| Literature DB >> 31695316 |
Filomena J Ribeiro1,2, Tiago B Ferreira1,3, Catarina Relha1, Carina Esteves1, Sylvia Gaspar1.
Abstract
PURPOSE: To assess the efficacy of five calculators for toric intraocular lenses (IOL).Entities:
Keywords: astigmatism; toric IOL; toric IOL calculation
Year: 2019 PMID: 31695316 PMCID: PMC6718254 DOI: 10.2147/OPTH.S213132
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Mean absolute error in predicted residual astigmatism for each calculator
| Calculator | Mean ± SD (range) Diopters | 95% CI for mean | |
|---|---|---|---|
| 0.77±0.40 (0.01, 1.91) | 0.65–0.86 | - | |
| 0.35±0.29 (0.00, 1.36) | 0.27–0.44 | <0.001 | |
| 0.40±0.29 (0.00, 1.07) | 0.34–0.51 | <0.001 | |
| 0.72±0.41 (0.02, 1.87) | 0.64–0.85 | 0.656 | |
| 0.49±0.30 (0.00, 1.07) | 0.37–0.60 | 0.002 |
Abbreviations: SD, Standard deviation; TCRP, Total Corneal Refractive power; TCA, Total corneal astigmatism.
Centroid error in predicted residual astigmatism for each calculator
| Calculator | Centroid (diopters ± SD @ angle) 95% CI | ||
|---|---|---|---|
| 0.39±0.41 @ 166 0.33–0.45 | - | - | |
| 0.05±0.34 @ 167 0.00–0.10 | <0.001 | <0.001 | |
| 0.07±0.28 @ 160 0.03–0.11 | <0.001 | <0.001 | |
| 0.23±0.56 @ 158 0.16–0.31 | 0.134 | 0.078 | |
| 0.10±0.44 @ 156 0.04–0.16 | 0.009 | 0.042 |
Abbreviations: SD, Standard deviation; TCRP, Total Corneal Refractive power; TCA, Total corneal astigmatism.
Figure 1Double-angle plots illustrating the centroid prediction errors in residual astigmatism for each calculation method.
Mean absolute and centroid errors in predicted residual astigmatism for eyes with with-the-rule (WTR) and against-the-rule (ATR) corneal astigmatism
| Calculator | Eyes with WTR corneal astigmatism ( | Eyes with ATR corneal astigmatism ( | ||
|---|---|---|---|---|
| MAE ± SD diopters (range) 95% CI | Centroid (diopters ± SD @ angle) 95% CI | MAE ± SD diopters (range) 95% CI | Centroid (diopters ± SD @ angle) 95% CI | |
| 0.81±0.48 | 0.49±0.59 @ 168 | 0.64±0.43 | 0.38±0.44 @ 162 | |
| 0.31±0.23 | 0.08±0.39 @ 161 | 0.28±0.21 | 0.01±0.26 @ 175 | |
| 0.35±0.25 | 0.09±0.32 @ 155 | 0.26±0.19 | 0.01±0.25 @ 167 | |
| 0.73±0.42 | 0.39±0.61 @ 177 | 0.54±0.54 | 0.27±0.48 @ 179 | |
| 0.43±0.33 | 0.12±0.43 @ 145 | 0.35±0.29 | 0.07±0.39 @ 160 | |
Abbreviations: MAE, Mean absolute error; SD, Standard deviation; TCRP, Total Corneal Refractive power.
Figure 2Percentage of eyes within 0.25D–1.00D of absolute astigmatic prediction error with each calculation method.