| Literature DB >> 28497007 |
Sahar V Doctorvaladan1, Andrea T Jelks1, Eric W Hsieh2, Robert L Thurer2, Mark I Zakowski3, David C Lagrew4.
Abstract
Objective This study aims to compare the accuracy of visual, quantitative gravimetric, and colorimetric methods used to determine blood loss during cesarean delivery procedures employing a hemoglobin extraction assay as the reference standard. Study Design In 50 patients having cesarean deliveries blood loss determined by assays of hemoglobin content on surgical sponges and in suction canisters was compared with obstetricians' visual estimates, a quantitative gravimetric method, and the blood loss determined by a novel colorimetric system. Agreement between the reference assay and other measures was evaluated by the Bland-Altman method. Results Compared with the blood loss measured by the reference assay (470 ± 296 mL), the colorimetric system (572 ± 334 mL) was more accurate than either visual estimation (928 ± 261 mL) or gravimetric measurement (822 ± 489 mL). The correlation between the assay method and the colorimetric system was more predictive (standardized coefficient = 0.951, adjusted R2 = 0.902) than either visual estimation (standardized coefficient = 0.700, adjusted R2 = 00.479) or the gravimetric determination (standardized coefficient = 0.564, adjusted R2 = 0.304). Conclusion During cesarean delivery, measuring blood loss using colorimetric image analysis is superior to visual estimation and a gravimetric method. Implementation of colorimetric analysis may enhance the ability of management protocols to improve clinical outcomes.Entities:
Keywords: blood loss measurement; cesarean delivery; postpartum hemorrhage; quality improvement
Year: 2017 PMID: 28497007 PMCID: PMC5425292 DOI: 10.1055/s-0037-1601382
Source DB: PubMed Journal: AJP Rep ISSN: 2157-7005
Fig. 1Distribution of amniotic fluid volume recorded by marking the suction canister for volumetric assessment.
Fig. 2(A) Distribution of sponge blood content as measured by the assay method. (B) Distribution of canister blood content as measured by the assay method.
Fig. 3Distribution of blood loss as determined by the assay method.
Blood loss determinations
| Method | Extraction Assay | Visual EBL | QBLGrav (adjusted) | Colorimetric | |
|---|---|---|---|---|---|
| Sponge (mL) | Mean ± SD | 280 ± 222 | 759 ± 317 | 332 ± 255 | |
| Median (IQR) | 203 (138) | 695 (426) | 251 (181) | ||
| Bias (95% CI) | 480 (428–531) | 52 (32–71) | |||
|
| < 0.001 | < 0.001 | |||
| Canister (mL) | Mean ± SD | 190 ± 133 | 63 ± 335 | 240 ± 137 | |
| Median (IQR) | 142 (173) | −13 (285) | 199 (202) | ||
| Bias (95% CI) | −127 (−228 to −26) | 50 (26–74) | |||
|
| 0.014 | < 0.001 | |||
| Total blood loss per procedure (mL) | Mean ± SD | 470 ± 296 | 928 ± 261 | 822 ± 489 | 572 ± 334 |
| Median (IQR) | 384 (296) | 800 (200) | 651 (475) | 481 (332) | |
| Bias (95% CI) | 458 (396–520) | 352 (237–467) | 102 (72–132) | ||
|
| < 0.001 | < 0.001 | < 0.001 |
Abbreviations: CI, confidence interval; EBL, estimated blood loss; IQR, interquartile range; QBLGrav, quantitative blood loss; SD, standard deviation.
Adjusted by subtracting the measured amniotic fluid volume and irrigation volume.
Note: p Values reflect the statistical significance level of paired t-tests comparing each method with the extraction assay (reference standard).
Fig. 4Scatter plot of the blood content of each sponge comparing the assay and gravimetric methods.
Linear correlation of blood loss measurements versus extraction assay (reference standard)
| Method | Correlation coefficient (95% CI) | Standardized coefficient |
| Adjusted R2 |
|---|---|---|---|---|
| Visual EBL | 0.700 (0.523–0.819) | 0.700 | < 0.001 | 0.479 |
| QBLGrav (adjusted) | 0.564 (0.339–0.728) | 0.564 | < 0.001 | 0.304 |
| Colorimetric | 0.951 (0.915–0.972) | 0.951 | < 0.001 | 0.902 |
Abbreviations: CI, confidence interval; EBL, estimated blood loss; QBLGrav, quantitative blood loss.
Fig. 5Scatter plots of blood loss measured by (A) visual estimation, (B) gravimetric method, and (C) colorimetric method compared with the assay method. A line of unity representing perfect correlation is shown for comparison.
Assessment of agreement between methods of measuring blood loss and the extraction assay (reference standard)
| Visual EBL (mL) | QBLGrav (adjusted) (mL) | Colorimetric (mL) | |
|---|---|---|---|
| Bias (95% CI) (mL) | 458 | 352 | 102 |
| SD (error) (mL) | 218 | 405 | 106 |
| Upper limit of agreement (95% CI) (mL) | 886 | 1,145 | 309 |
| Lower limit of agreement (95% CI) (mL) | 31 | −441 | −105 |
| RMSE (mL) | 507 | 533 | 146 |
| CI (bias) | 62 | 115 | 30 |
| CI (LOAs)—calculated as | 105 | 194 | 51 |
Abbreviations: CI, confidence interval; EBL, estimated blood loss; LOA, limits of agreement; QBLGrav, quantitative blood loss; RMSE, root mean square error; SD, standard deviation.
Fig. 6Bland–Altman plot: Visual method.
Fig. 7Bland–Altman plot: Gravimetric method.
Fig. 8Bland–Altman plot: Colorimetric method.