over time and the subsequent
bone regeneration become the crucial aspect, e.g., in orthopedic and
maxillofacial applications, a noninvasive imaging modality (i.e.,
MRI) is alternatively suggested.[6]Multimodal imaging approaches (i.e., PET/CT, PET/MRI) have gained
clinical interest over the past years as two imaging modalities can
provide synergistic benefits.[23] In this
study a multimodal visible bone substitute (i.e., visible by either
CT or MRI) was developed. Such dual-labeled composite could give to
the radiologists the possibility to provide a patient-tailored imaging
workflow according to each specific clinical circumstance (i.e., CT
acquisition could be performed right after the surgery, while MRI
is applied for longitudinal monitoring). To date, a CT/MRI multimodal
approach is aimed only in preclinical or clinical studies related
to bone remodeling and biomaterials development. However, as MRI is
superior to CT for many aspects (i.e., it is radiation-free, provides
soft and hard tissues contrast, it gives 3D anatomical and functional
information) an ultimate long-term focus for research should still
be to make MRI as a preferential bone imaging screening tool.[8,22]Of course, it can also be critically regarded what is the
final
patient benefit of this labeling procedure. Adding a label to an existing
or newly developed medical device will make the device more complex
to understand and control in its biological behavior, more difficult
and timely to pass through legislation procedures, and moreover it
will inevitably make the treatment itself more costly. Thus, it could
be foreseen that the major importance of labeling procedures in general
is in the phases of preclinical and clinical research, and it is not
always necessary to incorporate labels into the finally developed
medical device for clinical use. The use of labeling might be very
dependent on the field of application. For instance, in dental treatment
patients who receive a bone filler in a periodontal lesion will be
repetitively subjected to imaging during regular follow-up anyways;
whereas for other types of surgery the follow-up period stops after
initial healing. Imaging agents would there only be necessary if special
indications necessitate a more long-term follow-up, for instance in
oncologic patients, or patients that are monitored regularly due to
systemic disease (e.g., severe osteoporosis or diabetes). Even then,
incorporation of a contrast is only necessary when the diagnostic
imaging sessions for the patient can also result in consequences for
the treatment, like early detection and planning of revision surgery.
Bone-Growth Assessment by Histology and Histomorphometry
To investigate the tissue reaction on the CPC composites with and
without imaging nanoparticles, we made histological sections at week
8. Because of the decalcification process during histological preparation,
it was not possible to detect CPC remnants in the sections. However,
the circular area of the original defect was still clearly recognizable,
and thus we used this area for qualitative and quantitative analysis
(Figure ). The empty
defect group shows incomplete healing proving that the performed defect
was a critical-size defect for the 8 week period. Furthermore, both
labeled and nonlabeled CPC materials did not induce any immune response
or fibrotic encapsulation. However, in several samples, a layer of
fibrous tissue was present at the interface between bone and implant.
This layer was not always continuous and appeared to contain fibrous
tissue and immature extracellular matrix fibers, yet often also showed
red-stained calcified nodules, indicating the onset of mineralization.
This effect was present in 1/8 samples of the CPC group and 4/8 samples
of the CPC/PFCE/gold group. χ2 testing demonstrated
that such frequencies are not significantly different (p > 0.05) between both groups. In all cases, there was direct contact
between bone and CPC, and uniform bone growth in all the groups was
observed. Such bone growth began at the edges and then proceeded throughout
the entire contact area toward the inside of the implant.
Figure 7
Histological
sections of the femora stained with EVG after 8 weeks
of implantation. The original bone structure in (A), the empty defect
in (B), the defect filled with CPC in (C), and the defect filled with
CPC/PFCE/gold composite in (D) are shown. For each specimen, a picture
of the general overview taken at 1× magnification (inset) and
the respective zoomed view (magnification 10×) are shown. The
markers correspond to 1 mm.
Histological
sections of the femora stained with EVG after 8 weeks
of implantation. The original bone structure in (A), the empty defect
in (B), the defect filled with CPC in (C), and the defect filled with
CPC/PFCE/gold composite in (D) are shown. For each specimen, a picture
of the general overview taken at 1× magnification (inset) and
the respective zoomed view (magnification 10×) are shown. The
markers correspond to 1 mm.Histomorphometrical analysis of a defined ROI (i.e., circle
of
2.5 mm in diameter) demonstrated the relative formation of the new
bone, with approximate filling of defect of 50%, which is typical
for this length of time.[17] No statistical
differences in bone content were found between the empty defect group,
the CPC group, and the CPC/PFCE/gold group (Figure ). This new bone formation is in the range
of the decrease of both MRI and CT signals. Aiming at clinical application,
further studies with longer experimental times and with materials
of differing degradation kinetics could help to develop a reliable
method for quantitative monitoring the degradation with MRI and CT
contrasts in vivo.
