A J Henningsson1, A Nilsson Bowers2, J Nordgren3, M Quttineh2, A Matussek2,4,5, S Haglund2.
Abstract
Noroviruses are a leading cause of epidemic and sporadic cases of acute gastroenteritis worldwide. The rapid diagnosis of norovirus infection is important for prompt infection control measures and may reduce the need for additional diagnostic testing. Here we evaluated the performance of the rapid Xpert Norovirus assay, and assessed the turn-around time (TAT) before and after the implementation of the analysis as a 24/7 service at all the three hospitals in Jönköping County, Sweden. We describe the implementation process which was performed in two steps during 2014. A total number of 276 clinical samples (stool and vomitus) from patients with symptoms of acute gastroenteritis were included in 2014-2015. The samples were analysed with the Xpert Norovirus assay and the already existing routine method: an in-house reverse transcription real-time PCR. Samples showing discrepant results with the two assays were further analysed by a third PCR method. The Xpert Norovirus assay performed well with a sensitivity of 100% and a specificity of 93% compared to the gold standard (defined as the result obtained by at least two of the three PCR methods). The median TAT decreased from 22 hours in 2013 to 2.4 hours in 2015 (p<0.001). We conclude that the performance of the Xpert Norovirus assay was excellent, and that the implementation of the analysis as a 24/7 service at all three hospitals in the county has greatly reduced the time to diagnosis which is beneficial for both patients and healthcare providers.Entities:
Mesh:
Year: 2017 PMID: 28540537 PMCID: PMC5602093 DOI: 10.1007/s10096-017-3005-9
Source DB: PubMed Journal: Eur J Clin Microbiol Infect Dis ISSN: 0934-9723 Impact factor: 3.267
Comparison between the gold standard test resulta and the Xpert Norovirus assay for detection of noroviruses (NoV)
| Method of detection | Gold standarda | |||
|---|---|---|---|---|
| Positive | Negative | Total | ||
| Xpert NoV | Positive | 84 | 13 | 97 |
| Negative | 0 | 179 | 179 | |
| Total | 84 | 192 | 276 | |
aGold standard: the result obtained by at least two of the three following methods for detection of NoVs: the Xpert Norovirus assay, the in-house method based on Kageyama et al. [12], and the method based on Nordgren et al. [13, 14].
The median turn-around time (TAT) before and after the GeneXpert® Norovirus assay was implemented at the laboratories of clinical chemistry as a 24/7 service at all three hospitals in the county
| Setting | 2013 | 2015 | P-value | ||
|---|---|---|---|---|---|
| TAT; hours (n) | IQR | TAT; hours (n) | IQR | ||
| All clients | 22 (1124) | 14–30 | 2.4 (1289) | 2.0–4.2 | *** |
| All clients excl. PHC | 20 (933) | 11–26 | 2.3 (1123) | 2.0–3.2 | *** |
| Jönköping | 19 (583) | 7.8–25 | 2.4 (778) | 2.0–3.4 | *** |
| Eksjö | 20 (140) | 14–26 | 2.0 (161) | 1.9–2.5 | *** |
| Värnamo | 24 (194) | 19–29 | 2.2 (195) | 2.0–3.6 | *** |
| PHC only | 38 (191) | 27–48 | 14 (166) | 6.4–22 | *** |
PHC primary healthcare centre, TAT median turn-around time from sampling to available test result, IQR interquartile range
*** P < 0.001
Fig. 1Percentage of test results reported within 4 hours from sample arrival at the laboratory before and after the GeneXpert® Norovirus assay was implemented at the laboratories of clinical chemistry as a 24/7 service at all three hospitals in the county. May 2014: analysis was performed during the daytime with GeneXpert. November 2014: analysis of NoV was implemented as a 24/7 service