| Literature DB >> 28553072 |
James A Davison1, Richard Potvin2.
Abstract
PURPOSE: This study aimed to assess the value of intraoperative aberrometry (IA) in determining the intraocular lens (IOL) sphere power in eyes with no previous ocular surgery. PATIENTS AND METHODS: We conducted a retrospective review of patients who underwent uncomplicated cataract surgery where standard preoperative (Preop) measurements and IA were performed. Calculated IOL sphere powers and postoperative refractions, both actual and theoretical, were compared based on the measurement method and lens type; lens types included multifocal, toric and aspheric single-vision non-toric IOLs.Entities:
Keywords: IOL power calculation; ORA; prediction error; refraction
Year: 2017 PMID: 28553072 PMCID: PMC5440073 DOI: 10.2147/OPTH.S135659
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
IOL characteristics by calculation group
| IOL type | Eyes, n | Same, n (%) | Preop, n (%) | IA, n (%) | Mixed n (%) |
|---|---|---|---|---|---|
| SN60WF | 14 | 6 (43) | 1 (7) | 5 (36) | 2 (14) |
| SN6ATx | 124 | 66 (53) | 33 (27) | 22 (18) | 3 (2) |
| SV25T0 | 22 | 1 (5) | 10 (45) | 4 (18) | 7 (32) |
|
| |||||
| All eyes | 160 | 73 (46) | 44 (28) | 31 (19) | 12 (8) |
Abbreviations: IOL, intraocular lens; Preop, preoperative; IA, intraoperative aberrometry.
Cumulative postoperative residual refractive error by group (“Same” is counted in both the IA and Preop groups)
| Group | Eyes | Within
| |||
|---|---|---|---|---|---|
| 0.25 D | 0.50 D | 0.75 D | 1.00 D | ||
| Same | 73 | 75% | 93.2% | 98.6% | 100% |
| Preop | 117 | 66% | 90.6% | 98% | 100% |
| IA | 104 | 74% | 89.4% | 97.1% | 99.0% |
| Mixed | 12 | 67% | 83.3% | 91.7% | 100.0% |
| All eyes | 160 | 67% | 88.1% | 96.9% | 99.4% |
Note:
Total is not the sum of all groups because Same is included in both Preop and IA groups.
Abbreviations: Preop, preoperative; IA, intraoperative aberrometry.
Difference in suggested lens power by IOL model
| IOL type | Eyes | IA suggests lower by
| No change | IA suggests higher by
| |||
|---|---|---|---|---|---|---|---|
| 1.5 D | 1.0 D | 0.5 D | 0.5 D | 1.0 D | |||
| SN60WF | 9 | 1 | 5 | 2 | 1 | ||
| SN6ATx | 124 | 1 | 4 | 38 | 66 | 14 | 1 |
| SV25T0 | 22 | 2 | 9 | 8 | 2 | 1 | |
| All lenses | 155 | 3 | 14 | 46 | 73 | 17 | 2 |
Abbreviations: IOL, intraocular lens; IA, intraoperative aberrometry.
Figure 1Prediction error by IOL calculation method.
Abbreviations: IOL, intraocular lens; IA, intraoperative aberrometry; Preop, preoperative; postop, postoperative; Rx, refraction.
Summary of results when Preop calculations and IA differed by 1.0 D or more
| IOL type | Sphere IOL power
| Postop Rx error | Expected IA impact | Percent of all IOLs in group | |||
|---|---|---|---|---|---|---|---|
| Pre | ORA | Implant | |||||
| SN60WF | 13.5 | 12.5 | 12.5 | −0.25 | Positive | 0.5 | 22 |
| 18 | 19 | 19 | −0.25 | Positive | |||
| SN6ATx | 9 | 7.5 | 8.5 | −0.13 | Positive | 1.0 | 4.70 |
| 9 | 8 | 8.5 | −0.13 | Positive | |||
| 12.5 | 11.5 | 12 | 0.25 | Negative | |||
| 23.5 | 24.5 | 24.5 | −1.63 | Negative | |||
| 25 | 24 | 25 | −0.63 | Positive | |||
| 34 | 33 | 34 | 0.25 | Negative | |||
| SV25T0 | 10.5 | 9.5 | 10 | 0.25 | Negative | 0.36 | 50 |
| 13.5 | 12.5 | 13.5 | −0.63 | Positive | |||
| 17.5 | 16 | 17 | −0.25 | Positive | |||
| 18.5 | 17.5 | 18 | −0.25 | Positive | |||
| 19.5 | 18.5 | 19.5 | −0.38 | Positive | |||
| 22 | 21 | 21.5 | −0.5 | Positive | |||
| 23 | 21.5 | 22.5 | −0.63 | Positive | |||
| 23 | 22 | 23 | −0.25 | Negative | |||
| 23.5 | 22.5 | 23.5 | −0.38 | Positive | |||
| 24 | 23 | 23.5 | 0.38 | Negative | |||
| 23.5 | 22.5 | 23.5 | 0.5 | Negative | |||
Notes:
SN60WF, Fisher’s exact test; SN6ATx and SV25T0, chi-square test.
Abbreviations: Preop, preoperative; IA, intraoperative aberrometry; IOL, intraocular lens; Postop Rx, postoperative refraction.
Surgeon choice of IOL power formula and “best” IOL power when IA and Preop calculations differ by 0.5 D
| Surgeon used | Eyes | Best IOL calculation
| |
|---|---|---|---|
| Preop | IA | ||
| Preop | 36 | 21 | 15 |
| IA | 27 | 7 | 20 |
| All | 63 | 28 | 35 |
Abbreviations: IOL, intraocular lens; IA, intraoperative aberrometry; Preop, preoperative.