Adaptation and validation of a commercial head phantom for cranial radiosurgery dosimetry end-to-end audit.

Alexis Dimitriadis1,2,3, Antony L Palmer1,4, Russell A S Thomas3, Andrew Nisbet1,2, Catharine H Clark1,3.   

Abstract

OBJECTIVE: To adapt and validate an anthropomorphic head phantom for use in a cranial radiosurgery audit.
METHODS: Two bespoke inserts were produced for the phantom: one for providing the target and organ at risk for delineation and the other for performing dose measurements. The inserts were tested to assess their positional accuracy. A basic treatment plan dose verification with an ionization chamber was performed to establish a baseline accuracy for the phantom and beam model. The phantom and inserts were then used to perform dose verification measurements of a radiosurgery plan. The dose was measured with alanine pellets, EBT extended dose film and a plastic scintillation detector (PSD).
RESULTS: Both inserts showed reproducible positioning (±0.5 mm) and good positional agreement between them (±0.6 mm). The basic treatment plan measurements showed agreement to the treatment planning system (TPS) within 0.5%. Repeated film measurements showed consistent gamma passing rates with good agreement to the TPS. For 2%-2 mm global gamma, the mean passing rate was 96.7% and the variation in passing rates did not exceed 2.1%. The alanine pellets and PSD showed good agreement with the TPS (-0.1% and 0.3% dose difference in the target) and good agreement with each other (within 1%).
CONCLUSION: The adaptations to the phantom showed acceptable accuracies. The presence of alanine and PSD do not affect film measurements significantly, enabling simultaneous measurements by all three detectors. Advances in knowledge: A novel method for thorough end-to-end test of radiosurgery, with capability to incorporate all steps of the clinical pathway in a time-efficient and reproducible manner, suitable for a national audit.

Entities:  

Mesh:

Year:  2017        PMID: 28452563      PMCID: PMC5602186          DOI: 10.1259/bjr.20170053

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  19 in total

1.  Current status of cranial stereotactic radiosurgery in the UK.

Authors:  Alexis Dimitriadis; Karen J Kirkby; Andrew Nisbet; Catharine H Clark
Journal:  Br J Radiol       Date:  2015-12-21       Impact factor: 3.039

2.  Spectral method for the correction of the Cerenkov light effect in plastic scintillation detectors: a comparison study of calibration procedures and validation in Cerenkov light-dominated situations.

Authors:  Mathieu Guillot; Luc Gingras; Louis Archambault; Sam Beddar; Luc Beaulieu
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

3.  A national dosimetric audit of IMRT.

Authors:  Geoff Budgell; Joe Berresford; Michael Trainer; Ellie Bradshaw; Peter Sharpe; Peter Williams
Journal:  Radiother Oncol       Date:  2011-05-05       Impact factor: 6.280

Review 4.  A contemporary review of stereotactic radiotherapy: inherent dosimetric complexities and the potential for detriment.

Authors:  Michael L Taylor; Tomas Kron; Rick D Franich
Journal:  Acta Oncol       Date:  2011-02-03       Impact factor: 4.089

5.  Multichannel film dosimetry with nonuniformity correction.

Authors:  Andre Micke; David F Lewis; Xiang Yu
Journal:  Med Phys       Date:  2011-05       Impact factor: 4.071

6.  A multicentre 'end to end' dosimetry audit for cervix HDR brachytherapy treatment.

Authors:  Antony L Palmer; Patricia Diez; Laura Gandon; Andrea Wynn-Jones; Peter Bownes; Chris Lee; Edwin Aird; Margaret Bidmead; Gerry Lowe; David Bradley; Andrew Nisbet
Journal:  Radiother Oncol       Date:  2015-01-09       Impact factor: 6.280

7.  A national dosimetry audit for stereotactic ablative radiotherapy in lung.

Authors:  Gail Distefano; Jonny Lee; Shakardokht Jafari; Clare Gouldstone; Colin Baker; Helen Mayles; Catharine H Clark
Journal:  Radiother Oncol       Date:  2017-01-20       Impact factor: 6.280

8.  Design, development, and implementation of the radiological physics center's pelvis and thorax anthropomorphic quality assurance phantoms.

Authors:  David S Followill; DeeAnn Radford Evans; Christopher Cherry; Andrea Molineu; Gary Fisher; William F Hanson; Geoffrey S Ibbott
Journal:  Med Phys       Date:  2007-06       Impact factor: 4.071

9.  Development of a modified head and neck quality assurance phantom for use in stereotactic radiosurgery trials.

Authors:  Austin M Faught; Stephen F Kry; Dershan Luo; Andrea Molineu; David Bellezza; Russell L Gerber; Scott E Davidson; Walter Bosch; Robert E Drzymala; Jim Galvin; Robert Timmerman; Jason Sheehan; Michael T Gillin; Geoffrey S Ibbott; David S Followill
Journal:  J Appl Clin Med Phys       Date:  2013-07-08       Impact factor: 2.102

10.  Evaluation and mitigation of potential errors in radiochromic film dosimetry due to film curvature at scanning.

Authors:  Antony L Palmer; David A Bradley; Andrew Nisbet
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

View more
  5 in total

Review 1.  Novel methodologies for dosimetry audits: Adapting to advanced radiotherapy techniques.

Authors:  Marlies Pasler; Victor Hernandez; Núria Jornet; Catharine H Clark
Journal:  Phys Imaging Radiat Oncol       Date:  2018-03-19

2.  The role of dosimetry audit in achieving high quality radiotherapy.

Authors:  Catharine H Clark; Núria Jornet; Ludvig P Muren
Journal:  Phys Imaging Radiat Oncol       Date:  2018-03-21

3.  Determination of commissioning criteria for multileaf-collimator, stereotactic radiosurgery treatments on Varian TrueBeam and Edge machines using a novel anthropomorphic phantom.

Authors:  Thomas A D Brown; Jessica M Fagerstrom; Caleb Beck; Connor Holloway; Krista Burton; Darryl G L Kaurin; Saikanth Mahendra; Marcus Luckstead; Kayla Kielar; James Kerns
Journal:  J Appl Clin Med Phys       Date:  2022-03-15       Impact factor: 2.243

4.  Development of a dedicated phantom for multi-target single-isocentre stereotactic radiosurgery end to end testing.

Authors:  Joel Poder; Ryan Brown; Harry Porter; Rashmi Gupta; Anna Ralston
Journal:  J Appl Clin Med Phys       Date:  2018-09-16       Impact factor: 2.102

5.  Evaluation of the RUBY modular QA phantom for planar and non-coplanar VMAT and stereotactic radiations.

Authors:  Daniela Poppinga; Jana Kretschmer; Leonie Brodbek; Jutta Meyners; Bjoern Poppe; Hui Khee Looe
Journal:  J Appl Clin Med Phys       Date:  2020-08-14       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.