Robert J Cionni1, Michael Breen2, Cody Hamilton3, Ravaughn Williams4.
Abstract
PURPOSE: To evaluate the spherical equivalent outcomes of intraoperative aberrometry (IA) power calculations compared with the surgeons' preoperative power calculations in eyes implanted with AcrySof® IQ T3 intraocular lenses (IOLs). PATIENTS AND METHODS: We assessed data collected by an IA system from multiple centers in the United States. Data was from patients who had undergone cataract extraction by phacoemulsification with the use of the Optiwave Refractive Analysis System and whose eyes had been implanted with model SN6AT3 (1.5 diopter [D] at IOL plane) aspheric lenses. The analysis was performed in 2 stages: hypothesis generation and confirmatory testing. Confirmatory endpoints were a comparison of absolute prediction errors for IA for the implanted IOL versus preoperative formula power calculations.Entities:
Keywords: astigmatism; intraocular lenses; low toric lenses; spherical equivalent outcomes
Year: 2019 PMID: 31496639 PMCID: PMC6689545 DOI: 10.2147/OPTH.S191887
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Preoperative lens power formulas used (N=6460)
| Formula | n | Percent |
|---|---|---|
| Haigis | 147 | 2.3 |
| Hoffer Q | 171 | 2.7 |
| Holladay 1 | 1038 | 16.1 |
| Holladay 2 | 3077 | 47.6 |
| Olsen | 3 | <0.01 |
| Other | 228 | 3.5 |
| SRK/T | 1796 | 27.8 |
Comparison of prediction errors by preoperative formula
| Formula (n [eyes]) | Preoperative formula | Intraoperative aberrometry | ||||
|---|---|---|---|---|---|---|
| Mean error | Median error | Percent of eyes with ≤0.5 D | Mean error | Median error | Percent of eyes with ≤0.5 D | |
| Holladay 1 (1038) | 0.35 | 0.28 | 77.3 | 0.30 | 0.23 | 82.8 |
| Holladay 2 (3077) | 0.36 | 0.28 | 76.2 | 0.29 | 0.24 | 83.9 |
| SRK/T (1796) | 0.36 | 0.29 | 75.2 | 0.29 | 0.23 | 83.1 |
Refractive outcomes in all eyes and in eyes in which power of the implanted IOL differed from that of the preoperatively planned lens, data for all patients
| All eyes | All patientsa | Patients with one eye only in the databaseb | ||||
|---|---|---|---|---|---|---|
| IA prediction error | Preoperative prediction error | Paired difference | IA prediction error | Preoperative prediction error | Paired difference | |
| MAE, D | 0.29 | 0.35 | −0.06c | 0.29 | 0.35 | −0.06c |
| MedAE, D | 0.23 | 0.28 | −0.04c | 0.24 | 0.28 | −0.03c |
| Eyes within ±0.5 D | ||||||
| n | 5388 | 4942 | - | 4212 | 3873 | - |
| % | 83.4 | 76.5 | 6.9c | 83.2 | 76.5 | 6.7c |
| MAE, D | 0.29 | 0.42 | −0.13c | 0.29 | 0.42 | −0.13c |
| MedAE, D | 0.23 | 0.34 | −0.15c | 0.24 | 0.34 | −0.14c |
| Eyes within ±0.5 D | ||||||
| n | 2155 | 1781 | - | 1692 | 1407 | - |
| % | 83.3 | 68.8 | 14.5c | 82.9 | 68.9 | 14.0c |
Notes: aN=6460 for all eyes, N=2587 for eyes in which power of the implanted IOL differed from that of plan. bN=5062 for all eyes, N=2042 for eyes in which power of the implanted IOL differed from that of plan. cp<0.0001.
Abbreviations: D, diopters; IOL, intraocular lens; MAE, mean absolute error; MedAE, median absolute error; IA, intraoperative aberrometry.
Figure 1Lens power implanted.
Abbreviation: IA, intraoperative aberrometry.