Figure 8
Relative bone formation quantification based on histology
performed
after 8 weeks from the surgery (n = 8). No statistical
differences were found.
Relative bone formation quantification based on histology
performed
after 8 weeks from the surgery (n = 8). No statistical
differences were found.One limitation of the presented experimental model was that
we
did not focus on the clearance of the applied labels from the body.
However, many previous investigations studied body uptake of either
perfluorocarbons (PFCs) or gold nanoparticles (AuNPs) as well as their
biological compatibility.[34,36,47−50] Specifically, it has been proven that PFCs are first removed from
the blood circulation by reticuloendothelial system filtration and
also can ultimately be exhaled through the respiration system.[34,36] Differently, AuNPs uptake seems to be size-dependent. Small AuNPs
(i.e., 5–15 nm) are excreted through kidney filtration, while
large particles (>200 nm) are captured by the immune system and
delivered
into the liver or the spleen.[47,48] After 24 h from intravenous
administration of 50 nm AuNPs in a rat model, particles showed accumulation
only in the blood, liver, and spleen.[49] Therefore, it could be speculated that the 40 nm AuNPs used in this
study would undergo a similar fate and finally be eliminated from
the body through the feces.[50] Another consideration
is the concentration of contrast agents used for the preparation of
the CPC/PFCE/gold composite. Specifically, the estimated PFCE concentration
is less than 5 mg per 100 mg of cement. Such concentration is considerably
low when compared to the large concentration that is requested when
PFCs are used in the blood circulation (i.e., about 10 g/kg).[36] Similarly, the used gold concentration (i.e.,
about 0.5 mg per 100 mg of cement) resulted to fall in the “therapeutic
window” within which AuNPs can be used without any side effects,
suggesting that an even higher concentration may be investigated for
further improvement of the CT contrast.[48]
Conclusions
Nowadays, one of the main
challenges in tissue engineering is to
develop detectable materials that can be followed longitudinally once
implanted in the body. Here, we propose an innovative ceramic material
with enhanced 19F MRI and CT contrast for bone application.
By labeling a CPC with PLGA/PFCE and AuNPs, we obtained a composite
that can be qualitatively and quantitatively detected until 8 weeks
postsurgery in an in vivo rat model of bone regeneration. Thus, our
CPC/PFCE/gold provided more correct anatomical and functional information,
with both MRI and CT, helping to overcome the imaging limitation of
ceramics materials. Successful development of this CPC composite has
substantial potential for clinical use.The combination of the
current contrast agents-based strategy is
suggested for all kinds of ceramic materials as well as for polymeric
scaffold or hydrogels. Certainly, for every different material or
application, a certain degree of customization (e.g., fine-tuning
the amount of nanoparticles) is required. For ceramics, PFCE could
be incorporated in specific cement-dedicated carriers (e.g., microrods,
fibers, bigger particles). Simultaneously, gold could be directly
combined with the carrier containing PFCE by following many strategies,
such as coating or incorporation in bigger beads. In this way, specific-cement-dedicated
carriers can easily be developed and scaled-up for clinical applications.
Furthermore, further improvement of the biological properties of the
dual-labeled CPC composite could be pursuit by adding specific growth
factors (i.e., BMP-2) which can be used to trigger new bone formation.
Authors: C I A van Houdt; R S Preethanath; B A J A van Oirschot; P H W Zwarts; D J O Ulrich; S Anil; J A Jansen; J J J P van den Beucken Journal: J Biomed Mater Res A Date: 2015-10-29 Impact factor: 4.396
Authors: Tian Liu; Pascal Spincemaille; Ludovic de Rochefort; Richard Wong; Martin Prince; Yi Wang Journal: Magn Reson Imaging Date: 2010-08-04 Impact factor: 2.546
Authors: J S Wang; J Diaz; A Sabokbar; N Athanasou; F Kjellson; K E Tanner; I D McCarthy; L Lidgren Journal: J R Soc Interface Date: 2005-03-22 Impact factor: 4.118
Authors: Erik J H Boelen; Gladius Lewis; Jie Xu; Tristan Slots; Leo H Koole; Catharina S J van Hooy-Corstjens Journal: J Biomed Mater Res A Date: 2008-07 Impact factor: 4.396
Authors: V Campana; G Milano; E Pagano; M Barba; C Cicione; G Salonna; W Lattanzi; G Logroscino Journal: J Mater Sci Mater Med Date: 2014-05-28 Impact factor: 3.896
Authors: Joice Maria Joseph; Maria Rosa Gigliobianco; Bita Mahdavi Firouzabadi; Roberta Censi; Piera Di Martino Journal: Pharmaceutics Date: 2022-02-09 Impact factor: 6.321