Literature DB >> 28625686

The Influence of Contamination and Cleaning on the Strength of Modular Head Taper Fixation in Total Hip Arthroplasty.

Annika Krull1, Michael M Morlock1, Nicholas E Bishop2.   

Abstract

BACKGROUND: Intraoperative interface contamination of modular head-stem taper junctions of hip implants can lead to poor fixation strength, causing fretting and crevice corrosion or even stem taper fracture. Careful cleaning before assembly should help to reduce these problems. The purpose of this study was to determine the effect of cleaning (with and without drying) contaminated taper interfaces on the taper fixation strength.
METHODS: Metal or ceramic heads were impacted onto titanium alloy stem tapers with cleaned or contaminated (fat or saline solution) interfaces. The same procedure was performed after cleaning and drying the contaminated interfaces. Pull-off force was used to determine the influence of contamination and cleaning on the taper strength.
RESULTS: Pull-off forces after contamination with fat were significantly lower than those for uncontaminated interfaces for both head materials. Pull-off forces after application of saline solution were not significantly different from those for uncontaminated tapers. However, a large variation in taper strength was observed, pull-off forces for cleaned and dried tapers were similar to those for uncontaminated tapers for both head materials.
CONCLUSION: Intraoperative contamination of taper interfaces may be difficult to detect but has a major influence on taper fixation strength. Cleaning of the stem taper with saline solution and drying with gauze directly before assembly allows the taper strength of the pristine components to be achieved. Not drying the taper results in a large variation in pull-off forces, emphasizing that drying is essential for sufficient and reproducible fixation strength.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  contamination; fat; impaction; pull-off force; taper connection; total hip arthroplasty (THA)

Mesh:

Substances:

Year:  2017        PMID: 28625686     DOI: 10.1016/j.arth.2017.05.009

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  9 in total

Review 1.  What Factors Drive Taper Corrosion?

Authors:  Robin Pourzal; Hannah J Lundberg; Deborah J Hall; Joshua J Jacobs
Journal:  J Arthroplasty       Date:  2018-03-30       Impact factor: 4.757

2.  Contact conditions for total hip head-neck modular taper junctions with microgrooved stem tapers.

Authors:  Maren Bechstedt; Jonathan A Gustafson; Steven P Mell; Julian Gührs; Michael M Morlock; Brett R Levine; Hannah J Lundberg
Journal:  J Biomech       Date:  2020-02-25       Impact factor: 2.712

Review 3.  What Surgeons Need to Know About Adverse Local Tissue Reaction in Total Hip Arthroplasty.

Authors:  Deborah J Hall; Robin Pourzal; Joshua J Jacobs
Journal:  J Arthroplasty       Date:  2020-01-15       Impact factor: 4.757

4.  Acute trunnion failure of a TMZF alloy stem with large diameter femoral heads.

Authors:  E J Fahey; E Pomeroy; F E Rowan
Journal:  J Orthop       Date:  2020-01-10

5.  Are Damage Modes Related to Microstructure and Material Loss in Severely Damaged CoCrMo Femoral Heads?

Authors:  Stephanie M McCarthy; Deborah J Hall; Mathew T Mathew; Joshua J Jacobs; Hannah J Lundberg; Robin Pourzal
Journal:  Clin Orthop Relat Res       Date:  2021-09-01       Impact factor: 4.755

Review 6.  What the Surgeon Can Do to Reduce the Risk of Trunnionosis in Hip Arthroplasty: Recommendations from the Literature.

Authors:  Claude B Rieker; Peter Wahl
Journal:  Materials (Basel)       Date:  2020-04-21       Impact factor: 3.623

7.  Influence of Different Damage Patterns of the Stem Taper on Fixation and Fracture Strength of Ceramic Ball Heads for Total Hip Replacement.

Authors:  Danny Vogel; Jessica Hembus; Mario Jackszis; Vera Bolte; Rainer Bader
Journal:  Biomed Res Int       Date:  2020-05-13       Impact factor: 3.411

8.  Retrieval Findings of Recalled Dual-Taper Hips.

Authors:  Anna Di Laura; Harry S Hothi; Johann Henckel; Young-Min Kwon; John A Skinner; Alister J Hart
Journal:  J Bone Joint Surg Am       Date:  2018-10-03       Impact factor: 5.284

9.  Importance of surgical assembly technique on the engagement of 12/14 modular tapers.

Authors:  A Wade; F Webster; A R Beadling; M G Bryant
Journal:  Proc Inst Mech Eng H       Date:  2021-10-25       Impact factor: 1.617

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.